← Product Code [NEW](/productcode/NEW) · K254238

# MINT MONO Sutures (K254238)

_Hans Biomed Corporation · NEW · Sep 18, 2026 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K254238

## Device Facts

- **Applicant:** Hans Biomed Corporation
- **Product Code:** [NEW](/productcode/NEW.md)
- **Decision Date:** Sep 18, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4840
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

MINT MONO Sutures are comprised of PDO and are indicated for use in soft tissue approximation where use of an absorbable suture is appropriate. The suture is not intended for lifting or filling aesthetic uses.

## Device Story

MINT MONO Sutures are synthetic, absorbable, sterile, monofilament surgical sutures made of polydioxanone (PDO). They are provided in various lengths and USP sizes (7-0, 6-0, 5-0) with stainless steel (SUS 304) needles or cannulas. Used by clinicians for soft tissue approximation; the device is implanted and degrades/resorbs over time. The device is not intended for aesthetic lifting or filling procedures. Substantial equivalence is established through non-clinical performance testing, including biocompatibility, tensile strength, and sterilization validation, confirming the device meets USP standards and performs similarly to existing absorbable polydioxanone sutures.

## Clinical Evidence

No clinical data was necessary to determine substantial equivalence. Evidence is based on bench testing, including biocompatibility (ISO 10993 series), sterilization validation (ISO 11135-1), shelf-life (ASTM F88/F88M, ASTM F1929), and mechanical performance (USP <881> tensile strength, dimension verification, degradation profile, and cannula testing).

## Technological Characteristics

Synthetic absorbable monofilament polydioxanone (PDO) suture. Stainless steel (SUS 304) needle/cannula. Sterilized via ethylene oxide. Compliant with USP <881> for tensile strength and USP standard suture sizes. Implantable, absorbable, contact duration >30 days.

## Regulatory Identification

An absorbable polydioxanone surgical suture is an absorbable, flexible, sterile, monofilament thread prepared from polyester polymer poly (p-dioxanone) and is intended for use in soft tissue approximation, including pediatric cardiovascular tissue where growth is expected to occur, and ophthalmic surgery. It may be coated or uncoated, undyed or dyed, and with or without a standard needle attached.

## Special Controls

*Classification.* Class II (special controls). The special control for the device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

## Predicate Devices

- DemeDIOX Absorbable Surgical Suture ([K181582](/device/K181582.md))

## Reference Devices

- MINT Sutures ([K220549](/device/K220549.md))
- PDREX Sutures ([K173779](/device/K173779.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD & DRUG ADMINISTRATION

September 18, 2026

Hans Biomed Corporation
% Sarah Fitzgerald
Senior Consultant
Emergo by UL
2500 Bee Caves Rd.
Bldg. 1, Suite 300
Austin, Texas 78746

Re: K254238

Trade/Device Name: MINT MONO Sutures
Regulation Number: 21 CFR 878.4840
Regulation Name: Absorbable Polydioxanone Surgical Suture
Regulatory Class: Class II
Product Code: NEW
Dated: August 14, 2026
Received: August 14, 2026

Dear Sarah Fitzgerald:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K254238 - Sarah Fitzgerald

Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the

{2}

K254238 - Sarah Fitzgerald

Page 3

Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

TEK N. LAMICHHANE -S

Tek N. Lamichhane, Ph.D.

Assistant Director

DHT4B: Division of Plastic and

Reconstructive Surgery Devices

OHT4: Office of Surgical and

Infection Control Devices

Office of Product Evaluation and Quality

Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

# **Indications for Use**

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K254238

Device Name
MINT MONO Sutures

Indications for Use (Describe)

MINT MONO Sutures are comprised of PDO and are indicated for use in soft tissue approximation where use of an absorbable suture is appropriate.

The suture is not intended for lifting or filling aesthetic uses.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)

☐ Over-The-Counter Use (21 CFR 801 Subpart C)

**CONTINUE ON A SEPARATE PAGE IF NEEDED.**

This section applies only to requirements of the Paperwork Reduction Act of 1995.

**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*

FORM FDA 3881 (8/23)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{4}

K254238

# 510(k) Summary -K254238

## MINT® MONO Sutures

### 1. Submission Sponsor

Hans Biomed Corporation

64, Yuseong-Daero 1628 Beon-Gil

Yuseong-Gu, Daejeon 34054 Korea

Contact: Mark Bae

Title: Regulatory Affairs Deputy Manager

Telephone: 82-02-466-2266

Email: mhbae@hansbiomed.com

### 2. Submission Correspondent

Emergo by UL

2500 Bee Cave Road

Building 1, Suite 300

Austin, TX 78746

Office Phone: (512) 327-9997

Email: LST.US.EmergoFDASubmissions@ul.com

Contact: Sarah Marie Fitzgerald

Title: Senior Consultant

### 3. Date Prepared

September 18, 2026

### 4. Device Identification

Trade/Proprietary Name: MINT MONO Sutures

Common/Usual and Classification Name: Polydioxanone absorbable surgical suture

Regulation Number(s): 878.4840

Product Code(s): NEW

Class: Class II

Classification Panel: General & Plastic Surgery

### 5. Legally Marketed Predicate Device

Device name: DemeDIOX Absorbable Surgical Suture

510(k) number: K181582

Manufacturer: Demetech Corporation

Page 1 of 4

{5}

K254238

# 6. Device Description

MINT® Product Family Sutures, including MINT® MONO Sutures, are synthetic, absorbable, sterile, surgical sutures comprised of polydioxanone. They are available in a range of lengths and thicknesses. They are provided sterile. They degrade and resorb over time.

MINT® MONO Sutures are smooth sutures. MINT® MONO Sutures are part of the MINT® Product Family of Sutures.

# 7. Indication for Use Statement

MINT® MONO Sutures are comprised of PDO and are indicated for use in soft tissue approximation where use of an absorbable suture is appropriate.

The suture is not intended for lifting or filling aesthetic uses.

# 8. Substantial Equivalence Discussion

MINT® MONO Sutures are equivalent to DemeDIOX Absorbable Surgical Sutures cleared in K181582. MINT® Sutures (K220549) and PDREX Sutures (K173779) are provided as reference devices.

The following table compares the MINT® MONO Sutures to the predicate device and forms the basis for the determination of substantial equivalence. The subject device does not raise any new questions of safety or effectiveness as compared to the predicate device.

Comparison of Characteristics

|  Attribute | Subject MINT® MONO Sutures | Predicate DemeDIOX Sutures K181582 | Comparison  |
| --- | --- | --- | --- |
|  Product Code / Regulation | NEW / 878.4840 | NEW / 878.4840 | Same  |
|  Intended Use | Absorbable surgical sutures for use in soft tissue | Absorbable surgical sutures for use in soft tissue | Same  |
|  Indications for Use | MINT® MONO Sutures are comprised of PDO and are indicated for use in soft tissue approximation where use of an absorbable suture is appropriate. The suture is not intended for lifting or filling aesthetic uses. | DemeDIOX Absorbable Surgical Suture is indicated for use in general soft tissue approximation including pediatric cardiovascular tissue where growth is expected to occur and ophthalmic surgery, but not for use in adult cardiovascular, microsurgery and neural tissue. These sutures are useful where absorbable sutures with extended | Similar Provides fewer details compared to predicate but equivalent details compared to reference devices.  |

Page 2 of 4

{6}

K254238

|  Attribute | Subject MINT® MONO Sutures | Predicate DemeDIOX Sutures K181582 | Comparison  |
| --- | --- | --- | --- |
|   |  | wound support (up to six weeks) is desirable. |   |
|  Suture Material | Polydioxanone | Polydioxanone | Same  |
|  Suture Characteristics | Synthetic absorbable monofilament | Synthetic absorbable monofilament | Same  |
|  Sterilization | Ethylene oxide | Ethylene oxide | Same  |
|  Size (USP) | Compliant with USP Standard Suture Sizes | Compliant with USP Standard Suture Sizes | Same  |
|  Contact Type | Implant, Absorbable | Implant, Absorbable | Same  |
|  Contact Duration | >30 days | >30 days | Same  |
|  Suture Diameter | 7-0, 6-0, and 5-0 USP sizes | 6-0 USP size | Similar Within range of reference devices.  |
|  Suture Tensile Strength | Compliant with USP <881> requirements | Compliant with USP <881> requirements except smallest is too short to perform knot pull test | Same  |
|  Accessories | Needle / Cannula | Needle / Cannula | Same  |
|  Accessory Material (Patient Contacting) | Stainless Steel (SUS 304) | Stainless Steel | Same  |

This submission is to add MINT® MONO Sutures, which are smooth, to cleared sutures in the MINT® Product Family of Sutures. All previously cleared MINT® Sutures are barbed.

### 9. Non-Clinical Performance Data

Testing was conducted in alignment with the FDA guidance “Class II Special Controls Guidance Document: Surgical Sutures,” confirming that the design output meets the design input and specifications for the sutures.

Testing conducted included biocompatibility evaluation and testing per ISO 10993-1, ISO 10993-3, ISO 10993-5, ISO 10993-6, ISO 10993-10, ISO 10993-11, ISO 10993-23, sterilization validation in accordance with ISO 11135-1, shelf-life testing in accordance with ASTM F88/F88M and ASTM F1929, tensile strength testing to confirm USP <881> monograph strength, dimension verification (length, diameter / gauge), degradation profile / degradation, residual tensile strength, cannula elasticity, cannula bending strength, and cannula pull-out testing.

The results of non-clinical performance testing demonstrates substantial equivalence between the subject and predicate device.

### 10. Animal Performance Data

No animal testing was conducted for this submission.

Page 3 of 4

{7}

K254238

# 11. Clinical Performance Data

No clinical data was necessary to determine the substantial equivalence of this device.

# 12. Statement of Substantial Equivalence

All testing and evidence confirms that the subject device does not raise any new questions related to the safety or effectiveness and that the device is at least as safe and at least as effective as the predicate device for the intended use and indications for use. Therefore, the device can be considered substantially equivalent to the predicate device.

Page 4 of 4

---

**Source:** [https://fda-staging.innolitics.com/device/K254238](https://fda-staging.innolitics.com/device/K254238)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).