K254186 · Philips Medical Systems B.V. · OWB · Jan 16, 2026 · Radiology
Device Facts
Record ID
K254186
Device Name
Azurion R3.1
Applicant
Philips Medical Systems B.V.
Product Code
OWB · Radiology
Decision Date
Jan 16, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Azurion series (within the limits of the used Operating Room table) are intended for use to perform: • Image guidance in diagnostic, interventional, and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular, and neuro procedures. • Cardiac imaging applications including diagnostics, interventional and minimally invasive surgery procedures. Additionally: • The Azurion series can be used in a hybrid Operating Room. • The Azurion series contain a number of features to support a flexible and patient centric procedural workflow.
Device Story
Interventional fluoroscopic X-ray system for real-time anatomical visualization during procedures. Inputs: X-ray radiation; outputs: real-time images, post-processed images, and reference/control images. Used in hybrid ORs by clinicians for image guidance in vascular, non-vascular, cardiovascular, and neuro procedures. Features include radiation safety controls and DICOM-based connectivity for HIS/PACS integration. Design change adds hook-and-loop (Velcro) fasteners to mattress accessories to prevent slippage on the AD7X patient table. Clinicians use visual output to guide minimally invasive interventions, improving procedural accuracy and patient safety.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical performance testing, including verification of functional/non-functional requirements, safety risk control measures, and compliance with FDA-recognized consensus standards (e.g., IEC 60601 series, ISO 14971).
Technological Characteristics
Interventional fluoroscopic X-ray system. Key modification: hook-and-loop (Velcro) fasteners added to mattress accessories for AD7X patient table. Compliance with IEC 60601-1, 60601-2-43, 60601-2-54, 60601-2-28, and 60601-1-2. Connectivity via DICOM. Software lifecycle per IEC 62304. Usability per IEC 62366-1.
Indications for Use
Indicated for image guidance in diagnostic, interventional, and minimally invasive surgery (vascular, non-vascular, cardiovascular, neuro) and cardiac imaging. Patient population: all human patients of all ages; weight limited by table specifications.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
January 16, 2026
Philips Medical Systems B.V.
% Supriya Dalvi
Principal Regulatory Affairs Specialist
Veenpluis 6
Bldg. Qp
Best, Noord-Brabant 5684PC
NETHERLANDS
Re: K254186
Trade/Device Name: Azurion R3.1
Regulation Number: 21 CFR 892.1650
Regulation Name: Image-intensified fluoroscopic x-ray system
Regulatory Class: Class II
Product Code: OWB, JAA
Dated: December 23, 2025
Received: December 23, 2025
Dear Supriya Dalvi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254186 - Supriya Dalvi
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K254186 - Supriya Dalvi
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K254186 | |
| Device Name Azurion R3.1 | |
| Indications for Use (Describe) The Azurion series (within the limits of the used Operating Room table) are intended for use to perform: • Image guidance in diagnostic, interventional, and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular, and neuro procedures. • Cardiac imaging applications including diagnostics, interventional and minimally invasive surgery procedures. Additionally: • The Azurion series can be used in a hybrid Operating Room. • The Azurion series contain a number of features to support a flexible and patient centric procedural workflow. Patient Population: All human patients of all ages. Patient weight is limited to the specification of the patient table. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
PHILIPS
Azurion R3.1
Special 510(k)
510(k) Summary
# 510(k) Summary
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.
Date Prepared: December 23rd, 2025
Manufacturer: Philips Medical Systems Nederland B.V.
Veenpluis 6,
5684 PC Best
The Netherlands
Establishment Registration Number: 3003768277
Primary Contact Person: Dr. Supriya A. Dalvi, PhD
Regulatory Affairs Specialist, IGT-S Fixed
Tel: +31 (0) 628945536
E-mail: supriya.dalvi@philips.com
Secondary contact: Daniela Zaczek
Person: Director of Regulatory Affairs, IGT-S Fixed
Tel: +31 (0) 628517824
Email: Daniela.Zaczek@philips.com
Device: Trade Name: Azurion R3.1
Classification Name: Image-intensified fluoroscopic x-ray system
Classification Regulation: 21 CFR §892.1650
Classification Panel: Radiology
Device Class: Class II
Product Code: Primary Code: OWB;
Subsequent Code: JAA
Predicate Device: Trade Name: Azurion R3.1
Manufacturer: Philips Medical Systems Nederland B.V.
510(k) Clearance: K251827 (cleared on 24 Oct 2025)
Classification Name: Image-intensified fluoroscopic x-ray system
Classification Regulation: 21 CFR §892.1650
Classification Panel: Radiology
Device Class: Class II
Product Code: Primary Code: OWB;
Subsequent Code: JAA
PHILIPS
Philips IGTS
Page 1 of 5
{5}
PHILIPS
Azurion R3.1
Special 510(k)
510(k) Summary
## Device description:
The Azurion R3.1 is classified as an interventional fluoroscopic X-Ray system. The primary performance characteristics of the Azurion R3.1 include:
- Real-time image visualization of patient anatomy during procedures
- Imaging techniques and tools to assist interventional procedures
- Post processing functions after interventional procedures
- Storage of reference/control images for patient records
- Compatibility with hospital information systems (HIS) and image archiving systems via DICOM
- Built in radiation safety controls.
The only changes to the subject device, Azurion R3.1 includes the design change to the mattress accessory for all the existing mattresses of the predicate device (Azurion R3.1, K251827, 24 October 2025) and introduction of new gray color mattress. The change includes the addition of a hook and loop fastener (Velcro) solution for use between the mattress and the system integrated patient table (AD7X), to ensure that the mattress does not slip from the patient table.
## Indications for Use:
The Azurion series (within the limits of the used Operating Room table) are intended for use to perform:
- Image guidance in diagnostic, interventional, and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular, and neuro procedures.
- Cardiac imaging applications including diagnostics, interventional and minimally invasive surgery procedures.
### Additionally:
- The Azurion series can be used in a hybrid Operating Room.
- The Azurion series contain a number of features to support a flexible and patient centric procedural workflow.
### Patient Population:
All human patients of all ages. Patient weight is limited to the specification of the patient table.
The indications for use of subject device, Azurion R3.1 with mattress changes are the same as the indications for use of the predicate device (Azurion R3.1, K251827, 24 October 2025).
## Technological characteristics:
Azurion R3.1 has the same fundamental technological characteristics as compared to the predicate device (Azurion R3.1, K251827, 24 October 2025).
The only change to the subject device, Azurion R3.1 includes the design change to the mattress accessory. It includes the addition of a hook and loop fastener (Velcro) solution for use between the mattress and the system integrated patient table (AD7X), to ensure that the mattress does not slip from the patient table. This hook and loop fastener solution will be added to all existing mattresses (Yellow, green, blue) of the predicate device (Azurion R3.1, K251827, 24 October 2025). In addition,
PHILIPS
Philips IGTS
Page 2 of 5
{6}
PHILIPS
Azurion R3.1
Special 510(k)
510(k) Summary
a new gray color mattress will be released also incorporating the same hook and loop fastener (Velcro) solution as the other mattresses.
The outcome of the technological characteristics comparison and risk assessment demonstrate that the change made to the subject device, Azurion R3.1, does not raise any new questions regarding safety or effectiveness thus demonstrating the substantial equivalence of the subject device, Azurion R3.1, with the predicate device (Azurion R3.1, K251827, 24 October 2025).
## Summary of Non-Clinical Performance Data:
Non-clinical performance testing has been performed on the subject device, Azurion R3.1 and the associated change. The test results demonstrates compliance with the following International and FDA-recognized consensus standards:
- IEC 60601-2-43 Medical electrical equipment - Part 2-43: Particular requirements for the basic safety and essential performance of X-ray equipment for interventional procedures, Edition 2.2 2019-10, Recognition Number 12-329.
- IEC 60601-2-28 Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis, Edition 3.0 2017, Recognition Number 12-309.
- IEC 60601-1-2 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic Compatibility – Requirements and tests, Edition 4.1 2020, Recognition number 19-36.
- IEC TS 60601-4-2 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance medical electrical equipment and medical electrical systems, Edition 1.0 2024, Recognition Number 19-50.
- IEC 62304 Medical device software - Software life cycle processes, Edition 1.1 2015, Recognition number 13-79.
- IEC 62366-1 Medical devices - Part 1: Application of Usability Engineering to Medical Devices, Edition 1.1 2020, Recognition number 5-129.
- ISO 14971 Medical devices - Application of risk management to medical devices, Edition 3.0 corrected version 2019, Recognition number 5-125.
- IEC 61910-1 Medical electrical equipment - Radiation dose documentation - Part 1: Radiation dose structured reports for radiography and radioscopy, Edition 1.0 2014, Recognition Number 12-290.
- NEMA XR-27 X-ray equipment for interventional procedures - User quality control mode, Edition 1.1 Amendment 1-2013, Recognition Number 12-286.
- ISO 15223-1 Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements, Fourth edition 2021-07, Recognition Number 5-134.
- ISO 20417 Medical devices - Information to be supplied by the manufacturer, First edition 2021-04 Corrected version 2021-12, Recognition Number 5-135.
PHILIPS
Philips IGTS
Page 3 of 5
{7}
PHILIPS
Azurion R3.1
Special 510(k)
510(k) Summary
- ISO 17664-2 Processing of health care products - Information to be provided by the medical device manufacturer for the processing of medical devices - Part 2: Non-critical medical devices, First edition 2021-02, Recognition Number 14-579.
- IEC 60601-2-54 Medical Electrical Equipment- Part 2-54: Particular Requirements for the Basic Safety and Essential Performance of X-Ray Equipment for Radiography and Radioscopy, Edition 1.1 2009 + A1:2015 + A2:2018.
- IEC 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance, Edition 3.1 2005 + C1:2006 + C2:2007 + A1:2012.
- IEC 60601-1-3 Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment, Edition 2.1 2008 + A1:2013.
- IEC 60601-1-6 Medical Electrical Equipment Part 1-6: General Requirements for Basic Safety and Essential Performance- Collateral Standard: Usability, Edition 3.1 2010 + A1:2013.
- DICOM Edition 2023
- 21CFR Subchapter J 1020.30 Diagnostic x-ray systems and their major components.
- 21CFR Subchapter J 1020.32 Fluoroscopic equipment.
Non-clinical verification testing of the functional and non-functional requirements as well as performance, reliability and safety has been successfully performed to verify that all the requirements of System Requirements Specification as well as the safety risk control measures from the Risk Management Matrix have been implemented.
Non-clinical validation testing has been performed to validate that the subject device, Azurion R3.1 conforms to the intended use, claims, user needs, effectiveness of safety measures and instructions for use.
All these tests were used to support substantial equivalence of the subject device and demonstrate that the subject device, Azurion R3.1:
- Complies with the aforementioned FDA recognized consensus standards.
- Meets the acceptance criteria and is adequate for its intended use.
Therefore, the subject device, Azurion R3.1 is substantially equivalent to the predicate device (Azurion R3.1, K251827, 24 October 2025) in terms of safety and effectiveness.
PHILIPS
Philips IGTS
Page 4 of 5
{8}
PHILIPS
Azurion R3.1
Special 510(k)
510(k) Summary
## Summary of Clinical Performance Data
The subject device, Azurion R3.1 did not require clinical study since substantial equivalence to the predicate device (Azurion R3.1, K251827, 24 October 2025) was demonstrated with the following attributes:
- Indications for use;
- Technological characteristics;
- Non-clinical performance testing; and
- Safety and effectiveness.
## Substantial Equivalence Conclusion:
The subject device, Azurion R3.1 is substantially equivalent to the predicate device (Azurion R3.1, K251827, 24 October 2025) in terms of indications for use, technological characteristics and safety and effectiveness.
Additionally, substantial equivalence was demonstrated by non-clinical performance tests provided in this 510(k) premarket notification. These tests demonstrate that Azurion R3.1 complies with the system level requirements, user needs and international and FDA-recognized consensus standards, and is as safe and effective as its predicate device (Azurion R3.1, K251827, 24 October 2025), without raising any new safety and/or effectiveness concerns.
PHILIPS
Philips IGTS
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.