K260142 · Motion Informatics ltd. · IPF · May 27, 2026 · Physical Medicine
Device Facts
Record ID
K260142
Device Name
StimelMD (SSMD) system
Applicant
Motion Informatics ltd.
Product Code
IPF · Physical Medicine
Decision Date
May 27, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
1. Relaxation of muscle spasms 2. Prevention of the retardation of disuse atrophy 3. Increasing local blood circulation 4. Stroke Rehabilitation by muscle re- education 5. Maintaining or increasing range of motion. 6. Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
Device Story
Portable, battery-powered powered muscle stimulator (SSMD) with biofeedback; used for stroke rehabilitation and muscle therapy. Inputs: EMG signals captured via surface electrodes. Transformation: Stimel unit processes EMG signals to deliver electrical stimulation to muscles; managed via companion web application (desktop/mobile) for protocol configuration, monitoring, and reporting. Output: Electrical stimulation pulses; AR animations via Meta Quest 3 headset for patient interface. Used in clinical or home settings under healthcare provider supervision. Provider uses app to review progress and adjust treatment. Benefits: Muscle re-education, atrophy prevention, and improved circulation.
Clinical Evidence
Bench testing only. Software verification and validation performed per FDA guidance. Electrical safety and EMC testing conducted per IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11, and IEC 60601-2-10. All tests passed.
Technological Characteristics
Powered muscle stimulator; 12V Ni-MH rechargeable battery; Type BF applied parts. Connectivity: Bluetooth (Stimel unit to app), Wi-Fi (app to cloud). Electrodes: 2"x2" RE-PLY TENS 654. Output: Single channel, synchronous or alternating modes. Software: Web-based application for configuration and reporting. Sterilization: Not specified.
Indications for Use
Indicated for patients requiring muscle spasm relaxation, prevention of disuse atrophy, increased local blood circulation, stroke rehabilitation via muscle re-education, maintenance/increase of range of motion, and post-surgical calf muscle stimulation to prevent venous thrombosis. Use is restricted to prescription under medical supervision.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 27, 2026
Motion Informatics ltd.
% Orly Maor
Company Regulatory Consultant
RA Regulation
25a Sirkin St.
Kfar Saba, 4442157
Israel
Re: K260142
Trade/Device Name: StimeIMD (SSMD) system
Regulation Number: 21 CFR 890.5850
Regulation Name: Powered Muscle Stimulator
Regulatory Class: Class II
Product Code: IPF
Dated: April 26, 2026
Received: April 27, 2026
Dear Orly Maor:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260142 - Orly Maor
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K260142 - Orly Maor
Page 3
Sincerely,
Tushar Bansal -S
Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
StimelMD (SSMD) system
Page 7 of 38
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260142 | ? |
| Please provide the device trade name(s). | | ? |
| StimelMD (SSMD) system | | |
| Please provide your Indications for Use below. | | ? |
| 1. Relaxation of muscle spasms 2. Prevention of the retardation of disuse atrophy 3. Increasing local blood circulation 4. Stroke Rehabilitation by muscle re- education 5. Maintaining or increasing range of motion. 6. Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) Summary
StimelMD (SSMD) system
K260142
Date Prepared: January 6, 2026
I. SUBMITTER
Motion Informatics Ltd.
Yarka Industrial Zone, 2496700
Tel: +972 4 849 1349
Fax: +972 4 849 2384
Contact Person
Orly Maor
25 Sirkin Street
Kfar Saba 4442157 Israel
Tel: +972-52-8901368
oram.ma@gmail.com
II. DEVICE
Name of Device: StimelMD (SSMD) system
Common or Usual Name: SSMD
Classification Name: Powered muscle stimulator
Regulatory Class: Class II, per 21 CFR 890.5850
Product Code: IPF
Classification Panel: Physical Medicine
III. PREDICATE DEVICE
The predicate device is the Spatial StimelMD (SSMD) (former STIMEL-03), cleared under K130424 (product code IPF Regulation No. 21 CFR 890.5850).
IV. DEVICE DESCRIPTION
The SSMD is a Powered Muscle Stimulator with biofeedback. The device is portable, battery powered muscular electrical stimulator with biofeedback used as a training system for rehabilitation of paralyzed muscles, mainly after stroke.
The same controls are implemented in the current modified device. The treatment parameters are managed by the healthcare provider/user via a companion (therapist) application.
The App is used by the healthcare provider to configure treatment protocols, monitor session performance, review patient progress, and generate analytical reports. It is a web application that is operated using a desktop or mobile platforms and interfaces with the Stimel Unit via Bluetooth and with the cloud backend via Wi-Fi communication.
Device Parts and accessories:
1. Stimel Unit – the main device box that produces the treatment.
Page 1 of 5
{5}
2. Disposable Electrodes – surface bi-directional electrodes to capture the EMG signal from the patient and transfer the electrical stimulation to the patient.
3. Patient cable – connecting the Stimel unit to the Disposable Electrodes.
4. Charger – AC/DC Battery Charger (110/220 VAC; 12V DC)
5. Meta Quest 3 AR headset – of-the-shelf part, provides AR animations and serves as patient’s interface during treatment.
## V. INDICATIONS FOR USE
1. Relaxation of muscle spasms
2. Prevention or retardation of disuse atrophy
3. Increasing local blood circulation
4. Stroke Rehabilitation by muscle re-education
5. Maintaining or increasing range of motion.
6. Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| | Motion Informatics Ltd. Spatian StimelMD (SSMD) Subject device | Motion Informatics Ltd. Spatian StimelMD (SSMD) (STIMEL-03) | SE determination |
| --- | --- | --- | --- |
| K number | K260142 | K130424 | |
| Regulation | 890.5850 | 890.5850 | Same |
| Product Code | IPF | IPF | Same |
| Indications for Use | 1. Relaxation of muscle spasms 2. Prevention of the retardation of disuse atrophy 3. Increasing local blood circulation 4. Stroke Rehabilitation by muscle re-education 5. Maintaining or increasing range of motion. 6. Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis Powered muscle stimulators should only be used under medical supervision for | 1. Relaxation of muscle spasms 2. Prevention of the retardation of disuse atrophy 3. Increasing local blood circulation 4. Stroke Rehabilitation by muscle re-education 5. Maintaining or increasing range of motion. 6. Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis Powered muscle stimulators should only be | Same |
| | Motion Informatics Ltd. Spatian StimelMD (SSMD) Subject device | Motion Informatics Ltd. Spatian StimelMD (SSMD) (STIMEL-03) | SE determination |
Page 2 of 5
{6}
| | adjunctive therapy for the treatment of medical diseases and conditions | used under medical supervision for adjunctive therapy for the treatment of medical diseases and conditions | |
| --- | --- | --- | --- |
| Power Source during Treatment | Battery Type 12V (Rechargeable Ni-MH battery 12V, 2200mAh) | Battery Type 12V (Rechargeable Li-ion battery, INP18650-4S1P, 2600mAh) | Similar No influence on safety |
| | Method of Line Current Isolation- Not physically possible to connect the charge adapter and the electrodes at the same time | Method of Line Current Isolation- Not physically possible to connect the charge adapter and the electrodes at the same time | Same |
| | Patient Leakage Current: Not physically cable is completely isolated | Patient Leakage Current: Not physically cable is completely isolated | Same |
| | Normal condition 0 μA | Normal condition 0 μA | Same |
| | Single fault Condition 0 μA | Single fault Condition 0 μA | Same |
| Power Source during Charging | Input Voltage100-240VAC | Input Voltage100-240VAC | Same |
| | Frequency 50-60 Hz | Frequency 50-60 Hz | Same |
| | Input Current330 mA (in battery charge mode from charger) | Input Current330 mA (in battery charge mode from charger) | Same |
| | Classification -Internally powered, continuous operation, type BF applied parts | Classification -Internally powered, continuous operation, type BF applied parts | Same |
| Number of Output Modes | 1 Mode Pulse Waveform Intensity Maximum Voltage Positive/Negative Pulse Duration Total Pulse Duration Min/Max Load Pulse Repetition Rate | 1 Mode Pulse Waveform Intensity Maximum Voltage Positive/Negative Pulse Duration Total Pulse Duration Min/Max Load Pulse Repetition Rate | Same |
| Number of Output Channels | One output channel that can be operated either Synchronous or | One output channel that can be operated either Synchronous or | Same |
| | Motion Informatics Ltd. Spatian StimelMD (SSMD) Subject device | Motion Informatics Ltd. Spatian StimelMD (SSMD) (STIMEL-03) | SE determination |
| | Alternating. | Alternating. | |
Page 3 of 5
{7}
Page 4 of 5
| | Method of Channel Isolation Channel Isolated by Transformer. | Method of Channel Isolation Channel Isolated by Transformer. | Same |
| --- | --- | --- | --- |
| Indicator Display | - On/Off Status - Low Battery - Voltage/Current Level | - On/Off Status - Low Battery - Voltage/Current Level | Same |
| Electrode type | 2"X2" RE-PLY TENS 654 | 2"X2" RE-PLY TENS 654 | Same |
| Additional features | - | Headset (off the shelf) | Different however not a safety issue as the treatment parameters were not changed |
| | - | Software App The App is a replica of the physical buttons that were on the previous cleared device | Different however not a safety issue as the treatment parameters were not changed |
## PERFORMANCE DATA
The following performance data were conducted. The SSMD System met the predetermined acceptance criteria ensuring substantial equivalence to the predicate. No new safety or effectiveness issues were raised during testing:
## Software Validation
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA’s Guidance for Industry and FDA Staff, “Content of Premarket Submissions for Device Software Functions”. Cybersecurity was also evaluated.
The tests passed.
In addition, cybersecurity information was provided.
## Electrical Safety and EMC
Electrical Safety per IEC 60601-1 and Electromagnetic compatibility (EMC) per IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11 and IEC 60601-2-10 were conducted on the SSMD System. The tests passed.
## VII. CONCLUSIONS
Motion Informatics Ltd. has demonstrated that the modified SSMD System is substantially equivalent in fundamental design, technology, function, device materials, packaging, operating
{8}
principal, and intended use/ indication for use to the predicate device, the cleared STIMEL-03 System.
The minor differences do not raise any new questions of safety or effectiveness. Performance data demonstrates that the modified SSMD is as safe and effective as the cleared STIMEL-03.
Thus, the SSMD System is substantially equivalent to its predicate device.
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.