K254197 · CAO Group, Inc. · GEX · Mar 11, 2026 · General, Plastic Surgery
Device Facts
Record ID
K254197
Device Name
Picasso Pro Diode Laser (002-00460)
Applicant
CAO Group, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Mar 11, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Picasso Pro Diode Laser is indicated for dentistry and oral soft tissue procedures of: 1) The removal of lesions, excision, incision, vaporization, ablation, hemostasis, and photocoagulation on soft tissue including abscess treatment, contouring, curettage, sulcular debridement, pulp extirpation, pulpotomy, frenectomy, gingivectomy, troughing, and removal of inflamed edematous tissue. 2) Temporary relief of minor muscle and joint pain, stiffness, minor arthritis pain, muscle spasm, temporary increase in local blood circulation, and temporary relaxation of muscles by means of topical elevated tissue temperature from infrared spectral emissions; 3) Light activation of bleaching materials for teeth whitening and laser-assisted whitening/bleaching of teeth.
Device Story
Picasso Pro Diode Laser delivers 810nm laser energy to oral/dental soft tissue; solid-state diodes generate energy; optical fiber system transmits energy to surgical site via metal handpiece with disposable single-use tips; foot-actuated switch controls beam activation; features 630nm aiming beam, adjustable power, and continuous/pulse delivery modes; used by licensed dental professionals in dental operatory; provides surgical precision for soft tissue procedures, pain relief via infrared thermal effects, and teeth whitening activation; benefits include consistent energy delivery, reduced collateral tissue damage, and versatile clinical utility.
Clinical Evidence
Bench testing only. Device validated for compliance with IEC 60601-1, 60601-1-6, 60601-2-22, 60825-1, and 62366-1. Software verification and validation conducted per FDA guidance. Biocompatibility testing not required as patient-contacting materials (handpiece, fiber tips) are identical to those in the cleared Picasso Plus (K152939).
Indicated for dental and oral soft tissue procedures (lesion removal, excision, incision, vaporization, ablation, hemostasis, photocoagulation, abscess treatment, contouring, curettage, sulcular debridement, pulp extirpation, pulpotomy, frenectomy, gingivectomy, troughing, removal of inflamed edematous tissue), temporary relief of minor muscle/joint pain/stiffness/arthritis/spasm, and light activation for teeth whitening.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 11, 2026
CAO Group, Inc.
Robert Larsen
Regulatory Affairs Manager
4628 W. Skyhawk Dr.
West Jordan, Utah 84084
Re: K254197
Trade/Device Name: Picasso Pro Diode Laser (002-00460)
Regulation Number: 21 CFR 878.4810, 872.6475
Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Regulatory Class: Class II
Product Code: GEX, NVK, EEG
Dated: December 23, 2025
Received: December 23, 2025
Dear Robert Larsen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254197 - Robert Larsen
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K254197 - Robert Larsen
Page 3
Sincerely,
TANISHA
L. HITHE-S
Digitally signed by
TANISHA L. HITHE-S
Date: 2026.03.11
23:22:03 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K254197 | |
| Device Name Picasso Pro Diode Laser (002-00460) | |
| Indications for Use (Describe) The Picasso Pro Diode Laser is indicated for dentistry and oral soft tissue procedures of: 1) The removal of lesions, excision, incision, vaporization, ablation, hemostasis, and photocoagulation on soft tissue including abscess treatment, contouring, curettage, sulcular debridement, pulp extirpation, pulpotomy, frenectomy, gingivectomy, troughing, and removal of inflamed edematous tissue. 2) Temporary relief of minor muscle and joint pain, stiffness, minor arthritis pain, muscle spasm, temporary increase in local blood circulation, and temporary relaxation of muscles by means of topical elevated tissue temperature from infrared spectral emissions; 3) Light activation of bleaching materials for teeth whitening and laser-assisted whitening/bleaching of teeth. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary – K254197
## Applicant Information:
Company Name: CAO Group, Inc.
Company Address: 4628 West Skyhawk Drive
West Jordan, Utah 84084 U.S.A.
Company Phone: 1-801-256-9282
Company Fax: 1-801-256-9287
Contact Person: Robert K. Larsen
Preparation Date: March 6, 2026
## Device Name:
Trade Name: Picasso Pro Diode Laser
Common Name: Powered Laser Surgical Instrument
Product Code: GEX, NVK, EEG
Regulation: 878.4810, 872.6475
Product Classification: Class II
510(k) Number: K254197
## Legally Marketed Predicate Device for Substantial Equivalence:
Precise SHP Diode Laser, manufactured by CAO Group, Inc. (K181601)
## Legally Marketed Reference Device for Substantial Equivalence:
Picasso Plus, Picasso Lite Plus, manufactured by AMD Group, LLC (K152939)
NeoLas 810 Soft Tissue Laser, manufactured by Sheaumann Laser, Inc. (K150664)
## Description of Submitted Device:
The Picasso Pro Diode Laser is a device for delivering laser energy to human oral and dental soft tissue for a variety of surgical procedures and treatments. This energy is generated by solid-state diodes, which provide a consistent and reliable generation of laser energy at $810 \pm 20\mathrm{nm}$ for a maximum of 7 watts of energy output. The laser energy is delivered to the surgical site by means of an optical fiber system, which allows for the safe transmission of laser energy to the site without creating undue risk to the patient or operatory staff by errant or collateral laser emissions. The device features some user definable settings, including a switchable $630\mathrm{nm}$ aiming beam, adjustable power output for both the working beam and aiming beam, and continuous delivery or pulse delivery options. The working end of the delivery fiber is contained within a metal handpiece with an attached disposable single-use tip for surgical procedures. This handpiece system is incorporated into the device. The activation of the working beam diodes is completed by use of a foot-actuated switch.
{5}
# Indications for Use of the Submitted Device:
The Picasso Pro Diode Laser is indicated for dentistry and oral soft tissue procedures of:
1) The removal of lesions, excision, incision, vaporization, ablation, hemostasis, and photocoagulation on soft tissue including abscess treatment, contouring, curettage, sulcular debridement, pulp extirpation, pulpotomy, frenectomy, gingivectomy, troughing, and removal of inflamed edematous tissue.
2) Temporary relief of minor muscle and joint pain, stiffness, minor arthritis pain, muscle spasm, temporary increase in local blood circulation, and temporary relaxation of muscles by means of topical elevated tissue temperature from infrared spectral emissions;
3) Light activation of bleaching materials for teeth whitening and laser-assisted whitening/bleaching of teeth.
# Rationale for Substantial Equivalence:
The submitted device is identical in materials and construction of the overall laser unit as the predicate Precise SHP Diode Laser. The submitted device incorporates the exact same handpiece design, handpiece materials of construction, and single use fiber tips as the reference Picasso Plus Diode Laser. The submitted device features the exact same indications for use as the predicate Precise SHP Diode Laser, with the addition of an indication of pulp extirpation that is featured in the reference NeoLas 810 Soft Tissue Laser. The submitted device is used by the same licensed dental professional, in the same dental operatory environment, with the same operating principle, same laser energy source at the same wavelength and same output intensity, and the same energy delivery method as the cited predicate devices. The submitted device uses the same control principle, same software architecture, similar user input-output interface, and same hardware control mechanisms as the predicate Precise SHP Diode Laser. The submitted device was designed by and will be manufactured by the same manufacturer as the predicate Precise SHP Diode Laser.
| Present Submission (pending) | Predicate Device (K181601) | Predicate Device (K150664) | Reference Device (K152939) |
| --- | --- | --- | --- |
| DEVICE CHARACTERISTICS/SPECIFICATIONS | | | |
| Working Beam Output Wavelength: 810±10nm | Working Beam Output Wavelength: 810±10nm | Working Beam Output Wavelength: 808±10nm | Working Beam Output Wavelength: 810±10nm |
| Working Beam Output Power Range: 0.5 - 7.0 watts | Working Beam Output Power Range: 0.5 - 3.0 watts | Working Beam Output Power Range: Up to 2.5 watts | Working Beam Output Power Range: 0.5 -7.0 watts |
| Aiming Beam Output Wavelength: 650±20nm | Aiming Beam Output Wavelength: 650±20nm | Aiming Beam Output Wavelength: Not Available | Aiming Beam Output Wavelength: 650±20nm |
| Aiming Beam Output Power Range: < 3 milliwatts | Aiming Beam Output Power Range: < 3 milliwatts | Aiming Beam Output Power Range: Not Available | Aiming Beam Output Power Range: < 3 milliwatts |
| Working Beam Activation: Wireless | Working Beam Activation: Wireless | Working Beam Output Power Range: Not Available | Working Beam Output Power Range: < 3 milliwatts |
{6}
| Present Submission (pending) | Predicate Device (K181601) | Predicate Device (K150664) | Reference Device (K152939) |
| --- | --- | --- | --- |
| momentary foot switch | momentary foot switch | Working Beam Activation: | Working Beam |
| Emission Frequency: Continuous or Pulsed | Emission Frequency: Continuous or Pulsed | Not Available | Activation: Wireless momentary foot switch |
| Electrical Power Input (Power Supply): 100-240VAC switchable, 47-63Hz, 1.5A | Electrical Power Input (Power Supply): 100-240VAC switchable, 47-63Hz, 1.5A | Emission Frequency: Not Available | Emission Frequency: Continuous or Pulsed |
| Main Unit Dimensions: 10” x 7.5” x 5.5” | Main Unit Dimensions: 5.5” x 4.5” x 4.5” | Electrical Power Input (Power Supply): Not Available | Electrical Power Input (Power Supply): 100-240VAC switchable, 47-63Hz, 1.5A |
| Main Unit Weight: 4 lbs. | Main Unit Weight: 1.25 lbs. | Unit Dimensions: Not Available | Main Unit Dimensions: 9” x 6.25” x 6” |
| Factory-defined Presets: Yes | Factory-defined Presets: Yes | Unit Weight: Not Available | Main Unit Weight: 2 lbs. |
| Laser Stop Button: Yes | Laser Stop Button: Yes | Factory-defined Presets: Not Available | Factory-defined Presets: Yes |
| Patient-contacting Materials: | Patient-contacting Materials: | Laser Stop Button: Not Available | Laser Stop Button: Yes |
| – Handpiece: Anodized aluminum | – Handpiece: Anodized aluminum | Patient-contacting Materials: | Patient-contacting Materials: |
| – Single-use Laser Fiber Tip: Polypropylene housing, Type 316 stainless steel canula, with silica glass optical fiber protruding from the canula | – Single-use Laser Fiber Tip: ABS plastic housing, with silica glass optical fiber protruding from the canula | – Not Available | – Handpiece: Anodized aluminum |
| Single-use Articles: Single-Use Laser Fiber Tip; | Single-Use Laser Fiber Tip; | | – Single-use Laser Fiber Tip: Polypropylene housing, Type 316 stainless steel canula, with silica glass optical fiber protruding |
| Sterilization of Articles: | Sterilization of Articles: | | |
| – Single-Use Tip: Immersion in approved liquid chemical sterilant, then rinsed with clean water | – Single-Use Tip: Immersion in approved liquid chemical sterilant, then rinsed with clean water | Single-use Articles: Not Available | |
{7}
| Present Submission (pending) | Predicate Device (K181601) | Predicate Device (K150664) | Reference Device (K152939) |
| --- | --- | --- | --- |
| | | Sterilization of Articles: Not Available | from the canula
Single-use Articles: Single-Use Laser Fiber Tip;
Sterilization of Articles:
- Single-Use Tip: Immersion in approved liquid chemical sterilant, then rinsed with clean water |
**Key Differences:**
The submitted device is identical in design, materials, and construction to the Picasso Plus reference device, except for the detachable display and the therapy and whitening handpiece attachments which are identical in materials to the Precise SHP predicate device.
Information on the NeoLas 810 reference device could not be obtained, relative to the configuration or components of the device in order to make a determination of similarities or differences. It should be noted that the submitted device and the reference NeoLas 810 device share identical laser source and laser wavelength, and statements in the 510(k) clearance letter for this reference device suggest a similar range of output power for the infrared laser beam.
The submitted device differs from the Precise SHP predicate device and Picasso Plus reference device with regards to physical dimensions, and the user interface for the current device being larger than was utilized for the Precise SHP predicate device.
**INDICATIONS FOR USE**
The Picasso Pro Diode Laser is indicated for dentistry and oral soft tissue procedures of:
1) The removal of lesions, excision, incision, vaporization, ablation, hemostasis, and photocoagulation on soft tissue including abscess treatment,
The Precise SHP Diode Laser is indicated for dentistry and oral soft tissue procedures of:
1) The removal of lesions, excision, incision, vaporization, ablation, hemostasis, and photocoagulation on soft tissue including abscess treatment,
The NeoLas 810nm Soft Tissue laser and its accessories are intended for use on adult and pediatric patients in dentistry, dermatology and other surgical areas in the following procedures:
The Picasso Plus /Picasso Lite Plus is generally indicated for incision, excision, vaporization, ablation and coagulation of oral soft tissues including the following:
- Gingival troughing for crown impression
{8}
| Present Submission (pending) | Predicate Device (K181601) | Predicate Device (K150664) | Reference Device (K152939) |
| --- | --- | --- | --- |
| contouring, curettage, sulcular debridement, pulpotomy, pulp extirpation frenectomy, gingivectomy, troughing, and removal of inflamed edematous tissue. | contouring, curettage, sulcular debridement, pulpotomy, frenectomy, gingivectomy, troughing, and removal of inflamed edematous tissue. | Soft Tissue Indications including Pulpal Tissues* Incision, excision, vaporization, ablation and coagulation or oral soft tissues, including: | - Gingivectomy - Gingivoplasty - Gingival incision and excision - Hemostasis and coagulation - Excisional and incisional biopsies - Exposure of unerupted teeth - Fibroma removal - Frenectomy and frenotomy - Implant recovery - Incision and drainage of abscess - Leukoplakia - Opercullectomy - Oral papillectomies - Pulpotomy - Pulpotomy as an adjunct to root canal therapy - Reduction of gingival hypertrophy - Soft tissue crown lengthening - Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa - Vestibuloplasty Laser Periodontal Procedures, including: |
| 2) Temporary relief of minor muscle and joint pain, stiffness, minor arthritis pain, muscle spasm, temporary increase in local blood circulation, and temporary relaxation of muscles by means of topical elevated tissue temperature from infrared spectral emissions; | 2) Temporary relief of minor muscle and joint pain, stiffness, minor arthritis pain, muscle spasm, temporary increase in local blood circulation, and temporary relaxation of muscles by means of topical elevated tissue temperature from infrared spectral emissions; | - Excisional and incisional biopsies - Exposure of unerupted teeth - Fibroma removal - Flap preparation—incision of soft tissue to prepare a flap and expose the bone. - Flap preparation—incision of soft tissue to prepare a flap and expose unerupted teeth (hard and soft tissue impactions) - Frenectomy and frenotomy - Gingival troughing for crown impressions - Gingivectomy - Gingivoplasty - Gingival incision and excision | - Sulcular debridement (curettage, removal of diseased, infected, inflamed and necrosed soft tissue in the periodontal pocket to improve clinical indices including: gingival index, |
| 3) Light activation of bleaching materials for teeth whitening and laser-assisted whitening/bleaching of teeth. | 3) Light activation of bleaching materials for teeth whitening and laser-assisted whitening/bleaching of teeth. | | |
{9}
| Present Submission (pending) | Predicate Device (K181601) | Predicate Device (K150664) | Reference Device (K152939) |
| --- | --- | --- | --- |
| | | - Hemostasis and coagulation
- Implant recovery
- Incision and drainage of abscesses
- Laser soft tissue curettage of the post-extraction tooth sockets and the periapical area during apical surgery
- Leukoplakia
- Operculectomy
- Oral papillectomies
- Pulpotomy
- Pulp extirpation
- Pulpotomy as an adjunct to root canal therapy
- Reduction of gingival hypertrophy
- Removal of pathological tissues (i.e. cysts, neoplasm or abscess) and hyperplastic tissues (i.e. granulation tissue) from around the apex
- Root canal debridement and cleaning
- Soft tissue crown lengthening
- Treatment of canker sores, herpetic and | gingival bleeding index, probe depth, attachment loss and tooth mobility)
- Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium
- Picasso assisted new attachment procedure (cementum-mediated periodontal ligament new-attachment to the root surface in the absence of long junctional epithelium). |
{10}
| Present Submission (pending) | Predicate Device (K181601) | Predicate Device (K150664) | Reference Device (K152939) |
| --- | --- | --- | --- |
| | | aphthous ulcer of the oral mucosa
- Vestibuloplasty
NOTE: Any tissue growth (i.e. cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation | |
**Key Differences:**
The indications for use of the submitted device are exactly identical to those of the Precise SHP predicate device except that the submitted device adds a procedure of ‘pulp extirpation’, as found in the indications for use of the NeoLas 2 reference device. The submitted device does not borrow from or make any additional comparisons to the indications for use of either the NeoLas 2 or the Picasso Plus reference devices.
**Performance Data:**
This device has been validated and certified compliant to the following recognized standards:
- 60601-1 Edition 3.2 2020-08. Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
- 60601-1-6 Edition 3.2 2020-07. Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability.
- 60601-2-22 Edition 3.1 2012-10. Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment.
- 60825-1 Edition 2.0 2007-03. Safety of laser products - Part 1: Equipment classification, and requirements.
- 62366-1 Edition 1.1 2020-06. Medical devices - Part 1: Application of usability engineering to medical devices.
Software verification and validation testing consistent with basic documentation were conducted in accordance with the recommendations provided in the FDA guidance document "Content of Premarket Submissions for Device Software Functions".
The optical fibers, handpieces, and single-use tips utilized in the Picasso Pro, in its final finished form, are identical to the same components of optical fibers, handpieces, and single-use tips utilized in the predicate
{11}
device, the "Picasso Plus" in formulation, processing, sterilization, and geometry, and no other chemicals have been added (e.g., plasticizers, fillers, additives, cleaning agents, mold release agents). The predicate device was cleared under K152939. Therefore, based on this information, no biocompatibility testing was needed to support substantial equivalence of this device.
Considering the similarities in design, materials, construction, user profile, environment of use, and indications for use between the submitted device and the predicate and reference devices, as indicated above and throughout the submission, no performance testing or comparison was conducted.
## Conclusion:
The Picasso Pro Diode Laser is substantially equivalent to the listed predicate devices without raising any new issues of safety or effectiveness. This device shares a very similar intended use, and similar principle of operation. Performance data supports that the device is capable of achieving the intended effect without raising substantial concerns for safety.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.