The Embrace Neonatal MRI System is indicated for use as a magnetic resonance imaging device for producing axial, sagittal, coronal and oblique images that display the internal structure of neonatal head with circumference of up to 38 cm and weight between 1Kg and 4.5 Kg. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
Device Story
Embrace Neonatal MRI System is a 1 Tesla permanent magnet MRI; captures MR images of neonatal head. Device utilizes static magnetic field, RF pulses, and gradient magnetic fields to induce precession of magnetization; signals sampled and processed into image data. Used in hospital settings; operated by trained clinicians. System includes integrated patient bed with temperature control (20.5°C–36.5°C). Clinicians view images on 24-inch LED display via workstation to assist in diagnostic decision-making. Benefits include non-invasive visualization of neonatal internal head structures. Revision E introduces software enhancements for usability and cybersecurity, plus updated imaging sequences (e.g., Propeller FSE, TOF).
Clinical Evidence
Bench testing only. Performance validated via NEMA MS-1, MS-2, MS-3, MS-5, MS-12, MS-14, MS-4, and MS-8 standards. Software verification and validation, biocompatibility testing (ISO 10993-1), and simulated use validation performed. No clinical data presented.
Technological Characteristics
1 Tesla permanent magnet MRI; 150 mT/m gradient strength; 500 T/m/sec slew rate. Includes integrated patient bed with air-based temperature control. Connectivity via Windows PC workstation. Software life-cycle per IEC 62304. Biocompatibility per ISO 10993-1. Standards: IEC 60601-1, 60601-1-2, 60601-2-33, 60601-2-20, 60601-1-6.
Indications for Use
Indicated for neonates with head circumference up to 38 cm and weight between 1 kg and 4.5 kg requiring MR imaging of the head.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 13, 2026
Aspect Imaging, Ltd.
% Alex Cadotte
Vice President, Digital Health, AI, and Radiology
Mcra
803 7th St. NW
Washington, District of Columbia 20001
Re: K254277
Trade/Device Name: Embrace Neonatal MRI System
Regulation Number: 21 CFR 892.1000
Regulation Name: Magnetic Resonance Diagnostic Device
Regulatory Class: Class II
Product Code: LNH, MOS
Dated: December 9, 2025
Received: December 30, 2025
Dear Alex Cadotte:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254277 - Alex Cadotte
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K254277 - Alex Cadotte
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Daniel M. Krainak, Ph.D.
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K254277 | |
| Device Name Embrace Neonatal MRI System | |
| Indications for Use (Describe) The Embrace Neonatal MRI System is indicated for use as a magnetic resonance imaging device for producing axial, sagittal, coronal and oblique images that display the internal structure of neonatal head with circumference of up to 38 cm and weight between 1Kg and 4.5 Kg. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K254277
# 510(k) Summary
Device Trade Name: Embrace Neonatal MRI System
Manufacturer: Aspect Imaging LTD
27 Shaked St.
Industrial Area Hevel Modi'in
Shoham 60850, Israel
Contact Person: Yaron Eshel
VP, Quality Assurance and
Regulatory Affairs
Yaron.Eshel@aspectimaging.com
Prepared by: MCRA, LLC
803 7th Street, NW, 3rd Floor
Washington, DC 20001
Office: 202.552.5800
Date Prepared: March 10, 2026
Classifications: 21 CFR 892.1000
Class: II
Product Codes: LNH, MOS
Primary Predicate: Embrace Neonatal MRI System (Aspect Imaging, K170978)
Page 1 of 8
{5}
Page 2 of 8
# Indications For Use:
The Embrace Neonatal MRI System is indicated for use as a magnetic resonance imaging device for producing axial, sagittal, coronal and oblique images that display the internal structure of neonatal head with circumference of up to 38 cm and weight between 1Kg and 4.5 Kg. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
# Device Description
The Embrace Neonatal MRI System is a 1 Tesla Permanent magnet MRI system producing MR images of the neonatal head. During MRI scan, body parts to be imaged are held within a uniform static magnetic field, and are subject to sequences of RF pulses and gradient magnetic fields. The signal from the precession of the magnetization created by these fields is sampled and processed to produce image data.
# Predicate Device:
The purpose of this Traditional 510(k) is to demonstrate that the Embrace Neonatal MRI System (Revision E) is substantially equivalent to predicate device for the purpose of introducing the system to interstate commerce. It has been determined that the Embrace Neonatal MRI System (Revision E) are substantially equivalent to the following predicate:
- Primary Predicate: Embrace Neonatal MRI System (Aspect Imaging, K170978)
| Specification | Predicate Device Embrace Neonatal MRI System (K170978) | Subject Device Embrace Neonatal MRI System | Discussion of Differences |
| --- | --- | --- | --- |
| Product Code | LNH, MOS | LNH, MOS | Same |
| Classification / Regulation | Magnetic Resonance Diagnostic Device (21 CFR.892.1000) | Magnetic Resonance Diagnostic Device (21 CFR 892.1000) | Same |
{6}
| Specification | Predicate Device Embrace Neonatal MRI System (K170978) | Subject Device Embrace Neonatal MRI System | Discussion of Differences |
| --- | --- | --- | --- |
| Intended Use/Indication for Use | The Embrace Neonatal MRI System is indicated for use as a magnetic resonance imaging device for producing axial, sagittal, coronal and oblique images that display the internal structure of neonatal head with circumference of up to 38 cm and weight between 1Kg and 4.5 Kg. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis. | The Embrace Neonatal MRI System is indicated for use as a magnetic resonance imaging device for producing axial, sagittal, coronal and oblique images that display the internal structure of neonatal head with circumference of up to 38 cm and weight between 1Kg and 4.5 Kg. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis. | Same |
| Patient Population | Patients requiring MR images of the Neonatal Head | Patients requiring MR images of the Neonatal Head | Same |
| Anatomical Sites | Neonatal Head | Neonatal Head | Same |
| Environment of Use | Hospital setting | Hospital setting | Same |
| Energy Used and/or delivered | Magnetic Resonance | Magnetic Resonance | Same |
| Human Factors | The Embrace Neonatal MRI System is designed similar to other commercially available MRI systems and therefore is familiar and easy for use for the user. Furthermore, the device contains a user-friendly software interface through which the user may easily access all device functions. | The Embrace Neonatal MRI System is designed similar to other commercially available MRI systems and therefore is familiar and easy for use for the user. Furthermore, the device contains a user-friendly software interface through which the user may easily access all device functions. | Same |
| Hardware Specifications: | | | |
| Magnet -Physical | | | Same |
Page 3 of 8
{7}
| Specification | Predicate Device Embrace Neonatal MRI System (K170978) | Subject Device Embrace Neonatal MRI System | Discussion of Differences |
| --- | --- | --- | --- |
| Diminutions | 171×145×229 cm | 171×145×229 cm | |
| -Bore Opening | 184×260mm | 184×260mm | |
| -Weight | 5500 (5680 with patient bed) Kg | 5500 (5680 with patient bed) Kg | |
| -Field Strength | 1 Tesla Permanent Magnet | 1 Tesla Permanent Magnet | |
| Gradient | | | |
| -Strength | 150 mT/m | 150 mT/m | Same |
| -Rise Time | 0.300mSec | 0.300mSec | |
| -Slew Rate | 500 T/m/Sec | 500 T/m/Sec | |
| IT infrastructure | Windows PC on workstation table | Windows PC on workstation table | Revision E contains software enhancements for usability and strengthened cybersecurity |
| Computer Display | 24” LED Display | 24” LED Display | Same |
| RF Coils | 1 Head Coils | 1 Head Coils | |
| -Coil Type | TX/RX | TX/RX | |
| -Coil Geometry | Cylindrical | Cylindrical | Same |
| -Inner dimensions (mm) | 143 Diameter | 143 Diameter | |
| -Coil Design | Linear Volume | Linear Volume | |
| Patient Bad | Integrated patient bed | Integrated patient bed | Improvements to the bed for usability and serviceability are included in revision E |
| Target population | Neonates with head circumference of up to 38 cm and weight | Neonates with head circumference of up to 38 cm and weight | Same |
Page 4 of 8
{8}
| Specification | Predicate Device Embrace Neonatal MRI System (K170978) | Subject Device Embrace Neonatal MRI System | Discussion of Differences |
| --- | --- | --- | --- |
| | between 1Kg and 4.5Kg | between 1Kg and 4.5Kg | |
| -Dimensions | 60.6cm W x 120cm H x 140cm L | 60.6cm W x 120cm H x 140cm L | |
| -Patient weight capacity | 4.5 Kg Max | 4.5 Kg Max | Same |
| -Compartment length | 58 cm | 58 cm | |
| -Set Temperature | 20.5°C - 36.5°C | 20.5°C - 36.5°C | |
| -Warm Up time | 45 minutes | 45 minutes | |
| -Temperature control | Air | Air | Same |
| -Humidity control | No | No | |
| -MRI compatibility | Aspect Embrace 1T | Aspect Embrace 1T | |
| Performance Testing | | | |
| Imaging Performance: (Recognition no. in brackets) | • NEMA MS 1 - Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging (12-188)
• NEMA MS 12 - Quantification and Mapping of Geometric Distortion for Special Applications (12-306)
• NEMA MS 2 - Determination of Two-Dimensional Geometric Distortion in Diagnostic Magnetic Resonance Images (12-196)
• NEMA MS 3 - Determination of Image Uniformity in Diagnostic Magnetic Resonance Images (12-187)
• NEMA MS 5 - Determination of Slice | • NEMA MS 1 - Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging (12-188)
• NEMA MS 12 - Quantification and Mapping of Geometric Distortion for Special Applications (12-306)
• NEMA MS 14 - Characterization of Radiofrequency (RF) Coil Heating in Magnetic Resonance Imaging Systems (12-331)
• NEMA MS 2 - Determination of Two-Dimensional Geometric Distortion in Diagnostic Magnetic Resonance Images (12-196)
• NEMA MS 3 - | MS-14 was added to conform with FDA applicable standards |
Page 5 of 8
{9}
| Specification | Predicate Device Embrace Neonatal MRI System (K170978) | Subject Device Embrace Neonatal MRI System | Discussion of Differences |
| --- | --- | --- | --- |
| | Thickness in Diagnostic Magnetic Resonance Imaging (12-322) | Determination of Image Uniformity in Diagnostic Magnetic Resonance Images (12-187)
• NEMA MS 5 - Determination of Slice Thickness in Diagnostic Magnetic Resonance Imaging (12-322) | |
| Safety Performance: (Recognition no. in brackets) | | • NEMA MS 4 - Acoustic Noise Measurement Procedure for Diagnostic Magnetic Resonance Imaging Devices (12-232)
• NEMA MS 8 - Characterization of the Specific Absorption Rate (SAR) for MRI Systems (12-315) | MS-4 and MS -8 were added to conform with FDA safety standards |
| Electromagnetic Compatibility, Electrical, Mechanical, and Thermal Safety | | | |
| Voluntary consensus standards (Recognition no. in brackets) | • IEC 60601-1 - Medical equipment/medical electrical equipment - Part 1: General requirements for basic safety and essential performance (19-37)
• IEC 60601-1-2 - Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard:
Electromagnetic compatibility - Requirements and tests (19-36)
• IEC 60601-2-33 - Medical electrical equipment - Part 2-33: Particular requirements for the basic | • IEC 60601-1 - Medical equipment/medical electrical equipment - Part 1: General requirements for basic safety and essential performance (19-37)
• IEC 60601-1-2 - Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard:
Electromagnetic compatibility - Requirements and tests (19-36)
• IEC 60601-2-33 - Medical electrical equipment - Part 2-33: Particular requirements for the | Voluntary consensus standards (Recognition no. in brackets) |
Page 6 of 8
{10}
| Specification | Predicate Device Embrace Neonatal MRI System (K170978) | Subject Device Embrace Neonatal MRI System | Discussion of Differences |
| --- | --- | --- | --- |
| | safety and essential performance of magnetic resonance equipment for medical diagnosis (12-347)
• IEC 60601-2-20 - Medical Electrical Equipment Part 2-20: Particular requirements for the basic safety and essential performance of infant transport incubators (6-462)
• IEC 60601-1-6 - Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability (5-132)
• IEC 62304 - Software life-cycle processes (13-79) | basic safety and essential performance of magnetic resonance equipment for medical diagnosis (12-347)
• IEC 60601-2-20 - Medical Electrical Equipment Part 2-20: Particular requirements for the basic safety and essential performance of infant transport incubators (6-462)
• IEC 60601-1-6 - Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability (5-132)
• IEC 62304 - Software life-cycle processes (13-79) | |
| Sequence available to the User | • Gradient Echo
• Gradient Echo 3D
• Spin Echo
• Fast Spin Echo
• ADC Map Single Shot Splice Axial
• ADC Map Spin Echo | • 2D Spin Echo
• (SE) 2D Fast Spin Echo (FSE)
• 3D Fast Spin Echo (FSE)
• 3D Gradient Echo (GRE)
• ADC map (Diffusion Weighted imaging with ADC mapping)
• 2D Time of Flight (TOF)
• 3D Time of Flight (TOF) 2D
• Propeller FSE
• 2D ADC Propeller | Modification of sequences for the user |
Software: In addition, system and software verification and Validation testing were performed to demonstrate performance of the device as part of design controls activity.
Biocompatibility: Patient-contacting materials were assessed per FDA Guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". Patient contact surfaces were classified based on type of contact and duration. All surfaces are considered surface contacting for durations less than or equal to 24
Page 7 of 8
{11}
hours. Where required, evidence of testing was provided for endpoints including cytotoxicity, sensitization, and irritation.
All patient contact surfaces of the Embrace Neonatal MRI System (Revision E) meet biocompatibility requirements consistent with contact type and duration.
The following design control activities were applied to the development of the system:
- Risk Management
- Requirements Management
- Design Reviews
- Unit level module verification
- System Integration verification
- Performance verification (see table above for specific testing performed)
- Safety and EMC testing
- Simulated use validation testing
- Sample Phantom images
## Substantial Equivalence:
The subject device was demonstrated to be substantially equivalent to the predicate cited above with respect to indications, design materials, function, manufacturing, and performance. The non-clinical tests performed by the company demonstrated that any differences in the Embrace Neonatal MRI System (Revision E) do not raise new questions of safety and effectiveness.
## Conclusion:
The subject device and the predicate device have the same intended use, have similar technological characteristics, and are made of the same materials. The subject and predicate devices have similar software and EMC characteristics that were validated via new testing. The data included in this submission demonstrates substantial equivalence to the predicate device listed above. The Embrace Neonatal MRI System (Revision E) is as safe, as effective, and performs as well as, or better, than the predicate device.
Page 8 of 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.