PHT-70CHS is intended to produce panoramic, cephalometric, or 3D digital x-ray images. It provides diagnostic details of the dento-maxillofacial, sinus, and TMJ for adult and pediatric patients. The system also utilizes carpal images for orthodontic treatment. The device is to be operated by healthcare professionals.
Device Story
PHT-70CHS is a 4-in-1 digital X-ray system for 2D (panoramic, cephalometric) and 3D (cone beam CT) dental imaging. It captures X-ray signals via digital detectors (Xmaru1404CF-PLUS Eth, Xmaru2602CF Eth, or Jupi0606X1) and processes them into diagnostic images using FDK (back projection) and CS (iterative) reconstruction algorithms. The system includes metal artifact reduction (MAR) capabilities. Operated by healthcare professionals in dental clinical settings, the device outputs images to viewing software (EzDent-i, Ez3D-i). Clinicians use these images to visualize dento-maxillofacial, sinus, and TMJ anatomy, aiding in diagnosis and orthodontic treatment planning. The system benefits patients by providing high-quality diagnostic imaging with optimized radiation dose performance.
Clinical Evidence
Bench testing only. Performance evaluated per IEC 61223-3-4:2000 and IEC 61223-3-7:2021. Non-clinical testing compared MTF, DQE, NPS, and NED of new detectors against predicates. Dosimetric performance (DAP) confirmed equivalence under identical exposure conditions. Image quality verified via phantom testing.
Technological Characteristics
Digital X-ray system; 60-99 kV tube voltage; 4-16 mA tube current; 0.5 x 0.5 mm focal spot. Detectors: Xmaru1404CF-PLUS Eth, Xmaru2602CF Eth, Jupi0606X1. Connectivity: Ethernet. Software: DICOM 3.0 compatible. Standards: IEC 60601-1, IEC 60601-1-3, IEC 60601-2-63, IEC 60601-1-2, NEMA PS 3.1-3.18.
Indications for Use
Indicated for adult and pediatric patients requiring panoramic, cephalometric, or 3D digital x-ray imaging for diagnostic assessment of dento-maxillofacial, sinus, and TMJ structures, and carpal imaging for orthodontic treatment.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 31, 2026
VATECH Co., Ltd.
% Dave Kim
RA Consultant
Mtech Group, LLC
7505 Fannin St. Suite 610
HOUSTON, TX 77054
Re: K260093
Trade/Device Name: Green X 12 VE (PHT-70CHS); Green X VE (PHT-70CHS)
Regulation Number: 21 CFR 892.1750
Regulation Name: Computed Tomography X-Ray System
Regulatory Class: Class II
Product Code: OAS
Dated: March 3, 2026
Received: March 4, 2026
Dear Dave Kim:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260093 - Dave Kim
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K260093 - Dave Kim
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

for
Lu Jiang Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K260093 | |
| Device Name Green X 12 VE (Model: PHT-70CHS) Green X VE (Model: PHT-70CHS) | |
| Indications for Use (Describe) PHT-70CHS is intended to produce panoramic, cephalometric, or 3D digital x-ray images. It provides diagnostic details of the dento-maxillofacial, sinus, and TMJ for adult and pediatric patients. The system also utilizes carpal images for orthodontic treatment. The device is to be operated by healthcare professionals. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
VATECH Co., Ltd.
Premarket Notification 510(k)
# K260093 510(k) Summary
1. Special 510(k) Summary
This summary of 510(k) information is being submitted in accordance with requirements of 21 CFR Part 807.92.
2. Date 510K Summary prepared: Dec. 19, 2025
3. Administrative Information
Official Correspondent: Dave Kim / Mtech Group
Address: 7505 Fannin Street, Suite 610, Houston, TX 77054
Tel: +713-467-2607
Contact person: Mr. Dave Kim (davekim@mtechgroupllc.com)
510(k) Submitter: VATECH Co., Ltd.
Address: 13, Samsung 1-ro 2-gil, Hwaseong-si, Gyeonggi-do, 18449, Korea
Tel: +82-31-379-9492
Fax: +82-31-379-9400
Contact person: Daniel Kim / Manager (daniel.kim@vatech.co.kr)
4. Device Information
Type of 510(k) Submission: Special
Trade or Proprietary Name: Green X 12 VE (Model: PHT-70CHS)
Green X VE (Model: PHT-70CHS)
Common or Usual Name: System, X-ray, Computed tomography, Dental
Regulation Classification: Computed tomography x-ray system (21 CFR 892.1750)
Product Code: OAS
Class of Device: Class II
Panel: Radiology
5. Predicate Device Information
5-1. Predicate Device for Green X 12 VE
Manufacturer: VATECH Co., Ltd.
Trade or Proprietary Name: Green X 12 (Model: PHT-75CHS)
Common or Usual Name: System, X-ray, Computed tomography, Dental
Regulation Classification: Computed tomography x-ray system (21 CFR 892.1750)
Product Code: OAS
Class of Device: Class II
Panel: Radiology
510(k) Number: K231796
Page 1 / 9
{5}
VATECH Co., Ltd.
Premarket Notification 510(k)
## 5-2. Predicate Device for Green X VE
Manufacturer: VATECH Co., Ltd.
Trade or Proprietary Name: Green X 21 (Model : PHT-90CHO)
Common or Usual Name: System, X-ray, Computed tomography, Dental
Regulation Classification: Computed tomography x-ray system (21 CFR 892.1750)
Product Code: OAS
Class of Device: Class II
Panel: Radiology
510(k) Number: K243081
## 6. Device Description
The PHT-70CHS is a 4-in-1 digital X-ray system designed for both 2D and 3D dental radiographic imaging. The system integrates panoramic imaging, optional cephalometric imaging, dental computed tomography, and model imaging functions into a single unit. It is intended for dental diagnostic purposes and is capable of acquiring and processing multi-field-of-view digital radiographic images.
The PHT-70CHS is a complete digital radiographic imaging system that includes an X-ray generator, dedicated image receptors, and compatible image viewing software. The system supports acquisition of both 2D diagnostic images, including panoramic and cephalometric images, and 3D diagnostic images using cone beam computed tomography.
The materials, safety characteristics, X-ray source, indications for use, and image reconstruction including metal artifact reduction algorithms of the subject device are the same as those of the predicate devices PHT-75CHS (K231796) and PHT-90CHO (K243081).
Green X 12 VE and Green X VE are differentiated by the configuration of the CT and panoramic image receptors, which is reflected in their respective trade naming. Green X 12 VE is equipped with the Xmaru1404CF-PLUS Eth detector, while Green X VE is equipped with the Jupi0606X1 detector. Both configurations utilize the Xmaru2602CF Eth detector for cephalometric imaging.
## 7. Indication for use
PHT-70CHS is intended to produce panoramic, cephalometric, or 3D digital x-ray images. It provides diagnostic details of the dento-maxillofacial, sinus, and TMJ for adult and pediatric patients. The system also utilizes carpal images for orthodontic treatment. The device is to be operated by healthcare professionals.
Page 2 / 9
{6}
VATECH Co., Ltd.
Premarket Notification 510(k)
# 8. Substantial Equivalence Chart
8-1. Green X 12 VE - Substantial Equivalence Chart
| | | Subject Device | | Predicate Device | | Comparison |
| --- | --- | --- | --- | --- | --- | --- |
| Device Name | | Green X 12 VE (Model : PHT-70CHS) | | Green X 12 (Model : PHT-75CHS) | | - |
| Applicant Name | | VATECH Co., Ltd. | | VATECH Co., Ltd. | | - |
| 510(k) Number | | - | | K231796 | | - |
| Device Classification Name | | X-Ray, Tomography, Computed, Dental | | X-Ray, Tomography, Computed, Dental | | Same |
| Classification Product Code | | OAS | | OAS | | Same |
| Regulation Number | | 21 CFR 892.1750 | | 21 CFR 892.1750 | | Same |
| Indications for Use | | Green X 12 VE (Model: PHT-70CHS) is intended to produce panoramic, cephalometric, or 3D digital x-ray images. It provides diagnostic details of the dento-maxillofacial, sinus, and TMJ for adult and pediatric patients. The system also utilizes carpal images for orthodontic treatment. The device is to be operated by healthcare professionals. | | Green X 12 (Model: PHT-75CHS) is intended to produce panoramic, cephalometric or 3D digital x-ray images. It provides diagnostic details of the dento-maxillofacial, ENT, sinus and TMJ for adult and pediatric patients. The system also utilizes carpal images for orthodontic treatment The device is to be operated by healthcare professionals. | | Similar |
| Performance Specification | | Panoramic, Cephalometric and computed tomography | | Panoramic, Cephalometric and computed tomography | | Same |
| Input Voltage | | AC 100 - 240 V | | AC 100 - 240 V | | Same |
| X-Ray source | | D-052SB | | D-052SB | | Same |
| Tube Voltage | | 60 - 99 kV | | 60 - 99 kV | | Same |
| Tube Current | | 4 - 16 mA | | 4 - 16 mA | | Same |
| Focal Spot Size | | 0.5 x 0.5 mm | | 0.5 x 0.5 mm | | Same |
| Exposure Time | | Max. 16.9 s | | Max. 16.9 s | | Same |
| Slice Width | | Min. 0.1 mm | | Min. 0.1 mm | | Same |
| Total Filtration | | Min. 2.5 mm Al | | Min. 2.5 mm Al | | Same |
| Mechanical | | Compact design | | Compact design | | Same |
| Electrical | | LDCP logic circuit | | LDCP logic circuit | | Same |
| Software | | DICOM 3.0 Format compatible | | DICOM 3.0 Format compatible | | Same |
| 2D Image Viewing Program | | EzDent-i (K241114) | | EzDent-i (K223820) | | Similar |
| 3D Image Viewing Program | | Ez3D-i (K231757) | | Ez3D-i (K222069) | | Similar |
| Anatomical Sites | | Maxillofacial | | Maxillofacial | | Same |
| Image Receptor | CT&PANO | Xmaru1404CF-PLUS Eth | | Xmaru1404CF-PLUS | | Comparable 1) |
| | CEPH | Xmaru2602CF Eth | | Xmaru2602CF | | Comparable 2) |
| Size of Imaging Volume | | Xmaru1404CF-PLUS Eth | Max. 120 x 85 mm | Xmaru1404CF-PLUS | Max. 120 x 85 mm | Same |
| Pixel Size | CT&PANO | 99 μm -2X2 binning (system spec)
198 μm - 4X4 binning (system spec) | | 99 μm -2X2 binning (system spec)
198 μm - 4X4 binning (system spec) | | Same |
| | CEPH | 100 μm- Non binning (detector spec)
200 μm -2X2 binning (system spec) | | 100 μm- Non binning (detector spec)
200 μm -2X2 binning (system spec) | | Same |
Page 3 / 9
{7}
VATECH Co., Ltd.
Premarket Notification 510(k)
# Difference Summary:
The Green X 12 VE (Model: PHT-70CHS) dental computed tomography X-ray system described in this 510(k) is identical to the predicate device, Green X 12 (Model: PHT-75CHS), with respect to intended use, materials, safety characteristics, X-ray source, image reconstruction, and MAR (Metal Artifact Reduction) algorithms.
The following two things have changed:
1) The communication method of the X-ray detector has been changed.
Comparable 1)
The Xmaru1404CF-PLUS Eth maintains the same sensor structure, imaging performance, and clinical functionality as the existing Xmaru1404CF-PLUS, with the communication method between the detector and the system changed from optical communication to Ethernet communication.
Comparable 2)
The Xmaru2602CF Eth maintains the same sensor structure, imaging performance, and clinical functionality as the existing Xmaru2602CF, with the communication method between the detector and the system changed from optical communication to Ethernet communication.
The above Xmaru1404CF-PLUS Eth and Xmaru2602CF Eth detectors have already been cleared under 510(k) K243088, and this change does not affect the safety or performance of the device.
2) The latest version of the 2D/3D image viewing software (3rd party software) has been applied.
- Predicate Device: EzDent-i (K223820), version 3.4 / Ez3D-i (K222069), version 5.4
- Subject Device: EzDent-i (K241114), version 3.5 / Ez3D-i (K231757), version 5.5
The 3rd party software does not directly affect the safety or performance of the Subject Device. In addition, the scope of functions available for use with the Subject Device remains within the FDA 510(k)-cleared indications and functionality of EzDent-i and Ez3D-i, respectively.
Page 4 / 9
{8}
VATECH Co., Ltd.
Premarket Notification 510(k)
8-2. Green X VE - Substantial Equivalence Chart
| | | Subject Device | | Predicate Device | | Comparison |
| --- | --- | --- | --- | --- | --- | --- |
| Device Name | | Green X VE (Model : PHT-70CHS) | | Green X 21 (Model : PHT-90CHO) | | - |
| Applicant Name | | VATECH Co., Ltd. | | VATECH Co., Ltd. | | - |
| 510(k) Number | | - | | K243081 | | - |
| Device Classification Name | | X-Ray, Tomography, Computed, Dental | | X-Ray, Tomography, Computed, Dental | | Same |
| Classification Product Code | | OAS | | OAS | | Same |
| Regulation Number | | 21 CFR 892.1750 | | 21 CFR 892.1750 | | Same |
| Indications for Use | | Green X VE (Model PHT-70CHS) is intended to produce panoramic, cephalometric, or 3D digital x-ray images. It provides diagnostic details of the dento-maxillofacial, sinus, and TMJ for adult and pediatric patients. The system also utilizes carpal images for orthodontic treatment. The device is to be operated by healthcare professionals. | | Green X 21 (Model: PHT-90CHO) is intended to produce panoramic, cephalometric, or 3D digital X-ray images. It provides diagnostic details of the dento-maxillofacial, sinus, TMJ, and ENT for adult and pediatric patients. The system also utilizes carpal images for orthodontic treatment. The device is to be operated by healthcare professionals. | | Similar |
| Performance Specification | | Panoramic, Cephalometric and computed tomography | | Panoramic, Cephalometric and computed tomography | | Same |
| Input Voltage | | AC 100 - 240 V | | AC 100 - 240 V | | Same |
| X-Ray source | | D-052SB | | D-052SB | | Same |
| Tube Voltage | | 60 - 99 kV | | 60 - 99 kV | | Same |
| Tube Current | | 4 - 16 mA | | 4 - 16 mA | | Same |
| Focal Spot Size | | 0.5 x 0.5 mm | | 0.5 x 0.5 mm | | Same |
| Exposure Time | | Max. 18.0 s | | Max. 18.0 s | | Same |
| Slice Width | | Min. 0.1 mm | | Min. 0.1 mm | | Same |
| Total Filtration | | Min. 2.5 mm Al | | Min. 2.5 mm Al | | Same |
| Mechanical | | Compact design | | Compact design | | Same |
| Electrical | | LDCP logic circuit | | LDCP logic circuit | | Same |
| Software | | DICOM 3.0 Format compatible | | DICOM 3.0 Format compatible | | Same |
| 2D Image Viewing Program | | EzDent-i (K241114) | | EzDent-i (K241114) | | Same |
| 3D Image Viewing Program | | Ez3D-i (K231757) | | Ez3D-i (K231757) | | Same |
| Anatomical Sites | | Maxillofacial | | Maxillofacial | | Same |
| Image Receptor | CT&PANO | Jupi0606X1 | | Jupi1012X | | Comparable 3) |
| | CEPH | Xmaru2602CF Eth | | Venu1012VD | | Comparable 4) |
| Size of Imaging Volume | | Jupi0606X1 | Max. 180 x 100 mm | Jupi1012X | Max. 210 x 190 mm | Different - Comparable 3) |
| Pixel Size | CT&PANO | 100 μm (1x1 binning)
200 μm (2x2 binning) | | 100 μm (1x1 binning)
200 μm (2x2 binning) | | Same |
| | CEPH | 200 μm (2x2 binning) | | 125 μm | | Different - Comparable 4) |
Page 5 / 9
{9}
VATECH Co., Ltd.
Premarket Notification 510(k)
## Difference Summary:
The Green X VE (Model: PHT-70CHS) dental CT X-ray system described in this 510(k) is identical to the predicate device, Green X 21 (Model: PHT-90CHO), with respect to intended use, materials, safety characteristics, X-ray source, image reconstruction, and MAR (Metal Artifact Reduction) algorithms.
However, differences exist in the image receptor configuration, resulting in variations in the image acquisition method and maximum field of view (FOV). In addition, there are differences in the versions of the 2D/3D image viewing software. These differences are described below.
## Comparable 3)
There is a difference in the maximum available field of view (FOV) due to differences in the CT/PANO image detectors. The subject device is equipped with the Jupi0606X1 detector, providing a maximum FOV of 180 × 100 mm, whereas the predicate device is equipped with the Jupi1012X detector, providing a maximum FOV of 210 × 190 mm.
This difference is attributable to the physical size of the detectors and does not affect the fundamental imaging acquisition principles or the intended diagnostic purpose of the system.
## Comparable 4)
As specifications of Green X VE CEPH mode is identical with that of Green X 12 VE, it is compared with Green X 12. Xmaru2602CF Eth maintains the same sensor structure, imaging performance, and clinical functionality as the existing Xmaru2602CF, with the communication method between the detector and the system changed from optical communication to Ethernet communication.
These differences are attributable to variations in detector configuration and physical specifications and do not adversely affect the safety or performance of the subject device. Also, Xmaru2602CF Eth detector has already been cleared under 510(k) K243088. Accordingly, the fundamental technological characteristics of the subject device are substantially equivalent to those of the predicate device.
Additionally, the latest version of the 2D/3D image viewing software (3rd party software) has been applied.
- Predicate Device: EzDent-i (K223820), version 3.4 / Ez3D-i (K222069), version 5.4
- Subject Device: EzDent-i (K241114), version 3.5 / Ez3D-i (K231757), version 5.5
The 3rd party software does not directly affect the safety or performance of the Subject Device. In addition, the scope of functions available for use with the Subject Device remains within the FDA 510(k)-cleared indications and functionality of EzDent-i and Ez3D-i, respectively.
Page 6 / 9
{10}
VATECH Co., Ltd.
Premarket Notification 510(k)
## 9. Performance Data
### - Summary of Performance Testing
The following information further substantiates the substantial equivalence between the subject device and the predicate device: The fundamental technological characteristics of the subject and predicate device are identical. The imaging modes are also equivalent; PANO, CEPH (Optional), CBCT and MODEL. All viewing software programs have been cleared with previous 510k submissions; EzDent-i (K241114) and Ez3D-i (K231757). The sponsor tested the subject device in a laboratory and provided a non-clinical performance report. The same test protocol was used to test the performance of the subject and the predicate device for comparison. The sponsor certifies that adequate design and development controls (according to 21 CFR 820.30) were in place for manufacturing the subject device.
For the subject device and predicate device, the differences are as follows.
1) The subject device is equipped with the Xmaru1404CF-PLUS Eth detector (Green X 12 VE) for CT/PANO mode and Xmaru2602CF Eth (Green X 12 VE, Green X VE) detector for CEPH mode whereas predicate device (Green X 12) is equipped with Xmaru1404CF-Plus and Xmaru2602CF detector. 'Eth' means a model with an Ethernet communication port.
2) The subject device (Green X VE) is equipped with the Jupi0606X1 detector for CT/PANO mode, providing a maximum field of view (FOV) of 180 mm × 100 mm, whereas the predicate device (Green X 21) is equipped with the Jupi1012X detector, providing a larger maximum FOV of 210 mm × 190 mm.
### New X-ray detectors
The subject device is equipped with New X-ray detectors as follows.
Green X 12 VE: Xmaru1404CF-PLUS Eth for CT/PANO mode and Xmaru2602CF Eth for CEPH mode
Green X VE: Jupi0606X1 for CT/PANO mode and Xmaru2602CF Eth for CEPH mode
The subject device was tested and evaluated in a laboratory using the same test protocol as the predicate device to compare non-clinical and clinical performance. Non-clinical tests were conducted for the subject device's new X-ray detectors, comparing their performance with the predicate device in terms of Modulation Transfer Function (MTF), Detective Quantum Efficiency (DQE), Noise Equivalent Dose (NED), and Noise to Power Spectrum (NPS).
For CT/PANO/CEPH detector of Green X 12 VE and CEPH detector of Green X VE, according to the results of non-clinical tests, the pixel size of the new detector (Xmaru1404CF-PLUS Eth / Xmaru2602CF Eth) was found to be equal to that of the predicate device (Xmaru1404CF-PLUS / Xmaru2602CF). Technically, there are no differences between the detectors besides communication methods. Consequently, the image test patterns of Xmaru1404CF-PLUS Eth clearly demonstrate test objects across the same spatial frequency range as Xmaru1404CF-PLUS without any aliasing artifacts. Both Xmaru1404CF-PLUS Eth and Xmaru2602CF Eth showed equivalent performance compared to Xmaru1404CF-PLUS and Xmaru2602CF in terms of DQE, MTF, and NPS.
For CT/PANO detector of Green X VE, according to the results of non-clinical tests, the pixel size of the new detector (Jupi0606X1) was found to be very similar to that of the predicate device (Jupi1012X). Consequently, the image test patterns of the Jupi0606X1 clearly demonstrate test objects across the same spatial frequency range as the Jupi1012X without any aliasing artifacts. Jupi0606X1 showed similar or better performance compared to Jupi1012X in terms of DQE, MTF, and NED.
In conclusion, the diagnostic image quality of the new X-ray detectors was equivalent to or better than that of the predicate device, with no significant differences in efficiency and safety.
Page 7 / 9
{11}
VATECH Co., Ltd.
Premarket Notification 510(k)
The acceptance test was conducted in accordance with the requirements of 21 CFR Part 1020.30, 1020.31, 1020.33, IEC 61223-3-7:2021 (Evaluation and routine testing in medical imaging departments - Part 3-7: Acceptance and constancy tests - Imaging performance of X-ray equipment for dental cone beam computed tomography), and IEC 61223-3-4:2000 (Evaluation and routine testing in medical imaging departments - Part 3-4: Acceptance tests - Imaging performance of dental X-ray equipment).
The CT modality was evaluated through quantitative testing to verify and measure four key parameters representing CT image quality: noise, contrast, CNR, and MTF 10%. These evaluations were conducted using the provided reconstruction algorithms, FDK (back projection) and CS (iterative).
For PANO and CEPH modalities, quantitative assessments were performed with an emphasis on line pair resolution and low contrast performance using phantoms. The evaluation also included the software function additions, all of which met the requirements specified in IEC 61223-3-4:2000 and IEC 61223-3-7:2021.
In addition, the dosimetric performance of the subject device and the predicate device was compared in X-ray Dose Performance Comparison Evaluation Report, in terms of DAP.
For Green X 12 VE, With the identical FDD (Focal Spot to Detector Distance), exposure area, and exposure conditions, DAP measurement in the PANO, CEPH, and CBCT mode of each device was found to be equivalent. For Green X VE, PANO and CEPH mode share identical FDD (Focal Spot to Detector Distance), exposure area, and exposure conditions with its predicate devices (Green X 21 for PANO and Green X 12 for CEPH) and DAP measurement of each device was found to be equivalent. There are some differences in FOVs and exposure conditions for CBCT mode between the two devices but applying similar exposure areas and irradiation conditions (tube voltage, tube current), the subjective device showed similar DAP measurements compared to its predicate device.
Moreover, the clinical considerations and Image Quality Evaluation Report further demonstrated that the general image quality of the subject device is equivalent or better than the predicate device.
- Software Verification and Validation Testing
Software verification and validation were conducted and documented as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions." The software for this device was considered as a "Basic Documentation Level", since a failure or latent flaw in the software would not present a hazardous situation with a probable risk of death or serious injury to either a patient, user of the device, or others in the environment of use, prior to the implementation of risk control measures.
Green X 12 VE (Model: PHT-70CHS) and Green X VE (Model: PHT-70CHS) provides the following imaging viewer programs;
- 2D Image viewing program: EzDent-i (K241114)
- 3D Image viewing program: Ez3D-i (K231757)
- Safety, EMC and Performance Data
Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1:2005+AMD1:2012+AMD2:2020 (Edition 3.2), IEC 60601-1-3:2008+AMD1:2013+AMD2:2021 (Edition 2.2), IEC 60601-2-63:2012+AMD1:2017+AMD2:2021 (Edition 1.2) were performed, and EMC testing were conducted in accordance with standard IEC 60601-1-2:2014+AMD1:2020 (Edition 4.1). The manufacturing facility is in conformance with the relevant EPRC standards as specified in 21 CFR 1020.30, 31, and 33 and the records are available for review. PHT-70CHS conforms to the provisions of NEMA PS 3.1-3.18, Digital Imaging and Communications in Medicine (DICOM) Set.
Page 8 / 9
{12}
VATECH Co., Ltd.
Premarket Notification 510(k)
## Bench Testing
Non-clinical consideration report was provided in accordance with the FDA guidelines "Guidance for the submissions of 510(k)'s for Solid State X-ray Imaging Devices". Bench testing according to FDA Guidance "Format for Traditional and Abbreviated 510(k)s, Performance Testing – Bench" were performed. Image Quality evaluation report according to IEC 61223-3-4:2000 and IEC 61223-3-7:2021 were also performed.
## Applied FDA Guidance
"Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submission"
"Off-the-Shelf Software Use in Medical Devices"
"Pediatric Information for X-ray Imaging Device Premarket Notifications"
All test results were satisfactory.
## 10. Conclusions
The subject device and the predicate device have the same indications for use and exhibit similar technical characteristics. The performance data for the subject device demonstrates equivalence or superiority to the predicate device in several aspects.
The newly applied X-ray detector has been proven in performance tests to perform equivalently or better compared to the predicate device across various performance parameters, including DQE, MTF, NPS, and NED. Additionally, the image quality of the new X-ray detectors has been evaluated in compliance with IEC 61223-3-4:2000 and IEC 61223-3-7:2021. Both standard requirements were met.
Safety and effectiveness of new-detector application have been clarified through each verification. Despite the differences in detector models, it was found that total dose level and image quality of subjective device is equivalent to the predicate device when similar exposure areas and exposure conditions are applied. Subject device has been evaluated according to the international standard and U.S. code and proved to be equivalent to the predicate device.
Quality assurance procedures are adhered to, and the specifications and functional requirements have been verified.
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, VATECH Co., Ltd. concludes that Green X 12 VE (Model: PHT-70CHS) and Green X VE (Model: PHT-70CHS) are substantially equivalent to the predicate device as described herein.
Page 9 / 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.