The SMART Osteotomy plates are intended for use in open and closed wedge osteotomies of the medial proximal tibia, lateral proximal tibia, medial distal femur and lateral distal femur, treatment of bone and joint deformities, fractures and malalignment caused by injury or disease such as osteoarthritis. The system is intended for skeletally mature patients only.
Device Story
SMART Osteotomy System consists of anatomically shaped, machined titanium plates and bone screws; used to stabilize osteotomy gaps in proximal tibia or distal femur until union. Plates feature holes for variable angle locking screws (prevent back-out) and non-locking screws (compress plate to bone). System implanted by surgeons in clinical settings; requires standard reusable surgical instruments. Plates/screws are single-use, gamma-sterilized. Device maintains correction angle and reduces fractures by fixing plate position relative to bone. Benefits include stabilization of bone fragments and correction of malalignment, facilitating healing in skeletally mature patients.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993-1), sterilization validation, cleaning validation for reusable instruments, usability, shelf-life, MRI evaluation, packaging testing, and mechanical testing per ASTM F543 and ASTM F382.
Technological Characteristics
Materials: Medical grade titanium alloy. Principle: Anatomically shaped locking compression plates with variable angle locking and non-locking screws. Dimensions: Plates range 85.3mm-103mm length; screws 4.5mm diameter. Connectivity: None. Sterilization: Gamma radiation (plates/screws); steam (reusable instruments). Standards: ASTM F543 (screws), ASTM F382 (plates), ISO 10993-1 (biocompatibility).
Indications for Use
Indicated for skeletally mature patients requiring open/closed wedge osteotomies, treatment of bone/joint deformities, fractures, or malalignment of the medial/lateral proximal tibia and medial/lateral distal femur caused by injury or disease (e.g., osteoarthritis).
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 10, 2026
Actis Medical Pty., Ltd.
% Joseph Azary
Regulatory Consultant
Aztech Regulatory & Quality, LLC
543 Long Hill Ave.,
Shelton, CT 06484
Re: K260090
Trade/Device Name: SMART Osteotomy System
Regulation Number: 21 CFR 888.3030
Regulation Name: Single/multiple component metallic bone fixation appliances and accessories
Regulatory Class: Class II
Product Code: HRS, HWC
Dated: October 10, 2025
Received: January 12, 2026
Dear Joseph Azary :
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260090 - Joseph Azary
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260090 - Joseph Azary
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Thomas Mcnamara -S
For: Christopher Ferreira, M.S.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260090 | ? |
| Please provide the device trade name(s). | | ? |
| SMART Osteotomy System | | |
| Please provide your Indications for Use below. | | ? |
| The SMART Osteotomy plates are intended for use in open and closed wedge osteotomies of the medial proximal tibia, lateral proximal tibia, medial distal femur and lateral distal femur, treatment of bone and joint deformities, fractures and malalignment caused by injury or disease such as osteoarthritis.
The system is intended for skeletally mature patients only. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K260090
1 of 7
# 510(k) Summary
## Traditional 510(k)
## Orthopeasia Spinal Orthopedic Fixation System
1. **SUBMITTER/510(k) HOLDER**
Actis Medical Pty. Ltd
3/18 Ashwin Parade
Torrensville, South Australia
5031
Australia
**Submission Contact**
Joseph Azary
Aztech Regulatory & Quality LLC
543 Long Hill Avenue
Shelton, CT 06484
Email: jazary@erols.com
Alternate Email: jazary@aztechregulatory.com
Telephone: (203) 242-6670
**Date Prepared:** January 9, 2026
**Reason for 510(k):** New device / Initial Submission
2. **DEVICE NAME**
Proprietary Name: SMART Osteotomy System
Classification Name: Single / Multiple Component Metallic Bone Fixation Appliance and Accessories
Common Name: Metallic Bone Fixation Appliance and Accessories
Classification Regulation: 21 CFR 888.3030
Product Code: HRS
Secondary Product Codes: HWC
Classification: Class 2
Medical Specialty (Panel): Orthopedic
Actis Medical
January 2026
Traditional 510(k) for SMART Osteotomy System
{5}
K260090
2 of 7
# 3. PREDICATE DEVICES
| Device Name | Manufacturer | 510(k) Number |
| --- | --- | --- |
| DePuy Synthes TOMOFIX | Synthes (USA) Products | K141796 |
| Osteotomy System | LLC | |
# 4. DEVICE DESCRIPTION
The SMART Osteotomy system consists of profiled plates, designed to fit the proximal tibia or distal femur. Plates are anatomically shaped machined titanium, with holes for placement of screws that hold the plate to the bone. They are designed to be implanted on the surface of the proximal tibia or distal femur to stabilize the osteotomy gap, maintaining the correction angle until union. Bone screws fix the plate in position relative to the bone to locate fragments, reduce the fracture, and prevent the plate from shifting. Variable angle locking screws are fixed to the plate, preventing screw back-out. Non-locking screws compress the plate to the bone surface, featuring threads that penetrate deeper into cancellous bone.
The plates and screws are single use, terminally sterilized by the manufacturer via gamma radiation. The devices are packaged individually in a heat sealed inner Tyvek pouch then packaged in an heat sealed outer pouch, The double packaged device is placed inside a cardboard box that is labeled and shrink wrapped.
The devices are used with a standard armamentarium of reusable surgical instruments that are to be sterilized by the user.
# 5. INTENDED USE / INDICATIONS FOR USE
The SMART Osteotomy plates are intended for use in open and closed wedge osteotomies of the medial proximal tibia, lateral proximal tibia, medial distal femur and lateral distal femur, treatment of bone and joint deformities, fractures and malalignment caused by injury or disease such as osteoarthritis.
The system is intended for skeletally mature patients only.
# 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The subject device has similar and/or identical characteristics as the predicate devices including material composition, principles of operation, similar indications for use, system components, configurations, tapered ends, holes for screws, similar sized screws, sterile, single use and standard armamentarium of instruments.
Actis Medical
Traditional 510(k) for SMART Osteotomy System
January 2026
{6}
K260090
3 of 7
The minor differences are that the devices have different shaped heads and the dimensions of the subject device plates are slightly smaller compared to the predicate.
## Substantial Equivalence Comparison Table
| Technological Characteristic | Subject Device | TOMOFIX K141796 | Comparison |
| --- | --- | --- | --- |
| Indications for Use | The SMART Osteotomy plates are intended for use in open and closed wedge osteotomies of the medial proximal tibia, lateral proximal tibia, medial distal femur and lateral distal femur, treatment of bone and joint deformities, fractures and malalignment caused by injury or disease such as osteoarthritis. The system is intended for skeletally mature patients only. | The DePuy Synthes TOMOFIX Osteotomy System is intended for osteotomies, treatment of bone and join deformities, fixation of fractures, and malalignment caused by injury or disease, such as osteoarthritis, of the distal femur and proximal tibia.
Specifically,
The TOMOFIX Medial Proximal Tibia Plates are indicated for open and closed wedge osteotomies fixation of fractures and malalignment caused by injury or disease, such as osteoarthritis, of the medial proximal tibia
The TOMOFIX Lateral Proximal Tibia Plates are indicated for open and closed wedge osteotomies fixation of fractures and malalignment caused by injury or disease, such as osteoarthritis, of the lateral proximal tibia
The TOMOFIX Lateral Distal Femur Plates are indicated for open and closed wedge osteotomies fixation of fractures and malalignment caused by injury or disease, such as osteoarthritis, of the Actis Medical
Traditional 510(k) for SMART Osteotomy System | EQUIVALENT
The devices are both used for treatment of bone and joint deformities, fixation of fractures and malalignment in Tibia and Femur. |
January 2026
{7}
K260090
4 of 7
| | | lateral distal femur
The TOMOFIX Medial Distal Femur Plates are indicated for open and closed wedge osteotomies fixation of fractures and malalignment caused by injury or disease, such as osteoarthritis , of the medial distal femur | |
| --- | --- | --- | --- |
| System Components | Plates and Screws
Instruments | Plates and Screws
Instruments | IDENTICAL |
| Configurations | Offered in
Medial High Tibia
Lateral High Tibia
Medial Distal Femur
Lateral Distal Femur | Offered in
Medial High Tibia
Lateral High Tibia
Medial Distal Femur
Lateral Distal Femur | IDENTICAL |
| Classification Regulation / Product Code | HRS
Class II
21 CFR 888.3030 | HRS
Class II
21 CFR 888.3030 | IDENTICAL |
| Material | Medical Grade Titanium Alloy | Medical Grade Titanium Alloy | IDENTICAL |
| Sterility | Plates and Screws are provided sterile using Gamma Radiation
Instruments must be sterilized using steam sterilization by user | Plates and Screws are provided sterile using Gamma Radiation
Instruments must be sterilized using steam sterilization by user | IDENTICAL |
| Single Use | Plates and Screws are Single Use | Plates and Screws are Single Use | IDENTICAL |
| Pediatric Usage | No, not indicated for pediatric usage | No, patients where growth plates have fused | IDENTICAL |
| Principles of Operation | Anatomically shaped plates with holes for placement of screws that hold the plate to the bone. They are designed to be implanted on the surface of the proximal tibia to stabilize the osteotomy gap, | Pre-contoured to fit the anatomy of the medial or lateral proximal tibia. Locked internal fixator providing angular stable fixation for osteotomy gap stabilization until union is achieved. The fixed-angle locking | IDENTICAL |
Actis Medical
Traditional 510(k) for SMART Osteotomy System
January 2026
{8}
K260090
5 of 7
| | maintaining the correction angle until union.
Bone screws fix the plate in position relative to the bone to locate fragments, reduce the fracture, and prevent the plate from shifting. Locking screws are fixed to the plate, preventing screw back-out. Non-locking screws compress the plate to the bone surface, featuring threads that penetrate deeper into cancellous bone. | holes provide multiple fixed-angle constructs throughout the plate, improving retention of screws in the plate and in cortical bone. Dynamic compression can be achieved by eccentric insertion titanium cortex screws in the dynamic compression unit portion of the hole. | |
| --- | --- | --- | --- |
| Shape | L-shaped locking compression plate
Designed to bridge osteotomy gap with enough strength to maintain correction angle
Tapered ends
Plates able to accommodate both locking and compression screws | T-Shaped locking compression plate
Design to bridge osteotomy gap with enough strength to maintain correction angle
Tapered ends
Plates have combi holes for either axial compression or locking capability | Devices have different shaped “heads”, but all are design to match the anatomy. |
| Plate Dimensions | Lateral Distal Femur Plate: Length 103mm, width 32mm, Right and Left
Lateral High Tibia Plate: Length 89mm, width 34mm, Right and Left
Medial High Tibia Plate: Length 85.3mm, width 35mm, Right and Left
Medial Distal Femur Plate: Length 102.5mm, width 29.6mm, Right and Left | Lateral Distal Femur Plate: Length 141mm, Right and Left
Lateral High Tibia Plate: Length 102mm, Right and Left
Medial High Tibia Head Plate: Length 112mm, Right and Left
Medial Distal Femur Plate: Length 113.5mm, Right and Left | SIMILAR
Both have plates for Lateral Distal Femur, Lateral High Tibia, Medial High Tibia and Medial Distal Femur
The subject device has slightly shorter lengths compared to the predicate devices. |
| Screw Dimensions | Locking Screws Diameter 4.5mm, lengths: 29mm – 90mm | 4.5mm diameter, non-locking
Lengths 24, 26, 28, 30, | SIMILAR
Both have screws with 4.5mm diameter and similar ranges of lengths. |
Actis Medical
Traditional 510(k) for SMART Osteotomy System
January 2026
{9}
K260090
6 of 7
| | Non-Locking Screws Diameter 4.5mm, lengths 29mm to 60mm | 32, 34, 36, 38, 40, 42, 44, 46, 48, 50, 52mm
4.5mm diameter, non-locking
Lengths 14, 16, 18, 20, 22, 54, 56, 58, 60, 62, 64, 66, 68, 70, 72, 76, 80, 85, 90mm
5.0mm diameter, locking
Lengths 24, 26, 28, 30, 32, 34, 36, 38, 40, 42, 44, 46, 48, 50, 60, 65, 70, 75, 80 85
5.0 diameter, locking
Lengths: 14, 16, 18, 20, 22, 90 | |
| --- | --- | --- | --- |
| Instruments | Standard Armamentarium of Instruments
Drills, Drill Guides, Depth Gage, Screwdriver, Bone Tamp | Standard Armamentarium of instruments
Drills, Drill Guides, Depth Gages, Screwdriver, Bone Spreader, Guide Wire | EQUIVALENT |
## 7. PERFORMANCE TESTING
The SMART Osteotomy System was subjected to a variety of testing including:
- Biocompatibility Evaluation per ISO 10993-1
- Sterilization Validation
- Cleaning Validation (for reusable instruments)
- Usability
- Stability / Shelf Life
- MRI evaluation
- Packaging testing
- Mechanical Testing per ASTM F543 and ASTM F382
Actis Medical
Traditional 510(k) for SMART Osteotomy System
January 2026
{10}
K260090
7 of 7
# 8. SAFETY AND EFFICACY
The subject device has been tested to and complies with applicable consensus standards. All of the testing passed and there were no identified issues that would impact the safety, performance, or efficacy of the subject device.
# 9. CONCLUSION
The information presented supports substantial equivalence of the SMART Osteotomy System to the predicate device based on similarities in intended use, design, principles of operation and performance specifications.
The company concluded that based on testing, compliance to consensus standards, the indications for use, technological characteristics, and comparison to predicate device the SMART Osteotomy System found that the differences do not impact the ability of the devices to be safe and effective for their intended use.
Actis Medical
Traditional 510(k) for SMART Osteotomy System
January 2026
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.