K254276 · Q’Apel Medical, Inc. · NRY · May 7, 2026 · Cardiovascular
Device Facts
Record ID
K254276
Device Name
Lynx Aspiration Catheter System
Applicant
Q’Apel Medical, Inc.
Product Code
NRY · Cardiovascular
Decision Date
May 7, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Lynx Aspiration Catheter with the Aspiration Tubing and a compatible suction pump is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who fail thrombolytic drug therapy are candidates for treatment. The Aspiration Tubing is indicated to connect the Lynx Aspiration Catheter to the compatible suction pump.
Device Story
Lynx Aspiration Catheter System removes thrombus from neurovasculature via aspiration. System includes single-lumen, variable-stiffness catheter, aspiration tubing, and compatible suction pump. Catheter features hydrophilic coating, distal radiopaque marker, and proximal luer hub. Operated by physicians in clinical settings; introduced via guide catheter or long sheath to target occlusion. Aspiration applied through catheter lumen to pull thrombus into device. Output is visual confirmation of clot removal via fluoroscopy. Benefits include revascularization of occluded intracranial vessels in acute ischemic stroke patients.
Clinical Evidence
No clinical data. Evidence includes bench testing (dimensional, friction, radiopacity, simulated use, lumen collapse, particulate/coating integrity, leakage, tensile, torque, burst, kink resistance) and a GLP porcine animal study (n=6) comparing the 072 model to the Penumbra RED 72. Animal study endpoints included angiographic, gross necropsy, and histopathological assessment of vascular injury; results showed comparable safety profiles.
Indicated for revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (internal carotid, M1/M2 middle cerebral, basilar, and vertebral arteries) within 8 hours of symptom onset. Candidates include patients ineligible for or failing thrombolytic drug therapy.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Penumbra System (Reperfusion Catheter RED 72) (K242104)
Penumbra System (Reperfusion Catheter RED 62) (K203440)
Penumbra System (Reperfusion Catheter RED 43) (K222808)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 7, 2026
Q'Apel Medical, Inc.
Jennifer Mateus
Vice President, Regulatory, Quality and Clinical Affairs
4245 Technology Drive
Fremont, California 94538
Re: K254276
Trade/Device Name: Lynx Aspiration Catheter System
Regulation Number: 21 CFR 870.1250
Regulation Name: Percutaneous Catheter
Regulatory Class: Class II
Product Code: NRY
Dated: December 29, 2025
Received: December 30, 2025
Dear Jennifer Mateus:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254276 - Jennifer Mateus
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K254276 - Jennifer Mateus
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
NAIRA MURADYAN -S
Naira Muradyan, Ph.D.
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K254276 | |
| Device Name Lynx™ Aspiration Catheter System | |
| Indications for Use (Describe) The Lynx Aspiration Catheter with the Aspiration Tubing and a compatible suction pump is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who fail thrombolytic drug therapy are candidates for treatment. The Aspiration Tubing is indicated to connect the Lynx Aspiration Catheter to the compatible suction pump. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Page 1 of 8
510(k) Summary
K254276
Lynx™ Aspiration Catheter System
Date Prepared: May 6, 2026
I. SUBMITTER
Q'Apel Medical, Inc.
4245 Technology Drive
Fremont, CA 94538 United States
Contact Person
Jennifer Mateus
Vice President, Regulatory, Quality and Clinical Affairs
4245 Technology Drive
Fremont, CA 94538 United States
408-896-4940
jmateus@qapelmedical.com
II. DEVICE
Name of Device: Lynx™ Aspiration Catheter System
Common or Usual Name: Catheter, Thrombus Retriever
Classification Regulation: 21 CFR § 870.1250
Regulation Name: Percutaneous Catheter
Product Codes: NRY
Regulatory Class: II
III. PREDICATES
- Penumbra System (Reperfusion Catheter RED 72), K242104
- Penumbra System (Reperfusion Catheter RED 62), K203440
- Penumbra System (Reperfusion Catheter RED 43), K222808
IV. DEVICE DESCRIPTION
The Lynx™ Aspiration Catheter (Lynx) is intended for the removal of thrombus from the neurovasculature under aspiration when used in conjunction with the Aspiration Tubing and a commercially available compatible suction pump. The Lynx is a single-lumen, variable-stiffness catheter with a hydrophilic coating to minimize friction during navigation. The catheter features a radiopaque marker at the distal end for angiographic visualization and a luer hub at the proximal end for flushing and aspiration connections. The Lynx is introduced through a guide catheter or long sheath into the intracranial vasculature and navigated to the target occlusion with the aid of common interventional devices.
The Lynx is offered in various working lengths, nominal inner diameters (IDs) and outer diameters (ODs), and is designed for either individual or coaxial use to assist in accessing the target vasculature (aspiration is solely applied through the inner catheter when used coaxially).
The Lynx is provided sterile, non-pyrogenic, and is intended for single use only.
{5}
The Lynx is packaged with the following accessories: Rotating Hemostasis Valve (RHV), Aspiration Tubing with Flow Switch and Aspiration Tubing Extension, and Scale Card.
# V. INDICATIONS FOR USE
The Lynx Aspiration Catheter with the Aspiration Tubing and a compatible suction pump is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who fail thrombolytic drug therapy are candidates for treatment.
The Aspiration Tubing is indicated to connect the Lynx Aspiration Catheter to the compatible suction pump.
# VI. COMPARISION OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATES
The Lynx and the predicate devices, Penumbra System with Reperfusion Catheters RED 72 (K242104), RED 62 (K203440) and RED 43 (K222808), share the same fundamental design and technological characteristics, similar materials, and sterilization method.
The following table compares key features of the subject and predicate devices.
| Characteristic | Subject Device Lynx™ Aspiration Catheter System | Predicate Device Penumbra System (Reperfusion Catheter RED 72) K242104 | Predicate Device Penumbra System (Reperfusion Catheter RED 62) K203440 | Predicate Device Penumbra System (Reperfusion Catheter RED 43) K222808 |
| --- | --- | --- | --- | --- |
| Indications for Use | The Lynx Aspiration Catheter with the Aspiration Tubing and a compatible suction pump is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who fail thrombolytic drug therapy are candidates for treatment. The Aspiration Tubing is indicated to connect the Lynx Aspiration Catheter to the compatible suction pump. | Penumbra Reperfusion Catheters and Separators As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment. Penumbra 3D Revascularization Device As part of the Penumbra | Penumbra Reperfusion Catheters and Separators As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment. | Penumbra Reperfusion Catheters and Separators As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment. |
{6}
| Characteristic | Subject Device Lynx™ Aspiration Catheter System | Predicate Device Penumbra System (Reperfusion Catheter RED 72) K242104 | Predicate Device Penumbra System (Reperfusion Catheter RED 62) K203440 | Predicate Device Penumbra System (Reperfusion Catheter RED 43) K222808 |
| --- | --- | --- | --- | --- |
| | | System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment. **Penumbra Aspiration Tubing** As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump. **Penumbra Aspiration Pump** The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems. | **Penumbra 3D Revascularization Device** As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment. **Penumbra Aspiration Tubing** As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump. **Penumbra Aspiration Pump** The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems. | **Penumbra 3D Revascularization Device** As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment. **Penumbra Aspiration Tubing** As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump. **Penumbra Aspiration Pump** The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspirati on Systems. |
| Product Code | NRY | NRY | NRY | NRY |
| Classification | Class II | Class II | Class II | Class II |
Page 3 of 8
{7}
| Characteristic | Subject Device Lynx™ Aspiration Catheter System | Predicate Device Penumbra System (Reperfusion Catheter RED 72) K242104 | Predicate Device Penumbra System (Reperfusion Catheter RED 62) K203440 | Predicate Device Penumbra System (Reperfusion Catheter RED 43) K222808 |
| --- | --- | --- | --- | --- |
| Materials | Stainless Steel laser cut tube LCT, Nitinol LCT, Nylon, Pebax, polytetrafluoroethylene (PTFE), Polycarbonate, Polyurethane (NEUsoft), 25 Denier Vectran Fiber (072 Aspiration Catheter only) | Stainless Steel, PTFE, Polyurethane, Polyether Block Amide, Nylon 12, Nitinol, Platinum/Iridium | Stainless Steel, PTFE, Polyurethane, Polyether Block Amide, Nylon 12, Nitinol, Platinum/Iridium | Stainless Steel, PTFE, Polyurethane, Polyether Block Amide, Nylon 12, Nitinol, Platinum/Iridium |
| Aspiration Catheter Design | Single lumen, variable stiffness | Single lumen, variable stiffness | Single lumen, variable stiffness | Single lumen, variable stiffness |
| Distal Radiopaque Marker | Yes | Yes | Yes | Yes |
| Marker Band | Platinum (90%)/Iridium (10%) | Platinum (90%) /Iridium (10%) | Platinum (90%) /Iridium (10%) | Platinum (90%) /Iridium (10%) |
| ID Band Color | Proflex [green] | Polyolefin, PET black [white foil] | Polyolefin, PET black [white text] | Polyolefin, PET black [white text] |
| Inner Diameter | 072: 0.072 in. 058: 0.058 in. 044: 0.044 in. | RED 72: 0.072 in | RED 62: 0.062 in. | RED 43: 0.043 in. |
| Outer Diameter | 072: 0.084 in 058: 0.070 in 044: 0.056 in | RED 72: 0.085 in | RED 62: 0.076 in. | RED 43: Proximal OD 0.060 in. Distal OD 0.056 in. |
| Effective Length | 072: 132 cm 058: 138 cm, 149 cm 044: 145 cm, 160 cm | RED 72: 115cm, 120cm, 125cm, 127cm, 132 cm | RED 62: 115, 120, 125, 127,132, 138, 160 cm | RED 43: 115, 120, 125, 127, 132, 136, 138, 145, 150, 153, 155, 160, 162, 167 cm |
| Hydrophilic Coating | Hydrophilic (proprietary) | Hydrophilic (proprietary) | Hydrophilic (proprietary) | Hydrophilic (proprietary) |
| Hydrophilic Coating Length | 072 and 058: 30 cm 044: 50 cm | RED 72: 30 cm | RED 62: 30 cm | RED 43: 60 cm |
| Tip Shapes Offered | Straight | Shapeable | Shapeable | Shapeable |
| Accessories Supplied | RHV, Aspiration Tubing with Flow Switch and Aspiration Tubing Extension, Scale Card | RHV, Peelable Sheath, Shaping Mandrel, SENDit (optional), Aspiration Tubing with Flow Switch (integral to the tubing) | RHV, Peelable Sheath, Shaping Mandrel, Aspiration Tubing with Flow Switch (integral to the tubing) | RHV, Peelable Sheath, Shaping Mandrel, Aspiration Tubing with Flow Switch (integral to the tubing) |
| Sterile Aspiration Tubing | Yes | Yes | Yes | Yes |
| Aspiration Tubing Inner Diameter | 0.110 in. | 0.110 in. | 0.110 in. | 0.110 in. |
| Aspiration Tubing Effective Length | 98 in. | 112 in. | 112 in. | 112 in. |
| Aspiration Pump Vacuum Pressures | Between -20 inHg and -29 inHg | Between -20 inHg and -29 inHg | Between -20 inHg and -29 inHg | Between -20 inHg and -29 inHg |
| Packaging Materials | Pouch: Tyvek (polyethylene), Mylar (Polyester) Backer Card: PETG (Polyethylene | Pouch: Polyester / Polyethylene/Tyvek Tray: Polystyrene Carton: SBS Paperboard | Pouch: Polyester / Polyethylene/ Tyvek Tray: Polystyrene Carton: SBS Paperboard | Pouch: Polyester /Polyethylene/ Tyvek Tray: Polystyrene Carton: SBS Paperboard |
Page 4 of 8
{8}
| Characteristic | Subject Device Lynx™ Aspiration Catheter System | Predicate Device Penumbra System (Reperfusion Catheter RED 72) K242104 | Predicate Device Penumbra System (Reperfusion Catheter RED 62) K203440 | Predicate Device Penumbra System (Reperfusion Catheter RED 43) K222808 |
| --- | --- | --- | --- | --- |
| | terephthalate glycol) Carton: SBS Paperboard | | | |
| Condition Supplied | Sterile | Sterile | Sterile | Sterile |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide |
Page 5 of 8
{9}
# VII. PERFORMANCE DATA
The results of verification and validation testing conducted on the Lynx demonstrate that the device performs as intended and is substantially equivalent to the predicate. The following performance data were provided in support of the substantial equivalence determination:
| Test Name | Objective | Result |
| --- | --- | --- |
| Dimensional/Visual Test | To demonstrate the device meets all dimensional and visual product specifications. | The device met the acceptance criteria. |
| Friction Test (Coating Lubricity) | To evaluate the catheter coating lubricity (frictional forces). | The device met the acceptance criteria. |
| Radiopacity Test | To demonstrate that the catheter's distal marker band is clearly visible under typical fluoroscopic imaging conditions. | The device met the acceptance criteria. |
| Simulated Use Test | To demonstrate the device can remove clots using a clinically relevant benchtop model. | The device performs as intended under simulated use conditions. |
| Delivery and Retrieval Test | To demonstrate the catheter meets delivery and retrieval force specifications. | The device met the acceptance criteria. |
| Lumen Collapse | To demonstrate the device lumen (catheter and aspiration tubing) does not collapse under aspiration pressures expected in clinical use. | The device met the acceptance criteria. |
| Particulate and Coating Integrity Test | To demonstrate the quantity and size of particles generated during simulated use (including multiple deployment cycles) are comparable to the predicates and reference devices. To demonstrate the catheter hydrophilic coating has sufficient integrity after simulated use. | The device met the acceptance criteria. |
| Hub Liquid Leakage and Air Leakage Tests | To demonstrate the device (catheter) has no liquid leakage or air leakage per test methods described in ISO 10555-1. | The device met the acceptance criteria. |
| Functionality and Leakage Test (Aspiration Tubing) | To demonstrate functionality of the flow switch to control fluid flow of aspiration tubing. To demonstrate resistance of the aspiration tubing to leakage under vacuum. | The device met the acceptance criteria. |
| Tensile Test | To demonstrate the device (catheter, aspiration tubing) has adequate tensile strength including all bonds and joints. | The device met the acceptance criteria. |
| Elongation Test | To demonstrate the device achieves adequate elongation before breaking under tensile test conditions. | The device met the acceptance criteria. |
| Corrosion Resistance Test | To demonstrate that the catheter has no visual evidence of corrosion when tested per methods described in ISO 10555-1. | The device met the acceptance criteria. |
| Torque Strength Test | To demonstrate the catheter has sufficient torque strength in the simulated use model. | The device met the acceptance criteria. |
| Burst Pressure Test | To demonstrate the catheter can withstand sufficient pressure for clinical use. | The device met the acceptance criteria. |
| Distal Tip Stiffness Test | To demonstrate the catheter distal tip stiffness is comparable to the predicate. | The device met the acceptance criteria. |
| Kink Resistance Test | To demonstrate the catheter is resistant to kinking. | The device met the acceptance criteria. |
| Luer Hub Tests | To demonstrate conformance with ISO 80369-7 for small bore connectors for intravascular use. | The device met the acceptance criteria. |
| Simulated Use | To demonstrate the device usability in removing | The device performs as intended under |
| | simulated device specifications. | simulated device specifications. |
| Tensile Test | To demonstrate the device is sensitive to the pressure of the catheter. | The device met the acceptance criteria. |
| Friction Test (Coating Lubricity) | To evaluate the catheter coating lubricity (frictional forces). | The device met the acceptance criteria. |
| Radiopacity Test | To demonstrate the device is relatively sensitive to the pressure of the catheter. | The device met the acceptance criteria. |
| Simulated Use Test | To demonstrate the device is sensitive to the pressure of the catheter. | The device met the acceptance criteria. |
{10}
| Test Name | Objective | Result |
| --- | --- | --- |
| Physician Usability Test | clots using a clinically relevant benchtop model, with comparison to the predicate. | simulated use conditions. |
| Shelf Life | To demonstrate that the device performance and packaging integrity is maintained over the proposed shelf-life (6 months). | The device met the acceptance criteria. |
| Packaging Validation | To demonstrate the packaging of the device meets all product specifications. | The device met the acceptance criteria. |
# VIII. BIOCOMPATIBILITY
The subject Lynx Aspiration Catheter is categorized as limited exposure ( $\leq 24$ hours), externally communicating device with circulating blood contact in accordance with ISO 10993-1. The Aspiration Tubing and RHV are categorized as limited exposure ( $\leq 24$ hours), externally communicating devices with indirect contact with circulating blood. The following testing supports biocompatibility and substantial equivalence of the subject device.
| Test Name | Results | Conclusion |
| --- | --- | --- |
| Cytotoxicity: MEM Elution | Non-cytotoxic | Pass |
| Sensitization: Magnusson-Kligman Method | Non-sensitizing | Pass |
| Irritation: Intracutaneous Reactivity | Non-irritant | Pass |
| Acute Systemic Toxicity: Acute Systemic Injection | Non-toxic | Pass |
| Systemic Toxicity: Material Mediated Pyrogenicity | Non-pyrogenic | Pass |
| Hemocompatibility: Hemolysis - Direct and Indirect | Non-hemolytic | Pass |
| Hemocompatibility: Platelet and Leukocyte Count | Comparable to the predicate | Pass |
| Hemocompatibility: Complement Activation | Non-activator of the complement system | Pass |
| Hemocompatibility: In-vivo Thrombogenicity | Comparable to the predicate | Pass |
| Hemocompatibility: Partial Thromboplastin Time (PTT) | Comparable to the predicate | Pass |
# IX. STERILITY
The subject Lynx is sterilized with ethylene oxide (EO). The sterilization process has been validated in accordance with ISO 11135:2014 to achieve a sterility assurance level (SAL) of $1 \times 10^{-6}$ . EO and Ethylene Chlorohydrin (ECH) residuals were below the limits specified in ISO 10993-7:2008. Bacterial endotoxin levels were below the level of 2.15 endotoxin units (EU)/device. Both baseline and accelerated aged shelf-life testing were conducted demonstrating the device and packaging will perform as intended to support the proposed 6-month shelf-life.
# X. ANIMAL STUDY
A Good Laboratory Practices (GLP) animal study was conducted to evaluate the safety and performance of the Lynx Aspiration Catheter (072 model) System in comparison to the predicate Penumbra System (Reperfusion Catheter RED 72) in a porcine model. Six animals were evaluated at 3-day and 30-day timepoints. The complete test system (catheter, aspiration tubing, pump) was compared to the complete control system. Study procedures included aspiration of soft and firm experimental clots from ascending pharyngeal arteries and renal artery branches, and wedge assessments under maximum vacuum conditions in lingual arteries. Three passes were performed per vessel. Vessels were evaluated via angiography, gross necropsy, and histopathology by a board-certified veterinary pathologist. Performance was comparable between test and control devices. Histopathology revealed minimal to mild vascular injury at 3 days with appropriate healing by 30 days. The pathologist concluded that the test and control devices demonstrated comparable safety profiles with no evidence of excess or unexpected pathology in the test group.
{11}
Page 8 of 8
XI. CLINICAL
Clinical testing was not deemed necessary to support the substantial equivalence of the Lynx Aspiration Catheter System.
XII. CONCLUSION
The subject device, the Lynx Aspiration Catheter System is substantially equivalent to the predicate devices considering the same intended use, similar technological characteristics, and results of the non-clinical testing. The Lynx Aspiration Catheter System incorporates similar design features and procedural steps and the differences do not raise any new or different questions of safety or effectiveness when compared to the predicate devices. Safety and performance testing demonstrates that the Lynx Aspiration Catheter System performs as intended.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.