This Single-use Digital Hysteroscope is intended to be used for viewing of the adult cervical canal and uterine cavity for the purpose of performing diagnostic and operative procedures. It can only be used with Geyi Medical endoscope camera system GY-AIO-121.
Device Story
Single-use digital hysteroscope; consists of operation section, irrigation/suction tubes, and rigid insertion section. Distal end integrates LED light source and CMOS image sensor (400x400 or 1280x720 resolution). Instrument channel (1.95mm–2.35mm) allows passage of biopsy/scissor/foreign body forceps. Operated in hospitals/clinics by physicians; connects to Geyi Medical endoscope camera system (GY-AIO-121) via cable. Provides real-time visualization of uterine cavity; assists in diagnostic/operative decision-making. Benefits include reduced cross-contamination risk due to single-use design and high-resolution imaging for improved visualization.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including mechanical performance, image quality parameters (FOV, resolution, SNR, etc.), electrical safety (IEC 60601-1/2-18), EMC (IEC 60601-1-2), biocompatibility (ISO 10993-1), photobiological safety (IEC 62471), and sterilization validation (ISO 11135).
Technological Characteristics
Materials: PEBAX, stainless steel, PI, silicone gel, ABS, TPU, PC. Sensing: CMOS image sensor. Illumination: LED. Connectivity: Wired to Geyi Medical camera host. Sterilization: Ethylene Oxide (EO). Standards: ISO 10993, ISO 11135, IEC 60601 series, IEC 62471, ASTM F1980, ASTM D4169.
Indications for Use
Indicated for viewing the adult cervical canal and uterine cavity to perform diagnostic and operative procedures.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
Predicate Devices
SeleneView® Single-Use Digital Hysteroscope (K240978)
Reference Devices
Single-Use Video Hysteroscope: RP-G-C24, RP-G-C0101 (K232003)
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
September 8, 2026
Zhejiang Geyi Medical Instrument Co., Ltd.
Cathy Tan
Director, Regulatory Affairs
# 1, 2 Factory, # 5, Hutang Rd. Xiaya Town
Jiande City, Zhejiang 311606
CHINA
Re: K254294
Trade/Device Name: Single-use Digital Hysteroscope (GYUT-A 12.6- 0,
GYUT-A12.6- 12, GYUT-A 12.6- 22, GYUT-A 12.6- 30,
GYUT-A 14.4- 0, GYUT-A 14.4- 12,GYUT-A 14.4- 22,
GYUT-A 14.4- 30, GYUT-B 15- 0, GYUT-B 15- 12,
GYUT-B 15-22, GYUT-B 15- 30)
Regulation Number: 21 CFR 884.1690
Regulation Name: Hysteroscope and accessories
Regulatory Class: II
Product Code: HIH
Dated: August 9, 2026
Received: August 10, 2026
Dear Cathy Tan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254294 - Cathy Tan
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn
{2}
K254294 - Cathy Tan
Page 3
(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Reginald K. Avery -S
for
Jason R. Roberts, Ph.D.
Assistant Director
DHT3B: Division of Reproductive,
Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K254294 | ? |
| Please provide the device trade name(s). | | ? |
| Single-use Digital Hysteroscope (GYUT-A 12.6- 0, GYUT-A 12.6- 12, GYUT-A 12.6- 22, GYUT-A 12.6- 30, GYUT-A 14.4- 0, GYUT-A 14.4- 12, GYUT-A 14.4- 22, GYUT-A 14.4- 30, GYUT-B 15- 0, GYUT-B 15- 12, GYUT-B 15- 22, GYUT-B 15- 30) | | |
| Please provide your Indications for Use below. | | ? |
| This Single-use Digital Hysteroscope is intended to be used for viewing of the adult cervical canal and uterine cavity for the purpose of performing diagnostic and operative procedures. It can only be used with Geyi Medical endoscope camera system GY-AIO-121. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K254294
Page 1 of 10
# 510(k) SUMMARY
510(k) Number:K254294
## I. SUBMITTER
Name: Zhejiang Geyi Medical Instrument Co., LTD.
Address: No.1,2 Factory,No.5 hutang Road, Xiaya, Jiande City, 311606, Zhejiang
Province.P.R.China
Name of contact person: Cathy Tan
Tel:+86-15158157854
Fax: +86-15158157854
E-mail:lilyxu@zjgeyi.com
Date prepared: September 7, 2026
## II. Proposed Device
Submission Number: K254294
Trade/Device Name: Single-use Digital Hysteroscope (GYUT-A 12.6- 0, GYUT-A
12.6- 12, GYUT-A 12.6- 22, GYUT-A 12.6- 30, GYUT-A 14.4- 0, GYUT-A 14.4- 12,
GYUT-A 14.4- 22, GYUT-A 14.4- 30, GYUT-B 15- 0, GYUT-B 15- 12, GYUT-B 15-
22, GYUT-B 15- 30)
Regulation Number: 21 CFR 884.1690
Regulation Name: Hysteroscope and accessories
Regulation class: II
Review Panel: Obstetrics/Gynecology
Product code: HIH
## III. Predicate Device
510(k) number: K240978
Device name: SeleneView® Single-Use Digital Hysteroscope (HC29, HC29-G, HC35,
HC35-G, HO35, HO35-G, HO42, HO42-G)
Applicant: MacroLux Medical Technology Co., Ltd.
Regulation Number: 21 CFR 884.1690
Regulation Name: Hysteroscope and accessories
{5}
K254294
Page 2 of 10
Regulation class: II
Review Panel: Obstetrics/Gynecology Product code: HIH
Note: This predicate device has not been subject to a design-related recall.
#### IV. Reference Device
510(k) number: K232003
Device name: Single-Use Video Hysteroscope: RP-G-C24, RP-G-C0101
Applicant: Guangzhou Red Pine Medical Instrument Co., Ltd.
Regulation Number: 21 CFR 884.1690 Regulation Name: Hysteroscope and accessories Regulation class: II
Review Panel: Obstetrics/Gynecology Product code: HIH
Note: This reference device has not been subject to a design-related recall.
#### V. Indications for use
This Single-use Digital Hysteroscope is intended to be used for viewing of the adult cervical canal and uterine cavity for the purpose of performing diagnostic and operative procedures.
It can only be used with Geyi Medical endoscope camera system, GY-AIO-121.
#### VI. Description of the Device
The Single-use Digital Hysteroscope consists of three main parts: an operation section with a connecting cable, water irrigation and suction tube section and insertion section with a rigid insertion tube. The head of the insertion section integrates an LED light source, a 400*400 pixel (GYUT-A) or 1280*720-pixel (GYUT-B) CMOS image sensor and an instrument channel tube with an inner diameter of 2.15 mm, 2.35 mm, or 1.95 mm. During use, the LED light source provides illumination, the CMOS image sensor captures the image pixels of the tissue area, and the instrument channel tube serves as a passage for biopsy forceps, scissor forceps, foreign body forceps. The operation section is equipped with two function shortcut keys corresponding to the preset functions of the endoscope camera host. Pressing the key, the host will have the corresponding function. The water irrigation and suction tube is connected to the corresponding irrigation tubing/suction tubing of the fluid management system. The water irrigation tube provides a pathway for irrigation fluid to the distal end of the hysteroscope, while the suction tube
{6}
K254294
Page 3 of 10
provides a pathway for removal of fluids from the surgical site through the fluid management system. The plug at the tail end of the connecting cable is connected to the handle extension cable to transmit the image pixels obtained by the image sensor to the host, thereby achieving the surgical imaging function.
It is ethylene oxide (EO) sterilized. It's single use only.
There are twelve models of the Single-Use Video Hysteroscope. The models are identical except for the width of insertion port, Bending Angle, and diameter of instrument channel, as shown in the table below:
| Models | Maximum insertion portion width | Minimum instrument channel diameter | Bending Angle | Working length |
| --- | --- | --- | --- | --- |
| GYUT-A 12.6-0 | 4.5mm | 2.15mm | 0° | 220mm±10mm |
| GYUT-A 12.6-12 | 4.5mm | 2.15mm | 12°±5° | 220mm±10mm |
| GYUT-A 12.6-22 | 4.5mm | 2.15mm | 22°±5° | 220mm±10mm |
| GYUT-A 12.6-30 | 4.5mm | 2.15mm | 30°±5° | 220mm±10mm |
| GYUT-A 14.4-0 | 5.1mm | 2.35mm | 0° | 220mm±10mm |
| GYUT-A 14.4-12 | 5.1mm | 2.35mm | 12°±5° | 220mm±10mm |
| GYUT-A 14.4-22 | 5.1mm | 2.35mm | 22°±5° | 220mm±10mm |
| GYUT-A 14.4-30 | 5.1mm | 2.35mm | 30°±5° | 220mm±10mm |
| GYUT-B 15-0 | 5.3mm | 1.95mm | 0° | 220mm±10mm |
| GYUT-B 15-12 | 5.3mm | 1.95mm | 12°±5° | 220mm±10mm |
| GYUT-B 15-22 | 5.3mm | 1.95mm | 22°±5° | 220mm±10mm |
| GYUT-B 15-30 | 5.3mm | 1.95mm | 30°±5° | 220mm±10mm |
The materials used in construction of the insertion portion of Hysteroscope are Polyether block polyamide(PEBAX), Stainless steel, polyimide(PI), Silicone gel and
Acrylonitrile Butadiene Styrene(ABS). The materials used in construction of the water irrigation and suction portion of Hysteroscope are Thermoplastic polyurethane rubber(TPU), Polycarbonate(PC).
The handle does not contact the patient.
{7}
K254294
Page 4 of 10
## VII. Comparison with the Predicate Device
### Comparison of Indications with the Predicate Device
The indications for use are the same for the predicate device.
### Comparison of Technological Characteristics with the Predicate Device
Subjective device and the predicate device have the same intended use. The subject and predicate device have similar technological characteristics as evidenced by the table below. All differences have been discussed in detail and concluded that these differences in technological characteristics do not raise new questions regarding safety or effectiveness of the subject device.
a. The following basic technological elements are the same or similar for the subject and predicate device
| Item | Subject DeviceSingle-use Digital Hysteroscope | Predicate deviceSeleneView®Single-Use Digital Hysteroscope(HC29, HC29-G, HC35, HC35-G, HO35, HO35-G, HO42, HO42-G) | Discussion |
| --- | --- | --- | --- |
| 510(k) number | K254294 | K240978 | |
| Classification Product Code | HIH | HIH | same |
{8}
K254294
Page 5 of 10
| Indications for use | Single-use Digital Hysteroscope is intended to be used for viewing of the adult cervical canal and uterine cavity for the purpose of performing diagnostic and operative procedures. | The SeleneView® (model: HC29, HC29-G, HC35,HC35-G) is used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic procedures. The SeleneView® (model: HO35, HO35-G, HO42, HO42-G) is used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and therapeutic procedures. | Same |
| --- | --- | --- | --- |
| Route of Advancement | Advanced to uterine cavity via the cervical canal | Advanced to cervical canal and uterine cavity | Same |
| Site of use | Hospitals and physician offices | SeleneView® is intended for use in professional healthcare facility environments such as hospitals and clinics. SeleneView® is designed for use in adults. | Same |
| Components | The hysteroscope system is composed of Single-use Digital Hysteroscope and endoscopic video image processor. | SeleneView®Single-Use Digital Hysteroscope consists of a handle, insertion portion and endoscope cable. The SeleneView® is intended to be used combined with ViewHub® Video Processor to perform its intended function and the ViewHub® has been 510(k) cleared in K233779. | Same |
{9}
K254294
Page 6 of 10
| Cannula Outer Diameter | GYUT-A 12.6- 0/ GYUT-A 12.6- 12/GYUT-A 12.6- 22/ GYUT-A 12.6- 30: 4.5mm. GYUT-A 14.4- 0/ GYUT-A 14.4- 12/GYUT-A 14.4- 22/ GYUT-A 14.4- 30: 5.1mm. GYUT-B 15-0/GYUT-B 15-12/ GYUT-B 15-22/GYUT-B 15-30: 5.3mm. | HC29/HC29-G: 2.9mm HC35/HC35-G: 3.85mm HO35/HO35-G: 3.8mm HO42/HO42-G: 4.3mm | Different 1 |
| --- | --- | --- | --- |
| Minimum instrument diameter | GYUT-A 12.6- 0/ GYUT-A 12.6- 12/GYUT-A 12.6- 22/ GYUT-A 12.6- 30: 2.15mm. GYUT-A 14.4- 0/ GYUT-A 14.4- 12/GYUT-A 14.4- 22/ GYUT-A 14.4- 30: 2.35mm. GYUT-B 15-0/GYUT-B 15-12/ GYUT-B 15-22/GYUT-B 15-30: 1.95mm. | HC29/HC29-G: none HC35/HC35-G: none HO35/HO35-G: ≥1.8mm HO42/HO42-G: ≥2.0mm | Different 2 |
| --- | --- | --- | --- |
| Cannula Total Length | 220mm±10mm | 220 mm | Same |
| Illumination light source | LED | LED | Same |
| Image transmission | Images captured by the CMOS chip at the distal end are transmitted through the endoscopes and the video processor to the user. | The images are transmitted through the endoscope with a CMOS image module at the distal end and video processor. | Same |
| Optical Image | CMOS | CMOS | Same |
{10}
K254294
Page 7 of 10
| Sterilization | The Single-use Digital Hysteroscope is sterile following exposure to ethylene oxide (EO) | The Hysteroscope is sterile after sterilization with ethylene oxide (EO) | Same |
| --- | --- | --- | --- |
| Tissue contact materials | Compliant with ISO 10993 | Compliant with ISO 10993 | Same |
| Field of View | GYUT-A 12.6- 0/ GYUT-A 12.6- 12/ GYUT-A 12.6- 22/ GYUT-A 12.6- 30/GYUT-A 14.4- 0/ GYUT-A 14.4- 12/GYUT-A 14.4- 22/ GYUT-A 14.4- 30: 120°. GYUT-B 15-0/GYUT-B 15- 12/GYUT-B 15- 22/ GYUT-B 15- 30: 140°. | HC29/HC29-G/HO35/HO35-G: 120° HC35/HC35-G/HO42/HO42-G: 140° | Same |
| Depth of field | GYUT-A 12.6- 0/ GYUT-A 12.6- 12/GYUT-A 12.6- 22/ GYUT-A 12.6- 30/GYUT-A 14.4- 0/ GYUT-A 14.4- 12/GYUT-A 14.4- 22/ GYUT-A 14.4- 30: 3-50mm. GYUT-B 15-0/GYUT-B 15- 12/GYUT-B 15- 22/ GYUT-B 15-30: 3-100mm. | 3mm~100mm | same |
| Direction of view | 0° | 15° | Different 3 |
| Image Resolution | GYUT-A 12.6- 0,GYUT-A 12.6- 12/GYUT-A 12.6- 22, GYUT-A 12.6- 30,GYUT-A 14.4- 0, GYUT-A 14.4- 12,GYUT-A 14.4- 22, GYUT-A 14.4- 30: 400*400 pixels GYUT-B 15-0/GYUT-B-12/GYUT-B 15-22/GYUT-B 15-30: 1280*720 pixels | HC29/HC29-G/HO35/HO35-G: 400*400 pixels HC35/HC35-G/HO42/HO42-G: 1280*720 pixels | Same |
{11}
K254294
Page 8 of 10
**b. Discussion on differences between the subject and the predicate device**
Differences 1, 2 and 3 between the subject and predicate device include differences in outer diameter, minimum instrument channel diameter and direction of view are minor and do not raise different questions of safety and effectiveness. Differences have been evaluated through performance testing.
**VIII. Summary of Non-clinical Performance**
**Performance testing**
Mechanical testing was performed on the proposed device and included the following:
- Appearance
- Maximum insert section width
- Minimum Working channel width
- Working length
- Bending Angle
- Water Supply System
- Water Supply System sealing performance
- Performance of Flow Regulator in Water Supply System
- Suction system
- Suction system sealing performance
- Performance of Flow Regulator in suction System
- Tensile strength
- Luer connector
- Handle plug attachment force test
- Handle plug detachment force test
Comparative testing related to image quality parameters including the following list was performed for the proposed device and the predicate and reference devices to support substantial equivalence
- Color performance
- Field of view (FOV)
{12}
K254294
Page 9 of 10
- Direction of view (DOV)
- Uniformity
- Depth of field (DOF)
- Resolution
- Distortion
- Signal-to-noise ratio (SNR)
- Dynamic range
### Electrical safety and electromagnetic compatibility (EMC)
Electrical Safety and EMC data include in the submission meet the following standards.
- IEC60601-1:2012, and AAMI/ANSI ES60601-1:2005/(R)2012 And A1:2012, C1:2009/(R)2012 And A2:2010/(R)2012
- IEC 60601-2-18:2009: Part 2-18
- IEC 60601-1-2:2014 and A1:2020
### Biocompatibility
Biocompatibility evaluation of the Single-use Digital Hysteroscope was conducted in accordance with the FDA Guidance document “Use of International Standard ISO 10993-1, "Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process”. The device is considered to contact a breached or compromised surface for a duration of less than 24 hours.
The following biocompatibility studies were conducted on the final, finished, EO sterilized device and in accordance with the GLP requirements: cytotoxicity, intracutaneous reactivity, sensitization, acute systemic toxicity and pyrogenicity. All evaluation acceptance criteria were met.
### Photobiological safety
The optical performance include following items:
The LEDs in the proposed device were tested according to IEC 62471:2006 Medical electrical equipment, Photobiological safety of lamps and lamp systems.
{13}
K254294
Page 10 of 10
### Sterilization Validation and Shelf Life
The sterilization method has been validated to ISO 11135:2014, which has thereby determined the routine control and monitoring parameters. The Single-use Digital Hysteroscope is sterilized with EO to a sterility level of SAL 10⁻⁶, and EO/ECH residual tests were performed according to ISO 10993-7:2008.
Shelf life is determined based on stability studies which includes ageing test according to ASTM F1980-21 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices and simulated shipping distribution test according to ASTM D4169-22 Standard Practice for Performance Testing of Shipping Containers and Systems.
Samples were subjected to accelerated aging and simulated shipping distribution to evaluate the package integrity and device performance throughout the labeled shelf life.
### IX. Substantially Equivalent (SE) Conclusion
Performance testing presented in this premarket notification demonstrated that the proposed devices are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.