The I.T.S. PRS Phoenix II is indicated to stabilize one or more pelvic bone fractures in the pelvis in an adult patient. Indications for use of the I.T.S. PRS Phoenix II include: • Symphyseal Disruptions & Parasymphyseal Fractures • Disruptions of the Sacroiliac Joint • Fractures involving the Anterior Column of the Acetabulum • Fractures involving the Quadrilateral Surface • Fractures involving the Posterior Wall • Fractures involving the Posterior Column • Dorsal neutralization plating for Posterior Pelvic Ring Fractures • Fractures of the Acetabulum, Fractures of the Pelvic Ring, Fractures of the Ilium • Fracture Revision surgery of pseudoarthrosis, non-unions and mal-unions The I.T.S. PRS Phoenix II System is not intended for spinal use.
Device Story
I.T.S. PRS Phoenix II is a metallic bone fixation system comprising various anatomically shaped plates and screws; used by surgeons in clinical settings to stabilize pelvic fractures. The system provides mechanical support to bone segments during the healing process. It is not intended for spinal applications. The device is implanted via surgical procedure; the surgeon selects appropriate plates and screws based on fracture location and geometry. The system functions as a rigid fixation construct to maintain bone alignment, facilitating union. Benefits include stabilization of complex pelvic ring and acetabular fractures, aiding patient recovery.
Clinical Evidence
No clinical data was required for this submission. Evidence is based on non-clinical bench testing, including FEA-simulation, mechanical screw testing, biocompatibility (cytotoxicity), sterility, and MRI safety evaluations.
Technological Characteristics
Metallic bone fixation system (plates and screws) made of Titanium (ASTM F67, ASTM F136, Type II anodized). Includes various plate geometries (Symphysis, W-Plate, H-Plate, etc.) and cortical/cancellous screws. Mechanical performance tested per ASTM F382 and ASTM F543. MRI safety tested per ASTM F2182, F2052, F2213, and F2119. Sterilization per ISO 11137. Standalone mechanical device.
Indications for Use
Indicated for adult patients requiring stabilization of pelvic bone fractures, including symphyseal/parasymphyseal fractures, sacroiliac joint disruptions, acetabular fractures (anterior column, quadrilateral surface, posterior wall/column), posterior pelvic ring fractures, and revision surgery for pseudoarthrosis, non-unions, or mal-unions. Not for spinal use.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
I.T.S. Pelvic Reconstruction System PRS RX & PRS Phoenix (K210935)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 24, 2026
I.T.S. GmbH
% Mindy McCann
VP US Operations & Principal Consultant
Qserve Group US
350 S Main St., Suite 309
Doylestown, Pennsylvania 18901
Re: K260054
Trade/Device Name: I.T.S. PRS Phoenix II
Regulation Number: 21 CFR 888.3030
Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories
Regulatory Class: Class II
Product Code: HRS, HTN, HWC
Dated: March 27, 2026
Received: March 27, 2026
Dear Mindy McCann:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260054 - Mindy McCann
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260054 - Mindy Mccann
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# CHRISTOPHER FERREIRA -S
Christopher Ferreira, M.S.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K260054 | ? |
| Please provide the device trade name(s). | | ? |
| I.T.S. PRS Phoenix II | | |
| Please provide your Indications for Use below. | | ? |
| The I.T.S. PRS Phoenix II is indicated to stabilize one or more pelvic bone fractures in the pelvis in an adult
patient. Indications for use of the I.T.S. PRS Phoenix II include:
• Symphyseal Disruptions & Parasymphyseal Fractures
• Disruptions of the Sacroiliac Joint
• Fractures involving the Anterior Column of the Acetabulum
• Fractures involving the Quadrilateral Surface
• Fractures involving the Posterior Wall
• Fractures involving the Posterior Column
• Dorsal neutralization plating for Posterior Pelvic Ring Fractures
• Fractures of the Acetabulum, Fractures of the Pelvic Ring, Fractures of the Ilium
• Fracture Revision surgery of pseudoarthrosis, non-unions and mal-unions
The I.T.S. PRS Phoenix II System is not intended for spinal use. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K260054
510(k) Submission
I.T.S. GmbH
I.T.S. PRS Phoenix II
In Accordance with 21 CFR 807.92 of the Federal Code of Regulations
510(k) Summary
NAME OF FIRM:
I.T.S. GmbH
Autal 28
Lassnitzhoehe, 8301
Austria
www.its-implant.com
510(k) FIRM CONTACT:
Mindy Mccan
Qserve Group US, Inc.
350 S Main Street, Suite 309
Doylestown, Pennsylvania, 18901, United States
Tel. No. 424-271-8169
e-mail: usagent@qservegroup.com
TRADE NAME:
I.T.S. PRS Phoenix II
DATE:
April 23th, 2026
COMMON NAME:
Plate, fixation, bone, washer, bolt nut,
Screw, fixation, bone
REGULATORY CLASS:
Class II 21 CFR 888.3030 - Single/multiple component metallic bone fixation appliance and accessories
DEVICE PRODUCT CODE:
HRS
SUBSEQUENT PRODUCT CODE:
HTN, HWC
SUBSTANTIAL EQUIVALENCE:
PRIMARY PREDICATE
I.T.S. Pelvic Reconstruction System PRS RX & PRS Phoenix (K210935)
ADDITIONAL PREDICATES
I.T.S. Pelvic Reconstruction System (K063166)
Stryker PRO Plating System, Stryker Trauma Pelvic Set (Matta) (K223772)
DePuy Synthes 3.5mm Intrapelvic Acetabular System (K221809)
Acumed Pelvic Bone Plate System (K122538)
DEVICE DESCRIPTION:
The I.T.S. PRS Phoenix II consists of the following implants:
- Symphysis Plate, 4-Hole
- Symphysis Plate, 6-Hole
- Biplanar Symphysis Plate
- W-Plate 4-Hole (Small Frag)
- W-Plate 4-Hole (Large Frag)
- Extended Anterior Column Plate, left/right
- H-Plate
- Anterior Column Plate, left/right
- Bridge Plate
- Base Plate
- Infrapectineal Buttress Plate, left/right
- Helical Infrapectineal Plate, left/right
- Biplanar Plate, 21-Hole, left/right
- Biplanar Plate, 23-Hole, left/right
- Spring Plate
- Rim Stabilization Plate, 4-Hole
- Rim Stabilization Plate, 6-Hole
K260054 - Page 1/3
{5}
K260054 - Page 2/3
# 510(k) Submission
## I.T.S. GmbH
## I.T.S. PRS Phoenix II
- Posterior Wall Plate, left/right
- Global Posterior Stabilization Plate, left/right
- Posterior Column Plate, left/right
- Straight Plates
- Curved Plates
- Cortical Screw, D=2.7mm, Non-Locking
- Cortical Stabilization Screw, D=3.0mm, RH, Locking
- Cancellous Stabilization Screw, D=3.0mm, D=3.0mm, RH, Locking
- Cortical Screw, D=3.5mm, Non-locking
- Cancellous Screw, D=4.2mm, Locking
- Cancellous Screw, D=5.9mm, Non-Locking
- Cancellous Screw, D=5.9mm, Locking
- Cortical Screw, D=4.5mm, Non-Locking
- Cortical Screw, D=4.5mm, Locking
## INDICATION FOR USE:
The I.T.S. PRS Phoenix II is indicated to stabilize one or more pelvic bone fractures in the pelvis in an adult patient. Indications for use of the I.T.S. PRS Phoenix II include:
- Symphyseal Disruptions & Parasymphyseal Fractures
- Disruptions of the Sacroiliac Joint
- Fractures involving the Anterior Column of the Acetabulum
- Fractures involving the Quadrilateral Surface
- Fractures involving the Posterior Wall
- Fractures involving the Posterior Column
- Dorsal neutralization plating for Posterior Pelvic Ring Fractures
- Fractures of the Acetabulum, Fractures of the Pelvic Ring, Fractures of the Ilium
- Fracture Revision surgery of pseudoarthrosis, non-unions and mal-unions
The I.T.S. PRS Phoenix II system is not intended for spinal use
## CLINICAL TESTING:
Clinical data was not required for this submission.
## NON-CLINICAL TESTING:
The following non-clinically tests were performed:
- FEA-simulation according to ASTM F382 to test that the subject devices can withstand the same or more load than the cleared device.
- Screws were tested according to ASTM F543 and compared to the guideline: "Orthopedic Non-Spinal Metallic Bone Screws and Washers – Performance Criteria for Safety and Performance Based Pathway", issued on November 22, 2024
- On worst case products: Bioburden-, Sterility-, (both according to ISO 11137-1, ISO 11137-2, ISO, 11137-3) Endotoxin- (Ph. Eur. 2.6.14 / USP <85>) and Cytotoxicity-testing (ISO 10993-5) was performed to ensure sterility and an acceptable endotoxin values.
- RF-heating (according to ASTM F2182-19e2), magnetically induced displacement Force (according to ASTM F2052 - 21), magnetically induced Torque (according to ASTM F2213-17) and image Artifact evaluation and testing were performed according to ASTM F2213-17 (according to ASTM F2119 - 24) with worst case products. The subject device was compared to the evaluated worst-case products for torque, force and image artefacts. For RF-heating to products were compared to
{6}
K260054 - Page 3/3
# 510(k) Submission
## I.T.S. GmbH
### I.T.S. PRS Phoenix II
similar geometry and size products to prove they don't present the worst case.
Testing demonstrated that the I.T.S. PRS Phoenix II is substantially equivalent in mechanical performance to the predicate device I.T.S. Pelvic Reconstruction System PRS RX & PRS Phoenix (K210935)
## SUMMARY OF TECHNOLOGICAL CHARACTERISTICS:
The I.T.S.PRS Phoenix II is **substantially equivalent** in material (all plates are made from made from the same material (Titanium according to ASTM F67 and ASTM F136, Type II anodized), geometry and design (anatomically shaped plates with similar characteristics), indications and operational principles to the primary predicate systems legally marketed in the US listed above.
## CONCLUSIONS:
Based on the **similarity** in material, geometry, design, indications and operational principles, as well as both the Engineering Analysis and Performance Testing, the I.T.S.PRS Phoenix II has been demonstrated to be **substantially equivalent** (SE) to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.