The Aerloom system is intended for use by medical professionals and patients who are under medical supervision, for the treatment of the following conditions: • Chronic edema • Lymphedema • Venous insufficiency • Wound healing
Device Story
Aerloom is a portable, battery-powered pneumatic compression system for clinical or home use. It consists of a controller and multi-chambered garment worn over a liner. The controller inflates/deflates garment chambers in a distal-to-proximal sequence to apply graduated compression (20–80 mmHg), facilitating fluid movement from impaired lymphatic regions. The device features Bluetooth connectivity for an optional mobile application interface. It is operated by patients or clinicians to manage edema and promote wound healing. The system provides sequential calibrated gradient pressure, mirroring the therapeutic action of the predicate device.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including pressure verification, biocompatibility (ISO 10993), electrical safety (ANSI/AAMI ES 60601-1), and electromagnetic compatibility (IEC 60601-1-2).
Technological Characteristics
Pneumatic compression system; controller with rechargeable Li-ion battery; 7-chambered garment (nylon/polyester/polyurethane); pressure range 20-80 mmHg. Connectivity via Bluetooth Low Energy (BLE). Compliant with ANSI/AAMI ES 60601-1, IEC 60601-1-11, and IEC 60601-1-2. Software/firmware controlled.
Indications for Use
Indicated for adults (22+ years) requiring treatment for chronic edema, lymphedema, venous insufficiency, or wound healing under medical supervision.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
{0}
**U.S. FOOD & DRUG**
ADMINISTRATION
August 31, 2026
Evolym, LLC
Alex Chang
Regulatory Consultant
1209 Orange St.
Wilmington, Delaware 19801
Re: K254217
Trade/Device Name: Aerloom
Regulation Number: 21 CFR 870.5800
Regulation Name: Compressible Limb Sleeve
Regulatory Class: Class II
Product Code: JOW
Dated: December 24, 2025
Received: December 29, 2025
Dear Alex Chang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K254217 - Alex Chang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K254217 - Alex Chang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Meaghan Erlewein -S
For Nicole Gillette
Assistant Director
DHT2B: Division of Circulatory Support,
Structural, and Vascular Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K254217 | ? |
| Please provide the device trade name(s). | | ? |
| Aerloom | | |
| Please provide your Indications for Use below. | | ? |
| The Aerloom system is intended for use by medical professionals and patients who are under medical supervision, for the treatment of the following conditions: • Chronic edema • Lymphedema • Venous insufficiency • Wound healing | | |
| Please select the types of uses. | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{4}
Evolym LLC
Aerloom 510(k) Submission
K254217
# 510(k) Summary
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
I. SUBMITTER: Evolym LLC
1209 Orange St. Wilmington, DE 19801 USA
Establishment Registration: 33-4391469
CONTACT: Alex Chang
Regulatory Consultant
P: 408-839-5826
E-mail: info@evolym.com
DATE PREPARED: December 23 2025
## II. DEVICE:
TRADE NAME: AERLOOM
CLASSIFICATION NAME: Compressible limb sleeve
DEVICE CLASSIFICATION: CLASS II
PRODUCT CODE: JOW
## III. PREDICATE DEVICE:
Manufacturer: Tactile Systems Technology Inc.
Trade Name: entré, Model PD08-U
510(k): K143185
510(k) Summary
Page 1
{5}
Evolym LLC
Aerloom 510(k) Submission
# IV. DEVICE DESCRIPTION:
The Aerloom system consists of two primary components: a controller and a multi-chambered compression garment. The system delivers sequential pneumatic compression therapy to the affected limb to support the treatment of chronic edema, lymphedema, venous insufficiency, and wound healing.
The garment is worn over lightweight clothing or a liner to avoid direct skin contact and is wrapped securely around the treatment area. It contains independently controlled air chambers that inflate and deflate in a programmed sequence, applying graduated compression in the range of 20–80 mmHg.
The controller, which houses the compression logic and pump system, is compact, portable, and powered by a rechargeable lithium-ion battery, allowing for limited mobility and untethered use during treatment. Aerloom is designed to be easy to use and suitable for use in both clinical and home settings under medical supervision.
# V: INDICATIONS FOR USE:
The Aerloom system is intended for use by medical professionals and patients who are under medical supervision, for the treatment of the following conditions:
- Chronic edema
- Lymphedema
- Venous insufficiency
- Wound healing
510(k) Summary
Page 2
{6}
Evolym LLC
Aerloom 510(k) Submission
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE:
The Aerloom device and the predicate entré device are pneumatic compression devices that use the same fundamental technology, mode of action, and operating principles.
| Feature | Subject Device (Aerloom) | Predicate Device (entré) |
| --- | --- | --- |
| Indications for Use | The Aerloom system is intended for use by medical professionals and patients who are under medical supervision, for the treatment of the following conditions: - Chronic edema - Lymphedema - Venous insufficiency - Wound healing | The entré System is intended for use by medical professionals and patients who are under medical supervision, for the treatment of the following conditions: - Chronic edema - Lymphedema - Venous insufficiency - Wound healing |
| Product Type | Pneumatic Compression Device | Pneumatic Compression Device |
| Components | Controller Garments Hoses Power Adaptor | Controller Garments Hoses Power adapter |
| Mode of Action | Helps direct and move excess fluid from impaired lymphatic regions to healthier areas, where it can be naturally absorbed and processed by the body. | Helps direct and move excess fluid from impaired lymphatic regions to healthier areas, where it can be naturally absorbed and processed by the body. |
| Principles of Operation | The controller inflates the connected garment chambers in a pre-programmed sequence from distal to proximal, using a pressure gradient that delivers higher pressure distally. Each chamber is inflated in sequence and held at pressure, after which all chambers are deflated simultaneously. | The controller inflates the connected garment chambers in a pre-programmed sequence from distal to proximal, using a pressure gradient that delivers higher pressure distally. Each chamber is inflated in sequence and held at pressure, after which all chambers are deflated simultaneously. |
| Environment of Use | Hospital, clinical, and home settings | Hospital, clinical, and home settings |
| Expected Service Life | 5 Years | 5 Years |
| Controller unit size and weight | 5.7" x 1.5" x 1.6" 0.75 lbs | 11.0" x 6.0" x 8.0" 4 lbs |
| Electrical Requirements | Rechargeable Li-ion Battery Pack, with max 1.00-0.60 A input from 90-264 VAC | 100-240 VAC 50/60 Hz to AC Adapter with output voltage of 12.0V DC and 3.0A |
510(k) Summary
Page 3
{7}
Evolym LLC
Aerloom 510(k) Submission
| | 50/60 Hz to AC Adapter, with output voltage of 5.0V DC and 5.0A | |
| --- | --- | --- |
| Controller Enclosure Material | Plastic | Plastic |
| User Interface | Pushbuttons and a mobile application interface for use on a mobile device. | Pushbuttons |
| Wireless Functions | Bluetooth Low Energy (BLE) | None |
| Software / Hardware | Analog and digital electronic with microprocessor | Analog and digital electronic with microprocessor |
| Software Concern Level | Basic Documentation Level | Moderate Level of Concern |
| Device Total Pressure Range | 20-80 mmHg | 20-80 mmHg |
| Output | Sequential calibrated gradient Pressure | Sequential calibrated gradient Pressure |
| Treatment Time | 1 hour maximum | 1 hour maximum |
| Garment-Controller Connection | Hose attached to garments | Hose attached to garments |
| Garment Port | Plastic push to connect | Plastic push to connect |
| Garment Material | Nylon and polyester fabric Polyurethane coated polyester and nylon fabric | Nylon and polyester fabric Polyurethane coated polyester and nylon fabric |
| Garment Chambers | 7 | 8 |
| Device Cleanability | Can be cleaned using commonly available mild household cleaners and disinfectants. | Can be cleaned using commonly available mild household cleaners and disinfectants. |
Technological differences from the predicate device, include:
- Bluetooth connectivity for optional use with a mobile application.
- An internal battery to support portability.
- A smaller controller to support portability.
These differences do not raise new questions of safety or effectiveness for the Aerloom system compared to the predicate device.
510(k) Summary
Page 4
{8}
Evolym LLC
Aerloom 510(k) Submission
### VII. PERFORMANCE DATA:
The following performance data were provided in support of substantial equivalence determination.
#### Biocompatibility Testing
The Aerloom device can be supplied with either upper or lower garments that are applied over the provided circular-knit liners. These liners are biocompatible and have been tested in accordance with ISO 10993-1, ISO 10993-5, and ISO 10993-10. All patient-contacting materials are compliant with applicable biocompatibility safety standards.
#### Sterilization & Shelf-life Testing
The subject device is non-sterile, and components are unlikely to deteriorate with age.
#### Electrical Safety and Electromagnetic Compatibility (EMC)
The subject device was evaluated based on the following applicable performance and safety standards: o ANSI/AAMI ES 60601-1:2005/(R) 2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) [Including Amendment 2 (2021)], IEC 60601-1-11:2015, EN 60601-1-2:2015/A1:2021, BS EN 60601-1-2:2015+A1:2021, and IEC 60601-1-2:2014/AMD1:2020. Results demonstrated that the subject device was compliant to all applicable performance and safety standards.
#### Software Verification and Validation Testing
The subject device includes embedded firmware in the controller which has the ability to connect with custom software installed on a mobile device. The system software exhibits a moderate level of concern and because there is no risk of death or serious injury, this corresponds to the Basic recommended documentation level.
Software lifecycle planning and documentation as well verification and validation testing were performed in accordance with IEC 62304:2015 and as recommended by the following FDA Guidance documents for Industry and FDA Staff:
- Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
- Guidance for Industry, FDA Reviewers and Compliance on Off-The-Shelf Software Use in Medical Devices
- Guidance for Industry and Food and Drug Administration Staff Content of Premarket Submissions for Management of Cybersecurity in Medical Devices
- Guidance for Industry, FDA Reviewers and Compliance on Postmarket Management of Cybersecurity in Medical Devices
#### Mechanical Testing
The subject device was evaluated based on the following benchtop performance tests.
| Performance Study | Overview | Status |
| --- | --- | --- |
| Packaging Transportation Simulation | ASTM D4169-23e1 | Completed / Pass |
| Pressure Verification Testing | Pressure capable of delivering 20-80 mmHg compression pressures. | Completed / Pass |
#### Animal Study
Animal performance testing was not required to demonstrate safety and effectiveness of the Aerloom system.
510(k) Summary
Page 5
{9}
Evolym LLC
Aerloom 510(k) Submission
# Clinical Study
Clinical testing was not required to demonstrate the safety and effectiveness of the Aerloom system.
# VIII. CONCLUSION:
The data included in this submission demonstrates that the Aerloom system is substantially equivalent to the primary predicate device, entré (K143185).
510(k) Summary
Page 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.