Inno-Hook™ Disposable Clip Appliers (Ligating Clips) and Inno-Hook™ Ligating Clips
K254231 · Taiwan Surgical Corporation · FZP · Jul 20, 2026 · General, Plastic Surgery
Device Facts
Record ID
K254231
Device Name
Inno-Hook™ Disposable Clip Appliers (Ligating Clips) and Inno-Hook™ Ligating Clips
Applicant
Taiwan Surgical Corporation
Product Code
FZP · General, Plastic Surgery
Decision Date
Jul 20, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Inno-Hook™ ligating clips are non-absorbable polymer clips indicated for ligation of vessels or tissue structures. The Inno-Hook™ Disposable Clip Applier (Ligating Clips) is indicated as the delivery device for the Inno-Hook™ ligating clips. These clip appliers are designed for use with 5mm and 10mm cannulas. Surgeons should apply the appropriate size clip for the size of the vessel or tissue structure to be ligated so that the clip completely encompasses the vessel or tissue structure.
Device Story
Inno-Hook™ system consists of non-absorbable polyoxymethylene (POM) polymer ligating clips preloaded into a disposable mechanical clip applier. Used in general surgery to ligate vessels or tissue structures to prevent hemorrhage. Operated by surgeons via manual trigger squeeze; device automatically advances new clip upon actuation. Available in 5mm (Green) and 10mm (Purple) shaft diameters for use through trocar sleeves. Provides secure compression on tubular tissues. Benefits include reliable vessel ligation and procedural efficiency through preloaded, disposable delivery mechanism.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench performance testing, including pull-out force, slip force, airtight capability, clipped gap, trigger force, and mechanical function (ratchet, counter, rotation). Biocompatibility evaluated per ISO 10993-1; sterilization validated to SAL 10^-6 per ISO 11135; shelf-life validated per ASTM F1980.
Indicated for ligation of vessels or tissue structures in patients requiring surgical intervention. Designed for use with 5mm and 10mm cannulas; surgeons must select appropriate clip size to ensure complete vessel/tissue encompassment.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
Weck Auto Endo5 5mm Automatic Endoscopic 35cm Applier (AE05ML) (K230480)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 20, 2026
Taiwan Surgical Corporation
Ken Chen
3f., # 12, Sec. 12, Sheng Yi Rd., Hsinchu County
Zhubei City, TW 30261
Taiwan
Re: K254231
Trade/Device Name: Inno-Hook™ Disposable Clip Appliers (Ligating Clips)
Regulation Number: 21 CFR 878.4300
Regulation Name: Implantable clip
Regulatory Class: Class II
Product Code: FZP, GDO
Dated: December 22, 2025
Received: December 29, 2025
Dear Ken Chen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254231 - Ken Chen
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K254231 - Ken Chen
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TEK N. LAMICHHANE -S
Tek N. Lamichhane, Ph.D.
Assistant Director
DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices
OHT4: Office of Surgical
and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K254231
Device Name
Inno-Hook™ Disposable Clip Appliers (Ligating Clips)
Indications for Use (Describe)
The Inno-Hook™ ligating clips are non-absorbable polymer clips indicated for ligation of vessels or tissue structures.
The Inno-Hook™ Disposable Clip Applier (Ligating Clips) is indicated as the delivery device for the Inno-Hook™ ligating clips. These clip appliers are designed for use with 5mm and 10mm cannulas. Surgeons should apply the appropriate size clip for the size of the vessel or tissue structure to be ligated so that the clip completely encompasses the vessel or tissue structure.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
# 510(k) Summary
(K254231)
K254231 Page 1 of 4
This 510(k) summary of safety and effectiveness information is submitted as part of the Premarket Notification in compliance with requirements of CFR Part 807, Subpart E and Section 807.92
Submitter:
| Submitter: | TAIWAN SURGICAL CORPORATION |
| --- | --- |
| Mailing Address: | 3F., No.12, Sec.2, ShengYi Rd., Zhubei City, Hsinchu County 302, Taiwan |
| Phone Number: | +886-3-6588129 ext. 100 |
| Fax Number: | +886-3-6588355 |
| Contact Person: | Ken Chen |
| Email: | ernest.chen@twsc.com.tw |
| Date Prepared: | February 25, 2026 |
Device
| Trade Name: | Inno-Hook™ Disposable Clip Appliers (Ligating Clips) |
| --- | --- |
| Common Name: | Implantable clip |
| Classification Name: | Clip, Implantable |
| Classification Name: | 878.4300 |
| Classification Product Code: | FZP, GDO |
| Device Class: | II |
Predicate Device
| K230480, Weck Auto Endo5 5mm Automatic Endoscopic 35cm Applier (AE05ML) (Primary) |
| --- |
| K133202, Hem-o-lok Ligating Clips (Reference) |
{5}
K254231 Page 2 of 4
# Device Description
The Inno-Hook™ Ligating Clips are the primary therapeutic component of the device system, designed for use in general surgical procedures that require vessel or tissue ligation. These clips are non-absorbable, non-active implantable devices manufactured from a polymer material (polyoxymethylene/POM) and preloaded in the Inno-Hook™ Disposable Clip Applier during manufacture. They are designed to securely close around and maintain compression on tubular tissues to prevent hemorrhage. To accommodate different anatomical needs, the clips preloaded in the Disposable Clip Applier are available in specific sizes: the Medium-Large (ML) clips are intended for a vessel/tissue size range of 3mm to 10mm, and the Large (L) clips are intended for a vessel/tissue size range of 5mm to 13mm.
The Inno-Hook™ Disposable Clip Applier serves as the delivery mechanism specifically for the Inno-Hook™ Ligating Clips. These devices are provided sterile via Ethylene Oxide (EO) sterilization for single use, and are supplied preloaded with 9, 12, or 15 Inno-Hook™ Ligating Clips. The applier automatically advances a new ligating clip when the trigger is squeezed. To correspond with the clip sizes, the clip appliers provide two shaft diameter sizes and color-coded components: a 5mm shaft (Green) for delivering ML clips, and a 10mm shaft (Purple) for delivering L clips. These are designed for introduction through applicable trocar sleeves, or larger sized trocar sleeves with a fitted converter.
# Indications for Use
The Inno-Hook™ ligating clips are non-absorbable polymer clips indicated for ligation of vessels or tissue structures.
The Inno-Hook™ Disposable Clip Applier (Ligating Clips) is indicated as the delivery device for the Inno-Hook™ ligating clips. These clip appliers are designed for use with 5mm and 10mm cannulas. Surgeons should apply the appropriate size clip for the size of the vessel or tissue structure to be ligated so that the clip completely encompasses the vessel or tissue structure.
# Indication for Use Comparison
The intended use of the subject device, Inno-Hook™ Disposable Clip Applier (Ligating Clips), is identical to that of the primary predicate device (Weck Auto Endo5, K230480). Both instruments are used as delivery devices for non-absorbable polymer ligating clips. Additionally,
{6}
K254231 Page 3 of 4
both ligating clips are intended for use in procedures involving the ligation of vessels or tissue structures.
There are minor wording differences in the Indications for Use statements. These differences primarily reflect the specific proprietary names of the devices and accommodate the subject device's inclusion of an additional 10mm size model (indicated for use with 10mm cannulas), whereas the primary predicate specifies use with 5/5.5mm cannulas.
These minor differences do not constitute a new intended use because the fundamental therapeutic purpose—delivering non-absorbable polymer clips to completely encompass and ligate vessels or tissue structures—is exactly the same. The addition of the 10mm model merely accommodates different clinical and anatomical needs, and its functional equivalence has been fully established through comparative testing with a legally marketed reference device (Hem-o-lok Ligating Clips, K133202). Therefore, the differences in the indications for use statement do not alter the intended therapeutic effect and do not raise any new or different questions of safety and effectiveness.
### Technological Comparison
The subject device, Inno-Hook™ Disposable Clip Applier (Ligating Clips), has fundamentally similar technological characteristics to the primary predicate device (Weck Auto Endo5, K230480). Both devices operate on the same mechanical principle and the implantable ligating clips for both devices are manufactured from the same non-absorbable polymer material, Polyoxymethylene (POM).
There are minor technological differences between the subject device and the primary predicate device, such as size offering, applier materials, and design feature. However, these differences do not raise any new or different questions of safety and effectiveness.
To address the 10mm size model, a legally marketed reference device (Hem-o-lok Ligating Clips, K133202) was utilized to establish functional and dimensional equivalence. A comprehensive non-clinical bench testing demonstrates that the subject device performs comparably to both the primary predicate and the reference device. Therefore, the Inno-Hook™ Disposable Clip Applier is substantially equivalent to the legally marketed devices.
{7}
K254231 Page 4 of 4
# **Non-Clinical and/or Clinical
Tests Summary & Conclusion**
# Biocompatibility Testing:
Biocompatibility testing was conducted in accordance with ISO 10993-1:2018, Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process. The Inno-Hook™ Disposable Clip Applier (Ligating Clips) includes two types of patient-contacting components based on the nature and duration of body contact. The first contact type consists of the clip applier jaw and shaft, which have limited contact with tissue for a duration of less than 24 hours during the surgical procedure. The second contact type consists of the non-absorbable polymer ligating clips, which are intended to remain permanently implanted in the body and have long-term contact with tissue (>30 days).
# Sterilization Validation:
The sterilization process was validated to achieve a Sterility Assurance Level (SAL) of 10-6 in accordance with ISO11135:2014.
# Shelf-life:
The 3-years shelf-life for Inno-Hook™ Disposable Clip Applier (Ligating Clips) was established and validated through accelerated and real-time aging study conducted in accordance with ASTM F1980 and ISO11607-1/-2.
# Bench Performance Testing:
Bench performance testing included, evaluation of clipped pull-out force, clipped slip force, airtight capability, clipped gap, stagger clip test, torque of the rotation knob, full clip test, width of the Ligating Clip after Loading, two stage sound, function of ratchet, empty indicator, clip counter, and trigger force. The test results showed that the subject device demonstrated similar performance compared to the predicate device.
# **Conclusion**
Based on the supporting data provided and summarized above, the Inno-Hook™ Disposable Clip Applier (Ligating Clips) demonstrates substantial equivalence to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.