K254303 · Karl Storz SE & CO. KG · KNF · Sep 24, 2026 · Obstetrics/Gynecology
Device Facts
Record ID
K254303
Device Name
KARL STORZ SUPRALOOP
Applicant
Karl Storz SE & CO. KG
Product Code
KNF · Obstetrics/Gynecology
Decision Date
Sep 24, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.4160
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The KARL STORZ SUPRALOOP is a monopolar electrosurgical device intended for cutting and removal of soft tissue in gynecological procedures involving endoscopic hysterectomy in adult women. It is used with standard electrosurgical generators
Device Story
Modular monopolar electrosurgical system for laparoscopic supracervical hysterectomy; consists of reusable handle, reusable 5 mm outer sheath, and sterile single-use monopolar loop electrode. Physician operates handle to advance/retract loop through sheath, opening/closing loop around target tissue. System connects to standard electrosurgical generator; HF energy activates loop to circumferentially cut and coagulate tissue; current returns via separate patient return electrode. Device used in clinical/OR settings. Benefits include mechanical tissue removal and coagulation during endoscopic procedures.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench verification testing, including electrode attachment/detachment force, system interlocking, bending force, HF tissue/durability testing, compression, and tensile strength. Biocompatibility, sterility, and electrical safety/EMC testing met all predefined acceptance criteria.
Indicated for cutting and removal of soft tissue in endoscopic hysterectomy procedures in adult women.
Regulatory Classification
Identification
A unipolar endoscopic coagulator-cutter is a device designed to destroy tissue with high temperatures by directing a high frequency electrical current through the tissue between an energized probe and a grounding plate. It is used in female sterilization and in other operative procedures under endoscopic observation. This generic type of device may include the following accessories: an electrical generator, probes and electrical cables, and a patient grounding plate. This generic type of device does not include devices used to perform female sterilization under hysteroscopic observation.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 24, 2026
Karl Storz SE & CO. KG
Mario Trujillo
Regulatory Affairs Specialist
2151 E. Grand Avenue
El Segundo, CA 90245-5017
Re: K254303
Trade/Device Name: KARL STORZ SUPRALOOP
Regulation Number: 21 CFR§ 884.4160
Regulation Name: Unipolar Endoscopic Coagulator-Cutter and Accessories
Regulatory Class: II
Product Code: KNF
Dated: August 25, 2026
Received: August 25, 2026
Dear Mario Trujillo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254303 - Mario Trujillo
Page 2
Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K254303 - Mario Trujillo
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# JASON ROBERTS -S
Jason R. Roberts, Ph.D.
Assistant Director
DHT3B: Division of Reproductive,
Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K254303 | ? |
| Please provide the device trade name(s). | | ? |
| KARL STORZ SUPRALOOP | | |
| Please provide your Indications for Use below. | | ? |
| The KARL STORZ SUPRALOOP is a monopolar electrosurgical device intended for cutting and removal of soft tissue in gynecological procedures involving endoscopic hysterectomy in adult women. It is used with standard electrosurgical generators | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
STORZ
KARL STORZ — ENDOSKOPE
KARL STORZ Premarket Notification
KARL STORZ Supraloop
510(k) Summary
### 510(k) Summary
This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92 and the FDA guidance document titled "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]" issued on July 28, 2014. All data included in this document is accurate and complete to the best of KARL STORZ SE & Co. KG knowledge.
| Submitter: | KARL STORZ SE & Co. KG Dr.-Karl-Storz-Straße 34 78532 Tuttlingen, Germany |
| --- | --- |
| Contact: | Mario Trujillo Regulatory Affairs Specialist Tel.: (424) 218-8481 Email: Mario.Trujillo@karlstorz.com |
| Date of Preparation: | September 23, 2026 |
| Type of 510(k) Submission: | Traditional |
| Device Identification: | Trade Name: KARL STORZ SUPRALOOP Classification Name: Coagulator-Cutter, Endoscopic, Unipolar (And Accessories) (21 CFR Part 884.4160); |
| Regulatory Class: | II |
| Product Code: | KNF |
| Guidance Document: | Not applicable |
| Predicate Devices: | LINA LOOP, MODELS: EL-160-4, EL-160-8, EL-200-4, EL-200-8 (K070315) The predicate has been subject to a design related recall. Due to the potential for the predicate device to break during a procedure, additional warnings were added to the instructions for use. KARL STORZ has adequately addressed the design related issues of the predicate through labeling and performance testing of the subject device. |
| Device Description: | The KARL STORZ SUPRALOOP is a modular monopolar electrosurgical system for laparoscopic supracervical hysterectomy. The system consists of a reusable handle (26183MA), a reusable 5 mm outer sheath with a 300 mm working length (26183MB), and a sterile, single-use monopolar loop electrode. Two loop sizes are available: 120 mm x 85 mm (26183MC-S) and 200 mm x 150 mm (26183MD-S). The patient-contacting materials of the loop include nickel-titanium alloy (NiTi), polytetrafluoroethylene (PTFE), stainless steel 1.4305, and black PTFE. The patient-contacting materials of the outer sheath include black polyphenylsulfone (PPSU), zirconium oxide (ZrO2), black perfluoroalkoxy alkane (PFA), and gray silicone The loop electrode is connected through the reusable handle and sheath to a commonly available electrosurgical generator with a monopolar output. When HF energy is activated, current is conducted through the loop electrode to circumferentially cut and coagulate tissue, with current returning through a separate patient return electrode. The handle advances and retracts the loop through the sheath to open and close the loop around the target tissue. |
| Indications For Use: | The KARL STORZ SUPRALOOP is a monopolar electrosurgical device intended for cutting and removal of soft tissue in gynecological procedures involving endoscopic hysterectomy in adult women. It is used with standard electrosurgical generators |
{5}
STORZ
KARL STORZ—ENDOSKOPE
KARL STORZ Premarket Notification
KARL STORZ Supraloop
510(k) Summary
Substantial equivalence
Comparison Table: Subject vs. Predicate Device
| | **Subject Device** KARL STORZ SUPRALOOP | **Predicate Device, K070315** LiNA Loop |
| --- | --- | --- |
| **Indication for Use** | | |
| **Indications for Use** | The KARL STORZ SUPRALOOP is a monopolar electrosurgical device intended for cutting and removal of soft tissue in gynecological procedures involving endoscopic hysterectomy in adult women. It is used with standard electrosurgical generators. | The LiNA Gold Loop is a 5mm monopolar electrosurgical device intended for cutting and removal of soft tissue in gynecological procedures involving endoscopic hysterectomy. It is used with standard electrosurgical generators. |
| **Product Code** | KNF | KNF |
| **Target Population** | Adult women | Adult women |
| **Physical & Technological Characteristics** | | |
| **Components** | Handle, outer sheath, loop | Handle, outer sheath, loop |
| **Product Assembly and disassembly** | Requires user assembly | Provided Assembled |
| **Sheath Working Length** | 300 mm | Information not available |
| **Sheath Total Length** | 413 mm | Information not available |
| **Outer Sheath Total Length** | 0.5 mm | 0.5 mm |
| **Loop Working Length** | 120mm x 85mm 200mm x 150mm | 160mm x 80mm 200mm x 100mm |
| **Loop Total Length** | 593 mm 680 mm | Information not available |
| **Method of Action** | The loop opens when the handle is pushed forward. The loop closed when the handle is pushed backwards. | The loop opens when the handle is pushed forward. The loop closed when the handle is pushed backwards. |
| **Energy Source** | Monopolar | Monopolar |
| **HF Generator Compatibility** | Compatible with all standard electrosurgical generators with monopolar outlet. | Compatible with all standard electrosurgical generators with monopolar outlet. |
| **Cleaning and Sterilization Methods** | | |
| **Reusable Components** | Handle, outer sheath | None |
| **Cleaning** | Manual | N/A |
| **Sterilization** | Steam, EtO | Steam, EtO |
| **Sterile Single-Use Components** | Loop | All |
| **Shelf Life** | Loops: 5 years | Information not publicly available |
The subject and predicate devices have the same indications for use. Differences in design, dimensions, materials, assembly, and reprocessing were evaluated through bench, biocompatibility, sterility, packaging, shelf-life, electrical safety, and EMC testing. These differences do not raise different questions of safety or effectiveness.
{6}
STORZ
KARL STORZ—ENDOSKOPE
KARL STORZ Premarket Notification
KARL STORZ Supraloop
510(k) Summary
| Non-Clinical Performance Data: | Bench verification performance testing has been performed for the following parameters: - Attaching and Detaching Force of the Electrode - System Interlocking Test - Bending Force - HF Tissue Test and HF Durability - Compression Testing - Tensile Strength Testing - Biocompatibility evaluation was performed to ISO 10993The results met the predefined acceptance criteria and support the mechanical integrity and functional performance of the subject deviceElectrical safety and electromagnetic compatibility (EMC)Electrical safety and EMC testing were provided in the submission. The system complies with the IEC 60601-1, IEC 60601-2-2 standards for safety and the IEC 60601-1-2 standard for EMC.BiocompatibilityThe loop electrode and outer sheath were evaluated as external communicating devices with tissue contact for a limited duration of less than 24 hours in accordance with ISO 10993-1. The biological evaluation included: cytotoxicity, sensitization, irritation or intracutaneous reactivity, acute systemic toxicity and material-mediated pyrogenicity. All applicable acceptance criteria were met.SterilityThe single-use loop electrodes are supplied sterile by ethylene oxide sterilization validated in accordance with ISO 11135:2014 to a sterility assurance level of 10^-6. EO and ECH residuals were evaluated in accordance with ISO 10993-7 and met the applicable allowable limits. A 5-year shelf life was supported by accelerated aging, climate conditioning, transportation simulation, seal-strength testing, package visual inspection, and package integrity testing. Packaging validation was performed in accordance with ISO 11607-1:2019, ASTM F88/F88M-15 and the bubble-leak method.The reusable handle and sheath are supplied non-sterile and require reprocessing prior to first use and in between uses. The submission includes validation of the cleaning and steam sterilization reprocessing instructions. |
| --- | --- |
| Clinical Performance Data: | Clinical testing was not required to demonstrate the substantial equivalence to the predicate device |
| Conclusion: | The results of the non-clinical performance testing support the substantial equivalence of the KARL STORZ SUPRALOOP to the legally marketed predicate device, LiNA Loop (K070315). The subject device has the same intended use and indications for use as the predicate. The identified technological differences do not raise different questions of safety or effectiveness, and the performance data demonstrate that the KARL STORZ SUPRALOOP is as safe and effective as the predicate device for its intended use. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.