easyEndo™ E-Lite Single Use Powered Endoscopic Staplers; easyEndo™ E-Lite Single Use Powered Endoscopic Staplers with Curved Tip; easyEndo™ E-Lite Single Use Cartridges
K254291 · Ezisurg Medical Co., Ltd. · GDW · Sep 25, 2026 · General, Plastic Surgery
Device Facts
Record ID
K254291
Device Name
easyEndo™ E-Lite Single Use Powered Endoscopic Staplers; easyEndo™ E-Lite Single Use Powered Endoscopic Staplers with Curved Tip; easyEndo™ E-Lite Single Use Cartridges
Applicant
Ezisurg Medical Co., Ltd.
Product Code
GDW · General, Plastic Surgery
Decision Date
Sep 25, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
easyEndo™ E-Lite Single Use Powered Endoscopic Staplers are intended for transection, resection, and/or creation of anastomoses. The instruments have applications in multiple open or minimally invasive general and thoracic surgical procedures. easyEndo™ E-Lite Single Use Powered Endoscopic Staplers with Curved Tip are intended for transection, resection, and/or creation of anastomoses. The instruments have applications in multiple open or minimally invasive general and thoracic surgical procedures. easyEndo™ E-Lite Single Use Cartridges are intended for transection, resection, and/or creation of anastomoses. The instruments have applications in multiple open or minimally invasive general and thoracic surgical procedures.
Device Story
The easyEndo™ E-Lite is a sterile, single-use, battery-powered endoscopic stapler system. It consists of a powered handle, shaft (260mm, 350mm, or 440mm), and interchangeable cartridges. The device features an articulation knob allowing jaw movement up to 60 degrees in either direction. During operation, the device simultaneously cuts and staples tissue, placing six staggered rows of staples (three on each side of the cut line). It includes a lockout mechanism to prevent firing without a cartridge or with a used cartridge. The device is used by surgeons in open or minimally invasive procedures to facilitate tissue transection, resection, and anastomosis. The powered firing system replaces manual force, potentially improving consistency in staple formation and tissue division. The healthcare provider operates the device manually, using the articulation and firing controls to manipulate tissue and deploy staples, which are then visualized or confirmed through standard surgical practice.
Clinical Evidence
No clinical data was necessary. Evidence includes non-clinical bench testing (staple formation, line strength, leak-burst pressure, clamping force, lockout, electrical safety/EMC per IEC 60601-1/1-2, biocompatibility per ISO 10993, and sterilization/shelf-life validation) and GLP-compliant acute and chronic animal studies comparing the device to predicates in tissue transection/resection/anastomosis. Human factors testing was also performed.
Indicated for patients undergoing open or minimally invasive general and thoracic surgical procedures requiring transection, resection, or creation of anastomoses.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
ECHELON, ECHELON ENDOPATH™ and ECHELON FLEX families of endoscopic linear cutters and reloads (K202665)
Echelon Flex™ Powered Vascular Stapler with Advanced Placement Tip (K141952)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG
ADMINISTRATION
September 25, 2026
Ezisurg Medical Co., Ltd.
% Giselle Zhang
Lead RA&QA Consultant
Emergo Global Consulting, LLC
2500 Bee Cave Rd.
Bldg. 1, Suite 300
Austin, Texas 78746
Re: K254291
Trade/Device Name: easyEndo™ E-Lite Single Use Powered Endoscopic Staplers
easyEndo™ E-Lite Single Use Powered Endoscopic Staplers with Curved Tip
easyEndo™ E-Lite Single Use Cartridges
Regulation Number: 21 CFR 878.4750
Regulation Name: Implantable Staple
Regulatory Class: Class II
Product Code: GDW, GAG
Dated: August 25, 2026
Received: August 25, 2026
Dear Giselle Zhang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}
K254291 - Giselle Zhang
Page 2
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K254291 - Giselle Zhang
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# TEK N. LAMICHHANE -S
Tek N. Lamichhane, Ph.D.
Assistant Director
DHT4B: Division of Plastic and
Reconstructive Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K254291
Device Name
easyEndoTM E-Lite Single Use Powered Endoscopic Staplers
easyEndoTM E-Lite Single Use Powered Endoscopic Staplers with Curved Tip
easyEndoTM E-Lite Single Use Cartridges
Indications for Use (Describe)
easyEndoTM E-Lite Single Use Powered Endoscopic Staplers are intended for transection, resection, and/or creation of anastomoses. The instruments have applications in multiple open or minimally invasive general and thoracic surgical procedures.
easyEndoTM E-Lite Single Use Powered Endoscopic Staplers with Curved Tip are intended for transection, resection, and/or creation of anastomoses. The instruments have applications in multiple open or minimally invasive general and thoracic surgical procedures.
easyEndoTM E-Lite Single Use Cartridges are intended for transection, resection, and/or creation of anastomoses. The instruments have applications in multiple open or minimally invasive general and thoracic surgical procedures.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
# 510(k) Summary
(K254291)
## easyEndo™ E-Lite Single Use Powered Endoscopic Staplers and Cartridges
### 1. Submission Sponsor
Ezisurg Medical Co., Ltd.
No.399, Miaoqiao Rd., Pudong New District,
Shanghai, P.R. China
Contact Name: Hong Zhu
Contact Title: RA&QA VP
### 2. Submission Correspondent
Emergo Global Consulting, LLC
2500 Bee Cave Road
Building 1, Suite 300
Austin, TX 78746
Office Phone: (512) 327-9997
Email: LST.AUS.ProjectManagement@ul.com, Giselle.Zhang@ul.com
Contact: Giselle Zhang
Title: Lead QA/RA Consultant
### 3. Date Prepared
09/23/2026
### 4. Device Identification
| Trade/Proprietary Name: | easyEndo™ E-Lite Single Use Powered Endoscopic Staplers easyEndo™ E-Lite Single Use Powered Endoscopic Staplers with Curved Tip easyEndo™ E-Lite Single Use Cartridges |
| --- | --- |
| Classification Name: | Staple, Implantable |
| Regulation Number(s): | 21 CFR 878.4750 |
| Product Code(s): | GDW |
| Class: | II |
| Classification Panel: | General & Plastic Surgery |
| Classification Name: | Stapler, Surgical |
K254291, Page 1 of 7
{5}
Regulation Number(s): 21 CFR 878.4800
Product Code(s): GAG
Class: II
Classification Panel: General & Plastic Surgery
### 5. Legally Marketed Predicate Device(s)
Device name: ECHELON, ECHELON ENDOPATH™ and ECHELON FLEX families of endoscopic linear cutters and reloads
510(k) number: K202665
Manufacturer: Ethicon Endo-Surgery, LLC
Device Name: Echelon Flex™ Powered Vascular Stapler with Advanced Placement Tip
510(k) Number: K141952
Manufacturer: Ethicon Endo-Surgery, LLC
### 6. Device Description
The easyEndo™ E-Lite (previously named: easyEndo™ Powered) Single Use Powered Endoscopic Staplers and Cartridges, hereafter E-Lite Staplers and Cartridges, are sterile, single patient use instruments that simultaneously cut and staple tissue through a battery powered firing system. The E-Lite Staplers and Cartridges are sterilized by irradiation. The instruments deliver six staggered rows of staples, three on either side of the cut line.
The staplers are available in three shaft lengths of 260mm, 350mm and 440mm. The Articulation Knob can articulate the jaws to 60 degrees, 45 degrees, 30 degrees, and 15 degrees in each set direction to facilitate lateral access to the operative site.
The device has two staple line lengths, 45 mm and 60 mm. The staplers are packaged with a primary lithium battery pack that must be installed prior to use. The staplers are packaged without a cartridge and must be loaded prior to use.
The lock out feature is designed to prevent a used or improperly installed Cartridge from being re-fired or the Stapler from being fired without a Cartridge.
### 7. Indication for Use Statement
easyEndo™ E-Lite Single Use Powered Endoscopic Staplers are intended for transection, resection, and/or creation of anastomoses. The instruments have applications in multiple open or minimally invasive general and thoracic surgical procedures.
easyEndo™ E-Lite Single Use Powered Endoscopic Staplers with Curved Tip are intended for transection, resection, and/or creation of anastomoses. The instruments have applications in multiple open or minimally invasive general and thoracic surgical procedures.
easyEndo™ E-Lite Single Use Cartridges are intended for transection, resection, and/or creation of anastomoses. The instruments have applications in multiple open or minimally invasive general and thoracic surgical procedures.
K254291, Page 2 of 7
{6}
## 8. Substantial Equivalence Discussion
The following table compares the easyEndo™ to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance, and forms the basis for the determination of substantial equivalence. The subject device shares the same intended use with the predicates for transection, resection, and/or creation of anastomoses, but the safety and effectiveness of this device have not been studied for transection and resection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney and spleen. Although the subject device has more limited indications for use and differs in some technological characteristics, safety and performance data demonstrates that these differences do not raise any new questions of safety or effectiveness compared to the predicate device
Table 1 - Comparison of Stapler Characteristics
| Item | easyEndo™ E-Lite Single Use Powered Endoscopic Staplers (Proposed device) | ECHELON, ECHELON ENDOPATH™ and ECHELON FLEX families of endoscopic linear cutters and reloads (Primary Predicate: K202665) | Echelon Flex™ Powered Vascular Stapler with Advanced Placement Tip (Secondary Predicate: K141952) | SE Discussion |
| --- | --- | --- | --- | --- |
| Manufacturer | Ezisurg Medical Co., Ltd. | Ethicon Endo-Surgery, LLC | Ethicon Endo-Surgery, LLC | / |
| Product Code | GDW | GDW | GDW | Same |
| Classification | Class II (21 CFR 878.4750 & 21 CFR 878.4740) | Class II (21 CFR 878.4750) | Class II (21 CFR 878.4750) | Same |
| Indications for Use | easyEndo™ E-Lite Single Use Powered Endoscopic Staplers are intended for transection, resection, and/or creation of anastomoses. The instruments have applications in multiple open or minimally invasive general and thoracic surgical procedures. | ECHELON, ECHELON ENDOPATH™ and ECHELON FLEX families of endoscopic linear cutters and reloads are intended for transection, resection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general, gynecologic, urologic, thoracic, and pediatric | The Echelon Flex™ Powered Vascular Stapler with Advanced Placement Tip and it's reloads are intended for transection and resection of tissue and vasculature. The instruments have application in multiple open or minimally invasive general gynecologic, urologic, | The subject device has a more limited indication for use than the predicate device which is acceptable. |
K254291, Page 3 of 7
{7}
| | | surgical procedures. They can be used with staple line or tissue buttressing materials. The instruments may also be used for transection and resection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney and spleen. | thoracic, and pediatric surgical procedures. | |
| --- | --- | --- | --- | --- |
| Operating Principle | Powered | Powered | Powered | Same |
| Cutting Mechanism | Stapler places two, triple staggered rows of staples and simultaneously cuts and divides the tissue between the two rows. | Stapler places two, triple staggered rows of staples and simultaneously cuts and divides the tissue between the two rows. | Four staggered rows of staples, two on either side of the cut line. | Equivalent |
| Jaw Structure | Linear, Curved | Linear | Curved | Equivalent |
| Lock-out Feature | Yes | Yes | Yes | Same |
| Main Components | Cutting Knife Anvil Jaw (with or without Curved Tip) Staple Battery Pack Shaft | Cutting Knife Anvil Jaw Staple Battery Pack Shaft | Cutting Knife Anvil Jaw (with Advanced Placement Tip) Staple Battery Pack Shaft | Equivalent |
| Suture Length | 45mm, 60mm | 45mm, 60mm | 35 mm | Equivalent |
| Articulation Degree | Total 120° (60° in each direction) | Total 90° (45° in each direction) | Total 100° (50° in each direction) | Equivalent |
| Sterilization | Irradiation sterilization | Irradiation sterilization | Irradiation sterilization | Same |
| Materials or substances contacted with patient | Anvil jaw: Stainless steel 05Cr17Ni4Cu4Nb, Stainless steel 06Cr19Ni10 (304) | Stainless steel | Stainless steel | Equivalent |
K254291, Page 4 of 7
{8}
| | Stainless steel 12Cr17Ni7 Curved Tip: Stainless steel 05Cr17Ni4Cu4Nb and Stainless steel 12Cr17Ni7 | | | |
| --- | --- | --- | --- | --- |
| | Shaft: stainless steel 06Cr19Ni10 | Stainless steel | Stainless steel | |
| | Cutting knife: stainless steel 20Cr13 | Stainless steel | Stainless steel | |
Table 2 - Comparison of Cartridge Characteristics
| Item | easyEndo™ E-Lite Single Use Cartridges (Proposed device) | ECHELON, ECHELON ENDOPATH™ and ECHELON FLEX families of endoscopic linear cutters and reloads (Primary Predicate: K202665) | Echelon Flex™ Powered Vascular Stapler with Advanced Placement Tip and reloads (Secondary Predicate: K141952) | SE Discussion |
| --- | --- | --- | --- | --- |
| Manufacturer | Ezisurg Medical Co., Ltd. | Ethicon Endo-Surgery, LLC | Ethicon Endo-Surgery, LLC | / |
| Product Code | GDW, GAG | GDW | GDW | Same |
| Classification | Class II (21 CFR 878.4750 & 21 CFR 878.4740) | Class II (21 CFR 878.4750) | Class II (21 CFR 878.4750) | Same |
| Indications for Use | easyEndo™ E-Lite Single Use Cartridges are intended for transection, resection, and/or creation of anastomoses. The instruments have applications in multiple open or minimally invasive general and thoracic surgical procedures. | ECHELON, ECHELON ENDOPATH™ and ECHELON FLEX families of endoscopic linear cutters and reloads are intended for transection, resection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general, gynecologic, urologic, thoracic, and pediatric surgical | The Echelon Flex™ Powered Vascular Stapler with Advanced Placement Tip and its reloads are intended for transection and resection of tissue and vasculature. The instruments have application in multiple open or minimally invasive general gynecologic, urologic, | The subject device has a more limited indication for use than the predicate device which is acceptable. |
K254291, Page 5 of 7
{9}
| | | | procedures. They can be used with staple line or tissue buttressing materials. The instruments may also be used for transection and resection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney and spleen. | thoracic, and pediatric surgical procedures. | |
| --- | --- | --- | --- | --- | --- |
| Staple height | Open staple height | 2.0 mm, 2.5 mm, 3.5 mm, 3.8 mm, 4.1 mm, 4.4 mm | 2.0 mm, 2.6 mm, 3.6 mm, 3.8 mm, 4.1 mm, 4.2 mm | 2.5 mm | Equivalent |
| | Closed staple height | 0.75 mm, 1.0 mm, 1.5 mm, 1.7 mm, 2.0 mm, 2.3 mm | 0.75 mm, 1.0 mm, 1.5 mm, 1.8 mm, 2.0 mm, 2.3 mm | 1.0mm | |
| Row Number of Staples | | 6 | 6 | 4 | Equivalent |
| Staple Shape | | Standard “B” shaped staple | Standard “B” shaped staple | Standard “B” shaped staple | Same |
| Materials or substances contacted with patient | Staples: Unalloyed Titanium (Ti Grade 2) | Staples: titanium (Ti3Al2.5V) alloy | Staples: titanium (Ti3Al2.5V) alloy | Equivalent | |
| | | Cartridge jaw: Stainless steel 05Cr17Ni4Cu4Nb, | Stainless steel | | Stainless steel |
| | | Cartridge: PA | Unknown | | Unknown |
### 9. Non-Clinical Performance Data
To demonstrate safety and effectiveness of easyEndo™ E-Lite Single Use Powered Endoscopic Staplers and Cartridges and to show substantial equivalence to the predicate device, Ezisurg completed the following non-clinical tests. Results confirm that the design inputs and performance specifications for the device are met. The easyEndo™ E-Lite Single Use Powered Endoscopic Staplers and Cartridges passed the testing in accordance with internal requirements, national standards, and international standards shown below, supporting its safety and effectiveness, and its substantial equivalence to the predicate device:
- Staple height and formation
- Staple line strength
- Staple line integrity
K254291, Page 6 of 7
{10}
- Leak-burst pressure testing
- Clamping Force
- Obstruction and lockout testing
- Functional testing
- Electrical safety and EMC testing in accordance with IEC 60601-1 and IEC 60601-1-2 standards
- Biocompatibility of materials according to ISO 10993 standards
- Sterilization, packaging, transportation and shelf-life validations.
Additionally, Human Factors testing was executed in accordance with the FDA Guidance Document Applying Human Factors and Usability Engineering to Medical Devices.
#### **10. Animal Study**
Animal testing was conducted in alignment with good laboratory practices (GLP) per 21 CFR §58.
The study is intended to evaluate the safety effectiveness and handling characteristics of easyEndo™ E-Lite in a chronic and acute animal study with the predicate devices. Animals have been divided into test and control groups and performed transection, resection, and/or the creation of anastomoses on target tissues or vasculature, according to the corresponding indications and specific tissue types.
In terms of safety and performance, there was no significant difference between the test and control groups, and all evaluation indicators met the acceptance criteria.
#### **11. Clinical Performance Data**
No clinical data was necessary to determine the substantial equivalence of this device.
#### **12. Statement of Substantial Equivalence**
The easyEndo™ E-Lite Single Use Powered Endoscopic Staplers and Cartridges have the same intended use as the predicate devices and similar technological characteristics. The differences in technological characteristics do not raise new or different questions of safety and effectiveness. Performance testing has demonstrated the easyEndo™ E-Lite Single Use Powered Endoscopic Staplers and Cartridges is as safe and effective as the predicate device. Therefore, the easyEndo™ E-Lite Single Use Powered Endoscopic Staplers and Cartridges is substantially equivalent to the predicate device.
K254291, Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.