The OsteoCoil™ GT Active Compression System is intended for use in bone fragment stabilization and boney fusion.
Device Story
Cannulated, Nitinol-based dynamic compression screw system; used for bone fragment stabilization and fusion in small bones (wrist, hand, foot). Device consists of a cannulated rod with a distal threaded region and a proximal head; features a helical cut/dynamic compression element allowing axial elongation to maintain compression during bone resorption. Includes a torque-bearing mechanism at the distal interface of the sleeve and internal rod to isolate the compression element from torsional loads during insertion. Operated by healthcare professionals in hospital settings; inserted over a guidewire. Provides continuous compression across bony fusion sites despite bone remodeling; benefits patient by maintaining stability during healing. Implants provided sterile (single-use); instruments provided non-sterile for reprocessing.
Clinical Evidence
No clinical data included. Bench testing and Finite Element Analysis were used to confirm performance. Biocompatibility, sterility, and shelf-life processes are identical to the predicate device (K233567).
Technological Characteristics
Material: Nitinol (ASTM). Principle: Dynamic compression via helical cut shaft with integrated torque-bearing mechanism. Form factor: Cannulated screw. Energy: None (mechanical). Sterilization: Gamma irradiation (SAL 10^-6). Shelf-life: 7 years. Connectivity: None.
Indications for Use
Indicated for bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation of small bones in the wrist, hand, and foot.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}
**FDA U.S. FOOD & DRUG**
ADMINISTRATION
August 24, 2026
Flower Orthopedics Corporation DBA Conventus Flower Orthopedics
% Jessica Czamanski
Regulatory, Quality, and Compliance Consultant
DuVal & Associates, P.A.
825 Nicollet Mall, Suite 1820
Medical Arts Bldg.
Minneapolis, Minnesota 55402
Re: K260106
Trade/Device Name: OsteoCoil™ GT Active Compression System
Regulation Number: 21 CFR 888.3040
Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener
Regulatory Class: Class II
Product Code: HWC
Dated: January 12, 2026
Received: January 14, 2026
Dear Jessica Czamanski:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260106 - Jessica Czamanski
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260106 - Jessica Czamanski
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**Shumaya Ali -S**
Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260106 | ? |
| Please provide the device trade name(s). | | ? |
| OsteoCoil™ GT Active Compression System | | |
| Please provide your Indications for Use below. | | ? |
| The OsteoCoil™ GT Active Compression System is indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, & fracture fixation of the small bones for wrist, hand, and foot. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
conventus
FLOWER
ORTHOPEDICS
Traditional 510(k)
OsteoCoil™ GT Active Compression System
K260106
# K260106 510(K) SUMMARY
# I. Submitter
Submitter's Name: Flower Orthopedics Corporation DBA Conventus Flower Orthopedics
Contact Person: Kevin Rzasa
Address: 100 Witmer Road
Suite 280,
Horsham, PA, 19044
Telephone: 610-202-1342
Email: kevin.rzasa@flowerortho.com
Date prepared: August 19, 2026
# II. Applicant Correspondent
Contact's Name: DuVal & Associates, P.A.
Contact Person: Jessica Czamanski
Telephone: 754-422-9101
Email: jczamanski@duvalfdalaw.com
# III. Device
Trade Name: OsteoCoil™ GT Active Compression System
Common Name: Screw, Bone, Fixation
Classification Name: Smooth or threaded metallic bone fixation fastener
Product Classification: Class II
Regulation Number: 21 CFR § 888.3040
Product Code: HWC
# IV. Predicate Device
Manufacturer: Conventus Flower Orthopedics
Conventus Flower Orthopedics
Page 1
{5}
conventus
FLOWER
ORTHOPEDICS
Traditional 510(k)
OsteoCoil™ GT Active Compression System
K260106
Device Name: OsteoCoil™ Nitinol Compression System
510(k) Number: K233567
Product Classification: Class II
Regulation Number: 21 CFR § 888.3040
# V. Device Description
The proposed device is comprised of dynamic compression screw and required accessories, intended to provide adequate initial compression and adequate compression over an extended period of time when used in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation of small bones. For example, if there is bone resorption, then the product will axially contract for continued compression.
The OsteoCoil GT implants are comprised of an implantable, cannulated rod with a threaded region at the distal end. The unthreaded portion of the rod contains a helical cut or dynamic compression feature which allows axial elongation. A thin-walled tube with a head at the proximal end is seated around the cannulated rod and attached at the head end. There is a torque bearing mechanism at the distal interface of the sleeve and internal rod, which allows torque to be transmitted without loading the compression element. The cannulation allows for insertion over a guidewire. The head engages with a proximal end of a first bone segment, and the threaded region engages with a second bone segment. The head also includes a drive feature for insertion. The system is comprised of sterile implants for single use only. Instruments are provided non-sterile and may be reprocessed.
# VI. Intended Use
The OsteoCoil™ GT Active Compression System is intended for use in bone fragment stabilization and boney fusion.
Conventus Flower Orthopedics
Page 2
{6}
conventus
FLOWER
ORTHOPEDICS
Traditional 510(k)
OsteoCoil™ GT Active Compression System
K260106
# VII. Indications for Use
The OsteoCoil™ GT Active Compression System is indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, & fracture fixation of small bones for wrist, hand, and foot.
# VIII. Comparison of Technological Characteristics with Predicate Device
The Subject and Predicate devices in this submission are almost identical, with differences limited to minor design characteristics.
Table 1: General Technological Characteristics Comparison
| Product Features | Subject Device Conventus' OsteoCoil™ GT Active Compression System (K260106) | Predicate Device Conventus' OsteoCoil™ Nitinol Compression System (K233567) |
| --- | --- | --- |
| Classification | Class II | SAME |
| Product Code | HWC | SAME |
| Regulation Number | §888.3040 | SAME |
| Regulation Name | Smooth or threaded metallic bone fixation fastener | SAME |
| Intended Use | The OsteoCoil™ GT Active Compression System is intended for use in bone fragment stabilization and boney fusion. | SAME |
| Indications for Use | The OsteoCoil™ GT Active Compression System is indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, & fracture | SAME |
Conventus Flower Orthopedics
Page 3
{7}
conventus
FLOWER
ORTHOPEDICS
Traditional 510(k)
OsteoCoil™ GT Active Compression System
K260106
Table 1: General Technological Characteristics Comparison
| Product Features | Subject Device Conventus' OsteoCoil™ GT Active Compression System (K260106) | Predicate Device Conventus' OsteoCoil™ Nitinol Compression System (K233567) |
| --- | --- | --- |
| | fixation for small bones for wrist, hand, and foot. | |
| **Materials** | Nitinol ASTM | SAME |
| **Environment of Use** | Hospital | SAME |
| **Provided Sterile** | Implants – Yes Instruments – No | SAME |
| **Principle of Operation** | The OsteoCoil™ GT Active Compression System supports the fracture site in the same way traditional bone screws do while taking advantage of the patient anatomy and bone healing occurring at the injury site. The proprietary compressive technology of the implant causes an opposing compression force between the implant head and threads. Typically, a bone screw provides an initial compression force which is lost as soon as there is bone length loss due to bone remodeling. The unique OsteoCoil implant results in a compression force applied across a bony fusion site even with loss of length throughout | The principle of operation is the same, except that the OsteoCoil Nitinol Compression System does not have the torque bearing mechanism in the Subject Device. Therefore, the transmission of torque is less efficient. |
Conventus Flower Orthopedics
Page 4
{8}
[LOGO]
Conventus
FLOWER
ORTHOPEDICS
Traditional 510(k)
OsteoCoil™ GT Active Compression System
K260106
Table 1: General Technological Characteristics Comparison
| Product Features | Subject Device Conventus' OsteoCoil™ GT Active Compression System (K260106) | Predicate Device Conventus' OsteoCoil™ Nitinol Compression System (K233567) |
| --- | --- | --- |
| | bone resorption. The OsteoCoil GT system adds a torque bearing mechanism to the implant shaft, which is distal to the helical shaft and reduces the torsional load on the helical compression feature. | |
| User Profile | Healthcare Professional Only | SAME |
| Single Use | Implants – Yes Instruments – No | SAME |
| Sterilization | Gamma Irradiation | SAME |
| SAL | 10^{-6} | SAME |
| Shelf-Life | 7 years | SAME |
Conventus Flower Orthopedics
Page 5
{9}
conventus
FLOWER
ORTHOPEDICS
Traditional 510(k)
OsteoCoil™ GT Active Compression System
K260106
# IX. Performance Data
# Bench Testing
No additional bench performance testing is included to support this Traditional 510(k). The OsteoCoil™ GT Active Compression System and its Predicate Device are significantly similar in design, materials, and manufacturing, and Finite Element Analysis was performed to further confirm expected performance of the device.
# Reprocessing, Sterility, and Shelf-Life
No additional reprocessing, sterility, and shelf-life testing is required as processes are identical to those in K233567. Routine pyrogen testing will be conducted on representative samples from each product lot.
# Biocompatibility
No additional biocompatibility testing is required as manufacturing materials and processes are identical to those in K233567.
# Electrical Safety and Electromagnetic Compatibility (EMC)
There are no electrical components in the proposed device. All components are components of Nitinol and have not been evaluated for electromagnetic compatibility.
# Performance Testing – Animal
This submission does not include animal testing.
# Performance Testing – Clinical
This submission does not include clinical testing.
# X. Conclusion
Base on the data and justifications presented in this submission, The OsteoCoil™ GT Active Compression System is considered substantial equivalent to the OsteoCoil Nitinol Compression System.
Conventus Flower Orthopedics
Page 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.