← Product Code [LNH](/productcode/LNH) · K254250

# MagicScan-1.5, GreenMR Serenity 1.5T (K254250)

_Xingaoyi Medical Equipment Co.,Ltd · LNH · Aug 17, 2026 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K254250

## Device Facts

- **Applicant:** Xingaoyi Medical Equipment Co.,Ltd
- **Product Code:** [LNH](/productcode/LNH.md)
- **Decision Date:** Aug 17, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The MagicScan-1.5, GreenMR Serenity 1.5T system is indicated for use as a Magnetic Resonance Imaging System that produces sagittal, transverse, coronal, and oblique cross sectional images, and that display internal anatomical structure and/or function of the head, body and extremities. These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist the diagnostic. Contrast agents may be used depending on the region of interest of the scan.

## Device Story

1.5T superconducting MRI system with 60cm bore; utilizes magnetic resonance physics to generate cross-sectional images of head, body, and extremities. Components include magnet, RF power amplifier, RF coils, gradient power amplifier, gradient coils, patient table, spectrometer, and vital signal module. Operated by trained clinicians in clinical settings. Inputs include patient positioning and scan parameters; system transforms signals into anatomical images. Output viewed by physicians to assist in diagnosis. Benefits include non-invasive visualization of internal structures and function. GreenMR Serenity 1.5T variant features a liquid-helium-free magnet design using closed-loop refrigeration.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993), electrical/mechanical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and NEMA performance standards for image quality and SAR characterization.

## Technological Characteristics

1.5T superconducting magnet; 60cm bore; active shielding. RF system: 63.87MHz, 20kW amplifier, 24 receive channels. Gradient system: water-cooled, 38 mT/m max amplitude, 169 T/m/s max slew rate. Connectivity: DICOM compliant. Software: IEC 62304 compliant. Sterilization: N/A (non-invasive). GreenMR Serenity 1.5T variant uses liquid-helium-free magnet with dual-compressor closed-loop refrigeration.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- United Imaging uMR 588 ([K183137](/device/K183137.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD & DRUG
ADMINISTRATION

August 17, 2026

Xingaoyi Medical Equipment Co.,Ltd
Zhengying ZUO
Regulatory Affair Manager
#555 Yeshan Rd.
Yuyao, Zhejiang
China

Re: K254250

Trade/Device Name: MagicScan-1.5, GreenMR Serenity 1.5T
Regulation Number: 21 CFR 892.1000
Regulation Name: Magnetic Resonance Diagnostic Device
Regulatory Class: Class II
Product Code: LNH
Dated: July 24, 2026
Received: July 24, 2026

Dear Zhengying ZUO:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

{1}

K254250 - Zhengying ZUO

Page 2

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn

{2}

K254250 - Zhengying ZUO

Page 3

(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Daniel M. Krainak, PhD
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K254250 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  MagicScan-1.5, GreenMR Serenity 1.5T  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  The MagicScan-1.5, GreenMR Serenity 1.5T system is indicated for use as a Magnetic Resonance Imaging System that produces sagittal, transverse, coronal, and oblique cross sectional images, and that display internal anatomical structure and/or function of the head, body and extremities. These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist the diagnostic. Contrast agents may be used depending on the region of interest of the scan.  |   |   |
|  Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |

{4}

[LOGO]

K254250

510(k) Premarket Notification Submission

510(k) Summary

MagicScan-1.5, GreenMR Serenity 1.5T

Magnetic Resonance Imaging System

{5}

[LOGO]

510(k) Premarket Notification Submission

### 510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

Date Prepared: Jul 21, 2026

Manufacturer: Xingaoyi Medical Equipment Co.,Ltd

No.555 Yeshan Road,Yuyao,Zhejiang,P.R.China

Contact person: ZhengYing ZUO

Regulatory Affair Manager

Xingaoyi Medical Equipment Co.,Ltd

Tel: +86 15867341527

email: 289892155@qq.com

### Identification of the Device:

Proprietary/Trade Name: MagicScan-1.5, GreenMR Serenity 1.5T

Classification Name: Nuclear Magnetic Resonance Imaging

Regulation Name: Magnetic resonance diagnostic device

Regulation Number: 21 CFR 892.1000

Regulatory Class: II

Product code: LNH

Review Panel: Radiology

### Primary Predicate Device:

Proprietary/Trade Name: United Imaging uMR 588

Classification Name: Nuclear Magnetic Resonance Imaging

Regulation Name: Magnetic resonance diagnostic device

Regulation Number: 21 CFR 892.1000

Regulatory Class: II

Product code: LNH

Review Panel: Radiology

Submitter/510(k) Holder: Shanghai United Imaging Healthcare Co., Ltd.

Clearance: K183137 (cleared Jan 11, 2019)

### Device Description:

The MagicScan-1.5/ GreenMR Serenity 1.5T Magnetic Resonance Imaging System is a 1.5T superconducting magnetic resonance diagnostic device with a 60cm size patient bore. It consists of components such as magnet, RF power amplifier, RF coils, gradient power amplifier, gradient coils, patient table, spectrometer, computer, equipment cabinets, power distribution system,

1

{6}

KGY

510(k) Premarket Notification Submission

internal communication system, and vital signal module etc. The MagicScan-1.5 / GreenMR Serenity 1.5T Magnetic Resonance Diagnostic Device is designed to conform to NEMA and DICOM standards.

# Indications for Use:

The MagicScan-1.5, GreenMR Serenity 1.5T system is indicated for use as a Magnetic Resonance Imaging System that produces sagittal, transverse, coronal, and oblique cross sectional images, and that display internal anatomical structure and/or function of the head, body and extremities.

These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist the diagnostic. Contrast agents may be used depending on the region of interest of the scan.

# Comparison with Predicate Device:

The MagicScan-1.5, GreenMR Serenity 1.5T and its predicate device, the United Imaging uMR 588 (K183137), have the equivalent intended use, and similar physical characteristics.

Table 1: Comparison with Predicate Device

|  Item | Subject Device MagicScan-1.5, GreenMR Serenity 1.5T | Predicate Device United Imaging uMR 588 (K183137) | Comments  |
| --- | --- | --- | --- |
|  **General**  |   |   |   |
|  Product Code | LNH | LNH | Same  |
|  Regulation No. | 21 CFR 892.1000 | 21 CFR 892.1000 | Same  |
|  Class | II | II | Same  |

2

{7}

KGY

510(k) Premarket Notification Submission

|  Item | Subject Device MagicScan-1.5, GreenMR Serenity 1.5T | Predicate Device United Imaging uMR 588 (K183137) | Comments  |
| --- | --- | --- | --- |
|  Indications for use | The MagicScan-1.5, GreenMR Serenity 1.5T system is indicated for use as a magnetic resonance diagnostic device that produces sagittal, transverse, coronal, and oblique cross sectional images, and that display internal anatomical structure and/or function of the head, body and extremities. These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist the diagnostic. Contrast agents may be used depending on the region of interest of the scan. | The uMR 588 system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces sagittal, transverse, coronal, and oblique cross sectional images, spectroscopic images, and that display internal anatomical structure and/or function of the head, body and extremities. These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist the diagnostic. Contrast agents may be used depending on the region of interest of the scan. | Equivalent Indications for use is Less than predicate device. The subject device does not support spectroscopic images. Note 1.  |
|  **Magnet system**  |   |   |   |
|  Type of Magnet | Superconducting | Superconducting | Same  |
|  Field Strength | 1.5 Tesla | 1.5 Tesla | Same  |
|  Patient-accessible bore dimensions | 60cm | 60 cm | Same  |
|  Type of Shielding | Actively shielded, EIS technology | Actively shielded, EIS technology | Same  |
|  Magnetic stability | <0.1ppm/h | <0.1ppm/h | Same  |

3

{8}

[LOGO]

510(k) Premarket Notification Submission

|  Item | Subject Device MagicScan-1.5, GreenMR Serenity 1.5T | Predicate Device United Imaging uMR 588 (K183137) | Comments  |
| --- | --- | --- | --- |
|  Magnet Homogeneity | ≤ 0.450 ppm @ 40cm DSV ≤ 0.040 ppm @ 30cm DSV ≤ 0.020 ppm @ 20cm DSV ≤ 0.003 ppm @ 10cm DSV | ≤ 0.450 ppm @ 40cm DSV ≤ 0.040 ppm @ 30cm DSV ≤ 0.020 ppm @ 20cm DSV ≤ 0.003 ppm @ 10cm DSV | Same  |
|  Type of coolant | MagicScan-1.5: Liquid helium in the magnet, Helium gas in the closed-loop compressor/cold head refrigeration circuit; GreenMR Serenity 1.5T: No liquid helium in the magnet, Helium gas in the closed-loop compressor/cold head refrigeration circuit. | Liquid helium in the magnet, Helium gas in the closed-loop compressor/cold head refrigeration circuit; | MagicScan-1.5: same as the predicate device; GreenMR Serenity 1.5T: different cooling mechanism from predicate. The magnet contains no liquid helium. See Note 2.  |
|  Liquid helium capacity | MagicScan-1.5: Liquid helium - 1000L in the magnet, Helium gas - in the closed-loop compressor/cold head refrigeration circuit.(one compressor and one cold head); GreenMR Serenity 1.5T: liquid helium - 0L in the magnet, Helium gas - in the closed-loop compressor/cold head refrigeration circuit. (two compressors and two cold heads). | Liquid helium - 1000L in the magnet; Helium gas - in the closed-loop compressor/cold head refrigeration circuit.(one compressor and one cold head). | MagicScan-1.5: same as the predicate device; GreenMR Serenity 1.5T: different cooling mechanism from predicate, the magnet contains no liquid helium. See Note 2.  |

4

{9}

KGY

510(k) Premarket Notification Submission

|  Item | Subject Device MagicScan-1.5, GreenMR Serenity 1.5T | Predicate Device United Imaging uMR 588 (K183137) | Comments  |
| --- | --- | --- | --- |
|  Liquid helium consumption (normal usage state) | 0 (Adopting 4K cooling head) | 0 (Adopting 4K cooling head) | Same  |
|  **Gradient system**  |   |   |   |
|  Max gradient amplitude | 38 mT/m | 33 mT/m | The higher gradient amplitude enables the system to achieve thinner slice thickness, smaller fields of view, and higher b-values in diffusion-weighted imaging. See Note 3.  |
|  Max slew rate | 169 T/m/s | 125 T/m/s | The higher slew rate enables the system to achieve thinner slice thickness, smaller fields of view, and higher b-values in diffusion-weighted imaging. See Note 4  |
|  Shielding | Active | Active | Same  |
|  Cooling | Water | water | Same  |
|  **RF system**  |   |   |   |
|  Transmission frequency | 63.87MHz | 63.87 MHz | Same  |
|  Transmission bandwidth | 600KHz | 600KHz | Same  |
|  Maximum launch site | 26uT | 26uT | Same  |

5

{10}

[LOGO]

510(k) Premarket Notification Submission

|  Item | Subject Device MagicScan-1.5, GreenMR Serenity 1.5T | Predicate Device United Imaging uMR 588 (K183137) | Comments  |
| --- | --- | --- | --- |
|  RF power amplifier type | Water cooling | Water cooling | Same  |
|  RF power Amplifier | 20kW | 18KW | Better, the larger the value, the clearer the image See Note 5.  |
|  Transmission coil without tuning | Yes | Yes | Same  |
|  Number of receive channels | 24 | 24 | Same  |
|  Adjustable receiving bandwidth | yes | yes | Same  |
|  Demodulation filtering | All digital orthogonal demodulation and filtering technology | All digital orthogonal demodulation and filtering technology | Same  |
|  Dynamic range (1Hz bandwidth) | 160dB | 160dB | Same  |
|  Noise coefficient | <0.5dB | <0.5dB | Same  |
|  **RF Coils**  |   |   |   |
|  Head-Neck Coil | 16 channels, receive, array, volume coil | 16 channels, receive, array, volume coil | Same  |
|  Body Coil | 16 channels, receive, array, volume coil (It can be imaged in combination with a spinal coil) | 6 channels, receive, array volume coil (It can be imaged in combination with a spinal coil) | Better, the number of channels is more, the coil better image quality and acceleration See Note 6.  |
|  Flex Coil | 4 channels, receive, array surface coil | 4 channels, receive, array surface coil | Same  |

6

{11}

[LOGO]

510(k) Premarket Notification Submission

|  Item | Subject Device MagicScan-1.5, GreenMR Serenity 1.5T | Predicate Device United Imaging uMR 588 (K183137) | Comments  |
| --- | --- | --- | --- |
|  Breast Coil | 8 channels, receive, array, volume coil | 10 channels, receive, array, volume coil | Channel Relatively small, but completely satisfied with the image quality See Note 7.  |
|  Spine Coil | 24 channels, Receive, array, surface coil | 24 channels, Receive, array, surface coil | Same  |
|  Ankle Coil | 8 channels, receive, array, volume coil | 4 channels, receive, array, surface coil (flex coil) | The intended use of the soft coil (Predicate device) includes the examination of the knee joint. See Note 8.  |
|  Knee Coil | 8 channels, receive, array, volume coil | 4 channels, receive, array, surface coil (flex coil) | The intended use of the soft coil (Predicate device) includes the examination of the knee joint See Note 8.  |
|  Shoulder coil | 8 channels, receive, array, volume coil | 4 channels, receive, array, surface coil (flex coil) | The intended use of the soft coil (Predicate device) includes the examination of the knee joint See Note 8  |
|  Hand Coil | 8 channels, receive, array, volume coil | 8 channels, receive, array, volume coil | Same  |
|  **Patient table**  |   |   |   |

7

{12}

KGY

510(k) Premarket Notification Submission

|  Item | Subject Device MagicScan-1.5, GreenMR Serenity 1.5T | Predicate Device United Imaging uMR 588 (K183137) | Comments  |
| --- | --- | --- | --- |
|  Dimensions | 2500 mm (L) × 600 mm (W) × (580mm-860mm) (H) | 2640 mm(L) X 880 mm(W)×(520-640) mm(H) | Similar See Note 9.  |
|  Maximum supported patient weigh | 250 Kg | 250 Kg | Same  |
|  Maximum speed of horizontal movement | 20cm/s | 20cm/s | Same  |
|  Support scanning range | 150cm | 150cm | Same  |
|  **Accessories**  |   |   |   |
|  Vital Signal Gating | Support RG signal triggering the scan | Support ECG/Respiratory/Pulse signal triggering the scan | Equivalent See Note 10.  |

Table 2 Comparison of the Application Software Features

|  Item | Subject Device MagicScan-1.5, GreenMR Serenity 1.5T | Predicate Device United Imaging uMR 588 (K183137) | Comments  |
| --- | --- | --- | --- |
|  **Standard sequences and parameters**  |   |   |   |
|  **2D spin echo sequence**  |   |   |   |
|  Scan matrix | 128X128 | 128X128 | Same  |
|  TR (min) | 6.8ms | 6.8ms | Same  |
|  TE (min) | 2.4ms | 2.4ms | Same  |
|  **2D fast spin echo sequence**  |   |   |   |
|  Scan matrix | 128X128 | 128X128 | Same  |
|  TR (min) | 6.8ms | 6.8ms | Same  |
|  TE (min) | 2.6ms | 2.6ms | Same  |
|  Slice thickness (min) | 0.1mm | 0.1mm | Same  |
|  Echo spacing (min) | 2.4ms | 2.4ms | Same  |

8

{13}

KGY

510(k) Premarket Notification Submission

|  Echo train length (ETL) | 1024 | 1024 | Same  |
| --- | --- | --- | --- |
|  3D gradient echo (3D GRE)  |   |   |   |
|  Scan matrix | 128X128 | 128X128 | Same  |
|  TR (min) | 1ms | 1ms | Same  |
|  TE (min) | 0.4ms | 0.4ms | Same  |
|  Slice thickness (min) | 0.05mm | 0.05mm | Same  |
|  Plane echo  |   |   |   |
|  Scan matrix | 128X128 | 128X128 | Same  |
|  TR (min) | 5ms | 5ms | Same  |
|  TE (min) | 1ms | 1ms | Same  |
|  Slice thickness (min) | 0.1mm | 0.1mm | Same  |
|  Echo spacing (min) | 0.4ms | 0.4ms | Same  |
|  TR (min) | 5ms | 5ms | Same  |
|  TE (min) | 1ms | 1ms | Same  |

Table 3: Comparison difference Analysis

|  Note 1 | The indications for use of the subject device are covered by the indications for use of the predicate device. Spectroscopic images is not intended to be used with the subject device. The less indication for use of the subject device does not impact safety and effectiveness.  |
| --- | --- |
|  Note 2 | The subject device of GreenMR Serenity 1.5T contains no liquid helium in the magnet. In the event of a quench, there is no cryogenic liquid boil-off or gas release into the examination room. The difference does not impact safety and effectiveness.  |
|  Note 3 | The max gradient amplitude of the proposed device is larger than the predicate device. Peripheral nerve stimulation and cardiac stimulation was controlled according to IEC 60601-2-33. The difference does not impact safety and effectiveness.  |
|  Note 4 | The max slew rate of the proposed device is larger than the predicate device. Peripheral nerve stimulation and cardiac stimulation was controlled according to IEC 60601-2-33. The difference does not impact safety and effectiveness.  |
|  Note 5 | The power of RF power amplifier for subject device is larger her than the predicate device, the subject device with larger power is benefit for image quality. The difference does not impact safety and effectiveness.  |

9

{14}

KGY

510(k) Premarket Notification Submission

|  Note 6 | The number of the receiving channels of subject device is more than the predicate device,the subject device with more channels is benefit for image quality.The difference does not impact safety and effectiveness.  |
| --- | --- |
|  Note 7 | The number of receiving channels is fewer than the predicate device. It is a little difference, it completely satisfied with the image quality. The effects was evaluated with clinical images.The difference does not impact safety and effectiveness.  |
|  Note 8 | The intended use of predicate device includes the subject device,the number of receiving coil of subject device is more than the predicate device,the more channels is benefit for image quality.The difference does not impact safety and effectiveness.  |
|  Note 9 | The dimensions of patient table of subject device are similar to the predicate device, which can satisfy clinical application.The difference does not impact safety and effectiveness.  |
|  Note 10 | The RG of subject device is covered by predicate device,the RG is same as predicate device.The difference does not impact safety and effectiveness.  |

### Technology:

The MagicScan-1.5, GreenMR Serenity 1.5T employs the same fundamental scientific technology as its predicated device.

### Determination of substantial equivalence:

The Proposed MagicScan-1.5, GreenMR Serenity 1.5T are substantially equivalent to the predicate United Imaging uMR 588 (K183137) with regards to intended use, image capabilities, technological characteristics, and safety effectiveness.

### Comparison Analysis

As Table 3: Comparison difference Analysis, The subjected device and the predicted device are identical in technological characteristics, including design, material, principle of operation, etc. They are similar in technical specification. The differences between the proposed device and the predicate device will not raise any new issues of safety or effectiveness. They can demonstrate substantial equivalence with predicate device.

### Summary of Non-clinical test:

The MagicScan-1.5, GreenMR Serenity 1.5T were evaluated for biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic, and mechanical safety, and have been found to

1

{15}

[LOGO]

510(k) Premarket Notification Submission

comply with applicable medical device safety standard, The MagicScan-1.5, GreenMR Serenity 1.5T complies with voluntary standards:

1. ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)]

2. IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests

3. IEC/TR 60601-4-2 Edition 1.0 2016-05 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems

4. ☐IEC 60601-2-33 Ed. 4:2022 Medical Electrical Equipment - Part 2-33: Particular Requirements for The Basic Safety and Essential Performance of Magnetic Resonance Equipment for Medical Diagnostic

5. ☐IEC 60825-1: 2014, Edition 3.0, Safety of laser products - Part 1: Equipment classification and requirements.

6. ISO10993-1 Fifth edition 2018-08 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process

7. ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity

8. ISO 10993-10 Fourth edition 2021-11 Biological evaluation of medical devices - Part 10: Tests for skin sensitization

9. ISO10993-23 First edition 2021-01 Biological evaluation of medical devices - Part 23: Tests for irritation

10. ISO14971 Third Edition 2019-12 Medical devices - Application of risk management to medical devices

11. NEMAPS 3.1 - 3.20 2022d Digital Imaging and Communications in Medicine (DICOM) Set.

12. IEC60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability

13. IEC62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION Medical devices - Part 1: Application of usability engineering to medical devices

14. IEC62304 Edition 1.1 2015-06 CONSOLIDATED VERSION Medical device software - Software life cycle processes

15. ISTA 3E 2017 Similar Packaged-Products in Unitized Loads of Truckload Shipment

16. ☐IEC 62464-1 Edition 2.0: 2018-12, Magnetic resonance equipment for medical imaging Part 1: Determination of essential image quality

1

{16}

[LOGO]

510(k) Premarket Notification Submission

parameters.

17. ☐ NEMA MS 1-2008(R2020), Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Images

18. ☐ NEMA MS 2-2008(R2020), Determination of Two-Dimensional Geometric Distortion in Diagnostic Magnetic Resonance Images

19. ☐ NEMA MS 3-2008(R2020), Determination of Image Uniformity in Diagnostic Magnetic Resonance Images

20. ☐ NEMA MS 4-2023, Acoustic Noise Measurement Procedure for Diagnosing Magnetic Resonance Imaging Devices

21. ☐ NEMA MS 5-2018, Determination of Slice Thickness in Diagnostic Magnetic Resonance Imaging

22. ☐ NEMA MS 6-2008(R2014, R2020), Determination of Signal-to-Noise Ratio and Image Uniformity for Single-Channel Non-Volume Coils in Diagnostic MR Imaging

23. ☐ NEMA MS 8-2016, Characterization of the Specific Absorption Rate (SAR) for Magnetic Resonance Imaging Systems

24. ☐ NEMA MS 9-2008(R2020), Standards Publication Characterization of Phased Array Coils for Diagnostic Magnetic Resonance Images

25. ☐ NEMA MS 10-2010, Determination of Local Specific Absorption Rate (SAR) in Diagnostic Magnetic Resonance Imaging.

26. ☐ NEMA MS 11-2010, Determination of Gradient-Induced Electric Fields in Diagnostic Magnetic Resonance Imaging

27. ☐ NEMA MS 12-2016 Quantification and Mapping of Geometric Distortion for Special Applications

28. IEEE Std 3333.2.1-2015 IEEE Recommended Practice for Three-Dimensional (3D) Medical Modeling

29. NEMA MS 14-2019 Standard for Characterization of Radiofrequency (RF) Coil Heating in Magnetic Resonance Imaging Systems ☐

### Summary of Clinical Tests:

The subject of this premarket submission, did not require clinical studies to support substantial equivalence.

### Conclusion:

For testing, all pre-determined acceptance criteria were met. Results of these tests show that the proposed MagicScan-1.5, GreenMR Serenity 1.5T Magnetic Resonance Imaging System meet their intended use.

The proposed MagicScan-1.5, GreenMR Serenity 1.5T Magnetic Resonance Imaging System are similar to the predicate United Imaging uMR 588 (K183137) in terms of indications for use, design, technological characteristics, and software functions. Xingaoyi Medical Equipment Co.,Ltd considers the GreenMR Serenity 1.5T Magnetic Resonance Imaging System to be

1

{17}

[LOGO]

510(k) Premarket Notification Submission

substantially equivalent, with respect to safety and effectiveness, to the predicate device.

1

---

**Source:** [https://fda-staging.innolitics.com/device/K254250](https://fda-staging.innolitics.com/device/K254250)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
