← Product Code [PZX](/productcode/PZX) · K254232

# ECLIPTIS Left Atrial Appendage Exclusion System (ECS135); ECLIPTIS Left Atrial Appendage Exclusion System (ECS140); ECLIPTIS Left Atrial Appendage Exclusion System (ECS145); ECLIPTIS Left Atrial Appendage Exclusion System (ECS150) (K254232)

_Edwards Lifesciences · PZX · Jun 26, 2026 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K254232

## Device Facts

- **Applicant:** Edwards Lifesciences
- **Product Code:** [PZX](/productcode/PZX.md)
- **Decision Date:** Jun 26, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The ECLIPTIS left atrial appendage exclusion system is indicated for the exclusion of the heart's left atrial appendage, performed under direct visualization and in conjunction with other cardiac surgical procedures. Direct visualization, in this context, requires that the surgeon is able to see the heart directly, with or without assistance from a camera, endoscope, etc., or other appropriate viewing technologies.

## Device Story

Implantable clip system for left atrial appendage (LAA) exclusion; used during cardiac surgery under direct visualization. System comprises sterile, single-use delivery tool pre-loaded with nitinol/PEEK/silicone clip. Surgeon uses malleable shaft and rotation knob to position clip at LAA base; lever controls opening/closing. Red trigger releases clip; design allows clip recapture and redeployment if placement is suboptimal. Output is permanent mechanical LAA exclusion. Benefits include reduced risk of thromboembolism from LAA; device is MR conditional.

## Clinical Evidence

Bench testing included mechanical, corrosion, fatigue, MR, particulate, and sterility testing. Biocompatibility assessed per ISO 10993-1 for permanent implant (clip) and limited contact (delivery system). In vivo animal study performed to validate performance. No human clinical trial data reported.

## Technological Characteristics

Implantable clip: nitinol spring, PEEK bars, silicone overmold. Sizes: 35, 40, 45, 50 mm. Delivery system: malleable shaft, rotation knob, lever-actuated clip deployment/recapture. MR Conditional. Sterilization: radiation. Biocompatibility: ISO 10993 compliant.

## Regulatory Identification

An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.

## Predicate Devices

- AtriCure AtriClip LAA Exclusion System ([K210293](/device/K210293.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

![img-0.jpeg](img-0.jpeg)

June 26, 2026

Edwards Lifesciences  
Corinne Yestrepsky  
Sr. Manager Regulatory Affairs  
One Edwards Way  
Irvine, CA 92614 USA

Re: K254232

Trade/Device Name: ECLIPTIS Left Atrial Appendage Exclusion System  
Regulation Number: 21 CFR 878.4300  
Regulation Name: Implantable Clip  
Regulatory Class: Class II  
Product Code: PZX  
Dated: May 26, 2026  
Received: May 27, 2026

Dear Corinne Yestrepsky:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device'

U.S. Food & Drug Administration  
10903 New Hampshire Avenue  
Silver Spring, MD 20993  
www.fda.gov

{1}

K254232 - Corinne Yestrepsky

Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}

K254232 - Corinne Yestrepsky

Page 3

Sincerely,

Katherine

N. Trivedi -S

Digitally signed by
Katherine N. Trivedi -S

Date: 2026.06.26
10:01:21 -06'00'

Katherine Trivedi

Assistant Director

DHT2B: Division of Circulatory Support,
Structural, and Vascular Devices

OHT2: Office of Cardiovascular Devices

Office of Product Evaluation and Quality

Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.

# Indications for Use

510(k) Number (if known)
K254232

Device Name
ECLIPTIS Left Atrial Appendage Exclusion System

Indications for Use (Describe)

The ECLIPTIS left atrial appendage exclusion system is indicated for the exclusion of the heart's left atrial appendage, performed under direct visualization and in conjunction with other cardiac surgical procedures.

Direct visualization, in this context, requires that the surgeon is able to see the heart directly, with or without assistance from a camera, endoscope, etc., or other appropriate viewing technologies.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)

☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{4}

# 510(k) Summary

|  **510(k) Submitter:** | Edwards Lifesciences LLC 1 Edwards Way Irvine, CA, 92614 800-424-3278  |
| --- | --- |
|  **Contact Person:** | Corinne Yestrepsky, PhD Sr. Manager Regulatory Affairs Edwards Lifesciences LLC Corinne_Yestrepsky@edwards.com Tel: 949-250-5095  |
|  **Date of Preparation:** | December 29, 2025  |
|  **Trade Name:** | ECLIPTIS Left Atrial Appendage Exclusion System  |
|  **Common Name:** | Implantable left atrial appendage clip and delivery system  |
|  **Classification:** | Regulation: 21 CFR 878.4300, implantable clip Device Class: II Product Code: PZX (Left Atrial Appendage Clip, Implantable)  |
|  **Predicate Device:** | AtriCure AtriClip LAA Exclusion System, K210293  |
|  **Device Description:** | The ECLIPTIS left atrial appendage exclusion system is an implantable clip pre-loaded onto a sterile, single-use delivery system. The clip is available in four sizes: 35, 40, 45, and 50 mm. The delivery system has a malleable shaft and rotation knob to enable access to the base of the left atrial appendage. The lever facilitates opening and closing of the clip. Once the clip is in position, it is released by pulling the red trigger and sliding the delivery system off of the clip. The delivery system can be used to recapture and redeploy the clip.  |
|  **Indications for Use:** | The ECLIPTIS left atrial appendage exclusion system is indicated for the exclusion of the heart's left atrial appendage, performed under direct visualization and in conjunction with other cardiac surgical procedures. Direct visualization, in this context, requires that the surgeon is able to see the heart directly, with or without assistance from a camera, endoscope, etc., or other appropriate viewing technologies.  |

## Technological Characteristics and Basis for Substantial Equivalence

The subject and predicate device share the same intended use and operating principle. A comparison of the technological characteristics of the subject and the predicate device is summarized in Table 1.

Page 1 of 3

{5}

**Table 1 Comparison of Technological Characteristics**

|  Technological Characteristic | Subject Device: ECLIPTIS Left Atrial Appendage Exclusion System | Predicate Device: AtriCure AtriClip LAA Exclusion System (K210293) | Substantial Equivalence Discussion  |
| --- | --- | --- | --- |
|  Indications for Use | The device is indicated for the exclusion of the heart's left atrial appendage, performed under direct visualization and in conjunction with other cardiac surgical procedures. Direct visualization, in this context, requires that the surgeon is able to see the heart directly, with or without assistance from a camera, endoscope, etc., or other appropriate viewing technologies. | Same | N/A  |
|  Device Classification | 21 CFR 878.4300; Class II; Product Code: PZX (Left Atrial Appendage Clip, Implantable) | Same | N/A  |
|  Contraindications | Do not use this device as a contraceptive tubal occlusion device. Do not use this device if the patient has a known allergy to nitinol (nickel titanium alloy). | Same | N/A  |
|  Clip Sizes Offered | 35 , 40 , 45 , and 50 mm | Same | N/A  |
|  Clip Mechanics | Curved mechanical clip with U-shape opening and tip-first closure. Constructed of nitinol spring, polyether ether ketone bars with silicone overmold. | Straight mechanical clip with closed loop design or V-shape opening and tip-first closure. Constructed of titanium with or without nitinol springs and covered in polyethylene terephthalate fabric – dependent on device model. | Difference in clip design does not raise new questions of safety and effectiveness. Bench and in vivo evaluations confirm subject device clip meets the intended use.  |
|  Clip to delivery system interface | Delivery system is pre-loaded with clip. | Same | N/A  |
|  Adjustment prior to deployment | Malleable shaft and rotation knob | Same | N/A  |

Page 2 of 3

{6}

**Table 1 Comparison of Technological Characteristics**

|  Technological Characteristic | Subject Device: ECLIPTIS Left Atrial Appendage Exclusion System | Predicate Device: AtriCure AtriClip LAA Exclusion System (K210293) | Substantial Equivalence Discussion  |
| --- | --- | --- | --- |
|  Adjustment after deployment | The clip is recapturable using the same delivery system. The clip can be re-deployed after recapture. | Not available. | The subject device clip can be recaptured to mitigate risks from suboptimal placement.  |
|  MR Safety | MR Conditional | Same | N/A  |
|  Biocompatibility | Materials have been assessed based on ISO 10993 and are commonly employed in tissue contacting devices and permanent implants. | Same | N/A  |
|  Pyrogen | Non-Pyrogenic | Same | N/A  |
|  Sterilization | Provided sterile for single use. Sterilized via radiation. | Same | N/A  |

### Performance Data

Design verification and validation were performed to ensure that the subject device meets its performance specifications and demonstrates substantial equivalence to the predicate device. There are no known performance standards for this device. The following tests were conducted to demonstrate equivalence to the predicate device:

- Mechanical Testing
- Corrosion and Fatigue Testing
- Magnetic Resonance Testing
- Pyrogenicity
- Particulate Characterization
- Sterility
- Animal Study

### Biocompatibility Testing

Biocompatibility of the subject device followed ISO 10993 and FDA Guidance, Use of International Standard ISO 10993-1 “Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process” dated September 8, 2023. The subject device clip is a permanent implant (contact duration > 30 days) and the subject device delivery system has limited contact (≤ 24 hours), biocompatibility endpoints were evaluated accordingly. The subject device passed all applicable testing.

### Conclusion

Based on the intended use, technological characteristics, and performance testing results, the subject device is substantially equivalent to the predicate device.

Page 3 of 3

---

**Source:** [https://fda-staging.innolitics.com/device/K254232](https://fda-staging.innolitics.com/device/K254232)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).