CustoMED Viewer is intended to be used as a software interface to assist in the visualization and communication of treatment options. The device displays three-dimensional anatomical models derived from CT imaging data for visualization and review by qualified healthcare professionals, and perform basic measurements on that data.
Device Story
CustoMED Viewer is a web-based software interface for surgical planning communication; used by surgeons and service providers. Input: DICOM files (CT) and 3D anatomical models (STL) pre-generated by third-party software. Operation: displays CT data and 3D overlays; enables user-driven annotations (drag-and-drop), landmark placement, and basic distance/angle measurements; allows surgeons to approve/decline design proposals. Output: visual feedback and annotated surgical plans. No autonomous algorithmic or AI/ML processing; does not modify source data. Benefit: facilitates effective communication between surgeons and designers to ensure clinical alignment during surgical planning.
Clinical Evidence
No clinical data. Performance supported by software verification and validation testing, including verification of linear (distance) and angular measurement accuracy (±1 mm or ±1°).
Technological Characteristics
Web-based software; DICOM viewer; 3D STL model visualization; manual annotation/measurement tools (distance, angle, landmark); no AI/ML; non-modifying of source data.
Indications for Use
Indicated for use by medical professionals and service providers as a software interface to assist in the visualization and communication of treatment options, including review of 3D anatomical models derived from CT imaging and performance of basic measurements.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 3, 2026
CustoMED, Ltd.
% Esther Kott
Director of Regulatory Affairs
LDP Consulting
44 Hey Beiyar St.
Tel Aviv, 6215017
Israel
Re: K260116
Trade/Device Name: CustoMED Viewer
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: LLZ
Dated: [NOTE: Use date of most recent supplement]
Received: April 30, 2026
Dear Esther Kott:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260116 - Esther Kott
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260116 - Esther Kott
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb, Ph.D.
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging Devices and
Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260116 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| CustoMED Viewer | | |
| Please provide your Indications for Use below. | | ? |
| CustoMED Viewer is intended to be used as a software interface to assist in visualization and communication of treatment options. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
# K260116 510(k) SUMMARY
# CustoMED Viewer
Company's Name and Address
CustoMED Ltd.
Derech Shiba 64
Ramat Gan
5266200
Israel
Contact Person
Esther Kott
Director of Regulatory Affairs
LDP Consulting
esther@ldp-consulting.com
+972-58-488-7454
Date Prepared
June 1, 2026
Device Tradename
CustoMED Viewer
Classification Name
System, Image processing, Radiological
Regulation Name
Medical Image Management and Processing System
Product Code
LLZ
Classification
21 CFR 892.2050
Predicate Device
SurgiCase Viewer (K170419)
Device Description
CustoMED Viewer was developed to allow medical professionals (surgeons) and service providers (designers or engineers assisting in planning the surgery) to communicate more effectively during procedures planning. It enables its end users, the medical professionals (e.g., ordering surgeons), to review and provide feedback on design proposals or intermediate outputs, ensuring alignment with clinical expectations.
The surgeon uploads a DICOM file to the CustoMED Viewer which is shared with the service provider. The service provider uses FDA-cleared third-party software for surgical planning and creates 3D data to upload to the viewer in stl format. The CustoMED Viewer is used only for visualization and does not include any algorithmic or AI/ML functionality.
The CustoMED Viewer does not alter the 3D data it imports, and its functioning is independent of the specific medical indication/situation or product/service it is used for. It is the responsibility of the service provider using the CustoMED Viewer to comply with the applicable medical device regulations.
Indications for Use
CustoMED Viewer is intended to be used as a software interface to assist in visualization and communication of treatment options.
Intended Use
CustoMED Viewer is intended to be used as a software interface to assist in the visualization and communication of treatment options. The device displays three-dimensional anatomical models derived from CT imaging data for visualization and review by qualified healthcare professionals, and perform basic measurements on that data.
Technological Characteristics
A comprehensive assessment demonstrates the subject device is substantially equivalent in intended use and indications for use, technological characteristics, and principles of operation to the
1
{5}
predicate device, as shown in the Technological Characteristics Table below.
**Technological Characteristics Table:**
| Comparison Feature | | Subject Device CustoMED Viewer (K260116) | Predicate Device SurgiCase Viewer (K170419) | Comparison |
| --- | --- | --- | --- | --- |
| Device Class | | Class II | Class II | Identical |
| Classification Panel | | Radiology | Radiology | Identical |
| Product Code | | LLZ | LLZ | Identical |
| Regulation Description | | Picture archiving and communications system | Picture archiving and communications system | Identical |
| Regulation Number | | 892.2050 | 892.2050 | Identical |
| Indications for Use | | CustoMED Viewer is intended to be used as a software interface to assist in visualization and communication of treatment options. | SurgiCase Viewer is intended to be used as a software interface to assist in visualization and communication of treatment options. | Identical |
| Clinical Characteristics | | | | |
| Indications of Use | | General indication for use | General indication for use | Identical |
| Type of scans | | CT | CT / MRI | Similar |
| Application/Anatomy | | General indication for use | General indication for use | Identical |
| Primary Imaging Modalities | | Upload by user: CT / STL | Upload by the service provider: CT / STL (by integration) | Similar |
| DICOM/CT Viewer | | Yes | Yes | Identical |
| 3D Reconstruction/ Segmentation | | Uploaded by the service provider: STL | Uploaded by the service provider: STL (by integration) | Identical |
| Intended User(s) | | - Medical professionals - Service providers | - Medical professionals - Service providers | Identical |
| Technological features | | | | |
| Segmentation Review and Annotation | | - View (CT Viewer) - 3D reconstruction visualization - Segmentation slice overlay - Annotation (Drag and Drop) - Approve/Decline by clicking on a button | - View (CT Viewer) - 3D reconstruction visualization - Segmentation slice overlay - Annotation (Comment) - Approve/Decline by comment | Similar – very minor difference in approve/decline process. This difference does not raise different questions of safety and effectiveness. |
| Landmarks Review and Annotation | Landmarks Annotation | - Landmark plane - Landmark line (2+ points) - Landmark point | - Commenting (text) - Draw line (text) | Different – does not raise new questions of safety or effectiveness |
| | Measurements Annotation | - Distances value (mm) + Visualization (3d) - Angles value (°) + Visualization (3d) | - Distances value (mm) + Visualization (3d) - Angles value (°) + Visualization (3d) - Draw and measure area (mm2) | Similar |
| Planning Review | Implants Selection | Template selection (list) | Template selection by comments addressed to the service provider (either by text in the Viewer, or using external tools) | Different – does not raise new questions of safety or effectiveness |
2
{6}
| Comparison Feature | Subject Device CustoMED Viewer (K260116) | Predicate Device SurgiCase Viewer (K170419) | Comparison |
| --- | --- | --- | --- |
| Surgical Plan Annotation | - Drag and drop or rotating by a gizmo (manipulator) elements (translation) - 3D Landmarks - 3D Implants - 3D Cutting plan - 3D Drilling cylinder | - Commenting (text) - Draw line (text) - Draw area (text) | Different – does not raise new questions of safety or effectiveness |
| Device/PSI Design | - View (CT Viewer) - 3D anatomy visualization - 3D PSI visualization - Landmark Point Annotation - Annotation (Drag and Drop) - Selection (list/dropdown) | - View (CT Viewer) - 3D anatomy visualization - 3D PSI visualization - Draw line (text) - Annotation (Comment) | Different – does not raise new questions of safety or effectiveness |
#### Performance Data
No Clinical tests
CustoMED Viewer is a software medical device for which performance testing demonstrates substantial equivalence to the predicate device and confirms that it is as safe and effective as the predicate device.
Performance of the CustoMED Viewer is supported by software verification and validation testing, including verification of linear (distance) and angular measurements accuracy ($\pm 1$ mm or $\pm 1^\circ$, respectively) of the measurements performed by the subject device.
#### Substantial Equivalence
The CustoMED Viewer is as safe and effective as its predicate device, the Materialise N.V. SurgiCase Viewer (K170419). Both devices have similar intended use and indications for use, technological characteristics, and principles of operation, as shown in the Technological Characteristics Table above.
Both CustoMED Viewer and SurgiCase Viewer share the same core technological characteristics: they are web-based Software that visualize pre-generated CT-based 3D anatomical data (.STL), support DICOM viewing with 3D overlays, allow user-driven measurements and annotations, and do not modify source data or generate treatment plans autonomously. Some minor technological differences exist between CustoMED Viewer and the predicate device, for example the predicate relies mainly on text-based annotations and comments attached to specific locations whereas the CustoMED Viewer allows clinicians to communicate changes by drag-and-drop annotation, SurgiCase supports MRI data and area measurements (mm²), which CustoMED does not etc. These minor technological differences between the CustoMED Viewer and its predicate device raise no new issues of safety or effectiveness.
#### Conclusion
Evaluation of intended use and indications for use, technological characteristics, and principles of operation, and performance testing, demonstrates that CustoMED Viewer is as safe and effective as its substantially equivalent predicate device.
3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.