The iNOS intraoral biosensor is intended to reduce nighttime snoring and mild to moderate obstructive sleep apnea (OSA) in adults. Additionally, the fully embedded oximeter sensor measures, displays, stores and transmits functional oxygen saturation of the arterial hemoglobin (SpO2). It is intended for continuous data collection during sleep. It can be used in sleep labs, long term care, hospitals and home use. The device is intended to be applied in the oral cavity. The device is intended for use under non-motion conditions in well perfused areas. The device is single patient, reusable.
Device Story
Intraoral biosensor; hermetically sealed pulse oximeter within maxillary/mandibular arch appliance. Measures red, green, infrared light absorption through oral soft tissue; calculates functional SpO2 based on light absorption ratios. Includes 6-axis IMU and 3.7V Li-ion battery. Used in sleep labs, hospitals, home; operated by patient. Data transmitted via encrypted BLE to companion Android app; app displays, stores, exports data. Healthcare professionals review app data to inform clinical decisions regarding sleep quality and OSA management. Benefits include continuous, non-invasive sleep monitoring integrated with snoring/OSA therapy.
Clinical Evidence
Bench testing only. Pulse oximetry accuracy evaluated in 25 patients across diverse demographics using co-oximetry reference. Compliance with ISO 80601-2-61, IEC 60601-1, IEC 60601-1-11, and ISO 10993 series (cytotoxicity, irritation, sensitization, systemic toxicity, genotoxicity, implantation). Wireless coexistence verified per ANSI C63.27.
Technological Characteristics
Medical grade copolyester/adhesive construction. Sensing: multi-channel PPG and 6-axis IMU. Energy: 3.7V 30mAh rechargeable Li-ion battery. Form factor: 40mm x 10mm x 2mm. Connectivity: BLE (2.4 GHz) with AES128 encryption. Ingress protection: IPx2. Non-sterile. Standards: IEC 60601-1, IEC 60601-1-2, ISO 80601-2-61, ISO 10993.
Indications for Use
Indicated for adults with mild to moderate obstructive sleep apnea (OSA) and nighttime snoring. Used for continuous SpO2 monitoring during sleep in clinical or home settings. Contraindicated for use under motion conditions or in poorly perfused areas.
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{0}
**FDA U.S. FOOD & DRUG**
ADMINISTRATION
August 25, 2026
Achaemenid, LLC
% Joseph Azary
Regulatory Consultant
Aztech Regulatory & Quality, LLC
543 Long Hill Ave.
Shelton, Connecticut 06484
Re: K254304
Trade/Device Name: iNOS intraoral biosensor
Regulation Number: 21 CFR 872.5570
Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea
Regulatory Class: Class II
Product Code: PLC, DQA, OUG, LQZ, LRK
Dated: July 29, 2026
Received: July 29, 2026
Dear Joseph Azary:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254304 - Joseph Azary
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K254304 - Joseph Azary
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and
ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K254304
Device Name
iNOS intraoral biosensor
Indications for Use (Describe)
The iNOS intraoral biosensor is intended to reduce nighttime snoring and mild to moderate obstructive sleep apnea (OSA) in adults.
Additionally, the fully embedded oximeter sensor measures, displays, stores and transmits functional oxygen saturation of the arterial hemoglobin (SpO2). It is intended for continuous data collection during sleep. It can be used in sleep labs, long term care, hospitals and home use.
The device is intended to be applied in the oral cavity. The device is intended for use under non-motion conditions in well perfused areas. The device is single patient, reusable.
Type of Use (Select one or both, as applicable)
☑
Prescription Use (Part 21 CFR 801 Subpart D)
☐
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
{4}
K254304
# 510(k) Summary
## iNOS INTRAORAL BIOSENSOR
# 1. SUBMITTER/510(k) HOLDER
Achaemenid, LLC
75 State Street
Suite 100
Boston, MA 02109
Contact Name: Joseph Azary (Aztech Regulatory & Quality LLC)
Email: jazary@aztechregulatory.com (alternate email) jazary@erols.com
Telephone: (203) 242-6670
Date: August 21, 2026
# 2. DEVICE NAME
| Proprietary Name: | iNOS Intraoral Biosensor |
| --- | --- |
| Common/Usual Name: | Intraoral device for snoring and/or obstructive sleep apnea with patient monitoring |
| Classification Name: | Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea |
| Classification Regulation: | 21 CFR 872.5570 |
| Classification Product code: | PLC |
| Classification: | Class 2 |
| Medical Specialty (Panel): | Division of Dental and ENT devices |
| Secondary Product Codes: | DQA, LQZ, LRK, OUG |
# 3. PREDICATE DEVICES
- Primary Predicate Device – Achaemenid – RADx Intraoral Appliance – K230532
- Secondary Predicate Device – ProSomnus RPMO2 OSA Device – K252765
# 4. DEVICE DESCRIPTION
The iNOS Intraoral Biosensor is a medical grade oximeter hermetically sealed inside an upper or lower arch oral appliance.
Achaemenid obtained 510(k) clearance for an intraoral appliance used for snoring and obstructive sleep apnea. The company used that design concept and materials to create an intraoral biosensor to be used as a pulse oximeter.
{5}
The iNOS Intraoral Biosensor is placed in the patient's mouth.
The device is used for oximetry reading during sleep from the oral soft tissue structure, when worn by an individual. The iNOS Intraoral Biosensor tracks blood oxygen saturation or SpO2 as a way to track sleep quality.
The device is paired with a smart electronic device that has the iNOS app. The data is transmitted via Bluetooth to the smart electronic device so that the data can be reviewed.
## Principles of Operation
The device measures red, green and infrared light through perfused tissue to detect fluctuating signals caused by arterial pulses. The pulse oximeter determines functional oxygen saturation of arterial hemoglobin from the color difference by measuring the ratio of absorbed red, green and infrared light as volume fluctuates with each pulse.
## Battery
The battery is a 3.7V lithium ion battery with a 30mAh capacity. A charged battery can be used for 8 hours. The battery and sensor are hermetically sealed within the oral appliance and cannot be removed or replaced.
## Physical Aspects
The iNOS Intraoral Biosensor is approximately 40mm x 10mm x 2mm in size to allow comfortable usage within the mouth.
## Electrical / Software Aspects
The iNOS Intraoral Biosensor includes an internal batter powered sensor to measure pulse oximetry. The device is low power and complies with applicable electrical safety requirements including IEC 60601-1, IEC 60601-1-2, IEC 60601-1-2 and ISO 80601-2-61.
The battery can be recharged.
The sensor has firmware installed and an internal antenna. The sensor communicates with the app that is installed on a smart electronic device.
The app on the smart electronic device receives and decrypts the transmitted information to allow display.
{6}
## Wireless Features
The Bluetooth transmissions are encrypted using ICS Certification AES128 encryption.
The Bluetooth operates in the 2.4 GHz ISM band (2400 to 2483.5 MHz). The Bluetooth utilizes Adaptive Frequency Hopping (AFH) as an inherent coexistence mechanism.
## Functionality with Smartphones and Tablets
The application gets sent to the customers after ordering a device. The app is installed on a smart electronic device and then gets paired with the iNOS Intraoral Biosensor.
## Description of the App
The App is used to display pulse oximetry measurements and it is up the Healthcare Professional to review, analyze and determine how to use that data.
## Built In Safety Features
The iNOS Intraoral Biosensor is a low power device that meets applicable electrical safety standards.
The device also meets the ingress protection requirements outlined in IPx2.
The materials used for the device are identical to the materials used in the RADx Intraoral Appliance cleared under K230532. These materials have been tested to biocompatibility test requirements.
## Biocompatibility
Biocompatibility was addressed in accordance with FDA's guidance document "Use of International Standard ISO 10993-1, Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process" and ISO 7405 "Dentistry – Evaluation of Biocompatibility of Medical Devices used in Dentistry".
The materials in contact with the patient for the subject device are identical to the materials and processes (e.g. housing material & fabrication, post-print processing / cleaning agents, mold release agents and housing cure / post-cure) used in the RADx Intraoral Appliance cleared under K230532. The encapsulation process does not introduce biological hazards when compared to the primary and secondary predicate devices. Therefore, the biocompatibility of the subject device is supported by the biocompatibility information for the primary predicate device.
{7}
# Accuracy
The accuracy of the pulse oximeter was evaluated at University of California San Francisco, where it was found to provide accurate results when tested with 25 patients with different genders, age groups and racial backgrounds.
# Software and Cybersecurity
Software documentation was provided in accordance with the FDA guidance “Content of Premarket Submissions for Device Software Functions”.
The iNOS device contains firmware that acquires multi-channel PPG and 6-axis IMU raw sensor data continuously during sleep monitoring sessions. Sensor data is transmitted to the companion Android application via encrypted BLE in real time. The firmware stores sensor data to an on-board flash memory during BLE disconnection events, provide up to 30 minutes of local buffering.
The iNOS Android Application is the companion software for the device. It provides BLE based connection, real time physiological signal display, signal processing algorithms (i.e. SpO2) and local data storage and export. The software falls under Class A.
Software Verification and Validation was conducted with passing results.
The submission includes cybersecurity documentation in accordance with FDA guidance “Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions” including cybersecurity testing (vulnerability assessment and penetration testing). The device meets the requirements of section 524(b) of the Food Drug & Cosmetic (FD&C) Act.
{8}
# 5. INDICATIONS FOR USE
The iNOS intraoral biosensor is intended to reduce nighttime snoring and mild to moderate obstructive sleep apnea (OSA) in adults.
Additionally, the fully embedded oximeter sensor measures, displays, stores and transmits functional oxygen saturation of the arterial hemoglobin (SpO2). It is intended for continuous data collection during sleep. It can be used in sleep labs, long term care, hospitals and home use.
The device is intended to be applied in the oral cavity. The device is intended for use under non-motion conditions in well perfused areas. The device is single patient, reusable.
# 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The similarities between the devices are as follows:
- All devices are indicated to reduce snoring and mild to moderate obstructive sleep apnea.
- The subject device and secondary predicate also include functionality to measure oxygen saturation.
- All devices are used in the oral cavity
- All devices are custom maxillary and mandibular arches composed of medical grade copolyester and adhesive.
- The subject device and secondary predicate collect data continuously during sleep.
- The subject device and secondary predicate both have SpO2 accuracy of 70-100%.
- The subject device and secondary predicate use Rechargeable Lithium Ion Batteries and meet IEC electrical safety standards.
- All of the devices are non-sterile
- The subject device and secondary predicate have wireless capabilities using Bluetooth Low Energy (BLE) with a range of 10 meters.
- All devices are single patient use, reusable.
The main differences are identified as:
- The primary predicate does not have a sensor to measure SpO2 and is not battery powered.
- The subject device was testing to IP22 for ingress protection whereas the secondary predicate was tested to IP68 for device and IP22 for charger.
- The subject device has a rechargeable battery that lasts approximately 8 hours
{9}
which is appropriate as the patient wears it while sleeping, whereas the secondary predicate has battery life of 4+ hours of continuous recording per session.
The main differences are not deemed significant based on the testing to electrical safety standards including IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11 and ISO 80601-2-61, as well as the oximetry accuracy study.
## 7. PERFORMANCE TESTING
The iNOS Intraoral Biosensor was subjected to a variety of testing including:
- Biocompatibility Evaluation per ISO 10993-1
- Electrical Safety Testing
- Performance of Pulse Oximeters per ISO 80601-2-61
- Software Verification and Validation
- Pulse Oximetry Accuracy
| Test Name | Test Description and acceptance criteria | Remarks - Conclusions |
| --- | --- | --- |
| Electromagnetic Disturbance (EMC) Testing | Testing for compliance to acceptance criteria in the EMC standard IEC 60601-1-2:2014 / AMD1:2021. The testing included mains terminal disturbance voltage, electromagnetic radiation disturbance, harmonic current emissions, voltage fluctuations, Flicker, and Immunity. | PASS All testing included as part of the EMC testing including immunity, voltage fluctuations, electromagnetic disturbance, harmonic current emissions all passed. |
| Electrical Safety | Testing for compliance to IEC 60601-1:2005/AMD2:2020 Testing included ingress, determination of accessible parts, rough surfaces/sharp corners, excessive temperature, humidity pretreatment, dielectric strength, push, impact, drop test, and mold stress relief. | PASS The testing concluded that the device complied with applicable portions of the IEC 60601-1 standard. |
{10}
| Test Name | Test Description and acceptance criteria | Remarks - Conclusions |
| --- | --- | --- |
| Electrical Safety - Home Healthcare Environment | Testing for compliance to IEC 60601-1-11:2015/AMD1:2020 | PASS The testing concluded that the device complied with applicable portions of the IEC 60601-1-11 standard. |
| Basic Safety and Essential Performance of Pulse Oximeter Equipment | Testing for compliance to ISO 80601-2-61:2017 and IEC 60529 Testing included ingress, settings and data storage after interruptions, faults and shock and vibration. | PASS The testing concluded that the device complied with applicable portions of the ISO 80601-2-61 and IEC 60529. Ref: 092-721010024-000 |
| Wireless Coexistence Testing | Testing to ensure the device will function properly in an environment with other electronic devices in use. | PASS The iNOS intraoral biosensor met the recommended test level as defined in ANSI C63.27 |
| Cytotoxicity | Testing for compliance to ISO 10993-5 | PASS The device was non-cytotoxic. Meets requirements. |
| Skin Irritation in Rabbits | Testing for compliance to ISO 10993-10 | PASS Non-Irritant. Meets requirements. |
| ISO Closed Patch Sensitization in Guinea Pigs | Testing for compliance to ISO 10993-10 | PASS Non-sensitizer and met requirements. |
| Acute Systemic Toxicity | Testing for compliance to USP 88 for Acute Systemic Toxicity | PASS No evidence of systemic toxicity and compliant with USP 88. |
| Implantation | Testing for compliance to USP 88 for Implantation. | PASS Macroscopic reaction was not significant as compared to the negative control article. |
| Systemic Toxicity | Testing for compliance to ISO 10993-11 | PASS The device was not toxic and was compliant to ISO 10993-11 |
| Genotoxicity | Testing for compliance to ISO 10993-3 | PASS The device was non-mutagenic and was compliant to ISO 10993-3. |
{11}
| Test Name | Test Description and acceptance criteria | Remarks - Conclusions |
| --- | --- | --- |
| Pulse Oximeter Testing Report | The testing included 25 patients. | PASS The testing was conducted using co-oximetry reference. |
# 8. CONCLUSION
Information presented supports substantial equivalence of the subject device to the predicate devices based on similarities.
The company believes that based on the indications for use, technological characteristics and comparison to predicate devices the subject device has been shown to be substantially equivalent to the predicate and the minor differences do not impact safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.