Fitbone Transport and Lengthening System is indicated for limb lengthening, open and closed fracture fixation, pseudoarthrosis, malunions, non-unions, or bone transport of the long bones. The Fitbone Transport and Lengthening System is indicated for adult only.
Device Story
Implantable intramedullary pulling transport nail; implanted in femur or tibia medullary canal; connected to subcutaneous receiver via bipolar feed line. External control set (electronics station/transmitter) induces electromagnetic field; receiver converts to DC voltage to power internal motor; motor drives telescope distraction. Used in operating theatre by surgeons; distraction activated by patient at home or in clinic. Retraction mode restricted to HCPs for over-distraction or plan adjustments. Provides controlled bone lengthening/transport; benefits patients with limb length discrepancies or bone defects.
Clinical Evidence
Bench testing only. Performance verified via mechanical testing including inverted load test, static 4-point bending, fatigue 4-point bending, and torsional testing per ASTM F1264-16.
Technological Characteristics
Intramedullary fixation rod; stainless steel 1.4441 (ASTM F138) and medical grade silicone. Distraction via internal motor powered by electromagnetic induction from external transmitter to subcutaneous receiver. Lengths 290-490mm; diameters 11-13mm. Sterilization: H2O2 plasma. Standalone implant system with external control unit.
Indications for Use
Indicated for adult patients requiring limb lengthening, open/closed fracture fixation, pseudoarthrosis, malunions, non-unions, or bone transport of long bones.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
Predicate Devices
FITBONE® Transport and Lengthening System (K232169)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 19, 2026
Orthofix Srl
Chiara Zuliani
Senior Technical Product RA Specialist
Via Delle Nazioni, 9
Bussolengo (Vr), IT 37012
Italy
Re: K260146
Trade/Device Name: Fitbone® Transport And Lengthening System
Regulation Number: 21 CFR 888.3020
Regulation Name: Intramedullary fixation rod
Regulatory Class: Class II
Product Code: HSB
Dated: April 16, 2026
Received: April 17, 2026
Dear Chiara Zuliani:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260146 - Chiara Zuliani
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260146 - Chiara Zuliani
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
FARZANA SHARMIN -S
Farzana Sharmin, PhD
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260146 | ? |
| Please provide the device trade name(s). | | ? |
| FITBONE® TRANSPORT AND LENGTHENING SYSTEM | | |
| Please provide your Indications for Use below. | | ? |
| Fitbone Transport and Lengthening System is indicated for limb lengthening, open and closed fracture fixation, pseudoarthrosis, malunions, non-unions, or bone transport of the long bones. The Fitbone Transport and Lengthening System is indicated for adult only. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
ORTHOFIX
K260146-Page 1
# 510(K) SUMMARY
# ORTHOFIX SRL
# FITBONE® Transport and Lengthening System
## Submitter information
| Company Name: | Orthofix S.r.l. |
| --- | --- |
| Address | Via Delle Nazioni, 9 37012 Bussolengo (VR) - Italy |
| Telephone | +39 045 6719000 |
| Fax | +39 045 6719380 |
| Contact Person | Chiara Zuliani Senior Technical Product RA Specialist |
| --- | --- |
| Address | Via Delle Nazioni, 9 37012 Bussolengo (VR) - Italy |
| Telephone | +39 045 6719000 |
| Fax | +39 045 6719380 |
| Email address | chiarazuliani@orthofix.it |
| Date of submission | 2026, January 19 |
## Trade Name, Common Name, Classification
Trade Name: FITBONE® Transport and Lengthening System
Common Name: Rod, Fixation, Intramedullary and Accessories
Classification Name: Intramedullary Fixation rod
Regulation Number: 21 CFR 888.3020
Product Code: HSB
Classification: Class II
Panel code: Orthopedic
## Predicate devices and additional reference device
| Primary Predicate | 510(k) Number | Manufacturer |
| --- | --- | --- |
| FITBONE® Transport and Lengthening System | K232169 | Orthofix s.r.l. |
| Reference Device | 510(k) Number | Manufacturer |
| FITBONE® TAA | K163368 | Orthofix s.r.l. |
| | K203399 | |
## Device description
The subject Fitbone® transport and lengthening system consists of the implantable intramedullary pulling transport nails and is a line extension of the existing primary predicate Fitbone® transport and lengthening nails (K232169).
The Subject device is implanted into the medullary canal of the femur or tibia and connected to the additional predicate/reference device intracutaneous Receiver (K203399 and K163368) by a bipolar feed line. The external FITBONE Control Set is identical to that previously
Via delle Nazioni 9 | 37012 Bussolengo | Verona, Italy
+39.045.6719000
orthofix.com
ORTHOFIX SRL
ORTHOFIX TECNOPIANISTICI Orthofix Medical Inc. Cintamano
SEDE LEGALE: VALVITTORI PISANI 90 | 20014 MILANO UTAIJA | C.F. e PISA DIRITENZI 100 | CZE: VIST NUMBER (TOPI/1N08750) Registro Imprese di Milano DIR70040/50 | R.E.A. M11257/89 | Cap. Soc. Eur. 1546.000 mt. vers.
SEDE SECONDARIA: VALDELLE NAZIONI, 9 | 37012 BUSSOLENGO, MERONALI ITALIA | Registro Imprese di Verona | R.E.A. VR 215209
{5}
ORTHOFIX®
K260146-Page 2
| | cleared for the additional predicate/reference device Fitbone TAA device (K203399, K163368) and consists of a control electronics station and transmitter. The power required for the distraction process is controlled by hermetically enclosed motor which draws the telescope apart. The electro-magnetic field sent from the Transmitter to the Receiver is converted in the Receiver into DC-Voltage to supply the motor of the subject Fitbone® transport and lengthening nails with voltage, when actioned. The nail is anchored to the bone by locking screws through medial-lateral and AP holes in the nail depending on the configuration holes in the nail. The energy needed for the distraction process is transmitted from the outside by placing the external transmitter over the implanted receiver, which is placed in the subcutaneous tissue during surgery. There is no transcutaneous contact between the implanted intramedullary nail and the outer surface of the patient's body. The subject nails, as the primary predicate Fitbone® transport and lengthening system, in same sterile conditions are made from implant grade stainless steel 1.4441, according to ASTM F138 “Standard Specification for Wrought 18Chromium-14Nickel-2.5Molybdenum Stainless Steel Bar and Wire for Surgical Implants (UNS S31673)”, and medical grade Silicone. The Subject, as the predicates, will be implanted only by Healthcare Professionals (HCP), with full awareness of the appropriate orthopedic procedures (including application and removal), in the operating theatre only. The limb distraction of the femur and tibia treatment will be activated in home by the patient or clinic theatres. The retraction mode of the intramedullary nail will be operated only by HCPs, in case of unintended over-distraction or if nail Rewind & Go is needed as per preoperative plan. |
| --- | --- |
| Indications for use | Fitbone Transport and Lengthening System is indicated for limb lengthening, open and closed fracture fixation, pseudoarthrosis, malunions, non-unions, or bone transport of the long bones. The Fitbone Transport and Lengthening System is indicated for adult only. |
| Technological Characteristics | The following table provides a comparison and assessment of fundamental scientific principles and technological characteristic of the subject, primary and additional reference device. Any differences have been demonstrated not to raise different issues of safety or performance by virtue of objective evidence (e.g., bench testing, process validations, etc.). Therefore, *technological characteristics* of the new Subject Fitbone® transport and lengthening nails are substantially equivalent to the primary predicate Fitbone® transport and lengthening system (K232169) and reference device. |
| Technological Characteristic | | Subject Device (Fitbone® transport and lengthening system) | Predicate Device Fitbone® transport and lengthening system, K232169) | Additional Reference Device (Fitbone TAA, K163368 and K203399) |
| --- | --- | --- | --- | --- |
| 1 | Nail Material | Stainless steel (1441-AISI 316LVM) | Stainless steel (1441-AISI 316LVM) | Stainless steel (1441-AISI 316LVM) |
| | | Assessment: The material is the same as the primary and reference predicates. Equivalent - no different questions of safety and effectiveness raised. | | |
| 2 | | H_{2}O_{2} Plasma | H_{2}O_{2} Plasma | H_{2}O_{2} Plasma |
Via delle Nazioni 9 | 37012 Bussolengo | Verona, Italy
+39.045.6719000
orthofix.com
ORTHOFIX® COORDINATING OTHER MEDICAL PROCEDURE
SEDE LEGALE: VIA VITTORI PISANI 9 | 37014 MILANO UTAVIA F.C.F. e P.IVA DI INTERIORO (CIT): VIA NUMERO TERRINO 0116 | Registro Imprese di Milano 0017000010 | RELA: M15757169 | Cap. Soc. Eur. 1546.000 mt. vers.
SEDE SECONDARIA: VIA DELLE NAZIONI 9 | 37012 BLISSOLENGO DIERONALI ITALIA | Registro Imprese di Verona RELA: VR 215209
{6}
ORTHOFIX®
K260146-Page 3
| | Sterilization Method | Assessment: the subject’s device sterilization method (H₂O₂ plasma) is equivalent to the primary predicate and to additional predicate/reference device. Equivalent - no different questions of safety and effectiveness raised | | |
| --- | --- | --- | --- | --- |
| 3 | Nail Size Range | 290-490mm in length; 11 and 13mm diameters in varying configurations | 290-490mm in length; 11 and 13mm diameters in varying configurations | Not applicable |
| | | Assessment: Size ranges of subject devices are equivalent to the primary. Equivalent - no different questions of safety and effectiveness raised | | |
| 4 | Maximum Bone Transport Stroke Available | 80 mm | 80 mm | Not applicable |
| | | Assessment: The transport stroke length of the subject device is the same of the primary predicate. Equivalent - no different questions of safety and effectiveness raised | | |
| 5 | Method of Distraction/Energy Source | Internal motor electro-magnetically induced by an external transmitter with signal received through a receiver placed just under skin | Internal motor electro-magnetically induced by an external transmitter with signal received through a receiver placed just under skin | Internal motor electro-magnetically induced by an external transmitter with signal received through a receiver placed just under skin |
| | | Assessment: Subject distraction methodology is equivalent to the primary and reference devices. Equivalent - no different questions of safety and effectiveness raised | | |
| Performance Analysis | | Mechanical testing/rationale of the subject devices to demonstrate equivalence in function and safety for the intended use of the subject Fitbone Transport Pulling nails in respect to the existing primary predicate Fitbone® transport and lengthening nails (K232169) have been performed for the following functional attributes: • Implantable rod: Inverted Load test for pulling nails with subject Fitbone Transport Pulling nails, • Implantable rod: Static 4-point bending test performed according to ASTM F1264-16 - Standard Specification and Test Methods for Intramedullary Fixation Devices on the primary predicate Fitbone® transport and lengthening nails (K232169) is deemed applicable to the subject devices, • Implantable rod: Fatigue 4-point bending test performed according to ASTM F1264-16 - Standard Specification and Test Methods for Intramedullary Fixation Devices on the primary predicate Fitbone® transport and lengthening nails (K232169) is deemed applicable to the subject devices, • Implantable rod: Torsional test performed according to ASTM F1264-16 - Standard Specification and Test Methods for Intramedullary Fixation Devices on the primary predicate Fitbone® transport and lengthening nails (K232169) is deemed applicable to the subject devices, | | |
Via delle Nazioni 9 | 37012 Bussolengo | Verona, Italy
+39.045.6719000
orthofix.com
ORTHOFIX® COORDINAMENTO Orthofix Medical Inc. Caltamano
SEDE LEGALE: VIA VITTORI PISANI 9 | 37014 MILANO ITALIA | C.F. e PISA DIRITENZI 10 | CZE: VAT NUMER ITO2011600700 | Registro Imprese di Milano 0017000350 | R.E.A. MI 5717169 | Cap. Soc. Eur. 1546.000 int. vero
SEDE SECONDARIA: VIA DELLE NAZIONI 9 | 37051 BUGGELENGO, MERENIGI ITALIA | Registro Imprese di Verona | R.E.A. VR 215209
{7}
ORTHOFIX®
K260146-Page 4
| Conclusion | Based upon equivalences in: intended use, patient population, site of application, conditions of use, operating principles, and the non-clinical performance data, the new nail model of FITBONE® Transport and Lengthening System have been shown to be safe and effective and to perform equivalently as compared to the legally marketed predicate device. Therefore, the subject new nail model of FITBONE® Transport and Lengthening System is substantially equivalent to the legally marketed predicate devices and the reference device chosen as basis for the scientific methodology. |
| --- | --- |
Via delle Nazioni 9 | 37012 Bussolengo | Verona, Italy
+39.045.6719000
orthofix.com
ORTHOFIX® COORDINAMENTO Orthofix Medical Inc. Elefawano
SEDE LEGALE: VIA VITTORI PISANI 90 | 22124 MILANO UTAIJA | C.F. A PISA 28970080100 LOTE: VAT NUMBER (TOININV00756) Registro Imprese di Milano 00970080100 | R.E.A. MI 5257769 / Cap. Soc. Eur. 15465000 mt. verb
SEDE SECONDARIA: VIA DELLE NAZIONI 9 | 37012 BUSSOLENGO, VERONA | ITALIA | Registro Imprese di Verona | R.E.A. VR 215209
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.