← Product Code [KNF](/productcode/KNF) · K254303

# KARL STORZ SUPRALOOP (K254303)

_Karl Storz SE & CO. KG · KNF · Sep 24, 2026 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K254303

## Device Facts

- **Applicant:** Karl Storz SE & CO. KG
- **Product Code:** [KNF](/productcode/KNF.md)
- **Decision Date:** Sep 24, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.4160
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Therapeutic

## Indications for Use

The KARL STORZ SUPRALOOP is a monopolar electrosurgical device intended for cutting and removal of soft tissue in gynecological procedures involving endoscopic hysterectomy in adult women. It is used with standard electrosurgical generators

## Device Story

Modular monopolar electrosurgical system for laparoscopic supracervical hysterectomy; consists of reusable handle, reusable 5 mm outer sheath, and sterile single-use monopolar loop electrode. Physician operates handle to advance/retract loop through sheath, opening/closing loop around target tissue. System connects to standard electrosurgical generator; HF energy activates loop to circumferentially cut and coagulate tissue; current returns via separate patient return electrode. Device used in clinical/OR settings. Benefits include mechanical tissue removal and coagulation during endoscopic procedures.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench verification testing, including electrode attachment/detachment force, system interlocking, bending force, HF tissue/durability testing, compression, and tensile strength. Biocompatibility, sterility, and electrical safety/EMC testing met all predefined acceptance criteria.

## Technological Characteristics

Modular system; reusable handle/sheath, single-use loop. Materials: NiTi, PTFE, stainless steel 1.4305, PPSU, ZrO2, PFA, silicone. Energy: Monopolar HF. Standards: IEC 60601-1, IEC 60601-2-2, IEC 60601-1-2, ISO 10993-1, ISO 11135:2014, ISO 11607-1:2019, ASTM F88/F88M-15. Sterilization: Steam (reusable components), EtO (single-use loops).

## Regulatory Identification

A unipolar endoscopic coagulator-cutter is a device designed to destroy tissue with high temperatures by directing a high frequency electrical current through the tissue between an energized probe and a grounding plate. It is used in female sterilization and in other operative procedures under endoscopic observation. This generic type of device may include the following accessories: an electrical generator, probes and electrical cables, and a patient grounding plate. This generic type of device does not include devices used to perform female sterilization under hysteroscopic observation.

## Predicate Devices

- LiNA Loop ([K070315](/device/K070315.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD & DRUG ADMINISTRATION

September 24, 2026

Karl Storz SE & CO. KG
Mario Trujillo
Regulatory Affairs Specialist
2151 E. Grand Avenue
El Segundo, CA 90245-5017

Re: K254303
Trade/Device Name: KARL STORZ SUPRALOOP
Regulation Number: 21 CFR§ 884.4160
Regulation Name: Unipolar Endoscopic Coagulator-Cutter and Accessories
Regulatory Class: II
Product Code: KNF
Dated: August 25, 2026
Received: August 25, 2026

Dear Mario Trujillo:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K254303 - Mario Trujillo

Page 2

Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

{2}

K254303 - Mario Trujillo

Page 3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# JASON ROBERTS -S

Jason R. Roberts, Ph.D.

Assistant Director

DHT3B: Division of Reproductive,

Gynecology, and Urology Devices

OHT3: Office of Gastrorenal, ObGyn,

General Hospital, and Urology Devices

Office of Product Evaluation and Quality

Center for Devices and Radiological Health

Enclosure

{3}

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K254303 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  KARL STORZ SUPRALOOP  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  The KARL STORZ SUPRALOOP is a monopolar electrosurgical device intended for cutting and removal of soft tissue in gynecological procedures involving endoscopic hysterectomy in adult women. It is used with standard electrosurgical generators  |   |   |
|  Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |

{4}

STORZ
KARL STORZ — ENDOSKOPE

KARL STORZ Premarket Notification

KARL STORZ Supraloop

510(k) Summary

### 510(k) Summary

This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92 and the FDA guidance document titled "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]" issued on July 28, 2014. All data included in this document is accurate and complete to the best of KARL STORZ SE & Co. KG knowledge.

|  Submitter: | KARL STORZ SE & Co. KG Dr.-Karl-Storz-Straße 34 78532 Tuttlingen, Germany  |
| --- | --- |
|  Contact: | Mario Trujillo Regulatory Affairs Specialist Tel.: (424) 218-8481 Email: Mario.Trujillo@karlstorz.com  |
|  Date of Preparation: | September 23, 2026  |
|  Type of 510(k) Submission: | Traditional  |
|  Device Identification: | Trade Name: KARL STORZ SUPRALOOP Classification Name: Coagulator-Cutter, Endoscopic, Unipolar (And Accessories) (21 CFR Part 884.4160);  |
|  Regulatory Class: | II  |
|  Product Code: | KNF  |
|  Guidance Document: | Not applicable  |
|  Predicate Devices: | LINA LOOP, MODELS: EL-160-4, EL-160-8, EL-200-4, EL-200-8 (K070315) The predicate has been subject to a design related recall. Due to the potential for the predicate device to break during a procedure, additional warnings were added to the instructions for use. KARL STORZ has adequately addressed the design related issues of the predicate through labeling and performance testing of the subject device.  |
|  Device Description: | The KARL STORZ SUPRALOOP is a modular monopolar electrosurgical system for laparoscopic supracervical hysterectomy. The system consists of a reusable handle (26183MA), a reusable 5 mm outer sheath with a 300 mm working length (26183MB), and a sterile, single-use monopolar loop electrode. Two loop sizes are available: 120 mm x 85 mm (26183MC-S) and 200 mm x 150 mm (26183MD-S). The patient-contacting materials of the loop include nickel-titanium alloy (NiTi), polytetrafluoroethylene (PTFE), stainless steel 1.4305, and black PTFE. The patient-contacting materials of the outer sheath include black polyphenylsulfone (PPSU), zirconium oxide (ZrO2), black perfluoroalkoxy alkane (PFA), and gray silicone The loop electrode is connected through the reusable handle and sheath to a commonly available electrosurgical generator with a monopolar output. When HF energy is activated, current is conducted through the loop electrode to circumferentially cut and coagulate tissue, with current returning through a separate patient return electrode. The handle advances and retracts the loop through the sheath to open and close the loop around the target tissue.  |
|  Indications For Use: | The KARL STORZ SUPRALOOP is a monopolar electrosurgical device intended for cutting and removal of soft tissue in gynecological procedures involving endoscopic hysterectomy in adult women. It is used with standard electrosurgical generators  |

{5}

STORZ
KARL STORZ—ENDOSKOPE

KARL STORZ Premarket Notification

KARL STORZ Supraloop

510(k) Summary

Substantial equivalence

Comparison Table: Subject vs. Predicate Device

|   | **Subject Device** KARL STORZ SUPRALOOP | **Predicate Device, K070315** LiNA Loop  |
| --- | --- | --- |
|  **Indication for Use**  |   |   |
|  **Indications for Use** | The KARL STORZ SUPRALOOP is a monopolar electrosurgical device intended for cutting and removal of soft tissue in gynecological procedures involving endoscopic hysterectomy in adult women. It is used with standard electrosurgical generators. | The LiNA Gold Loop is a 5mm monopolar electrosurgical device intended for cutting and removal of soft tissue in gynecological procedures involving endoscopic hysterectomy. It is used with standard electrosurgical generators.  |
|  **Product Code** | KNF | KNF  |
|  **Target Population** | Adult women | Adult women  |
|  **Physical & Technological Characteristics**  |   |   |
|  **Components** | Handle, outer sheath, loop | Handle, outer sheath, loop  |
|  **Product Assembly and disassembly** | Requires user assembly | Provided Assembled  |
|  **Sheath Working Length** | 300 mm | Information not available  |
|  **Sheath Total Length** | 413 mm | Information not available  |
|  **Outer Sheath Total Length** | 0.5 mm | 0.5 mm  |
|  **Loop Working Length** | 120mm x 85mm 200mm x 150mm | 160mm x 80mm 200mm x 100mm  |
|  **Loop Total Length** | 593 mm 680 mm | Information not available  |
|  **Method of Action** | The loop opens when the handle is pushed forward. The loop closed when the handle is pushed backwards. | The loop opens when the handle is pushed forward. The loop closed when the handle is pushed backwards.  |
|  **Energy Source** | Monopolar | Monopolar  |
|  **HF Generator Compatibility** | Compatible with all standard electrosurgical generators with monopolar outlet. | Compatible with all standard electrosurgical generators with monopolar outlet.  |
|  **Cleaning and Sterilization Methods**  |   |   |
|  **Reusable Components** | Handle, outer sheath | None  |
|  **Cleaning** | Manual | N/A  |
|  **Sterilization** | Steam, EtO | Steam, EtO  |
|  **Sterile Single-Use Components** | Loop | All  |
|  **Shelf Life** | Loops: 5 years | Information not publicly available  |

The subject and predicate devices have the same indications for use. Differences in design, dimensions, materials, assembly, and reprocessing were evaluated through bench, biocompatibility, sterility, packaging, shelf-life, electrical safety, and EMC testing. These differences do not raise different questions of safety or effectiveness.

{6}

STORZ
KARL STORZ—ENDOSKOPE

KARL STORZ Premarket Notification

KARL STORZ Supraloop

510(k) Summary

|  Non-Clinical Performance Data: | Bench verification performance testing has been performed for the following parameters: - Attaching and Detaching Force of the Electrode - System Interlocking Test - Bending Force - HF Tissue Test and HF Durability - Compression Testing - Tensile Strength Testing - Biocompatibility evaluation was performed to ISO 10993The results met the predefined acceptance criteria and support the mechanical integrity and functional performance of the subject deviceElectrical safety and electromagnetic compatibility (EMC)Electrical safety and EMC testing were provided in the submission. The system complies with the IEC 60601-1, IEC 60601-2-2 standards for safety and the IEC 60601-1-2 standard for EMC.BiocompatibilityThe loop electrode and outer sheath were evaluated as external communicating devices with tissue contact for a limited duration of less than 24 hours in accordance with ISO 10993-1. The biological evaluation included: cytotoxicity, sensitization, irritation or intracutaneous reactivity, acute systemic toxicity and material-mediated pyrogenicity. All applicable acceptance criteria were met.SterilityThe single-use loop electrodes are supplied sterile by ethylene oxide sterilization validated in accordance with ISO 11135:2014 to a sterility assurance level of 10^-6. EO and ECH residuals were evaluated in accordance with ISO 10993-7 and met the applicable allowable limits. A 5-year shelf life was supported by accelerated aging, climate conditioning, transportation simulation, seal-strength testing, package visual inspection, and package integrity testing. Packaging validation was performed in accordance with ISO 11607-1:2019, ASTM F88/F88M-15 and the bubble-leak method.The reusable handle and sheath are supplied non-sterile and require reprocessing prior to first use and in between uses. The submission includes validation of the cleaning and steam sterilization reprocessing instructions.  |
| --- | --- |
|  Clinical Performance Data: | Clinical testing was not required to demonstrate the substantial equivalence to the predicate device  |
|  Conclusion: | The results of the non-clinical performance testing support the substantial equivalence of the KARL STORZ SUPRALOOP to the legally marketed predicate device, LiNA Loop (K070315). The subject device has the same intended use and indications for use as the predicate. The identified technological differences do not raise different questions of safety or effectiveness, and the performance data demonstrate that the KARL STORZ SUPRALOOP is as safe and effective as the predicate device for its intended use.  |

---

**Source:** [https://fda-staging.innolitics.com/device/K254303](https://fda-staging.innolitics.com/device/K254303)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
