cobas Strep A Nucleic acid test for use on the cobas Liat System (07341911190); cobas Strep A Quality Control Kit (07402678190)
K260047 · Roche Molecular Systems, Inc. · PGX · Aug 28, 2026 · Microbiology
Device Facts
Record ID
K260047
Device Name
cobas Strep A Nucleic acid test for use on the cobas Liat System (07341911190); cobas Strep A Quality Control Kit (07402678190)
Applicant
Roche Molecular Systems, Inc.
Product Code
PGX · Microbiology
Decision Date
Aug 28, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.2680
Device Class
Class 2
Indications for Use
The cobas Strep A nucleic acid test for use on the cobas Liat System (cobas Strep A) is a qualitative in vitro diagnostic test for the detection of Streptococcus pyogenes (Group A β-hemolytic Streptococcus, Strep A) in throat swab specimens from patients with signs and symptoms of pharyngitis. The cobas Strep A assay utilizes nucleic acid purification and polymerase chain reaction (PCR) technology to detect Streptococcus pyogenes by targeting a segment of the Streptococcus pyogenes genome.
Device Story
The cobas Strep A test is a rapid, automated in vitro diagnostic assay for qualitative detection of Streptococcus pyogenes DNA in throat swab specimens. The system uses the cobas Liat Analyzer and single-use disposable assay tubes containing reagents for sample purification, nucleic acid extraction, and real-time PCR amplification. The user scans barcodes on the assay tube and sample, transfers the sample to the tube, and inserts it into the analyzer. The system automates all processing steps, including target enrichment and inhibitor removal. The analyzer provides qualitative results (Strep A Detected/Not Detected) on its display. The device is intended for clinical use to aid in the diagnosis of pharyngitis. This submission updates the system to software version 4.0, incorporating the Roche Kinetic Algorithm v2 (KA2) and enabling optional wireless network connectivity.
Clinical Evidence
No new clinical trials were conducted. Performance was established by re-analyzing raw fluorescence data from the original clinical study (K141338) using the updated SASA v1.40.0 script. The re-analysis confirmed that clinical sensitivity and specificity remained unchanged and consistent with existing performance claims.
Technological Characteristics
The system utilizes TaqMan probe-based real-time PCR. It consists of the cobas Liat Analyzer and single-use disposable assay tubes. Connectivity includes wired Ethernet and optional wireless (via USB Wi-Fi adapter). The software architecture uses gLASP (generic Liat Assay Specific Packages) and the Roche Kinetic Algorithm v2 (KA2). The system performs automated sample purification, nucleic acid extraction, and amplification.
Indications for Use
Indicated for patients with signs and symptoms of pharyngitis to detect Streptococcus pyogenes (Group A β-hemolytic Streptococcus) in throat swab specimens.
Regulatory Classification
Identification
A Streptococcus spp. nucleic acid-based assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify various Streptococcus spp. nucleic acids extracted directly from clinical specimens. The device detects specific nucleic acid sequences for organism identification. The identification aids in the diagnosis of diseases caused by bacteria belonging to the genus Streptococcus and provides epidemiological information on these diseases. Pathogenic streptococci are associated with infections, such as sore throat, impetigo (an infection characterized by small pustules on the skin), urinary tract infections, rheumatic fever, and kidney disease.
Special Controls
(b) Classification. Class II (special controls). The special controls for this device are: 1) Premarket notification submissions must include detailed device description documentation, including the device components, ancillary reagents required but not provided, and a detailed explanation of the methodology including primer/probe sequence, design, and rationale for sequence selection. 2) Premarket notification submissions must include detailed documentation from the following analytical and clinical performance studies: Analytical sensitivity (Limit of Detection), reactivity, inclusivity, precision, reproducibility, interference, cross reactivity, carry-over, and cross contamination. 3) Premarket notification submissions must include detailed documentation from a clinical study. The study, performed on a study population consistent with the intended use population, must compare the device performance to results obtained from well-accepted reference methods. 4) Premarket notification submissions must include detailed documentation for device software, including, but not limited to, software applications and hardwarebased devices that incorporate software. 5) Premarket notification submissions must include database implementation methodology, construction parameters and quality assurance protocols, as appropriate. 6) The device labeling must include limitations regarding the need for culture confirmation of negative specimens, as appropriate. 7) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling. 8) Premarket notification submissions must include details on an end user device training program that will be offered while marketing the device, as appropriate.
*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed device description documentation, including the device components, ancillary reagents required but not provided, and a detailed explanation of the methodology including primer/probe sequence, design, and rationale for sequence selection.
(2) Premarket notification submissions must include detailed documentation from the following analytical and clinical performance studies: Analytical sensitivity (Limit of Detection), reactivity, inclusivity, precision, reproducibility, interference, cross reactivity, carry-over, and cross contamination.
(3) Premarket notification submissions must include detailed documentation from a clinical study. The study, performed on a study population consistent with the intended use population, must compare the device performance to results obtained from well-accepted reference methods.
(4) Premarket notification submissions must include detailed documentation for device software, including, but not limited to, software applications and hardware-based devices that incorporate software.
(5) Premarket notification submissions must include database implementation methodology, construction parameters, and quality assurance protocols, as appropriate.
(6) The device labeling must include limitations regarding the need for culture confirmation of negative specimens, as appropriate.
(7) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling.
(8) Premarket notification submissions must include details on an end user device training program that will be offered while marketing the device, as appropriate.
Predicate Devices
cobas Strep A Nucleic acid test for use on the cobas Liat System (K200065)
Submission Summary (Full Text)
{0}
[LOGO]
FDA
U.S. FOOD & DRUG
ADMINISTRATION
### 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY AND INSTRUMENT
### I Background Information:
### A 510(k) Number
K260047
### B Applicant
Roche Molecular Systems, Inc.
### C Proprietary and Established Names
cobas Strep A Nucleic acid test for use on the cobas Liat System (07341911190); cobas Strep A Quality Control Kit (07402678190)
### D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| PGX | Class II | 21 CFR 866.2680 -Streptococcus spp. nucleic acid-based assay | MI - Microbiology |
| OOI | Class II | 21 CFR 862.2570 – Instrumentation for clinical multiplex test systems | CH - Clinical Chemistry |
### II Submission/Device Overview:
### A Purpose for Submission:
To demonstrate that the cobas Strep A Nucleic acid test run on a modified cobas Liat System with software version 4.0 is substantially equivalent to the previously cleared predicate device (i.e., K200065)
The updated cobas liat Analyzer Software (v4.0) incorporates the following changes:
- (a) Assay-related Changes: (i) update of cobas Strep A assay-specific script from the legacy “SASA v1.29.0” to “SASA v1.4.0 generic Liat Assay Specific Packages (gLASP) assay script” and (ii) replacement of the legacy result calculation algorithm with a new Roche Kinetic Algorithm v2 (“KA2,” incorporated in the updated cobas Strep A assay script).
- (b) Analyzer instrument-related Changes:(i) OS upgrade and (ii) enablement of optional wireless network connectivity.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
# **B Measurand:**
Group A *Streptococcus* DNA
# **C Type of Test:**
Real-time PCR assay for qualitative detection of Group A *Streptococcus* DNA in throat swab specimens.
# **III Intended Use/Indications for Use:**
# **A Intended Use(s):**
See Indications for Use below.
# **B Indication(s) for Use:**
The cobas Strep A nucleic acid test for use on the cobas Liat System (cobas Strep A) is a qualitative *in vitro* diagnostic test for the detection of *Streptococcus pyogenes* (Group A β-hemolytic *Streptococcus*, Strep A) in throat swab specimens from patients with signs and symptoms of pharyngitis.
The cobas Strep A assay utilizes nucleic acid purification and polymerase chain reaction (PCR) technology to detect *Streptococcus pyogenes* by targeting a segment of the *Streptococcus pyogenes* genome.
# **C Special Conditions for Use Statement(s):**
Rx – For Prescription Use Only
# **D Special Instrument Requirements:**
cobas Liat Analyzer with software version 4.0
# **IV Device/System Characteristics:**
# **A Device Description:**
The cobas Strep A test is a rapid, automated *in vitro* diagnostic test for the qualitative detection of *Streptococcus pyogenes* (Group A β-hemolytic *Streptococcus*, Strep A) DNA in throat swab specimens in Amies medium. The test is performed on the cobas Liat System which automates and integrates sample purification, nucleic acid amplification, and detection of the target sequence in biological samples using TaqMan probe-based real-time PCR assays. The system consists of an instrument and preloaded software for running tests and viewing the results. The system requires the use of a single-use disposable cobas Strep A assay tube that holds all the sample purification and PCR reagents and hosts the sample preparation and PCR processes. The test targets a well-conserved region of the spy1258 gene of Strep A. An internal control (IC) is also included. The IC is present to control for adequate processing of the target bacteria and to monitor the presence of inhibitors in the sample preparation and PCR.
K260047 - Page 2 of 8
{2}
# **B Principle of Operation:**
The cobas Strep A test is designed to identify and/or measure the presence of *Streptococcus pyogenes* by targeting a segment of the *Streptococcus pyogenes* genome. The test runs on the cobas Liat System, which consists of the cobas Liat Analyzer (hardware with integrated software) and single-use disposable assay tubes that contain all necessary sample preparation and PCR reagents in separate segments divided by frangible seals. The system fully automates all nucleic acid amplification test processes, including reagent preparation, target enrichment, inhibitor removal, nucleic acid extraction, amplification, real-time detection, and result interpretation in a rapid manner.
# **C Instrument Description Information:**
# 1. Instrument Name:
cobas Liat Analyzer with cobas Liat Software v4.0
# 2. Specimen Identification:
The cobas Liat System maintains positive identification of each sample and assay tube during processing and analysis by means of barcode labels.
Before starting a run, the user is required to scan the assay tube barcode and then scan the sample barcode. The user then transfers the sample into the assay tube and is required to scan the assay tube barcode again before inserting the assay tube into the analyzer and starting the run
# 3. Specimen Sampling and Handling:
Throat swab specimens are collected using a Liquid Amies Collection and Transport System, which includes a sterile swab and a tube containing 1 mL of Amies liquid transport medium. With the patient seated and mouth fully open, the swab is rubbed along the back of the throat, tonsils, and any visible white patches, taking care to avoid contact with the tongue, cheeks, teeth, or gums. The swab is then inserted into the transport medium tube, swirled three times against the inner wall, broken off, and capped securely. Once collected, specimens should be added to the cobas Strep A assay tube and processed on the cobas Liat Analyzer as soon as possible. If immediate testing is not feasible, specimens are stable for up to 48 hours either refrigerated at 2–8°C (preferred) or at room temperature (20–25°C), with refrigeration also recommended during transport in accordance with applicable biological agent transport regulations.
# 4. Calibration:
The Cobas Liat Analyzer hardware periodically performs automatic recalibration, and the Cobas Liat System does not run the assay until this auto-calibration step is performed.
# 5. Quality Control:
The following controls are used to validate the assay run in the cobas Liat Analyzer.
a. Internal Process control (IPC): Each Liat Tube contains a chemically-inactivated bacterium that is included to verify each step of Strep A sample processing before the PCR runs and to detect potential specimen-associated inhibition in the sample preparation or the PCR. The IPC should be positive in a negative sample and can be
K260047 - Page 3 of 8
{3}
negative or positive in a Strep A positive sample. The IPC is valid if it meets the acceptance criteria. An invalid result will be generated if the IPC fails to meet the acceptance criteria.
b. Positive Control: The Positive Control is a unit-dose of dried chemically-inactivated Strep A. The Report Result must be “Strep A Detected” for the positive control to pass. The positive control validates that the PCR reagents are working as expected.
c. Negative Control: A dilution Amies tube is used as the negative control sample. The Report Result must be “Strep A Not Detected” for the negative control to pass. The negative control is intended to monitor potential target contamination in the workflow or environment.
# V Substantial Equivalence Information:
# A Predicate Device Name(s):
cobas Strep A Nucleic Acid Test for use on the cobas Liat System
# B Predicate 510(k) Number(s):
K200065
# C Comparison with Predicate(s):
| Device & Predicate Device(s): | Device: K260047 | Predicate: K200065 |
| --- | --- | --- |
| Device Trade Name | cobas Strep A Nucleic Acid Test for use on the cobas Liat System | Same |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The cobas Strep A nucleic acid test for use on the cobas Liat System (cobas Strep A) is a qualitative in vitro diagnostic test for the detection of Streptococcus pyogenes (Group A β-hemolytic Streptococcus, Strep A) in throat swab specimens from patients with signs and symptoms of pharyngitis. The cobas Strep A assay | Same |
K260047 - Page 4 of 8
{4}
| | utilizes nucleic acid purification and polymerase chain reaction (PCR) technology to detect Streptococcus pyogenes by targeting a segment of the Streptococcus pyogenes genome. | |
| --- | --- | --- |
| Regulation | 21 CFR 866.2680 | Same |
| Product code | PGX | Same |
| Analyte | Group A Streptococcus pyogenes | Same |
| Target | Conserved region of Group A Streptococcus genome | Same |
| Sample Type | throat swab | Same |
| Instrument | cobas Liat Analyzer | Same |
| General Device Characteristic Differences | | |
| Software | cobas Liat Analyzer Software 4.0 | cobas Liat Analyzer Software 3.3 |
| Assay Script | Generic Liat Assay Specific Package (gLASP) SASA v1.40.0 | legacy IQuum Command Language (ICL) assay script (SASA v1.29) |
| Internet connectivity | Optional wireless connectivity via an off-the-shelf (OTS) USB Wi-Fi adapter is available in addition to the existing wired Ethernet connection | Only wired Ethernet connection is available |
### VI Standards/Guidance Documents Referenced:
Not applicable
### VII Performance Characteristics (if/when applicable):
#### A Analytical Performance:
1. Precision/Reproducibility:
Not applicable.
K260047 - Page 5 of 8
{5}
2. Linearity:
Not applicable.
3. Analytical Cross-reactivity/Interference:
Raw fluorescence data from the original Cross-reactivity/Interference studies (conducted to support K141338, the original clearance of the cobas Strep A Assay, which runs on ICL legacy script v.1.29) was recalculated using the updated Roche KA2 script (SASA v1.4.0 generic Liat Assay Specific Packages (gLASP) assay script). The re-analysis confirmed that the analytical cross-reactivity and interference remained consistent with the performance claims currently described in the Instructions for Use (IFU).
4. Detection Limit and Assay Reportable Range:
Previously obtained raw fluorescence data from the original limit of detection (LoD) study (conducted to support K141338, the original clearance of the cobas Strep A Assay, which runs on ICL legacy script v.1.29) was recalculated using the updated Roche KA2 script (SASA v1.4.0 generic Liat Assay Specific Packages (gLASP) assay script). The re-analysis confirmed that the analytical sensitivity remained consistent with the performance claims currently described in the Instructions for Use (IFU).
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Reanalysis of raw florescence data from the original shelf-life stability study was conducted with the updated script/algorithm, and no changes to the previous results (shelf-life of 24 months from the date of manufacture) were observed.
6. Assay Cut-Off:
Reanalysis of raw florescence data from the original Cut-off study was conducted with the updated script/algorithm, and no changes to the previous results were observed.
# B Comparison Studies:
1. Method Comparison with Predicate Device:
To demonstrate performance equivalency of the cobas Strep A nucleic acid test following migration from the legacy IQuum Command Language (ICL) assay script (SASA v1.29, running on software v3.5.1) to the generic Liat Assay Specific Package (gLASP) architecture (SASA v1.40.0, running on software v4.0.0), an equivalency study was conducted using ten cobas liat Analyzers with contrived samples prepared at ~2x LOD, and ~5x LOD concentrations, and with positive and negative controls. A total of 160 runs were performed (Table 1), of which 158 yielded valid results; one aborted run and one invalid result (both with the ICL script, and gLASP script) were repeated per protocol and produced valid results. All negative controls and negative samples correctly returned "Not Detected" results (0% hit rate) for both script versions. At ~2x LOD, detection rates were 97.5% for ICL and 95.0% for gLASP, both meeting the ≥95% acceptance criterion, while at ~5x LOD, both versions achieved a 100% detection rate. No deviations from the study protocol were observed. The study concluded that the migration to gLASP SASA v1.40.0 had no impact on assay performance, confirming equivalency between the two script versions.
Table 1: Equivalency Study Results Summary
K260047 - Page 6 of 8
{6}
| Test sample | ICL SASA Script v1.1.0 (legacy) | | | | SASA Script v1.4.0 (gLASP) | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Replicates | Target detected | Target not detected | Hit rate | Replicates | Target detected | Target not detected | Hit rate |
| Negative Control | 10 | 0 | 10 | 0% | 10 | 0 | 10 | 0% |
| Positive Control | 10 | 10 | 0 | 100% | 10 | 10 | 0 | 100% |
| Negative Sample | 10 | 0 | 10 | 0% | 10 | 0 | 10 | 0% |
| ~2× LOD (positive) Sample | 40 | 39 | 0 | 97.5% | 40 | 38 | 0 | 95% |
| ~5× LOD (positive) Sample | 10 | 10 | 0 | 100% | 10 | 10 | 0 | 100% |
#### Footnotes:
1. All 160 runs finally yielded valid results.
2. There were no deviations observed from the predetermined Equivalency Study Protocol.
3. All acceptance criteria outlined in Equivalency Study Protocol for cobas Liat assay's migration to gLASP Architecture were met.
#### 2. Matrix Comparison:
Not applicable.
### C Clinical Studies:
#### 1. Clinical Sensitivity:
Raw fluorescence data from the previous clinical study (conducted to support the original clearance of Strep A Assay, K141338) was recalculated using SASA v1.40.0. The re-analysis confirmed that clinical sensitivity and specificity remained unchanged and the performance claims in the Instructions for Use (IFU) remained supported.
#### 2. Clinical Specificity:
Please refer to section C1 above.
#### 3. Clinical Cut-Off
Not applicable
#### 4. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable
### D Expected Values/Reference Range:
Not applicable
### E Other Supportive Instrument Performance Characteristics Data:
EMC: Electrical safety and electromagnetic compatibility (EMC) testing were performed; the information was reviewed and was found to be acceptable.
Software and Cybersecurity: Software and cybersecurity documentation was reviewed and found to be acceptable.
K260047 - Page 7 of 8
{7}
### VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
### IX Conclusion:
The submitted information in this premarket notification is complete and a substantial equivalence decision.
K260047 - Page 8 of 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.