Drainage Catheter is designed for percutaneous drainage of abscess fluid, cyst, gall bladders nephrostomy, urinary and other fluids.
Device Story
Percutaneous drainage catheter; used for abscess, cyst, gallbladder, nephrostomy, and urinary fluid drainage. Available in Single Step and Seldinger types; features hydrophilic coating and radiopaque TPU shaft. Sizes 5 Fr-16 Fr; includes various pigtail configurations (string-locking, non-string-locking, mini, closed-loop, multi-side hole). Operated by clinicians in clinical settings. Device placed percutaneously to facilitate fluid collection drainage; operator uses radiopaque markers and depth printing for positioning. Benefits include effective fluid management and reduced risk of complications through biocompatible design. No clinical data; bench testing confirms performance and safety.
Clinical Evidence
No clinical data. Bench testing only. Compliance testing performed per ISO 11135, ISO 11607, ASTM F1980-21, ASTM F640-23, ASTM F1828, EN 1617, ISO 20697, ISO 80369, and ISO 10993 series. Testing included biocompatibility, package validation, transport simulation, 3-year accelerated aging, EO/ECH residuals, and mechanical performance.
Technological Characteristics
Materials: TPU (biocompatible, radiopaque). Configuration: Single Step or Seldinger; pigtail/closed-loop distal ends; hydrophilic coating. Sizes: 5-16 Fr. Sterilization: Ethylene Oxide. Standards: ISO 11135, ISO 11607, ASTM F1980-21, ASTM F640-23, ASTM F1828, EN 1617, ISO 20697, ISO 80369-7, ISO 10993.
Indications for Use
Indicated for percutaneous drainage of abscess fluid, cyst, gall bladders, nephrostomy, urinary, and other fluids in adult patients.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
BIOTEQ Drainage Catheter Set (One Step Type) (K200103)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG
ADMINISTRATION
September 18, 2026
Canyon Medical, Inc.
Huiqi Jiang
RA Manager
Bldg. 3, Phase 2 Accelerator, #11 Yaogu Ave.
Jiangbei New Area
Nanjing, Jiangsu 210032
China
Re: K254179
Trade/Device Name: Drainage Catheter
Regulation Number: 21 CFR 876.5010
Regulation Name: Biliary Catheter and Accessories
Regulatory Class: Class II
Product Code: FGE, GBO
Dated: August 18, 2026
Received: August 18, 2026
Dear Huiqi Jiang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254179 - Huiqi Jiang
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K254179 - Huiqi Jiang
Page 3
Sincerely,
STEPHANIE COLE -S
*for*
Anthony Lee, Ph.D., M.B.A.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K254179 | ? |
| Please provide the device trade name(s). | | ? |
| Drainage Catheter | | |
| Please provide your Indications for Use below. | | ? |
| Drainage Catheter is designed for percutaneous drainage of abscess fluid, cyst, gall bladders nephrostomy, urinary and other fluids. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K254179 - Page 1/8
CANYON
510(k) Summary
### 1. 510(k) Owner
Canyon Medical Inc.
### 2. Submission Correspondent
Submitter Name: Canyon Medical Inc.
Address: Building 3, Phase 2 Accelerator, No. 11 Yaogu Avenue, Jiangbei New Area, 210032, Nanjing, Jiangsu Province, People's Republic of China
Official Correspondent: Ms. Huiqi Jiang
Position: RA Manager
E-mail: jhq@canyonmedical.com.cn
Telephone: +86-13241196637
### 3. Date of Preparation: September 17, 2026
### 4. Subject Device Information
| Common Name: | Drainage Catheter |
| --- | --- |
| Trade Name: | Drainage Catheter |
| Regulation Number: | 21 CFR 876.5010 |
| Regulation Name: | Biliary catheter and accessories |
| Device Class: | II |
| Product Codes: | FGE, GBO |
### 5. Predicate Devices Information
| Subject device | Predicate device |
| --- | --- |
| Drainage Catheter (Seldinger Type) | K210419 BIOTEQ Drainage Catheter BT-PDS-series |
| Drainage Catheter (Single Step Type) | K200103 BIOTEQ Drainage Catheter Set (One Step Type) |
### 6. Device Description
The subject device, Drainage Catheter, is a percutaneous drainage catheter with hydrophilic coating and is used for drainage of abscess and fluid collections. Based on the application method, the Drainage Catheter is classified by puncture technique into Single Step Drainage Catheter and Seldinger Drainage Catheter. Both catheters are made from a soft, biocompatible plastic, a material that is radiopaque.
The operator can use different Drainage Catheters according to the type of accumulated fluid and place of accumulation. These sets include catheter sizes 5 Fr-16 Fr, pertaining to various fluid
2
{5}
K254179 - Page 2/8
[LOGO]
510(k) Summary
viscosity and according to the position of the accumulation.
### 7. Indications for Use
Drainage Catheter is designed for percutaneous drainage of abscess fluid, cyst, gall bladders, nephrostomy, urinary, and other fluids.
### 8. Summary and Comparison of Technological Characteristics
Drainage Catheter is described by puncture technique as the Single Step Drainage Catheter and Seldinger Drainage Catheter. Two predicate devices were selected for comparison. Equivalence comparison is identified with following summary of technical characteristics:
3
{6}
K254179 - Page 3/8
CANTON
510(k) Summary
Table 1. Comparison of Single Step Drainage Catheter to the predicate device
| Item | Subject Device | Predicate Device | Substantial equivalence determination |
| --- | --- | --- | --- |
| Proprietary Name | Drainage Catheter | BIOTEQ Drainage Catheter Set (One Step Type) | |
| 510(k) No. | K254179 | K200103 | |
| Classification | Class II | Class II | Same |
| Product Code | FGE, GBO | FGE, GBO | Same |
| Regulation Number | 21 CFR 876.5010 | 21 CFR 876.5010 | Same |
| Indications for Use | Drainage Catheter is designed for percutaneous drainage of abscess fluid, cyst, gall bladders nephrostomy, urinary and other fluids. | The BIOTEQ Drainage Catheter Set (One Step Type) is designed for percutaneous drainage of abscess fluid, cyst, gall bladders, nephrostomy, urinary, and other fluids. | Same |
| Type of Use | Prescription Use | Prescription Use | Same |
| Intended User | Adults | Adults and recommending clinicians should choose an appropriate catheter size for pediatric use. | Equivalent The intended user of the subject device is covered by the predicate device. |
| Catheter Shaft Material | TPU | TPU | Same |
| Distal Configuration | String Locking Pigtail, Non-String Locking Pigtail | String Locking Pigtail, Non-String Locking Pigtail | Same |
| Distal Shape | Pigtail, Close-loop Pigtail, Mini-Pigtail, Mini-Close loop Pigtail | Pigtail, Closed-Pigtail, Mini-Pigtail, Mini-closed Pigtail, J shape | Equivalent Except J shape, both devices have various pigtail and closed shape; only the names for closed shapes are different. The distal shape of the subject device is covered by the predicate device. |
| Distal Hydrophilic Coating | Yes | Yes | Same |
| Shaft Depth Printing | Yes | Yes | Same |
4
{7}
K254179 - Page 4/8
CANTON
510(k) Summary
| Item | | Subject Device | Predicate Device | Substantial equivalence determination |
| --- | --- | --- | --- | --- |
| Proprietary Name | | Drainage Catheter | BIOTEQ Drainage Catheter Set (One Step Type) | |
| 510(k) No. | | K254179 | K200103 | |
| Proximal Hub Assembly | | Hub (for String Lock Pigtail) | Hub (for String Lock Pigtail), F.L.L. Adapter | Same The proximal hub assembly of subject device is hub (the hub itself includes the adapter). Both devices have hub and adapter. |
| Size | Non-String Lock | Pigtail, Close-loop Pigtail: 5, 6, 7, 8, 10, 12, 14, 16 Fr | 5, 6, 7, 8, 10, 12, 14, 16 Fr | Equivalent |
| | String Lock | Pigtail, Close-loop Pigtail: 5, 6, 7, 8, 10, 12, 14, 16 Fr Mini-Pigtail, Mini-Close loop Pigtail: 6, 7, 8, 10, 12, 14 Fr | 6, 7, 8, 10, 12, 14, 16 Fr | Equivalent |
| Usable Length | | Pigtail, Close-loop Pigtail: 200, 250, 300, 350, 400, 450, 500 mm Mini-Pigtail, Mini-Close loop Pigtail: 200, 250, 300 mm | 200, 250, 300, 350, 400, 450, 500 mm | Equivalent |
| Included Insert Accessory | | • Stiffening Cannula • Trocar Stylet • Flexible Stiffening Cannula • Hub Clip (for String Lock Pigtail) • Thread (for String Lock Pigtail) • Split Straightener • Radiopaque band | • Trocar Needle (Metal Stiffening Cannula) • Trocar Stylet • Flexible (plastic) Stiffening Cannula • Wire cap (for string Lock Pigtail) • Suture Wire • Curve Straightener • Radiopaque band | Equivalent The accessories of subject device are same as the predicate device, only the descriptive names are different. |
| Packaging | | Tyvek/(PET/LDPE) pouch | Tyvek/(PET/LDPE) pouch | Same |
| Sterilization Method | | Ethylene Oxide | Ethylene Oxide | Same |
5
{8}
K254179 - Page 5/8
CAN'ON
510(k) Summary
Table 2. Comparison of Seldinger Drainage Catheter to the predicate device.
| Item | Subject Device | Predicate Device | Substantial equivalence determination |
| --- | --- | --- | --- |
| Proprietary Name | Drainage Catheter | BIOTEQ Drainage Catheter Set (Seldinger Type) | |
| 510(k) No. | K254179 | K210419 | |
| Classification | Class II | Class II | Same |
| Product Code | FGE, GBO | FGE, GBO | Same |
| Regulation Number | 21 CFR 876.5010 | 21 CFR 876.5010 | Same |
| Indications for Use | Drainage Catheter is designed for percutaneous drainage of abscess fluid, cyst, gall bladders nephrostomy, urinary and other fluids. | The BIOTEQ Drainage Catheter Set (Seldinger Type) is designed for percutaneous drainage of abscess fluid, cyst, gall bladders nephrostomy, urinary and other fluids. | Same |
| Type of Use | Prescription Use | Prescription Use | Same |
| Intended User | Adults | Adults and recommending clinicians should choose an appropriate catheter size for pediatric use. | Equivalent The intended user of the subject device is covered by the predicate device. |
| Catheter Shaft Material | TPU | TPU | Same |
| Distal Configuration | String Locking Pigtail, Non-String Locking Pigtail | String Locking Pigtail, Non-String Locking Pigtail | Same |
| Distal Shape | Pigtail, Mini-Pigtail, Close-loop Pigtail, Mini-Close loop Pigtail, Multi-Side holes Pigtail | Pigtail, Closed-Pigtail | Equivalent Both devices have pigtail and closed shapes. The subject device includes a multi-side holes pigtail configuration. |
| Distal Hydrophilic Coating | Yes | Yes | Same |
| Shaft Depth Printing | Yes | Yes | Same |
6
{9}
K254179 - Page 6/8
CAN'ON
510(k) Summary
| Item | | Subject Device | Predicate Device | Substantial equivalence determination |
| --- | --- | --- | --- | --- |
| Proprietary Name | | Drainage Catheter | BIOTEQ Drainage Catheter Set (Seldinger Type) | |
| 510(k) No. | | K254179 | K210419 | |
| Proximal Hub Assembly | | Hub (for String Lock Pigtail) | Hub (for String Lock Pigtail), F.L.L. Adapter | Same The proximal hub assembly of subject device is hub (the hub itself includes the adapter). Both devices have hub and adapter. |
| Size | Non-String Lock | Pigtail, Close-loop Pigtail: 5, 6, 7, 8, 10, 12, 14, 16 Fr Multi-Side holes Pigtail: 8, 10, 12, 14 Fr | 8, 10, 12, 14 Fr | Equivalent |
| | String Lock | Pigtail, Close-loop Pigtail: 5, 6, 7, 8, 10, 12, 14, 16 Fr Mini-Pigtail, Mini-Close loop Pigtail: 6, 7, 8, 10, 12, 14 Fr Multi-Side holes Pigtail: 8, 10, 12, 14 Fr | | |
| Usable Length | | Pigtail, Close-loop Pigtail: 200, 250, 300, 350, 400, 450, 500 mm Mini-Pigtail, Mini-Close loop Pigtail: 200, 250, 300 mm Multi-Side holes Pigtail: 400 mm | 400, 450, 500 mm | Equivalent |
| Included Insert Accessory | | • Stiffening Cannula • Flexible Stiffening Cannula • Hub Clip (for String Lock Pigtail) • Thread (for String Lock Pigtail) • Split Straightener • Radiopaque band | • Metal Stiffening Cannula • Flexible (plastic) Stiffening Cannula • Wire cap • Suture Wire • Curve Straightener • Radiopaque band | Equivalent The accessories of subject device are same as the predicate device, only the descriptive names are different. |
| Packaging | | Tyvek/(PET/LDPE) pouch | Tyvek/(PET/LDPE) pouch | Same |
| Sterilization | | Ethylene Oxide | Ethylene Oxide | Same |
7
{10}
K254179 - Page 7/8
CANTON
510(k) Summary
The technological characteristics of the subject device are substantially equivalent to those of predicate devices. The subject device has the same basic design as the predicate device. The comparison between the subject and predicate devices is based on the following:
- Same indications for use
- Same fundamental technology
- Same sterilization method
### 9. Non-Clinical Test Summary
The subject device has been subjected to compliance testing according to ISO 11135, ISO 11607-1, ISO 11607-2, ASTM F1980-21, ASTM F640-23, ASTM F1828, EN 1617, ISO 20697, and ISO 80369. Results from the testing confirm the design requirements.
1) Biocompatibility
The subject device is an externally communicating medical device with tissue contact of more than 30 days. Relevant tests are conducted as per ISO 10993 series standards.
2) Package Validation
The primary package of the proposed device is intended to maintain the sterility of the product during its shelf life.
3) Transport
The subject device is processed by compression, first vibration, shock and second vibration, and then tested for packaging integrity during transport.
4) Shelf Life
Accelerated aging was used to simulate storage of 3 years and the test results demonstrate that the aged samples complied with the pre-determined acceptance criteria.
5) EO/ECH Residual
Examination of the EO and ECH residuals was conducted in accordance with ISO 10993-7 to evaluate whether the sterilized proposed device comply with allowable limits for permanent patient contact (>30 days).
8
{11}
K254179 - Page 8/8
CANTON
510(k) Summary
# 6) Performance
The performance testing was conducted in accordance with EN 1617, ISO 20697, ASTM F640, ISO 80369-7 and ASTM F1828. All test results demonstrated that the Drainage Catheter met the requirements of the pre-defined acceptance criteria and intended use.
# 10. Clinical Test
None.
# 11. Conclusion
The testing demonstrates that the subject device is safe and effective, performs as well as the legally-marketed predicate devices and raises no new questions of safety or effectiveness.
9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.