AI/ML-enabled algorithms and locked neural network algorithms
—
adequate sensitivity and specificity
—
—
Plaque clinical study: 60 subjects
—
Gingival recession detection
AI/ML-enabled algorithms and locked neural network algorithms
—
adequate sensitivity and specificity
—
—
Gingival Recession clinical study: 109 subjects
—
Tooth wear detection
AI/ML-enabled algorithms and locked neural network algorithms
—
adequate sensitivity and specificity
—
—
Tooth Wear clinical study: 61 subjects
—
Surface caries detection
AI/ML-enabled algorithms and locked neural network algorithms
—
adequate sensitivity and specificity
—
—
Surface Caries clinical study: 34 subjects
—
Indications for Use
TRIOS Dx is intended as an aid in the diagnosis of oral health conditions and the assessment of changes in teeth and gingiva. TRIOS Dx is intended to aid in the diagnosis and assessment of: • caries • tooth wear • gingival recession • plaque
Device Story
Software-only medical device (SaMD) for dental clinics; operated by qualified dental professionals. Inputs: 3D intraoral scans from 3Shape or third-party scanners. Processing: AI/ML-enabled algorithms analyze scan data to detect/quantify oral health conditions. Outputs: Diagnostic aid assessments, comparative visualizations, and measurement tools displayed via graphical interface on Windows-based PCs. Clinical use: Professionals review results alongside traditional exams to communicate findings to patients; aids in diagnosis of caries, tooth wear, gingival recession, and plaque. Benefits: Enhanced visualization and objective assessment of oral health changes to support clinical decision-making.
Clinical Evidence
Clinical studies conducted in UK and Denmark under GCP. Plaque (n=60), Gingival Recession (n=109), Tooth Wear (n=61), Surface Caries (n=34). Studies were single-arm cross-sectional. Results demonstrated adequate sensitivity and specificity for all indications. No adverse events reported. Bench testing confirmed functionality for proximal caries. Data supports substantial equivalence.
Technological Characteristics
Software-only; runs on Windows-based generic computing platforms. Employs AI/ML-enabled locked neural network algorithms. Connectivity: Integrates with 3Shape Unite platform for case management. User interface: Mouse, keyboard, or touchscreen. No direct patient contact.
Indications for Use
Indicated for patients >18 years old to aid in the diagnosis and assessment of caries, tooth wear, gingival recession, and plaque.
Regulatory Classification
Identification
Dental Image Analyzer. A dental image analyzer is a prescription home use device that uses software intended to collect and analyze patient-specific, optical camera-based, intraoral images. The analyses are provided to dental health care professionals as an aid to diagnosis of oral health conditions and/or to monitor treatment progress.
Special Controls
In combination with the general controls of the FD&C Act. the Dental Image Analyzer is subject to the following special controls:
(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Non-clinical performance testing data must demonstrate that the device performs as intended under anticipated conditions of use and must include full specifications of the hardware requirements and testing to demonstrate the specified hardware ensures adequate data for validated and accurate measurements; (3) Clinical performance data must demonstrate the accuracy and precision of the diagnostic and monitoring device algorithms to perform as intended under anticipated conditions of use.
(4) Usability testing must demonstrate that the user can correctly use the device, based solely on reading the directions for use.
(5) Software verification, validation, and hazard analysis must be performed.
(6) Labeling must include:
(i) Information on the patient population for which the device has been demonstrated to be effective.
(ii) Information on how the device operates during the typical course of treatment
(iii) A detailed summary of the device technical parameters, including compatible operating systems, hardware, and/or accessories.
(iv) Instructions for cleaning of any reusable components.
(v) Warning that the device is not intended to be used independently for diagnosis and that the device does not replace clinical decision making.
- (7) Patient labeling must be provided and must include:
(i) Relevant contraindications, warnings, precautions, adverse effects/complications, and when to seek help from a dental healthcare professional.
(ii) Information on how the device operates during the typical course of treatment.
(iii) A detailed summary of the device technical parameters, including compatible operating systems, hardware, and/or accessories.
(iv) Information on the patient population for which there is clinical evidence of effectiveness.
- (v) Instructions for cleaning of any reusable components.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 10, 2026
3Shape TRIOS A/S
Jenny Axel
Senior Regulatory Affairs Specialist
Holmens Kanal 7, Copenhagen, 1060 DNK
Re: K260082
Trade/Device Name: TRIOS Dx (R1)
Regulation Number: 21 CFR 872.1770
Regulation Name: Dental Image Analyzer
Regulatory Class: Class II
Product Code: SHQ
Dated: January 12, 2026
Received: January 12, 2026
Dear Jenny Axel:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260082 - Jenny Axel
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260082 - Jenny Axel
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K260082 | ? |
| Please provide the device trade name(s). | | ? |
| TRIOS Dx (R1) | | |
| Please provide your Indications for Use below. | | ? |
| TRIOS Dx is intended as an aid in the diagnosis of oral health conditions and the assessment of changes in
teeth and gingiva.
TRIOS Dx is intended to aid in the diagnosis and assessment of:
• caries
• tooth wear
• gingival recession
• plaque | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
| 510(k) #: K260082 | 510(k) Summary | Prepared on: 2026-01-12 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | 3Shape TRIOS A/S | |
| Applicant Address | Holmens Kanal 7 Copenhagen 1060 Denmark | |
| Applicant Contact Telephone | +4550298535 | |
| Applicant Contact | Ms. Jenny Axel | |
| Applicant Contact Email | Jenny.Bathke@3Shape.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | TRIOS Dx (R1) | |
| Common Name | Dental image analyzer | |
| Classification Name | Dental Image Analyzer | |
| Regulation Number | 872.1770 | |
| Product Code(s) | SHQ | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| DEN230035 | Dental Monitoring | SBC |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| Device description summary | | |
| TRIOS Dx is a software-only medical device intended to aid dental professionals in the diagnosis of oral health conditions and assessing changes in teeth and gingiva. The device analyzes intraoral scan data using AI/ML-enabled algorithms and displays diagnostic aid results in the user interface for visualization and comparison of scans. Qualified dental professionals review these results alongside a traditional dental examination and communicate findings to the patient. | | |
| The software is intended to operate on a generic computing platform running Microsoft Windows and is intended for use in dental clinics by qualified dental professionals. TRIOS Dx is a diagnostic aid device, and it is not intended to replace clinical decision-making. The device has no direct patient contact. | | |
| TRIOS Dx accepts intraoral scans from 3Shape scanners or third-party scanners (with limited functionality) and integrates with the 3Shape Unite platform for case management. The user interacts with the device through a graphical interface using a computer mouse, keyboard, or touchscreen. Inputs include 3D dental scans; outputs include diagnostic aid assessments, comparative visualizations, and measurement tools for distances and surfaces. | | |
| The device provides features for patient communication, enabling dental professionals to present findings in a clear and understandable format. TRIOS Dx does not require accessories and is marketed in 'Standard' and 'Plus' configurations, based on the active user's license. | | |
| Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) |
{5}
TRIOS Dx is intended as an aid in the diagnosis of oral health conditions and the assessment of changes in teeth and gingiva.
TRIOS Dx is intended to aid in the diagnosis and assessment of:
- caries
- tooth wear
- gingival recession
- plaque
## Indications for Use Comparison
21 CFR 807.92(a)(5)
Both the subject and predicate devices are intended to aid in the diagnosis of oral health conditions and to assess changes in teeth and gingiva. Both the subject and predicate devices are intended to aid in the diagnosis of tooth wear, gingival recession, and plaque. Additionally, TRIOS Dx is intended to aid in the diagnosis of caries. The subject device's additional indication for use is claimed based on performance data. This additional indication falls within the same intended use as the predicate device and does not alter the intended therapeutic use of the device. While the indications for use for the subject device are not identical to the predicate device, the differences do not affect the safety and effectiveness of the device relative to the predicate.
In conclusion, all indications for use of the subject device fall within the intended use of the predicate device and, therefore, the two devices have the same intended use.
Three differences between the subject and predicate devices have been identified regarding intended users, intended patient profile, and intended use environment, which do not raise different questions of safety and effectiveness:
- Intended users: The subject device is intended to be operated by qualified dental professionals. The predicate device is intended to be operated by a patient, a non-healthcare professional, or a healthcare professional. The subject device's intended users are a subset of the predicate device's intended users. Software validation demonstrates that the device conforms to user needs in the specified use conditions.
- Intended patient profile: The subject device is intended for patients above 18 years of age. The predicate device is intended for patients above 6 years of age. The subject device's patient population is a subset of the predicate device's patient population. The subject device passed all clinical validation requirements for the intended patient profile.
- Intended use environment: The subject device is intended for use in dental clinics. The predicate device is intended for use in healthcare facilities or in a non-healthcare environment (patient's home). The subject device's clinical context or setting is a subset of the predicate device clinical context or setting. Software validation demonstrates that the device conforms to user needs in the specified use conditions.
## Technological Comparison
21 CFR 807.92(a)(6)
The subject and predicate device are technologically equivalent as follows:
- Both devices process digital intraoral scans
- Both are software devices with embedded machine learning and locked neural network algorithms aiding in the diagnosis of oral health conditions
- Both devices display oral health conditions within the user interface with a graphical overlay on the intraoral scan
- Both devices assess changes in teeth and gingiva
- Both workflows include a patient profile set-up, intraoral scan upload, analysis of uploaded intraoral scans, and communication of results
- Both devices' user interfaces include a computer, a display, a keyboard and a mouse or a touch screen
- Both are software devices designed to run on Windows operating systems
Three technological differences between the subject and predicate devices have been identified, which do not raise different questions of safety and effectiveness:
- Operating system: The predicate device is intended to operate on a broader range of platforms and operating systems than the subject device.
- Input specifications: The predicate device employs a broader range of input specifications than the subject device.
- Accessories and hardware components: The predicate device requires accessories and hardware components used to remotely track treatment progress, a feature which is not claimed by the subject device.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
{6}
Non-clinical testing included bench evaluations of the TRIOS Dx diagnostic aid functionality for proximal caries. All tests met the predefined acceptance criteria and demonstrated that TRIOS Dx performs as intended under anticipated conditions of use. These results support the additional indication for use (proximal caries) claimed for the subject device.
Clinical testing evaluated the TRIOS Dx performance for the following diagnostic aid indications:
- Plaque: n=60 subjects; single-arm cross-sectional clinical study; study location: United Kingdom
- Gingival Recession: n=109 subjects; single-arm cross-sectional clinical study; study location: United Kingdom
- Tooth Wear: n=61 subjects; single-arm cross-sectional clinical study; study location: United Kingdom
- Surface Caries: n=34 subjects; single-arm cross-sectional clinical study; study location: Denmark
Studies were conducted under Good Clinical Practice (GCP), registered on ISRCTN or clinicaltrials.gov, and adhered to ethical principles of the Declaration of Helsinki.
Clinical testing demonstrated adequate sensitivity and specificity for the diagnostic aid indications plaque, gingival recession, tooth wear, and surface caries. No adverse events or complications were reported.
Both the subject and predicate devices are intended to aid in the diagnosis of plaque, gingival recession, and tooth wear. Additionally, TRIOS Dx is intended to aid in the diagnosis of caries. The subject device's additional indication for use is claimed based on performance data. This additional indication falls within the same intended use as the predicate device and does not alter the intended therapeutic use of the device, nor does it affect the safety and effectiveness of the device relative to the predicate.
These results support the additional indications for use (surface caries) claimed for the subject device, confirm that the subject device performs as intended under anticipated conditions of use in the intended use population, and support substantial equivalence to the predicate device.
Non-clinical and clinical testing relied upon for substantial equivalence demonstrate that TRIOS Dx is as safe and as effective as the predicate device. Bench testing for proximal caries confirmed that the subject device performs as intended under anticipated conditions of use. Clinical testing showed adequate sensitivity and specificity for the plaque, gingival recession, tooth wear, and surface caries diagnostic aid indications without reported adverse events, and confirmed that the subject device performs as intended under anticipated conditions of use in the intended use population. These results support the conclusion that TRIOS Dx is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.