Disposable Sterile Skull Drill is a sterile, single-use device intended for cranium perforation during neurosurgical procedures to establish surgical access.
Device Story
Disposable Sterile Skull Drill is a single-use, sterile surgical instrument designed for cranium perforation; couples with surgical drive shanks via standard Hudson coupler. Device features three-fluted cutter and built-in clutch mechanism; automatically stops trephination upon piercing cranial bone of at least 3 mm thickness. Used by neurosurgeons in clinical/OR settings to establish surgical access. Output is physical perforation of skull; enables subsequent neurosurgical intervention. Benefits include controlled, automatic cessation of drilling to prevent unintended penetration beyond bone thickness.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing and biocompatibility testing.
Indicated for cranium perforation in patients with a cranial bone thickness of at least 3 mm during neurosurgical procedures requiring surgical access to the brain.
Regulatory Classification
Identification
Powered compound cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments employ a clutch mechanism to disengage the tip of the instrument after penetrating the skull to prevent plunging of the tip into the brain.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
September 11, 2026
Guizhou Zirui Technology Co., Ltd.
% Yoyo Chen
Senior Regulatory Affairs Manager
Nanos Medical Shanghai Limited
Rm. 1304, #88 Changning Rd., Changning District
Shanghai, 200042
China
Re: K254177
Trade/Device Name: Disposable Sterile Skull Drill (LZ)
Regulation Number: 21 CFR 882.4305
Regulation Name: Powered Compound Cranial Drills, Burrs, Trephines, And Their Accessories
Regulatory Class: Class II
Product Code: HBF
Dated: August 12, 2026
Received: August 12, 2026
Dear Yoyo Chen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254177 - Yoyo Chen
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K254177 - Yoyo Chen
Page 3
Sincerely,
YEN-CHIH
LIN-S
Digitally signed by
YEN-CHIH LIN-S
Date: 2026.09.11
11:11:23 -04'00'
for Jaime Raben, Ph.D.
Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K254177
Device Name
Disposable Sterile Skull Drill
Indications for Use (Describe)
Disposable Sterile Skull Drill is a sterile, single-use device intended for cranium perforation during neurosurgical procedures to establish surgical access.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
K254177, Page 1 of 5
# Traditional 510(k) SUMMARY - K254177
This summary of 510(K) safety and effectiveness information is submitted As Required by requirements of SMDA and 21 CFR §807.92.
## 1. Administrative Information
| **Submission Date** | August 19, 2026, |
| --- | --- |
| **Manufacturer information** | Guizhou Zirui Technology Co., Ltd. Address: 2F&3F, Building B, Phase 1,Guizhou Loushanguan High-Tech Industrial Development Zone, Chumi Town, Tongzi County, Zunyi, Guizhou, China, 563200 Contact: Mr. Changfeng Liu TEL: +86-0851-26628484 E-Mail: sunxiaoxiao@nanosmedical.com |
| **Submission Correspondent** | Nanos Medical Shanghai Limited Room 1304, No.88 Changning Road, Changning District, Shanghai, China, 200042 Contact: Ms. Yoyo Chen Senior Regulatory Affairs Manager E-Mail: yoyo_chen@nanosmedical.com |
## 2. General Information of Subject Device
| 510(k) Number | K254177 |
| --- | --- |
| Device Common Name: | Disposable Sterile Skull Drill |
| Proprietary name of the proposed devices | Disposable Sterile Skull Drill |
| Classification Name: | Drills, Burrs, Trephines & Accessories (Compound, Powered) (21 CFR 882. 4305) |
| Regulation Description: | Powered compound cranial drills, burrs, trephines, and their accessories (21 CFR 882. 4305) |
| Regulation Medical Specialty: | Neurology |
| Product Code(s): | HBF |
| Regulatory Class | Class 2 |
| Type of 510(k) Submission | Traditional 510(k) |
Page 1 of 5
{5}
K254177, Page 2 of 5
### 3. Legally Market Predicate Device
| Device Name: | evonos evoDrill Cranial Perforator |
| --- | --- |
| 510(k) Number | K231403 |
| Product Code: | HBF |
| Device Class | Class 2 |
| Manufacturer | evonos GmbH & CO. KG |
### 4. Device Description
Disposable Sterile Skull Drill (Product family code: LZ) is a removable drill and designed to couple with a surgical drive shank which fits a standard Hudson coupler. It is indicated for drilling access holes in the skull of patients with a cranial bone thickness of at least 3 mm, where treatment or diagnosis requires access to a specific zone of the brain.
The Disposable Sterile Skull Drill is available in different choices of inner and outer drill diameters to meet diverse surgical needs. The device incorporates a built-in clutch mechanism that stops the trephination process as soon as the skull bone has been pierced. Its three-fluted cutter automatically deploys a 3 mm stop-step, which enables automatic release and cessation of drilling once the cranial bone thickness reaches at least 3 mm.
### 5. Intended Use/Indication for Use
Disposable Sterile Skull Drill is a sterile, single-use device intended for cranium perforation during neurosurgical procedures to establish surgical access.
Page 2 of 5
{6}
K254177, Page 3 of 5
## 6. Substantial Equivalence Comparison
Table 1- Comparison with Predicate Device
| Description | | Subject Device (K254177) | Predicate Device (K231403) | Substantial Equivalence Discussion |
| --- | --- | --- | --- | --- |
| 1 | Device Name | Disposable Sterile Skull Drill | evoDrill Cranial Perforator | N/A |
| 2 | Product Code | HBF | HBF | Same |
| 3 | Indications for Use | Disposable Sterile Skull Drill is a sterile, single-use device intended for cranium perforation during neurosurgical procedures to establish surgical access. | The evoDrill Cranial Perforators are single-use surgical devices for cranium perforation. The perforators with the 3mm step automatically release and stop perforating at cranial bone thickness of at least 3mm. The perforators with the 1mm step automatically release and stop perforating at cranial bone thickness of at least 1mm. | Substantially equivalent. The Subject Device shares the same intended use and core automatic stopping technology as the Predicate Device for cranial bone thicknesses of at least 3mm. The Predicate Device's clearance for a broader range of bone thicknesses (including $$\geq 1\text{mm}$$) predicates the safety and effectiveness of the Subject Device's more specific indication. The difference does not raise new questions of safety or effectiveness, supporting a determination of substantial equivalence. |
| 4 | Prescription or OTC | Prescription | Prescription | Same |
| 5 | Compatible power source | Electric | Pneumatic or electric | Substantially equivalent. The difference in power source compatibility does not raise new questions regarding safety or effectiveness. The Subject Device operates entirely within the scope of the Predicate Device's cleared intended use and its validated electric mode. |
| 6 | Fittings type | Hudson | Hudson | Same |
| 7 | Bone Thicknesses | Indicated for use on cranial bone with a minimum thickness of at least 3 mm. | This device is designed to perforate both thin cranial bone (1-3 mm) and thicker bone segments (exceeding 3 mm). | Substantially equivalent The Subject Device operates within a subset of the Predicate Device's cleared range. Performance and safety for bone |
Page 3 of 5
{7}
K254177, Page 4 of 5
| | | | | thicknesses ≥3 mm are already established within the existing predicate clearance. |
| --- | --- | --- | --- | --- |
| 8 | Burr hole diameter (Outer/Inner) and Number of Cutting Flutes | 10/7mm; 12/9mm; 14/11mm. 3-flutes | 14/11mm; 13/9mm; 11/7mm; 9/6mm. 3 or 4 flutes | Substantially equivalent. The differences reflect minor design iterations intended to address specific surgical preferences without altering the core safety or functional profile of the device. |
| 9 | Rotation speed range | 970-1450 rpm | 800-1500rpm | Substantially equivalent. The Subject Device is fundamentally similar to the Predicate Device in terms of structural design and mechanism of action. Furthermore, the rotational speed range of the Subject Device falls within the validated range established for the Predicate Device. |
| 10 | Release Mechanism | The device designed with a built-in clutch mechanism to ensure the device skull drills with the 3mm step automatically release and stop perforating at cranial bone thickness of at least 3mm; | The device designed with a built-in clutch mechanism to ensure: The perforators with the 3mm step automatically release and stop perforating at cranial bone thickness of at least 3mm. The perforators with the 1mm step automatically release and stop perforating at cranial bone thickness of at least 1mm. | Substantially equivalent The Subject Device operates within a subset of the Predicate Device's cleared range. Performance and safety for bone thicknesses ≥3 mm are already established within the existing predicate clearance. |
| 11 | Sterilization | Sterilized by EO (Ethylene Oxide) | Sterilized by Radiation | Substantially equivalent. Sterilization method has been validated and confirmed to be safe and effective. This difference does not alter the safety and performance characteristics of the device, nor does it introduce any new risks. |
| 12 | Packaging system and shelf life | Disposable Sterile Skull Drill is individually packaged | Unknown | Substantially equivalent. The packaging system is designed as final sterile barrier |
Page 4 of 5
{8}
K254177, Page 5 of 5
| | | in Blister Tray. The validated shelf life of each product is 3 years. | | systems (SBS), and their design and validation strictly adhere to the international standard ISO 11607. |
| --- | --- | --- | --- | --- |
| 13 | Materials of Patient Contacting Components | Stainless steel | Plastic/ stainless steel /Aluminum | Substantially equivalent. Biocompatibility results demonstrate that the materials used in the device do not present any potential biocompatibility issues. Therefore, it is concluded that the material differences do not adversely affect the safety and effectiveness of the device. |
Based on the information summarized above, the proposed device is substantially equivalent the currently marketed predicate in indications for use, design, principle of operation and device material.
## 7. Summary of Non-Clinical Test Summary
Bench testing and biocompatibility testing were performed to support a determination of substantial equivalence. The results of these tests provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use. No new safety or performance issues were raised during the device testing.
## 8. Clinical Study Data
The substantial equivalence does not depend on clinical data.
## 9. Conclusion
The Disposable Sterile Skull Drill has been shown through comparison to be substantially equivalent to the identified predicate device.
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.