The UNIDRIVE S III system is intended for intrauterine use by trained gynecologists to hysteroscopically resect and remove tissue. The HOPKINS Telescope when used with the obturator and sheath is intended to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and surgical procedures.
Device Story
The UNIDRIVE SIII System is a motorized surgical device for hysteroscopic tissue resection. It consists of a motor control unit, DrillCut-X II Shaver handpiece, shaver blades, HOPKINS Telescope, obturator, and operating sheath. The control unit automatically detects the handpiece and sets parameters; the surgeon presets maximum speed via buttons. Shaver blades are inserted into the telescope working channel and activated by a one-pedal footswitch. The inner blade sheath oscillates within the outer sheath to resect tissue. The obturator facilitates telescope insertion; the sheath provides an outflow channel for fluid exchange. Used in clinical settings by gynecologists, the device provides real-time visualization and mechanical tissue removal, aiding in diagnostic and surgical intrauterine procedures. The system benefits patients by enabling minimally invasive access and efficient tissue resection.
Clinical Evidence
No clinical data was required. Substantial equivalence was established through non-clinical bench testing, including cutting performance comparisons and optical performance validation against the predicate device.
Technological Characteristics
Motorized surgical system; components include HOPKINS Telescope, obturator, operating sheath, and shaver blades. Materials tested for biocompatibility per ISO 10993 (cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity). Thermal safety per IEC 60601-2-18. Reprocessing validated per AAMI TIR12/30/39, ANSI/AAMI ST79, and ISO 17664/17655. Sterilization per ISO 11137. Connectivity: motor control unit with automatic handpiece detection. Form factor: endoscopic system with mechanical oscillation for tissue resection.
Indications for Use
Indicated for intrauterine tissue resection and removal, and visualization of the cervical canal and uterine cavity for diagnostic and surgical procedures in patients requiring hysteroscopy. Intended for use by trained gynecologists.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
{0}
[LOGO]
FDA
U.S. FOOD & DRUG
ADMINISTRATION
September 25, 2026
Karl Storz SE & CO.KG
Alita McElroy
Regulatory Affairs Manager
Dr.-Karl-Storz-Straße 34
Baden-Wurttemberg
Tuttlingen, 78532
Germany
Re: K254252
Trade/Device Name: UNIDRIVE SIII System; HOPKINS Telescope 0° (26205AMA) Operating Sheath (26205CD) Obturator (26205OC) Shaver Blade (26205SA) Shaver Blade (26205SB) Shaver Blade (091133-01/-10) Shaver Blade (091134-01/-10)
Regulation Number: 21 CFR§ 884.1690
Regulation Name: Hysteroscope and Accessories
Regulatory Class: II
Product Code: HIH
Dated: August 7, 2026
Received: August 10, 2026
Dear Alita McElroy:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254252 - Alita Mcelroy
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K254252 - Alita Mcelroy
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or (301)796-7100).
Sincerely,
Reginald K. Avery -S
for
Jason R. Roberts, Ph.D.
Assistant Director
DHT3B: Division of Reproductive,
Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K254252 | ? |
| Please provide the device trade name(s). | | ? |
| UNIDRIVE SIII System: HOPKINS Telescope 0° (26205AMA) Operating Sheath (26205CD) Obturator (26205OC) Shaver Blade (26205SA) Shaver Blade (26205SB) Shaver Blade (091133-01/-10) Shaver Blade (091134-01/-10) | | |
| Please provide your Indications for Use below. | | ? |
| The UNIDRIVE S III system is intended for intrauterine use by trained gynecologists to hysteroscopically resect and remove tissue. The HOPKINS Telescope when used with the obturator and sheath is intended to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and surgical procedures. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
STORZ
KARL STORZ—ENDOSKOPE
KARL STORZ Premarket Notification
UNIDIRVE SIII System
510(k) Summary
K254425
Page 1 of 4
## 510(k) Summary
This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.
| Submitter: | KARL STORZ Endoscopy-America, Inc. 2151 E. Grand Avenue El Segundo, CA 90245 |
| --- | --- |
| Contact: | Alita McElroy Regulatory Affairs Manager Phone: (424) 218-8376 Email: Alita.McElroy@karlstorz.com |
| Date of Preparation: | September 23, 2026 |
| Type of 510(k) Submission: | Traditional |
| Device Identification: | Trade Name: UNIDRIVE SIII System HOPKINS Telescope 0^{®} (26205AMA) Operating Sheath (26205CD) Obturator (26205OC) Shaver Blade (26205SA) Shaver Blade (26205SB) Shaver Blade (091133-01/-10) Shaver Blade (091134-01/-10) Common Name: Hysteroscope and accessories Classification Name: Hysteroscope and accessories (21 CFR 884.1690) |
| Regulatory Class: | II |
| Product Code: | HIH |
| Classification Panel: | Obstetrics/Gynecology |
| Predicate Device(s): | Predicate: UNIDRIVE SIII System (K223520) *The predicate has not been subject to a design related recall* |
1
{5}
STORZ
KARL STORZ—ENDOSKOPE
KARL STORZ Premarket Notification
UNIDIRVE SIII System
510(k) Summary
K254425
Page 2 of 4
| Device Description: | The UNIDRIVE® S III system is a motorized surgical device intended for intrauterine use to hysteroscopically resect and remove tissue. This submission adds the following new system components to the UNIDRIVE SIII System previously cleared in K223520: • HOPKINS Telescope 0° (26205AMA) • Obturator (26205OC) • Operating Sheath (26205CD) • Shaver Blades (26205SA, 26205SB) • Shaver Blades (091133-01/-10, 091134-01/-10) The UNIDRIVE S III system has a motor control unit and can be used with the DrillCut-X II Shaver handpiece GYN which houses the subject shaver blades. The DrillCut-X II Shaver handpiece GYN is connected to the motor connection socket of the control unit. The UNIDRIVE S III system control unit detects the handpiece automatically and loads the standard parameters. The control unit displays the available speed range on the screen, and the maximum speed can be preset using the +/- buttons. The shaver blade upon being connected to the DrillCut-X II Shaver handpiece GYN is inserted into the working channel of the compatible HOPKINS Telescope. The shaver blades are activated via a one-pedal footswitch that is connected to the UNIDRIVE SIII control unit. The shaver blades consist of two sheaths that fit into each other. The inner sheath oscillates within the outer sheath to enable tissue resection. The obturator is used to create access and facilitate insertion of the HOPKINS Telescope. The sheath is used optionally to provide an outflow channel if instruments other than the shaver blades are used to facilitate exchange of fluids. There are no changes to the system components of the UNIDRIVE SIII System cleared in K223520 with the exception of the DrillCut-X II Shaver handpiece GYN (26702050) that has updated manual cleaning instructions and sterilization dry time has been reduced from 60 minutes to 20 minutes. |
| --- | --- |
| Indications for Use: | The UNIDRIVE S III system is intended for intrauterine use by trained gynecologists to hysteroscopically resect and remove tissue. The HOPKINS Telescope when used with the obturator and sheath is intended to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and surgical procedures. |
2
{6}
STORZ
KARL STORZ—ENDOSKOPE
KARL STORZ Premarket Notification
UNIDIRVE SIII System
510(k) Summary
K254425
Page 3 of 4
| Technological Characteristics. | The subject and predicate devices are identical in their intended use and operating principle. The main differences between the subject and predicate devices include the following: - The subject device includes an operating sheath in addition to other components (control unit, footswitch shaver handpiece, shaver blade, endoscope, obturator) while the predicate does not include an operating sheath but includes other common components as the subject device. - The subject device includes shaver blades with outer diameter of 3mm and 4mm while the predicate includes shaver blades with outer diameter of 4mm. |
| --- | --- |
| | - The subject device includes sterile single use shaver blades as well as non-sterile reusable shaver blades while the predicate device includes only non-sterile reusable shaver blades. - The subject device includes an endoscope with 0° direction of view, outer diameter of 5.3mm and working channel diameter of 3.1mm while the predicate includes an endoscope with 6° direction of view, outer diameter of 7.3mm and working channel of 4.3mm. |
| Non-Clinical Performance Data: | There are no performance standards or special controls developed under Section 514 of the FD&C Act for the UNIDRIVE SIII System. However, the UNIDRIVE SIII System follows the FDA recognized consensus standards and is tested according to the following standards and FDA Guidance: **Hysteroscopes and Gynecologic Laparoscopes Guidance** - Hysteroscopes and Gynecologic Laparoscopes - Submission Guidance for a 510(k) **ISO Endoscopic Standards** - ISO 8600-1:2015 - ISO 8600-3:2019 - ISO 8600-5:2020 **Biocompatibility Summary** - Cytotoxicity (ISO 10993-5) - Sensitization (ISO 10993-10) - Irritation (ISO 10993-23) - Acute Systemic Toxicity (ISO 10993-11) - Material Mediated Pyrogenicity (Rabbit Pyrogen Study per USP) **Thermal Safety** - IEC 60601-2-18:2009 (3^{RD} Edition) |
3
{7}
**STORZ**
KARL STORZ—ENDOSKOPE
KARL STORZ Premarket Notification
UNIDIRVE SIII System
510(k) Summary
K254425
Page 4 of 4
| | **Reprocessing and Sterilization** - AAMI TIR12: 2020 - AAMI TIR30: 2011 - AAMI TIR39: 2009 - ANSI/AAMI ST79:2017 - ANSI/AAMI/ISO 17655-1:2006/2013 - ANSI/AAMI/ISO 17655-2:2009 - ISO17664-1:2021 - ISO 17664-2:2021 |
| --- | --- |
| | - Reprocessing Medical Device in Health Care Settings: Validation Methods and Labeling - ISO 11137-1:2025 - ISO 11137-2:2013 - ISO 11137-3:2017 The cutting performance of the subject device, UNIDRIVE SIII system has been successfully demonstrated by comparing to the predicate device UNIDRIVE SIII System cleared in K223520. Additional comparative bench testing between the subject and predicate devices was conducted to demonstrate the optical performance of the subject hysteroscope that is part of the subject UNIDRIVE SIII system to meet its intended use. |
| Clinical Performance Data: | Clinical Testing was not required to demonstrate substantial equivalence to the predicate device. Non-clinical testing was adequate to establish substantial equivalence. |
| Conclusion: | The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe and effective as the predicate device to support a substantial equivalence determination. |
4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.