K254208 · Expanse Medical, Inc. · QEZ · Mar 20, 2026 · Cardiovascular
Device Facts
Record ID
K254208
Device Name
FLOWRUNNER Aspiration System
Applicant
Expanse Medical, Inc.
Product Code
QEZ · Cardiovascular
Decision Date
Mar 20, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The FLOWRUNNER Aspiration System is indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems.
Device Story
Peripheral thrombectomy system for minimally invasive removal of thrombus from peripheral vasculature via aspiration. Components include aspiration catheter, aspiration sheath, vacuum fitting, high flow stopcock connector, Y-connector with hemostatic valve, and optional hand actuator clip. Device introduced to occlusion site; catheter advanced through sheath to target thrombus; suction applied via external vacuum source (-8.0 to -27.6 in Hg). Catheter extension and retraction cycles aspirate clot. Sterile saline infusion capability included. Used in clinical settings by trained interventionalists; fluoroscopic guidance required for placement. Output is physical removal of thrombus, improving vessel patency. Benefits include minimally invasive clot extraction.
Clinical Evidence
Bench testing only. Testing included air/liquid leak resistance, catheter bond/tip strength, fatigue, kink resistance, actuation/retraction, torsional bond strength, dimensional/visual inspection, vacuum compatibility, and radiopacity. Biocompatibility and package integrity leveraged from K234073.
Indicated for removal of fresh, soft emboli and thrombi from peripheral arterial and venous vessels.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 20, 2026
Expanse Medical, Inc.
Shiva Ardakani
Regulatory/ Quality and Clinical
7060 Koll Center Pkwy., Suite 300
Pleasanton, California 95466
Re: K254208
Trade/Device Name: FLOWRUNNER Aspiration System
Regulation Number: 21 CFR 870.5150
Regulation Name: Embolectomy Catheter
Regulatory Class: Class II
Product Code: QEZ, KRA
Dated: December 22, 2025
Received: December 29, 2025
Dear Shiva Ardakani:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254208 - Shiva Ardakani
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K254208 - Shiva Ardakani
Page 3
Sincerely,
GREGORY W. O'CONNELL - S
Digitally signed by
GREGORY W. O'CONNELL -S
Date: 2026.03.20 08:18:16
-04'00"
Gregory O'Connell
Assistant Director
DHT2C: Division of Coronary and Peripheral Intervention Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K254208 | |
| Device Name FLOWRUNNER Aspiration System | |
| Indications for Use (Describe) The FLOWRUNNER Aspiration System is indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Expanse Medical, Inc.
FLOWRUNNER Aspiration System Special 510(k)
# 510(k) Summary (K254208)
[As required by 21 CFR 807.92(c)]
1. Submitter's Name / Contact Person
Submitter: Expanse Medical, Inc.
7060 Koll Center Parkway, Suite 300
Pleasanton, CA 94566
Contact Person: Shiva Ardakani
Sr. VP QA/RA/CA
Phone: 925-931-1300 Ext. 209
Date Prepared: 18 March 2026
2. General Information
Trade Name: FLOWRUNNER Aspiration System
Regulation Number: 21 CFR 870.5150
Common/Usual Name: Aspiration Catheter
Classification Name: Embolectomy Catheter
Regulatory Class: Class II
Product Code: QEZ and KRA
Predicate Device: FLOWRUNNER Aspiration System (K251488)
3. Indication for Use
The FLOWRUNNER Aspiration System is indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems.
4. Device Description
The FLOWRUNNER Aspiration System is a peripheral thrombectomy system consisting of several components:
- FLOWRUNNER Aspiration Catheter
- FLOWRUNNER Aspiration Sheath
- Vacuum Fitting
- High Flow Stopcock Connector
- Y-Connector with Hemostatic Valve
- Hand Actuator Clip
The FLOWRUNNER Aspiration System family is designed for the minimally invasive removal of thrombus from the peripheral vasculature using aspiration. The extension of the device family includes smaller-diameter sizes of the FLOWRUNNER Aspiration System. The system is a single-use catheter-based device with the capability to infuse fluids. FLOWRUNNER SP Aspiration System is over the wire system.
Page -1
{5}
Expanse Medical, Inc.
FLOWRUNNER Aspiration System Special 510(k)
FLOWRUNNER Aspiration System consists of one aspiration catheter, one aspiration sheath, one vacuum fitting, one high flow stopcock connector, one Y-connector with hemostatic valve, and a hand actuator clip.
The FLOWRUNNER Aspiration System is introduced to the site of the primary occlusion. The Aspiration Catheter is advanced through the Aspiration Sheath and targets aspiration directly to the thrombus. The Aspiration Catheter is then retracted back into the Aspiration Sheath. This process of extension and retraction of the Aspiration Catheter is then repeated to fully aspirate the clot. Suction is applied directly to the Aspiration Catheter from an external vacuum source to aspirate thrombus from an occluded vessel (maximum pressure -27.6 in Hg and minimum pressure of -8.0 in Hg).
Sterile saline flows through the Aspiration Sheath and into the Aspiration Catheter when connected proximally. The Aspiration Catheter and Sheath are visible under fluoroscopy. The hand actuator is an optional, proximal clip attached to the Aspiration Catheter to assist the user.
The FLOWRUNNER Aspiration System comes in 7F, SP 7F, 12F, and 14F compatible diameters and 60cm, 80cm, 100cm, 120 and 130cm lengths.
## 5. Performance Data
Bench testing was performed to support a determination of substantial equivalence to the predicate. Results from this testing provide assurance that the subject device FLOWRUNNER has been designed and tested to assure conformance to the requirements for its intended use.
## 5.1 Summary of Non-Clinical Data
In accordance with Section 12, Part (a)(i)(3)(A) of the Safe Medical Devices Act of 1990, a summary of the information supporting the substantial equivalence of the is provided in the comparison table. This section identifies all testing performed to support the proposed extension of the device family size. Detailed summaries and results of the testing are provided within the applicable sections of this submission.
- Design Verification (Bench-Top Testing)
- Air Leak Resistance
- Liquid Leak Resistance
- Catheter Bond and Tip Strength
- Fatigue
- Kink Resistance
- Actuation/ Deployment and Retraction
- Torsional Bond Strength
- Dimensional Inspection
- Visual Inspection
- Compatibility with Vacuum Pump/ Generators
- Radiopacity
- Biocompatibility and Package Integrity Testing have been evaluated and leveraged using K234073.
Page -2
{6}
Expanse Medical, Inc.
FLOWRUNNER Aspiration System Special 510(k)
- Sterilization Validation has been conducted to support adoption of additional smaller 7F FLOWRUNNER Aspiration System.
The FLOWRUNNER smaller sizes met all established requirements.
## 6. Predicate Comparison
The predicate device for this submission is the FLOWRUNNER Aspiration System (K251488). The smaller sizes represent an extension of the FLOWRUNNER device family.
| Device Characteristic | Predicate | Subject |
| --- | --- | --- |
| | Expanse Medical, Inc.
FLOWRUNNER
Aspiration System | Expanse Medical, Inc.
FLOWRUNNER
Aspiration System Family |
| 510(k) Number | K251488 | K254208 |
| Classification | Class II, QEZ and KRA | Class II, QEZ and KRA |
| Intended Use | Embolectomy and Thrombectomy in Peripheral vasculature. | Same |
| Indication for Use | The FLOWRUNNER Aspiration System is indicated for the removal of fresh, soft, emboli and thrombi from vessels of the peripheral arterial and venous systems. | Same |
| Materials | Biocompatible, commonly utilized for interventional device | Same |
| Accessory Device provided | • FLOWRUNNER Aspiration Sheath
• Vacuum Fitting
• High Flow Stopcock Connector
• Y-Connector with Hemostatic Valve
• Hand Actuator Clip | Same |
| Guide Wire Compatibility | 0.035” | 14F – 0.035”
12F – 0.035”
7F – 0.014”
SP 7F – 0.014” |
| Catheter OD | 14F (4.5 mm)
12F (3.5 mm) | 14F (4.5 mm)
12F (3.5 mm)
7F (2.4 mm)
SP 7F (2.4 mm) |
| Effective Length | 60-120cm | 60-130cm |
| Packaging Configuration | Individually packaged | Same |
| Sterilization | EO | Same |
| Biocompatibility | Biocompatible | Same |
| Shelf Life | 13 Months | Same |
Page -3
{7}
Expanse Medical, Inc.
FLOWRUNNER Aspiration System Special 510(k)
## 7. Substantial Equivalence Comparison
For purposes of establishing substantial equivalence for the extended FLOWRUNNER Aspiration System product family, the previously cleared FLOWRUNNER Aspiration System was identified as the predicate device. The predicate device has the same intended use, and the introduction of new sizes does not raise new questions of safety or effectiveness. The predicate device received 510(k) clearance in 2023 (K234073) and 2025 (K251488) and has been actively marketed since that time.
## 8. Conclusion:
The FLOWRUNNER Aspiration System family is deemed substantially equivalent to the predicate device FLOWRUNNER in intended use, design, materials, packaging, fundamental scientific technology, manufacturing, important performance specifications, and sterilization. Performance testing demonstrated that the FLOWRUNNER Aspiration System raises no new questions of safety and effectiveness. Therefore, the FLOWRUNNER Aspiration System is considered substantially equivalent to the predicate device FLOWRUNNER.
Page -4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.