← Product Code [NAY](/productcode/NAY) · K254192

# Senhance Ultrasonic System (K254192)

_Asensus Surgical · NAY · May 12, 2026 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K254192

## Device Facts

- **Applicant:** Asensus Surgical
- **Product Code:** [NAY](/productcode/NAY.md)
- **Decision Date:** May 12, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Senhance Ultrasonic System and accessories are indicated for soft tissue surgical incisions when bleeding control and minimal thermal injury are important. The instruments are indicated for the sealing of vessels up to and including 5 mm in diameter. The Senhance Ultrasonic System and accessories are indicated for use with the Senhance Surgical System.

## Device Story

The Senhance Ultrasonic System is an energized surgical instrument system used with the Senhance Surgical System for laparoscopic procedures. It consists of a dissector, transducer, adapter, ultrasonic generator, and footswitch. The system converts electrical energy into ultrasonic energy (35-37 kHz) to incise soft tissue and seal vessels up to 5 mm in diameter. The surgeon operates the system via a footswitch while controlling the instrument through the Senhance Surgical System's robotic interface. The device provides hemostasis with minimal thermal injury. It is intended for use in clinical settings by surgeons. The output is the mechanical action of the ultrasonic dissector on tissue, which facilitates surgical cutting and vessel sealing, potentially reducing bleeding and thermal damage compared to traditional methods.

## Clinical Evidence

Bench and pre-clinical animal testing provided. Bench pressure testing on porcine arteries/veins (1.5-5.0 mm) confirmed burst pressures met criteria (≥240 mmHg for arteries, ≥50 mmHg for veins), non-inferior to predicate. Acute GLP animal study (porcine) evaluated thermal spread via histology; results were comparable to predicate. Chronic GLP animal study (min 3 weeks, n≥5) confirmed safe vessel sealing (0.1-5.0 mm) with no adverse structural changes or downstream tissue effects.

## Technological Characteristics

Ultrasonic energy system (35-37 kHz, torsional mode). Components: 12 mm curved shear dissector, transducer, adapter, generator, footswitch. Single-use disposable end effector (EO sterilized, SAL 10^-6). Shaft diameter 5.5 mm, length 349 mm. Standalone system; no network connectivity. Software is embedded in the generator.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- HARMONIC ACE® Shears + Adaptive Tissue Technology ([K120729](/device/K120729.md))

## Reference Devices

- Senhance Ultrasonic System ([K182421](/device/K182421.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

Asensus Surgical
Madhuvanthi Soundirarajan
Senior Regulatory Affairs Specialist
1 Tw Alexander Dr., Suite 160
Durham, North Carolina 27701

May 12, 2026

Re: K254192
Trade/Device Name: Senhance Ultrasonic System
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: NAY
Dated: April 10, 2026
Received: April 13, 2026

Dear Madhuvanthi Soundirarajan:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K254192 - Madhuvanthi Soundirarajan
Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}

K254192 - Madhuvanthi Soundirarajan
Page 3

Sincerely,

Colin K. Chen -S

Digitally signed by
Colin K. Chen -S
Date: 2026.05.12
17:06:31 -04'00'

Colin Kejing Chen, Ph.D.
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K254192 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  Senhance Ultrasonic System  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  The Senhance Ultrasonic System and accessories are indicated for soft tissue surgical incisions when bleeding control and minimal thermal injury are important. The instruments are indicated for the sealing of vessels up to and including 5 mm in diameter. The Senhance Ultrasonic System and accessories are indicated for use with the Senhance Surgical System.  |   |   |
|  Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |

{4}

510(k)#: K254192

# 510(k) Summary
## Senhance Ultrasonic System
### [In accordance with 21 CFR 807.92]

## Submitter Information

Submitter: Asensus Surgical, Inc.
1 TW Alexander Drive, Suite 160
Durham, NC 27703

Contact: Madhuvanthi Soundirarajan
Sr. Regulatory Affairs Specialist Email:
msoundirarajan@asensus.com
Ph.no: 3524339130

Date Summary Prepared: May 11, 2026

## Subject Device Information

Proprietary (Trade) Name: Senhance® Ultrasonic System
Common Name: System, Surgical, Computer Controlled Instrument
Classification: Class II
Classification Advisory Committee: General and Plastic Surgery
Regulation Number: 21 CFR 876.1500, Endoscope and Accessories
Product Codes: NAY (System, Surgical, Computer Controlled Instrument)

## Predicate Device Information

Primary Predicate Device: HARMONIC ACE® Shears + Adaptive Tissue Technology (K120729)
Reference Device: Senhance Ultrasonic System (K182421)

## Device Description:

The Senhance Ultrasonic System is an energized instrument system which delivers ultrasonic energy for soft tissue incisions and sealing of vessels up to and including 5 mm in diameter. It is designed to be used with the Senhance Surgical System, which precisely manipulates laparoscopically based instruments in surgery.

{5}

510(k)#: K254192

The Senhance Ultrasonic System is composed of five components:

1. The Dissector which interfaces with the tissue of interest
2. The Transducer which converts electrical energy into ultrasonic energy
3. The Senhance Adapter which physically attaches the instrument to the Senhance Surgical System manipulator arm
4. The Ultrasonic Generator which controls the energy settings to be delivered to the tissue
5. The Footswitch which attaches to the generator for the activation of ultrasonic energy by the surgeon's foot

The system components are presented in Figure 1 below:

![img-0.jpeg](img-0.jpeg)
Figure 1: The Senhance Ultrasonic System Components

# Intended Use/ Indications for Use:

The Senhance Ultrasonic System is intended to be used to incise soft tissues and seal vessels during laparoscopically based surgery using robotic control of the instrument provided by the Senhance Surgical System. The proposed indications for use statement is as follows:

The Senhance Ultrasonic System and accessories are indicated for soft tissue surgical incisions when bleeding control and minimal thermal injury are important. The instruments are indicated for the sealing of vessels up to and including 5 mm in diameter. The Senhance Ultrasonic System and accessories are indicated for use with the Senhance Surgical System.

The change to the indication statement is the addition of the underlined text.

# Summary of Technological Characteristics:

The Senhance Ultrasonic System has the same technological characteristics as the reference Senhance Ultrasonic System cleared under K182421. No changes to the technology have been made since the prior

{6}

510(k)#: K254192

clearance. The indications for use of Senhance Ultrasonic System and HARMONIC ACE® Shears + Adaptive Tissue Technology are the same.

The subject Senhance Ultrasonic system characteristics are outlined in the table below alongside the predicate and reference device system characteristics.

|  Characteristics | Subject Device Senhance Ultrasonic System | Predicate Device (K120729) HARMONIC ACE® Shears | Reference Device (K182421) Senhance Ultrasonic System  |
| --- | --- | --- | --- |
|  Manipulation Method | Robotic | Manual | Robotic  |
|  Method of Activation | Footswitch – not integrated to robotic system | Foot or Hand Switch | Footswitch – not integrated to robotic system  |
|  End Effector Design | 12 mm Curved Shear | Curved Shear | 12 mm Curved Shear  |
|  End Effector packaging and Sterility | Single Use Disposable (EO Sterilization) SAL 10^{-6} | Single Use Disposable (EO Sterilization) SAL 10^{-6} | Single Use Disposable (EO Sterilization) SAL 10^{-6}  |
|  End Effector Dimensions | Shaft Diameter: 5.5 mm Shaft Length: 349 mm | Shaft Diameter: 5 mm Shaft Length: 360 mm | Shaft Diameter: 5.5 mm Shaft Length: 349 mm  |
|  Ultrasonic Mode | Torsional | Linear | Torsional  |
|  Ultrasonic Wavelength | 35 to 37 kHz | Not Available | 35 to 37 kHz  |
|  Generator | Rebranded Lotus Series 4 generator | Generator G 11 | Rebranded Lotus Series 4 generator  |
|  Maximum Vessel Diameter for Vessel Sealing Indication | 5mm | 5mm | No vessel sealing indication claim  |

## Performance Data:

Bench and pre-clinical animal testing were conducted to demonstrate that the subject device is safe and effective for the expanded indications for use. This included the following tests:

- Bench Pressure Testing: A sufficient quantity of explanted porcine arteries and veins ranging from 1.5-5.0 mm were sealed and divided using both Senhance Ultrasonic devices and the predicate device. The acceptance criteria of the 90 percent lower tolerance bound burst pressure for arteries up to 5 mm in diameter is $\geq 240 \mathrm{mmHg}$, and the acceptance criteria of the 90 percent

{7}

510(k)#: K254192

lower tolerance bound burst pressure for veins up to 5 mm in diameter is ≥ 50 mmHg were satisfied. The burst pressure performance of the Senhance Ultrasonic System was not statistically inferior to the burst pressure performance of the predicate device.

- Acute Animal Study: To ensure there is no excessive tissue damage to the regions surrounding a sealed vessel when sealing a vessel using the Senhance Ultrasonic System, thermal spread testing via histological analysis was performed. The thermal spread test was conducted with the subject device and predicate device in a Good Lab Practices (GLP) environment using live porcine models. Each laparoscopic procedure was performed by a surgeon, during which both veins and arteries up to 5 mm in diameter were sealed using either Senhance Ultrasonic devices or control devices. The vessels were evaluated for cell viability and thermal spread by a board-certified pathologist via a histological analysis in order to identify any differences in performance between the two systems. The thermal spread with the use of the Senhance Ultrasonic System was noted to be comparable to the predicate, satisfying the primary objective of the acute study.

- Chronic Animal Study: Asensus Surgical performed a chronic animal study of a duration of a minimum of 3 weeks post-procedure in at least 5 animals to demonstrate that the Senhance Ultrasonic System can function properly to withstand normal body parameters for a variety of vessel types covering the entire range of sizes (0.1-5.0 mm diameter) for both veins and arteries. This study demonstrated that Senhance Ultrasonic System was able to safely seal appropriately sized abdominal cavity vessels while obtaining hemostasis and without structural changes at the treatment sites or in downstream tissues based on lack of procedural or in-life adverse events.

No change to any material or manufacturing process has been made since the clearance of the reference device, Senhance Ultrasonic System (K182421), and no change is being requested in this submission. All patient-contacting components remain biocompatible as demonstrated in K182421.

No changes to the reprocessing/cleaning and disinfection instructions are made as part of this submission. The shelf life of the dissector is labeled as 24 months, which was increased from 6 months as included in the K182421. This change is due to the completion of a new packaging validation study data showing sterility, package integrity, and device performance were maintained after 2 years' accelerated and real-time aging.

## Software and Cybersecurity:

The software for the Senhance Ultrasonic System is found within the Senhance Ultrasonic generator. No changes to the Senhance Ultrasonic System software have been made since the latest clearance on January 11, 2019, through K182421. Software verification and validation testing documentation, as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", was provided in K182421 and remains applicable to the subject device.

{8}

510(k)#: K254192

The documentation level for the software of this device was determined to be basic level, since no failure of any device software function could lead to a hazardous situation with a probable risk of death or serious injury to the patient, user of the device, or others in the environment of use, prior to the implementation of risk control measures. The generator software was assessed per the FDA guidance "Cybersecurity in Medical Device: Quality System Considerations and Content of Premarket Submissions" to identify any issues related to cybersecurity. Because the system is neither intended nor able to be connected to any private intranet or the public Internet, the risks attendant to networked devices are eliminated, and the risk of cyberattack is greatly reduced.

**Conclusion:**

The subject device, the Senhance Ultrasonic System, has the same intended use as the predicate, and its expanded indications for use do not affect the safety or effectiveness of the device. In addition, Senhance Ultrasonic System has the same technological characteristics as the reference Senhance Ultrasonic System cleared under K182421.

The bench testing and both the acute and chronic animal studies conducted in the GLP environment demonstrated that the subject device is safe and effective for sealing vessels up to and including 5 mm in diameter.

Thus, the subject Senhance Ultrasonic System is substantially equivalent to predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K254192](https://fda-staging.innolitics.com/device/K254192)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
