The LifeLens Physiological Monitor is intended for spot-check monitoring and continuous data storage of physiological data in home and healthcare settings. Data could include ECG, oxygen saturation of arterial hemoglobin (SpO2), heart rate (HR), heart rate variability (HRV), skin temperature (TEMP), respiration rate (RR). Data are securely transmitted from the device for storage, review, analysis, and display. Data from the device are intended as an aid to diagnosis, disease management, and treatment. The device is intended for use on adults (18 years of age or older). The LifeLens Physiological Monitor also reports wellness data including activity level, body position, step count, and gait analysis.
Device Story
Ambulatory physiological monitoring system; measures ECG, SpO2, HR, HRV, skin temperature, respiration rate, and wellness metrics (posture, balance, activity, steps). System components: disposable skin-contact patch, reusable battery-powered hub, reusable recharger, and mobile application. Hub records/stores/transmits data via Bluetooth to mobile app; includes microphones/tap detection for symptom annotation. Used in home/healthcare settings by adults. Data transmitted for storage, review, and display by clinicians/users as aid to diagnosis/management. No automated analysis, active monitoring, or alarm generation. Benefits: enables remote physiological data collection for disease management.
Clinical Evidence
Bench testing only. SpO2 validated in 29-participant study (70-100% SaO2 range) against arterial blood draw reference. HR and RR validated in 94-participant study against cleared reference devices. Testing verified compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-47, and ASTM E1112-00(2018).
Indicated for spot-check monitoring and continuous storage of physiological data (ECG, SpO2, HR, HRV, skin temperature, respiration rate) and wellness metrics (activity, posture, steps, gait) in adults (18+) in home and healthcare settings. Not intended for automated analysis, active monitoring, or alarm generation.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
UbiqVue 2A Multiparameter System (UX2550) (K242018)
Submission Summary (Full Text)
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
July 9, 2026
LifeLens Technologies, Inc.
% Endri Angjeli
Regulatory Consultant
MEDIcept, Inc.
200 Homer Ave.
Ashland, Massachusetts 01721
Re: K260157
Trade/Device Name: LifeLens Physiological Monitor
Regulation Number: 21 CFR 870.2800
Regulation Name: Electrocardiograph, ambulatory (without analysis)
Regulatory Class: Class II
Product Code: MWJ, DRG, DQA, BZQ, FLL,
Dated: June 6, 2026
Received: June 8, 2026
Dear Endri Angjeli:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260157 - Endri Angjeli
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K260157 - Endri Angjeli
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JENNIFER W. SHIH -S
Jennifer Kozen
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260157 | ? |
| Please provide the device trade name(s). | | ? |
| LifeLens Physiological Monitor | | |
| Please provide your Indications for Use below. | | ? |
| The LifeLens Physiological Monitor is intended for spot-check monitoring and continuous data storage of physiological data in home and healthcare settings. Data could include ECG, oxygen saturation of arterial hemoglobin (Sp02), heart rate (HR), heart rate variability (HRV), skin temperature (TEMP), respiration rate (RR). Data are securely transmitted from the device for storage, review, analysis, and display. Data from the device are intended as an aid to diagnosis, disease management, and treatment. The device is intended for use on adults (18 years of age or older). The LifeLens Physiological Monitor also reports wellness data including activity level, body position, step count, and gait analysis. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
# K260157 510(K) SUMMARY
510(k) Owner: LifeLens Technologies, Inc.
Address: 1 Ivybrook Blvd, Suite 100
Ivyland, PA 18974
Phone: (215) 622-4715
Contact Person: Matt Walenciak, VP of Quality & Project Management
Prepared by: Endri Angjeli, Regulatory Consultant
Prepared Date: January 16, 2025
Trade Name: LifeLens Physiological Monitor
Classification:
| Common Name | Regulations Number | Product Code |
| --- | --- | --- |
| Electrocardiograph, ambulatory (without analysis) | 21 CFR 870.2800 | MWJ |
| Radiofrequency physiological signal transmitter and receiver | 21 CFR 870.2910 | DRG |
| Oximeter | 21 CFR 870.2700 | DQA |
| Monitor, breathing frequency | 21 CFR 868.2375 | BZQ |
| Continuous measurement thermometer | 21 CFR 880.2910 | FLL |
Predicate Devices:
| Predicate 510(k) Number | Predicate Device Name | Device Type |
| --- | --- | --- |
| Primary K203168 | LifeLens Wireless ECG Monitor | DSH |
| Secondary K242018 | UbiqVue 2A Multiparameter System (UX2550) | DRG, DQA, MHX, FLL |
Page 1 of 7
{5}
K260157
## **Indications For Use:**
The LifeLens Physiological Monitor is intended for spot-check monitoring and continuous data storage of physiological data in home and healthcare settings. Data could include ECG, oxygen saturation of arterial hemoglobin (SpO2), heart rate (HR), heart rate variability (HRV), skin temperature (TEMP), respiration rate (RR). Data are securely transmitted from the device for storage, review, analysis, and display. Data from the device are intended as an aid to diagnosis, disease management, and treatment. The device is intended for use on adults (18 years of age or older). The LifeLens Physiological Monitor also reports wellness data including activity level, body position, step count, and gait analysis.
## **Device Description:**
The LifeLens Physiological Monitor is an ambulatory physiological monitoring system intended to measure oxygen saturation (SpO2), heart rate, respiration rate, electrocardiogram (ECG) data, skin temperature, and heart rate variability. In addition, the device also calculates and displays the following wellness metrics: Posture, Balance, Activity, and Steps.
The system consists of a disposable Patch, a reusable Hub, a reusable Recharger, and a Mobile Application.
The Skin Attachment Component, referred to as the SAC or Patch, is a single-use disposable, passive, hypoallergenic, thin, and breathable skin interface to which the Hub connects hermetically.
The Hub is a small, reusable, battery powered component including an electrical interface for coupling to the Patch, and hardware to record, store, and transmit physiological data from the subject during use. Additionally, the Hub includes microphones and tap detection circuitry to facilitate annotation (i.e., audio recordings) of symptoms by the user. The hub is a multi-patient use device.
The Recharger is a battery-powered, multi-purpose, reusable device for storing and charging two Hubs, interfacing with the Hub(s) during use, storing patient data, and allowing data offload after use. The recharger is a multi-patient use device.
The LifeLens mobile application connects to LifeLens Hub via Bluetooth to display device data output and status.
The device is not intended for automated analysis or active patient monitoring and is not intended to generate alarms or be used with alarm systems.
Page 2 of 7
{6}
K260157
Substantial Equivalence:
| | **Subject Device** LifeLens Physiological Monitor | **Primary Predicate K203168** LifeLens Wireless ECG Monitor | **Secondary Predicate K242018** UbiqVue 2A Multiparameter System (UX2550) | **Comparison to both Predicates** |
| --- | --- | --- | --- | --- |
| **Product Code** | DSH, DRG, DQA, BZQ, FLL | DSH | DRG, MHX, FLL, DQA | Similar |
| **Regulation Classification (Primary)** | 21 CFR 870.2800 Class II | 21 CFR 870.2800 Class II | 21 CFR 870.2910 Class II | Same |
| **Indications For Use** | The LifeLens Physiological Monitor is intended for spot-check monitoring and continuous data storage of physiological data in home and healthcare settings. Data could include ECG, oxygen saturation of arterial hemoglobin (SpO2), heart rate (HR), heart rate variability (HRV), skin temperature (TEMP), respiration rate (RRp) and wellness data (activity level, body position, step count, and gait analysis). Data are securely transmitted from the device for storage, review, analysis, and display. Data | LifeLens Wireless ECG Monitor is indicated for use on patients who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light headedness, pre-syncope, syncope, fatigue, chest pain and/or anxiety. The LifeLens Wireless ECG Monitor is intended for use by patients 18 years or | The UbiqVueTM 2A Multiparameter System is a wireless remote patient monitoring system intended for use by healthcare professionals for continuous collection of physiological data at home and in healthcare settings. This shall include electrocardiography, heart rate, SpO2%, respiration rate, pulse rate, photoplethysmography waveform, skin temperature, body temperature, body posture, body motion, R-R Interval, heart rate variability (HRV) and Blood Pressure (optional). Data is transmitted wirelessly near real time from UbiqVue 2A Wearable | Similar |
Page 3 of 7
{7}
K260157
| | Subject Device LifeLens Physiological Monitor | Primary Predicate K203168 LifeLens Wireless ECG Monitor | Secondary Predicate K242018 UbiqVue 2A Multiparameter System (UX2550) | Comparison to both Predicates |
| --- | --- | --- | --- | --- |
| | from the device are intended as an aid to diagnosis, disease management, and treatment. The device is intended for use on adults (18 years of age or older). | older. The LifeLens Wireless ECG Monitor is not intended to be worn during defibrillation. | Biosensor and 3rd party device (for Blood Pressure only) to Remote Central server for display, storage, and analysis. The UbiqVueTM 2A Multiparameter System is intended for non-critical, adult population. The UbiqVueTM 2A Multiparameter System shall include the ability to notify healthcare professionals through alerts when Physiological parameters fall outside the set limits, manual trigger by patient and to display multiple patient's physiological data for remote monitoring at home and with visual alarm for active monitoring at hospitals and out of hospital patient care settings (such as clinics, outpatient surgery facilities, long-term care facilities and physician offices) in which care is administered by healthcare professionals. | |
| **Intended Population** | Non- Critical patient 18 years or older | Non- Critical patient 18 years or older | Non- Critical adult patient | Same |
Page 4 of 7
{8}
K260157
| | Subject Device LifeLens Physiological Monitor | Primary Predicate K203168 LifeLens Wireless ECG Monitor | Secondary Predicate K242018 UbiqVue 2A Multiparameter System (UX2550) | Comparison to both Predicates |
| --- | --- | --- | --- | --- |
| Intended Use Environment | Home and Healthcare setting | Home and Healthcare setting | Home and Healthcare setting | Same |
| Monitored Parameters | | | | |
| ECG | 1 lead | 1 lead | N/A | Same |
| Heart Rate | Range: 30-220 bpm Accuracy: ± 3 bpm | N/A | 30-250 bpm Accuracy: ± 3 bpm or 5% whichever is greater | Similar |
| Respiration Rate | Range: 4-60 rpm Accuracy: ± 3 rpm | N/A | Range: 5-60 rpm Accuracy: ≤ 3 rpm | Similar |
| Skin Temperature | Range: 24°C to 44°C Accuracy: ± 0.1°C | N/A | Range: 15°C to 43°C Accuracy: As per ASTM E1112-00 | Similar |
| SpO2% | ± 3.2 % (100 to 70%) Less than 70% unspecified | N/A | ± 3 % (100 to 70%) Less than 70% unspecified | Similar |
### Summary of Technological Characteristics of Subject Devices Compared to Predicate Devices:
The subject device is a modified version of the primary predicate. The subject device incorporates additional physiological measurement capabilities. These additional measurement capabilities are similar to those of the secondary predicate. The subject device and both predicate devices have the same technological characteristics, including being battery powered, having dry-electrode electrocardiography (ECG), reflective photoplethysmography (PPG), and other comparable sensors. They are wireless, wearable devices with similar form factors, constructed with similar materials, which record and transmit physiological parameters. Any differences in features do not affect the safety or effectiveness of their intended use.
Page 5 of 7
{9}
K260157
# **Non-Clinical Performance Testing:**
| Electrical Safety and electromagnetic compatibility (EMC) | EMC, Electrical, Mechanical, and Thermal Safety were verified in accordance with IEC 60601-1 (ed. 3.2), IEC 60601-1-2: 2020 |
| --- | --- |
| Software Validation and Verification | Software verification and validation testing was conducted in accordance with FDA Guidance “Content of Premarket Submissions for Device Software Functions” |
| Multi-patient Use and Service Life testing | Multi-patient Use and Service Life testing was conducted in accordance with FDA Guidance “Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Guidance for Industry and Food and Drug Administration Staff” |
| Biocompatibility | The Patch, which is the skin-contact component, has been previously tested for cytotoxicity, irritation and skin sensitization, and has not changed from the predicate. |
| ECG | The ECG acquisition, has been previously tested in accordance with AAMI/ANSI/IEC 60601-2-47:2012 Medical Electrical Equipment – Part 2- 47: Particular Requirements for the Basic Safety and Essential Performance of Ambulatory Electrocardiographic Systems and has not changed from the predicate. |
| Cybersecurity | Cybersecurity testing and controls have been implemented in accordance with FDA Guidance “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions” |
| Human Factors | Human Factors verification was testing was conducted in accordance with FDA Guidance “Applying Human Factors and Usability Engineering to Medical Devices” |
| Heart Rate | Performance of heart rate measurements was verified in accordance to IEC 60601-2-47 Medical electrical equipment – Part 2-47: Particular requirements for the basic safety and essential performance of ambulatory electrocardiographic systems. |
| Skin Temperature | Performance of skin temperature measurement was verified with use of ASTM E111200(2018), Standard Specification for Electronic Thermometers for Intermittent Determination of Patient Temperature. |
Page 6 of 7
{10}
K260157
# Non-clinical Performance Validation
Oxygen saturation performance was validated in a study using arterial blood draw reference measurements from 29 participants, spanning an SaO2 range of 70% to 100%. The study included data from participants of diverse sets of demographic and skin colors/tones. The primary endpoint was met at each placement location.
Heart Rate and Respiration Rate performance was validated through a study evaluating data from 94 participants compared to cleared reference devices. The study included data from participants from diverse sets of demographic and various reported medical history components, consistent with intended use population. The primary endpoints were met at each placement location.
This medical device product has functions subject to FDA premarket review as well as functions that are not subject to FDA premarket review. For this application, if the product has functions that are not subject to FDA premarket review, FDA assessed those functions only to the extent that they either could adversely impact the safety and effectiveness of the functions subject to FDA premarket review or they are included as a labeled positive impact that was considered in the assessment of the functions subject to FDA premarket review.
# Conclusion:
Results of the testing show that the proposed subject device meets its intended use and support a determination that the proposed subject device does not raise new questions of safety or effectiveness as compared to the predicate devices. Therefore, the subject device is substantially equivalent to the predicate devices in terms of indications for use, design, technological characteristics, modes of operations, safety, and effectiveness.
Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.