Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Aqua Blue; Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Violet Blue; Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Dark Blue; Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Dark Violet Blue
K254285 · Hartalega NGC Sdn. Bhd. · LZA · Jul 20, 2026 · General Hospital
Device Facts
Record ID
K254285
Device Name
Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Aqua Blue; Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Violet Blue; Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Dark Blue; Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Dark Violet Blue
Applicant
Hartalega NGC Sdn. Bhd.
Product Code
LZA · General Hospital
Decision Date
Jul 20, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Aqua Blue is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs. Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Violet Blue is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs. Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Dark Blue is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs.
Device Story
Disposable, non-sterile, powder-free nitrile examination gloves; worn by healthcare examiners to prevent cross-contamination between patient and examiner. Device provides a physical barrier; tested for resistance to chemotherapy drug permeation per ASTM D6978. Used in clinical settings; operated by healthcare professionals. Output is the physical barrier provided by the glove material. Benefits include protection against patient-to-examiner contamination and specific chemotherapy agents. Subject device adds a 5-year expiration date labeling claim supported by real-time shelf-life testing.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including ASTM D6319 physical property testing, ASTM D5151 barrier integrity testing, and ASTM D6978 chemotherapy drug permeation testing performed on gloves aged through a 5-year real-time shelf life.
Technological Characteristics
Nitrile latex; powder-free; non-sterile; ambidextrous; single-use. Complies with ASTM D6319 (dimensions, physical properties), ASTM D5151 (barrier integrity), and ASTM D6978 (chemotherapy permeation). 5-year shelf life supported by ASTM D7160 real-time aging.
Indications for Use
Indicated for use as a non-sterile, disposable, ambidextrous, powder-free nitrile examination glove worn on the examiner's hand to prevent contamination between patient and examiner. Tested for resistance to permeation by specific chemotherapy drugs per ASTM D6978. Warning: Do not use with Carmustine and Thiotepa.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs (K162923)
Submission Summary (Full Text)
{0}

July 20, 2026
Hartalega NGC Sdn. Bhd.
Nor Akmar Yusoff
Manager - Regulatory Affairs
# 1, Persiaran Tanjung, Kawasan Perindustrian Tanjung
Sepang, Selangor Darul Ehsan 43900
Malaysia
Re: K254285
Trade/Device Name: Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs - Aqua Blue; Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs - Violet Blue; Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs - Dark Blue.
Regulation Number: 21 CFR 880.6250
Regulation Name: Non-Powdered Patient Examination Glove
Regulatory Class: Class I, reserved
Product Code: LZA, LZC, OPJ
Dated: June 25, 2026
Received: June 26, 2026
Dear Nor Akmar Yusoff:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254285 - Nor Akmar Yusoff
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn
{2}
K254285 - Nor Akmar Yusoff
Page 3
(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Bifeng Qian, M.D., Ph.D.
Assistant Director
DHT4C: Division of Infection
Control Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K254285
Device Name
Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Aqua Blue;
Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Violet Blue;
Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Dark Blue
Indications for Use (Describe)
Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Aqua Blue is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs.
The list of Chemotherapy Drugs tested (with breakthrough time) are:
| Chemotherapy Drug and Concentration | Minimum Breakthrough Detection Time Minutes |
| --- | --- |
| Carmustine (3.3 mg/ml) | 10.3 |
| Cisplatin (1.0 mg/ml) | >240 |
| Cyclophosphamide (20.0 mg/ml) | >240 |
| Dacarbazine (10.0 mg/ml) | >240 |
| Doxorubicin Hydrochloride (2.0 mg/ml) | >240 |
| Etoposide(Toposar)(20.0 mg/ml) | >240 |
| Fluorouracil (50.0 mg/ml) | >240 |
| Methotrexate (25.0 mg/ml) | >240 |
| Mitomycin C (0.5 mg/ml) | >240 |
| Paclitaxel (6.0 mg/ml) | >240 |
| Thiotepa (10.0 mg/ml) | 28.8 |
| Vincristine Sulfate (1.0 mg/ml) | >240 |
Caution: Testing showed a minimum breakthrough time of 10.3 minutes with Carmustine and 28.8 minutes with Thiotepa.
Warning: Do not use with Carmustine and Thiotepa.
Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Violet Blue is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs.
The list of Chemotherapy Drugs tested (with breakthrough time) are:
| Chemotherapy Drug and Concentration | Minimum Breakthrough Detection Time Minutes |
| --- | --- |
| Carmustine (3.3 mg/ml) | 10.2 |
| Cisplatin (1.0 mg/ml) | >240 |
| Cyclophosphamide (20.0 mg/ml) | >240 |
| Dacarbazine (10.0 mg/ml) | >240 |
| Doxorubicin Hydrochloride (2.0 mg/ml) | >240 |
| Etoposide(Toposar)(20.0 mg/ml) | >240 |
| Fluorouracil (50.0 mg/ml) | >240 |
| Methotrexate (25.0 mg/ml) | >240 |
| Mitomycin C (0.5 mg/ml) | >240 |
| Paclitaxel (6.0 mg/ml) | >240 |
| Thiotepa (10.0 mg/ml) | 30.2 |
| Vincristine Sulfate (1.0 mg/ml) | >240 |
FORM FDA 3881 (8/23)
Page 1 of 2
PSC Publishing Services (301) 443-6740
EF
{4}
Caution: Testing showed a minimum breakthrough time of 10.2 minutes with Carmustine and 30.2 minutes with Thiotepa.
Warning: Do not use with Carmustine and Thiotepa.
Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Dark Blue is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs.
The list of Chemotherapy Drugs tested (with breakthrough time) are:
| Chemotherapy Drug and Concentration | Minimum Breakthrough Detection Time Minutes |
| --- | --- |
| Carmustine (3.3 mg/ml) | 7.3 |
| Cisplatin (1.0 mg/ml) | >240 |
| Cyclophosphamide (20.0 mg/ml) | >240 |
| Dacarbazine (10.0 mg/ml) | >240 |
| Doxorubicin Hydrochloride (2.0 mg/ml) | >240 |
| Etoposide(Toposar)(20.0 mg/ml) | >240 |
| Fluorouracil (50.0 mg/ml) | >240 |
| Methotrexate (25.0 mg/ml) | >240 |
| Mitomycin C (0.5 mg/ml) | >240 |
| Paclitaxel (6.0 mg/ml) | >240 |
| Thiotepa (10.0 mg/ml) | 20.6 |
| Vincristine Sulfate (1.0 mg/ml) | >240 |
Caution: Testing showed a minimum breakthrough time of 7.3 minutes with Carmustine and 20.6 minutes with Thiotepa.
Warning: Do not use with Carmustine and Thiotepa.
Type of Use (Select one or both, as applicable)
☐
Prescription Use (Part 21 CFR 801 Subpart D)
☑
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 2 of 2
PSC Publishing Services (301) 443-6740
EF
{5}
# **SPECIAL 510(k) SUMMARY (K254285)**
**FOR**
**NITRILE POWDER FREE EXAMINATION GLOVES TESTED FOR USE WITH CHEMOTHERAPY DRUGS – AQUA BLUE**
**NITRILE POWDER FREE EXAMINATION GLOVES TESTED FOR USE WITH CHEMOTHERAPY DRUGS – VIOLET BLUE**
**NITRILE POWDER FREE EXAMINATION GLOVES TESTED FOR USE WITH CHEMOTHERAPY DRUGS – DARK BLUE**
(The information contained herein is being provided in accordance with the requirements of 21 CFR 807.92)
# **SUBMITTER INFORMATION**
Name : Hartalega NGC Sdn. Bhd.
Address : No. 1, Persiaran Tanjung,
Kawasan Perindustrian Tanjung,
43900 Sepang, Selangor Darul Ehsan,
Malaysia
Phone Number : (603) 8707 3000
Fax Number : (603) 6280 2533
Contact Person : Kuan Mun Leong
Date Prepared : June 22, 2026
# **CORRESPONDENT AND/OR PREPARER INFORMATION**
Contact Name : Nor Akmar Yusoff
Contact Title : Manager – Regulatory Affairs
Phone Number : (603) 8707 3000
Fax Number : (603) 6280 2533
Contact Email : akmar.yusoff@hartalega.com.my
# **DEVICE IDENTIFICATION**
Common Name of the Device : Examination Glove
Trade Name (Proprietary Name) : Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Aqua Blue;
Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Violet Blue;
Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Dark Blue
Device Class : 1
Product Code : LZA, LZC, OPJ
Regulation Number : 21 CFR 880.6250
{6}
# **PREDICATE DEVICE INFORMATION:**
| 510(k) Number | : | K162923 |
| --- | --- | --- |
| Manufacturer | : | Hartalega NGC Sdn. Bhd. |
| Trade Name (Proprietary Name) | : | Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Aqua Blue; Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Violet Blue; Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Dark Blue; Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Dark Violet Blue |
| Device Class | : | 1 |
| Product Code | : | LZA, LZC |
| Regulation Number | : | 21 CFR 880.6250 |
# **DESCRIPTION OF THE DEVICE:**
Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Aqua Blue is a disposable single-use, non- sterile, blue-colored and powder-free examination glove made from nitrile latex. Gloves meet the specification of ASTM D6319 and have been tested for resistance to permeation by chemotherapy drugs as per ASTM D6978.
Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Violet Blue is a disposable single-use, non- sterile, blue-colored and powder-free examination glove made from nitrile latex. Gloves meet the specification of ASTM D6319 and have been tested for resistance to permeation by chemotherapy drugs as per ASTM D6978.
Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Dark Blue is a disposable single-use, non- sterile, blue-colored and powder-free examination glove made from nitrile latex. Gloves meet the specification of ASTM D6319 and have been tested for resistance to permeation by chemotherapy drugs as per ASTM D6978.
# **DEVICE MODIFICATION:**
The proposed modification to the predicate devices is the addition of an expiration date labeling claim only. The proposed devices remain identical to the predicate devices with respect to materials, manufacturing process, packaging configuration, and bench performance. All other aspects including the Intended Use, Indications for Use, and technological characteristics remain unchanged.
# **INDICATIONS FOR USE:**
1) Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Aqua Blue is a non-sterile disposable device intended for medical purpose that is worn on the examiner’s hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs.
{7}
The list of Chemotherapy Drugs tested (with breakthrough time) are:
| Chemotherapy Drug and Concentration | Minimum Breakthrough Detection Time in Minutes |
| --- | --- |
| Carmustine (3.3 mg/ml) | 10.3 |
| Cisplatin (1.0 mg/ml) | > 240 |
| Cyclophosphamide (20.0 mg/ml) | > 240 |
| Dacarbazine (10.0 mg/ml) | > 240 |
| Doxorubicin HCl (2.0 mg/ml) | > 240 |
| Etoposide (20.0 mg/ml) | > 240 |
| Fluorouracil (50.0 mg/ml) | > 240 |
| Methotrexate (25.0 mg/ml) | > 240 |
| Mitomycin C (0.5 mg/ml) | > 240 |
| Paclitaxel (6.0 mg/ml) | > 240 |
| Thiotepa (10.0 mg/ml) | 28.8 |
| Vincristine Sulfate (1.0 mg/ml) | > 240 |
Caution: Testing showed a minimum breakthrough time of 10.3 minutes with Carmustine and 28.8 minutes with Thiotepa.
Warning: Do not use with Carmustine and Thiotepa.
2) Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Violet Blue is a non-sterile disposable device intended for medical purpose that is worn on the examiner’s hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs.
The list of Chemotherapy Drugs tested (with breakthrough time) are:
| Chemotherapy Drug and Concentration | Minimum Breakthrough Detection Time in Minutes |
| --- | --- |
| Carmustine (3.3 mg/ml) | 10.2 |
| Cisplatin (1.0 mg/ml) | > 240 |
| Cyclophosphamide (20.0 mg/ml) | > 240 |
| Dacarbazine (10.0 mg/ml) | > 240 |
| Doxorubicin HCl (2.0 mg/ml) | > 240 |
| Etoposide (20.0 mg/ml) | > 240 |
| Fluorouracil (50.0 mg/ml) | > 240 |
| Methotrexate (25.0 mg/ml) | > 240 |
| Mitomycin C (0.5 mg/ml) | > 240 |
| Paclitaxel (6.0 mg/ml) | > 240 |
| Thiotepa (10.0 mg/ml) | 30.2 |
| Vincristine Sulfate (1.0 mg/ml) | > 240 |
Caution: Testing showed a minimum breakthrough time of 10.2 minutes with Carmustine and 30.2 minutes with Thiotepa.
Warning: Do not use with Carmustine and Thiotepa.
{8}
3) Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Dark Blue is a non-sterile disposable device intended for medical purpose that is worn on the examiner’s hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs.
The list of Chemotherapy Drugs tested (with breakthrough time) are:
| Chemotherapy Drug and Concentration | Minimum Breakthrough Detection Time in Minutes |
| --- | --- |
| Carmustine (3.3 mg/ml) | 7.3 |
| Cisplatin (1.0 mg/ml) | > 240 |
| Cyclophosphamide (20.0 mg/ml) | > 240 |
| Dacarbazine (10.0 mg/ml) | > 240 |
| Doxorubicin HCl (2.0 mg/ml) | > 240 |
| Etoposide (20.0 mg/ml) | > 240 |
| Fluorouracil (50.0 mg/ml) | > 240 |
| Methotrexate (25.0 mg/ml) | > 240 |
| Mitomycin C (0.5 mg/ml) | > 240 |
| Paclitaxel (6.0 mg/ml) | > 240 |
| Thiotepa (10.0 mg/ml) | 20.6 |
| Vincristine Sulfate (1.0 mg/ml) | > 240 |
**Caution:** Testing showed a minimum breakthrough time of 7.3 minutes with Carmustine and 20.6 minutes with Thiotepa.
**Warning:** Do not use with Carmustine and Thiotepa.
{9}
# TECHNOLOGICAL CHARACTERISTICS COMPARISON TABLE:
| Characteristics and Parameters | Subject Device | Predicate Device (K162923) | Discussion |
| --- | --- | --- | --- |
| Trade Name | Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Aqua Blue; Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Violet Blue; Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Dark Blue | Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Aqua Blue; Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Violet Blue; Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Dark Blue; Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Dark Violet Blue | Same, except that the subject device does not include the Dark Violet Blue color variant. |
| Applicant | Hartalega NGC Sdn. Bhd. | Hartalega NGC Sdn. Bhd. | Same |
| Product Code | LZA, LZC, OPJ | LZA, LZC | Same |
| Classification | 1 | 1 | Same |
| Regulation Number | 21 CFR 880.6250 | 21 CFR 880.6250 | Same |
| Regulation Name | Non-Powdered Patient Examination Glove | Non-Powdered Patient Examination Glove | Same |
| Indications for Use | A non-sterile disposable device intended for medical purpose that is worn on the examiner’s hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs. | A non-sterile disposable device intended for medical purpose that is worn on the examiner’s hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs. | Similar, except that the subject device does not include the Dark Violet Blue color variant. |
| Chemotherapy Drug and Concentration | Minimum Breakthrough Detection Time in Minutes | Chemotherapy Drug and Concentration | Minimum Breakthrough Detection Time in Minutes |
| Aqua Blue | Violet Blue | Dark Blue | Aqua Blue | Violet Blue | Dark Blue | Dark Violet Blue |
| Carmustine (3.3 mg/ml) | 10.3 | 10.2 | 7.3 | Carmustine (3.3 mg/ml) | 10.3 | 10.2 | 7.3 | 0.2 |
| Cisplatin (1.0 mg/ml) | > 240 | > 240 | > 240 | Cisplatin (1.0 mg/ml) | > 240 | > 240 | > 240 | > 240 |
| Cyclophosphamide (20.0 mg/ml) | > 240 | > 240 | > 240 | Cyclophosphamide (20.0 mg/ml) | > 240 | > 240 | > 240 | > 240 |
| Dacarbazine (10.0 mg/ml) | > 240 | > 240 | > 240 | Dacarbazine (10.0 mg/ml) | > 240 | > 240 | > 240 | > 240 |
| Doxorubicin HCl (2.0 mg/ml) | > 240 | > 240 | > 240 | Doxorubicin HCl (2.0 mg/ml) | > 240 | > 240 | > 240 | > 240 |
| Etoposide (20.0 mg/ml) | > 240 | > 240 | > 240 | Etoposide (20.0 mg/ml) | > 240 | > 240 | > 240 | > 240 |
| Fluorouracil (50.0 mg/ml) | > 240 | > 240 | > 240 | Fluorouracil (50.0 mg/ml) | > 240 | > 240 | > 240 | > 240 |
| Methotrexate (25.0 mg/ml) | > 240 | > 240 | > 240 | Methotrexate (25.0 mg/ml) | > 240 | > 240 | > 240 | > 240 |
| Mitomycin C (0.5 mg/ml) | > 240 | > 240 | > 240 | Mitomycin C (0.5 mg/ml) | > 240 | > 240 | > 240 | > 240 |
| Paclitaxel (6.0 mg/ml) | > 240 | > 240 | > 240 | Paclitaxel (6.0 mg/ml) | > 240 | > 240 | > 240 | > 240 |
{10}
| Characteristics and Parameters | Subject Device | | | | Predicate Device (K162923) | | | | | Discussion |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Thiotepa (10.0 mg/ml) | 28.2 | 30.2 | 20.6 | Thiotepa (10.0 mg/ml) | 28.2 | 30.2 | 20.6 | 10.3 | |
| | Vincristine Sulfate (1.0 mg/ml) | > 240 | > 240 | > 240 | Vincristine Sulfate (1.0 mg/ml) | > 240 | > 240 | > 240 | > 240 | |
| | **Aqua Blue** Caution: Testing showed a minimum breakthrough time of 10.3 minutes with Carmustine and 28.2 minutes with Thiotepa. Warning: Do not use with Carmustine and Thiotepa **Violet Blue** Caution: Testing showed a minimum breakthrough time of 10.2 minutes with Carmustine and 30.2 minutes with Thiotepa. Warning: Do not use with Carmustine and Thiotepa **Dark Blue** Caution: Testing showed a minimum breakthrough time of 7.3 minutes with Carmustine and 20.6 minutes with Thiotepa. Warning: Do not use with Carmustine and Thiotepa | | | | **Aqua Blue** Caution: Testing showed an average breakthrough time of 10.3 minutes with Carmustine and 28.2 minutes with Thiotepa. Warning: Do not use with Carmustine Warning: Do not use with Thiotepa **Violet Blue** Caution: Testing showed an average breakthrough time of 10.2 minutes with Carmustine and 30.2 minutes with Thiotepa. Warning: Do not use with Carmustine Warning: Do not use with Thiotepa **Dark Blue** Caution: Testing showed an average breakthrough time of 7.3 minutes with Carmustine and 20.6 minutes with Thiotepa. Warning: Do not use with Carmustine Warning: Do not use with Thiotepa **Dark Violet Blue** Caution: Testing showed an average breakthrough time of 0.2 minutes with Carmustine and 10.3 minutes with Thiotepa. Warning: Do not use with Carmustine Warning: Do not use with Thiotepa | | | | | |
| **Type of use** | Over the counter use | | | | Over the counter use | | | | | Same |
| **Materials** | Nitrile | | | | Nitrile | | | | | Same |
| **Color** | Aqua Blue Violet Blue Dark Blue | | | | Aqua Blue Violet Blue Dark Blue Dark Violet Blue | | | | | Same, except that the subject device does not include the Dark Violet Blue color variant. |
| **Design** | Single Use Non-Sterile Powder-Free Ambidextrous | | | | Single Use Non-Sterile Powder-Free Ambidextrous | | | | | Same |
| **Sterility** | Non-Sterile | | | | Non-Sterile | | | | | Same |
| **Freedom from holes** | Meets ASTM D5151-19(2023): AQL 2.5 | | | | Meets D5151-06 (2011): AQL 2.5 | | | | | Same |
| **Length** | Meets ASTM D6319-19 (2023): ≥ 230 mm | | | | Meets ASTM D6319-10 (2015): ≥ 230 mm | | | | | Same |
| **Width** | Meets ASTM D6319-19 (2023): XS- 60-80 mm S – 70-90 mm M – 85-105 mm L – 100- 120mm XL – 110-130 mm XXL – 115-135 mm | | | | Meets ASTM D6319-10 (2015): XS- 60-80 mm S – 70-90 mm M – 85-105 mm L – 100- 120mm XL – 110-130 mm XXL – 115-135 mm | | | | | Same |
{11}
| Characteristics and Parameters | Subject Device | Predicate Device (K162923) | Discussion |
| --- | --- | --- | --- |
| Thickness | Meets ASTM D6319-19 (2023): Palm Thickness: ≥ 0.05 mm Finger Thickness: ≥ 0.05 mm | Meets ASTM D6319-10 (2015): Palm Thickness: ≥ 0.05 mm Finger Thickness: ≥ 0.05 mm | Same |
| Physical Properties | Meets ASTM D6319-19 (2023): Tensile Strength Before Aging: ≥ 14 MPa Tensile Strength After Aging: ≥ 14 MPa Ultimate Elongation Before Aging: ≥ 500 % Ultimate Elongation After Aging: ≥ 400 % | Meets ASTM D6319-10 (2015): Tensile Strength Before Aging: ≥ 14 MPa Tensile Strength After Aging: ≥ 14 MPa Ultimate Elongation Before Aging: ≥ 500 % Ultimate Elongation After Aging: ≥ 400 % | Same |
| Powder residual | Meets ASTM D6319-19 (2023) & ASTM D6124-06 (2022): Residual Powder: ≤ 2 mg per glove | Meets ASTM D6319-10 (2015) & ASTM D6124-06 (2011): Residual Powder: ≤ 2 mg per glove | Same |
| Primary Skin Irritation ISO 10993-10/ ISO 10993-23 | Under the conditions of the study, the device is not an irritant | Under the conditions of the study, the device is not an irritant | Same |
| Dermal Sensitization ISO 10993-10 | Under the conditions of the study, the device is not a sensitizer | Under the conditions of the study, the device is not a sensitizer | Same |
| Expiration date labelling claim | 5 years of shelf life | - | The subject and predicate devices are identical in all technological characteristics. The only difference is the addition of a 5-year expiration date labelling claim. Real-time aging data support this labeling change, and it does not affect the safety or effectiveness of the device. |
{12}
## SUMMARY OF NON-CLINICAL TESTING:
The subject devices and the predicate devices share the same intended use, identical material and identical manufacturing, and therefore share the same chemotherapy drug permeation testing, biocompatibility characteristics and compliance with ASTM D6319.
The expiration date labeling claim is supported by 5-year real-time shelf-life testing. Real-time stability testing was conducted in accordance with applicable ASTM standards, including ASTM D7160, ASTM D6319, and ASTM D5151, and demonstrated continued compliance with physical property and barrier integrity requirements through the labeled shelf life.
| Testing Performed | Purpose | Criteria | Result |
| --- | --- | --- | --- |
| Real-Time 5-Year Shelf-Life Testing (ASTM D7160)Physical Properties & Barrier IntegrityASTM D6319 & ASTM D5151 | To demonstrate that long-term aging does not impact physical properties or barrier integrity | - Conform to ASTM D6319 tensile strength of at least minimum 14 MPa and ultimate elongation of at least minimum 500%- Freedom from holes per ASTM D5151 (AQL ≤2.5) | Pass |
| *Chemotherapy Drug Permeation (ASTM D6978) | To confirm that 5-year real-time aging does not adversely affect chemotherapy drug resistance | Continued support of previously cleared chemotherapy drugs resistance labeling | No adverse impact on chemotherapy drug permeation performance was observed after 5 years of real-time aging; results continue to support the previously cleared labeling claims |
*Chemotherapy drug permeation testing in accordance with ASTM D6978 was performed on gloves aged through the full 5-year real-time shelf life.
The following testing was previously performed on the physically identical predicate device.
| Test | Purpose | Criteria | Result |
| --- | --- | --- | --- |
| Standard Test Method for Detection of Holes in Medical Gloves ASTM D6319-10 (2015) | To demonstrate glove integrity | Freedom from holes AQL 2.5 | Pass |
| Standard Test Method for Residual Powder on Medical Gloves ASTM D6124-06 (2011) | To demonstrate the gloves are ‘powder free’ | Average less than 2 mg/glove | Pass |
| Dimensional Conformance ASTM D6319-10 (2015) | To demonstrate appropriate dimensions for labeled sizes | Conforms to ASTM D6319 width, thickness, and length requirements for XS, S, M, L, XL and XXL AQL 4 | Pass |
| Tensile Performance ASTM D6319-10 (2015) | To demonstrate adequate tensile properties | Conforms to ASTM D6319 requirements with tensile strength of at least 14MPa and ultimate elongation of at least 500% prior to aging, and tensile strength of at least 14MPa and ultimate elongation of at least 400% after accelerated aging AQL 4 | Pass |
| Biocompatibility: Skin Irritation ISO 10993-10 | To demonstrate low potential for skin irritation | Under the conditions of the study, not an irritant | Pass |
| Biocompatibility: Skin Sensitization ISO 10993-10 | To demonstrate low potential for skin sensitization | Under the conditions of the study, not a sensitizer | Pass |
| ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs | Demonstrate barrier properties of gloves to permeation of chemotherapy drugs:Carmustine (3.3 mg/ml)Cisplatin (1.0 mg/ml)Cyclophosphamide (20.0 mg/ml)Dacarbazine (10.0 mg/ml) | N/A | Carmustine Minimum breakthrough time:Aqua Blue: 10.3 minViolet Blue: 10.2 minDark Blue: 7.3 min |
{13}
| Test | Purpose | Criteria | Result |
| --- | --- | --- | --- |
| | Doxorubicin HCl (2.0 mg/ml) Etoposide (20.0 mg/ml) Fluorouracil (50.0 mg/ml) Methotrexate (25.0 mg/ml) Mitomycin C (0.5 mg/ml) Paclitaxel (6.0 mg/ml) Thiotepa (10.0 mg/ml) Vincristine Sulfate (1.0 mg/ml) | | **Thiotepa Minimum breakthrough time:** Aqua Blue: 28.2 min Violet Blue: 30.2 min Dark Blue: 20.6 min No breakthrough detected during 240-minute test duration for remaining tested chemotherapy drugs |
### CLINICAL PERFORMANCE DATA:
Not applicable. Clinical performance data were not required to support the subject devices, as the indications for use are equivalent to those of the predicate devices.
### CONCLUSION:
The conclusion drawn from the non-clinical testing demonstrate that the subject devices are as safe, as effective, and performs as well as or better than the legally marketed predicate device K162923, Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Aqua Blue, Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Violet Blue and Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs – Dark Blue.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.