Eternal Art Porcelain Powders

K260067 · Shandong Huge Dental Material Corporation · EIH · Apr 9, 2026 · Dental

Device Facts

Record IDK260067
Device NameEternal Art Porcelain Powders
ApplicantShandong Huge Dental Material Corporation
Product CodeEIH · Dental
Decision DateApr 9, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.6660
Device ClassClass 2

Indications for Use

This device is indicated for use as a veneering material for fixed prosthesis in crowns and bridges. It is used for external staining and glazing in prosthetic dentistry by forming a porcelain veneer on to a ceramic substructure.

Device Story

Eternal Art Porcelain Powders are dental veneering materials used for color staining and glazing of zirconia or glass-ceramic restorations. The system consists of three components: porcelain powder, blending liquid, and pre-mixed stain/glaze paste. Dental technicians apply the material to the ceramic substructure to form a veneer. The device is used in a dental laboratory setting by trained technicians. The output is a finished porcelain veneer that provides aesthetic color and surface glazing to the dental prosthesis. The device benefits patients by improving the aesthetic appearance and surface finish of crowns and bridges.

Clinical Evidence

No clinical data. Bench testing performed according to ISO 6872:2024 (uniformity, radioactivity, flexural strength, chemical solubility, thermal expansion, glass transition temperature) and biocompatibility testing (cytotoxicity, skin sensitization, irritation, acute/subchronic systemic toxicity, genotoxicity) per ISO 10993 and ISO 7405. Results demonstrate substantial equivalence to the predicate.

Technological Characteristics

Type I ceramic material per ISO 6872:2024. Composed of glass powder (powder) and purified water/organic solvents (liquid). Non-sterile. Physical forms include powder, liquid, and pre-mixed paste. Performance characteristics include flexural strength, chemical solubility, and thermal expansion coefficient compliance with ISO 6872:2024.

Indications for Use

Indicated for use as a veneering material for fixed prosthesis in crowns and bridges, specifically for external staining and glazing of ceramic substructures in prosthetic dentistry.

Regulatory Classification

Identification

Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION April 9, 2026 Shandong Huge Dental Material Corporation Zheng Maggie Regulatory Affairs Manager # 68 Shanhai Rd., Donggang District Rizhao City, 276800 CHINA Re: K260067 Trade/Device Name: Eternal Art Porcelain Powders Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: January 9, 2026 Received: January 9, 2026 Dear Zheng Maggie: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K260067 - Zheng Maggie Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K260067 - Zheng Maggie Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, MICHAEL E. ADJODHA -S Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K260067 | | | Device Name Eternal Art Porcelain Powders | | | Indications for Use (Describe) This device is indicated for use as a veneering material for fixed prosthesis in crowns and bridges. It is used for external staining and glazing in prosthetic dentistry by forming a porcelain veneer on to a ceramic substructure. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} HUGE # K260067 # 510 (k) Summary This summary of 510(k) for the subjective device equivalence information is being submitted in accordance with the requirements of 21 C.F.R. 807.92. 1. Date Summary Prepared: April 8, 2026 2. Submitter Information: Owner's Name: Shandong Huge Dental Material Corporation Address: No. 68 Shanhai Road, Donggang District, Rizhao City, Shandong Province, 276800, P.R. China Telephone: 0086 633 2277268 Fax: 0086 633 2277298 Contact Person: Ms. Maggie Zheng Contact Title: Regulatory Affairs Manager E-mail: zhengxy@hugedent.com 3. Device Name Trade name: Eternal Art Porcelain Powders Common name: Porcelain Powders Classification name: Porcelain powder for clinical use (21 CFR 872.6660) Regulatory Class: Class II Product Code: EIH 4. Predicate and Reference Device Information | Owner/Operator | Device Trade Name | 510 (k) No. | Product Code | Predicate / Reference | | --- | --- | --- | --- | --- | | Liaoning Upcera Co., Ltd | Upcera Glaze Paste, Glaze Powder, and Glaze Liquid | K181167 | EIH | Predicate (Primary) | | Chengdu Besmile Medical Technology Co., Ltd. | Dental Porcelain Powder | K232676 | EIH | Reference | {5} HUGE | Owner/Operator | Device Trade Name | 510 (k) No. | Product Code | Predicate / Reference | | --- | --- | --- | --- | --- | | Aidite (Qinhuangdao) Technology Co., Ltd | Porcelain Powder | K250025 | EIH | Reference | Note: The main reason for inclusion of the reference devices, Dental Porcelain Powder (K232676) and Porcelain Powder (K250025), is to support similarities in composition of materials to the subject device. # 5. Description of Device Eternal Art Porcelain Powders are dental veneering materials used for color staining and glazing of the surfaces of restorations such as zirconia or glass-ceramic substrates. It has 3 models: Powder, Stain/Glaze and Blending Liquid. Powder is in the form of a porcelain dry powder, Blending Liquid is a material used to mix the Powder into paste form, and the Stain/Glaze is pre-mixed paste prepared by uniformly mixing the Powder and the Blending Liquid at a certain ratio, which can help dental technicians save the mixing process, and convenient to use. Eternal Art Porcelain Powders are classified as Type I Ceramic products that are provided as powders, pastes, or aerosols according to ISO 6872:2024 Dentistry—Ceramic materials, and designed into various specifications and shades to meet different needs. Meanwhile, the production process, processing technology, raw materials and performance of different specifications are consistent. # 6. Indications for Use This device is indicated for use as a veneering material for fixed prosthesis in crowns and bridges. It is used for external staining and glazing in prosthetic dentistry by forming a porcelain veneer on to a ceramic substructure. # 7. Summary of Physical and Chemical Properties Tests The physical properties of the subject device were determined and tested according to ISO 6872:2024 and internal standard. Bench testings were performed on the subject device and the predicate device, the test results demonstrated the substantial equivalence when compared to the predicate device. 13-2 {6} HUGE | Summary of Physical and Chemical Properties Test | | | | | --- | --- | --- | --- | | Model | Items per internal standards | Pass/fail criteria | Conclusion | | Powder and Stain/Glaze | Uniformity | Meet ISO 6872 | Satisfactory | | | Freedom from extraneous materials | Meet ISO 6872 | Satisfactory | | | Radioactivity | Meet ISO 6872 | Satisfactory | | | Mixing properties | Meet ISO 6872 | Satisfactory | | | Condensation properties | Meet ISO 6872 | Satisfactory | | | Flexural strength | Meet ISO 6872 | Satisfactory | | | Chemical solubility | Meet ISO 6872 | Satisfactory | | | Coefficient of thermal expansion | Meet ISO 6872 | Satisfactory | | | Glass transition temperature | Meet ISO 6872 | Satisfactory | | Blending Liquid | Meet internal standard | | Satisfactory | # 8. Technological Characteristics Comparison All components of the subject device are based upon industry well-known chemistry. The following table shows the significant technological characteristics for the subject device and indicates the following similarities and differences with the predicate and reference devices: | Technological Characteristics | Subject device K260067 | Predicate device K181167 | | --- | --- | --- | | Product Name | Eternal Art Porcelain Powders | Upcera Glaze Paste, Glaze Powder, and Glaze Liquid | | Company name | Shandong Huge Dental Material Corporation | Liaoning Upcera Co., Ltd | | Main Composition | The Powder is mainly composed of glass powder. The Blending Liquid is mainly composed of purified water and organic solvents. | The Powder is mainly composed of glass powder. The Blending Liquid is mainly composed of purified water and organic solvents. | | Physical Form | Powder, Liquid and Paste | Powder, Liquid and Paste | | Indications of Use | This device is indicated for use as a veneering material for fixed prosthesis in crowns and bridges. It is used for external staining and glazing in prosthetic dentistry | “Glaze Paste, Glaze Powder, and Glaze Liquid” are indicated for use as a veneering material for fixed prosthesis in crowns and bridges. This device is used in | {7} HUGE | Technological Characteristics | Subject device K260067 | Predicate device K181167 | | --- | --- | --- | | Product Name | Eternal Art Porcelain Powders | Upcera Glaze Paste, Glaze Powder, and Glaze Liquid | | Company name | Shandong Huge Dental Material Corporation | Liaoning Upcera Co., Ltd | | | by forming a porcelain veneer on to a ceramic substructure. | prosthetic dentistry by forming a porcelain veneer on to a ceramic substructure. | | Prescription/over-the-counter use | Prescription | Prescription | | Sterility | Non-sterile | Non-sterile | | Physical Properties | The subject device and predicate device have substantially equivalent physical properties as they all conform to the specifications set in ISO 6872:2024 and internal standard. | | The indications of the subject device are similar to those of 510(k) cleared predicate device. Based on the similarities in indications for use, the subject device has demonstrated substantial equivalence to the predicate device, and the minor difference in wording between the two devices does not affect the substantial equivalence. The subject device's main components are the same as those of the predicate device, with only minor difference in composition. These minor components have a history of safe use in legally marketed reference devices. Additionally, the subject device has successfully undergone biocompatibility testing. Therefore, these differences do not raise new safety risks and do not affect the substantial equivalence. Based on ISO 6872:2024 and internal standard, technological characteristics, physical properties, performance testing are carried to compare the subject device with predicate device. According to test results, the subject device has similar technological characteristics with the predicate device. # 9. Summary of Non-clinical testing The physical properties of the subject device were determined and tested according to ISO 6872:2024 and internal standard, and the test results demonstrated the qualification and substantial equivalence when compared to the predicate device. Additionally, the subject device is substantially equivalent to the predicate and reference devices. The compositions of the subject device do not contain any non-conventional chemicals compared to the legally marketed predicate and reference devices. 13-4 {8} HUGE Biocompatibility tests were performed fully following the ISO 7405 and ISO 10993 standards, including Cytotoxicity, Skin sensitization, Irritation, Acute systemic toxicity, Subchronic systemic toxicity and Genotoxicity, and test results are sufficient to prove the subject device is substantially equivalent to the predicate device. ## 10. Clinical Performance Data Clinical test is not applicable. ## 11. Conclusions Based on the indications for use, technological characteristics, performance testing and comparison to predicate device, the subject device has demonstrated substantial equivalence. Shandong Huge Dental Material Corporation concludes that the subject device is substantially equivalent to the predicate device described herein. 13-5
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...