← Product Code [NKG](/productcode/NKG) · K254165

# Infinity™ OCT System (K254165)

_Medtronic Sofamor Danek USA, Inc. · NKG · Jan 21, 2026 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K254165

## Device Facts

- **Applicant:** Medtronic Sofamor Danek USA, Inc.
- **Product Code:** [NKG](/productcode/NKG.md)
- **Decision Date:** Jan 21, 2026
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3075
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Infinity™ OCT System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7), and the thoracic spine from T1-T3. These devices are to be used with autograft and/or allograft. • Traumatic spinal fractures and/or traumatic dislocations. • Instability or deformity. • Failed previous fusions (e.g. pseudarthrosis). • Tumors involving the cervical spine. • Degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The Infinity™ OCT System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. The Infinity™ OCT System may be used with PASS OCT Patient Specific UNiD OCT rods. In order to achieve additional levels of fixation, the Infinity™ OCT System may be connected to the CD Horizon™ Spinal System and Vertex™ Reconstruction System rods with the Infinity™ OCT System rod connectors. Transition rods with differing diameters may also be used to connect the Infinity™ OCT System to the CD Horizon™ Spinal System. Refer to the CD Horizon™ Spinal System package insert, Vertex™ Reconstruction System and PASS OCT Spinal System package insert for a list of the indications of use. Note: The 3.0mm multi axial screw (MAS) requires the use of MAS CROSSLINK™ at each level in which the 3.0mm screw is intended to be used. The lateral offset connectors and MAS extension connectors are intended to be used with 3.5mm and larger diameter multi axial screws. The lateral offset connectors and MAS extension connectors are not intended to be used with 3.0mm screws. Note: Segmental fixation is recommended for these constructs.

## Device Story

Infinity™ OCT System comprises titanium alloy and cobalt chromium plates, rods, hooks, and screws for posterior occipitocervical-upper thoracic stabilization. Constructs are tailor-made for individual cases; components rigidly lock to rods. System used in spinal surgery by surgeons to provide immobilization/stabilization. Navigated referencing devices (spine clamps) provide rigid fixation between patient and reference frame, enabling assembly of patient reference frame for use with computer-assisted surgery systems (e.g., StealthStation™, Mazor™). Navigated instruments assist in precise anatomical localization during open or minimally invasive procedures. System components are non-sterile, reusable, and require cleaning/sterilization. Benefits include spinal column stabilization and restoration of integrity in tumor patients.

## Clinical Evidence

Bench testing only. Evaluated per FDA Spinal System 510(k) guidance. Testing included ASTM F2193 static and fatigue four-point bend testing. All pre-determined acceptance criteria met. No clinical data provided.

## Technological Characteristics

Materials: Medical grade titanium alloy and cobalt chromium. Components: Plates, rods, hooks, screws, multi-axial screws, connectors. Navigation: Reusable instruments compatible with StealthStation™/Mazor™ systems. Sterilization: Non-sterile, requires steam sterilization (validated wrap). Standards: ASTM F2193 for mechanical testing.

## Regulatory Identification

Posterior cervical screw systems are comprised of multiple, interconnecting components, made from a variety of materials that allow an implant system to be built from the occiput to the upper thoracic spine to fit the patient's anatomical and physiological requirements, as determined by preoperative cross-sectional imaging. Such a spinal assembly consists of a combination of bone anchors via screws (i.e., occipital screws, cervical lateral mass screws, cervical pedicle screws, C2 pars screws, C2 translaminar screws, C2 transarticular screws), longitudinal members (e.g., plates, rods, including dual diameter rods, plate/rod combinations), transverse or cross connectors, interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors), and closure mechanisms (e.g., set screws, nuts). Posterior cervical screw systems are rigidly fixed devices that do not contain dynamic features, including but not limited to: non-uniform longitudinal elements or features that allow more motion or flexibility compared to rigid systems.Posterior cervical screw systems are intended to provide immobilization and stabilization of spinal segments in patients as an adjunct to fusion for acute and chronic instabilities of the cervical spine and/or craniocervical junction and/or cervicothoracic junction such as: (1) Traumatic spinal fractures and/or traumatic dislocations; (2) deformities; (3) instabilities; (4) failed previous fusions (e.g., pseudarthrosis); (5) tumors; (6) inflammatory disorders; (7) spinal degeneration, including neck and/or arm pain of discogenic origin as confirmed by imaging studies (radiographs, CT, MRI); (8) degeneration of the facets with instability; and (9) reconstruction following decompression to treat radiculopathy and/or myelopathy. These systems are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

## Special Controls

*Classification.* Class II (special controls). The special controls for posterior cervical screw systems are:(1) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(2) Nonclinical performance testing must demonstrate the mechanical function and durability of the implant.
(3) Device components must be demonstrated to be biocompatible.
(4) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(5) Labeling must include the following:
(i) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(ii) Intended use and indications for use including levels of fixation;
(iii) Device specific warnings, precautions, and contraindications that include the following statements:
(A) “Precaution: Preoperative planning prior to implantation of posterior cervical screw systems should include review of cross-sectional imaging studies (
*e.g.,* CT and/or MRI) to evaluate the patient's cervical anatomy including the transverse foramen, neurologic structures, and the course of the vertebral arteries. If any findings would compromise the placement of these screws, other surgical methods should be considered. In addition, use of intraoperative imaging should be considered to guide and/or verify device placement, as necessary.”(B) “Precaution: Use of posterior cervical pedicle screw fixation at the C3 through C6 spinal levels requires careful consideration and planning beyond that required for lateral mass screws placed at these spinal levels, given the proximity of the vertebral arteries and neurologic structures in relation to the cervical pedicles at these levels.”
(iv) Identification of magnetic resonance (MR) compatibility status;
(v) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user, and;
(vi) Detailed instructions of each surgical step, including device removal.

## Predicate Devices

- Infinity™ OCT System ([K163375](/device/K163375.md))
- CD HORIZON® Spinal System ([K132639](/device/K132639.md))
- CD HORIZON® Spinal System ([K170679](/device/K170679.md))
- Stealth™ Spine Clamps ([K242464](/device/K242464.md))
- Navigated Interbody Trials and Inserter ([K131425](/device/K131425.md))
- VERTEX® Reconstruction System ([K123656](/device/K123656.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

January 21, 2026

Medtronic Sofamor Danek USA, Inc.
Wafa Mustafa
Sr. Regulatory Affairs Specialist
1800 Pyramid Place
Memphis, Tennessee 38132

Re: K254165
Trade/Device Name: Infinity™ OCT System
Regulation Number: 21 CFR 888.3075
Regulation Name: Posterior Cervical Screw System
Regulatory Class: Class II
Product Code: NKG, KWP, OLO
Dated: December 22, 2025
Received: December 22, 2025

Dear Wafa Mustafa:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K254165 - Wafa Mustafa
Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}

K254165 - Wafa Mustafa
Page 3

Sincerely,

EILEEN
CADEL-S
for
Colin O'Neill, M.B.E.
Assistant Director
DHT6B: Division of Spinal Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

K254165
Page 1 of 2

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K254165 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  Infinity™ OCT System  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  Infinity™ OCT System
The Infinity™ OCT System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7), and the thoracic spine from T1-T3. These devices are to be used with autograft and/or allograft.
• Traumatic spinal fractures and/or traumatic dislocations.
• Instability or deformity.
• Failed previous fusions (e.g. pseudarthrosis).
• Tumors involving the cervical spine.
• Degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability.
The Infinity™ OCT System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
The Infinity™ OCT System may be used with PASS OCT Patient Specific UNiD OCT rods. In order to achieve additional levels of fixation, the Infinity™ OCT System may be connected to the CD Horizon™ Spinal System and Vertex™ Reconstruction System rods with the Infinity™ OCT System rod connectors.
Transition rods with differing diameters may also be used to connect the Infinity™ OCT System to the CD Horizon™ Spinal System. Refer to the CD Horizon™ Spinal System package insert, Vertex™ Reconstruction System and PASS OCT Spinal System package insert for a list of the indications of use.
Note: The 3.0mm multi axial screw (MAS) requires the use of MAS CROSSLINK™ at each level in which the 3.0mm screw is intended to be used.
The lateral offset connectors and MAS extension connectors are intended to be used with 3.5mm and larger diameter multi axial screws. The lateral offset connectors and MAS extension connectors are not intended to be used with 3.0mm screws.
Note: Segmental fixation is recommended for these constructs.  |   |   |
|  Medtronic Navigated Reusable Instruments for use with StealthStation™ and IPC™ POWEREASE™ Systems
Medtronic Navigated Reusable Instruments are intended to be used during the preparation and placement of Medtronic screws during spinal surgery to assist in precisely locating anatomical structures in either open, or minimally invasive, procedures. Medtronic Navigated Reusable Instruments are specifically designed for use with the StealthStation™ System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra can be identified relative to a CT or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy.
Stealth™ Spine Clamps
When used with Medtronic computer assisted surgery systems, defined as including the Stealth™ System,  |   |   |

{4}

the following indications of use are applicable:

- The spine referencing devices are intended to provide rigid fixation between patient and patient reference frame for the duration of the surgery. The devices are intended to be reusable.
- The navigated instruments are specifically designed for use with Medtronic computer assisted surgery systems, which are indicated for any medical condition in which the use of stereotactic surgery may be appropriate or vertebra can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.
- The Stealth™ spine clamps are indicated for skeletally mature patients.

Please select the types of uses (select one or both, as applicable).

☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

K254165
Page 2 of 2

{5}

K254165
Page 1 of 6

510(k) Summary

MEDTRONIC
Infinity™ OCT System
December 22, 2025

|  I. Submitter | Medtronic Sofamor Danek, USA Inc.
1800 Pyramid Place
Memphis, Tennessee 38132  |
| --- | --- |
|  Contact Person | Wafa Mustafa
Sr. Regulatory Affairs Specialist
Email: wafa.mustafa@medtronic.com  |
|  II. Name of Device | Infinity™ OCT System  |
|  Common Name | Bone Screw, Stereotaxic Instrument  |
|  Classification Name | Posterior cervical screw system, Stereotaxic Instrument  |
|  Classification | Implants: Class II
Instruments: Class II  |
|  Product Codes | Implants: NKG (888.3075)
Implants: KWP (888.3050)
Instruments: OLO (882.4560)  |
|  III. Predicate Devices | **Primary Predicate:**
1. Infinity™ OCT System (K163375, S.E. 08/21/2017)

**Additional Predicate:**
2. CD HORIZON® Spinal System (K132639, S.E. 11/25/2013)
3. CD HORIZON® Spinal System (K170679, S.E. 05/11/2017)
4. Stealth™ Spine Clamps (K242464, S.E. 06/02/2025)
5. Navigated Interbody Trials and Inserter (K131425, S.E. 08/04/2023)
6. VERTEX® Reconstruction System (K123656, S.E. 02/25/2013)

The predicates have not been subject to a design related recall.  |

{6}

K254165
Page 2 of 6

IV. Device Description

**Infinity™ OCT System**

The Infinity™ OCT System is a posterior occipitocervical-upper thoracic system, which consists of a variety of shapes and sizes of plates, rods, hooks, screws, multi-axial screws, and connecting components, which can be rigidly locked to the rod in a variety of configurations, with each construct being tailor-made for the individual case.

The Infinity™ OCT System is fabricated from medical grade titanium alloy and medical grade cobalt chromium. Medical grade titanium alloy and medical grade cobalt chromium may be used together. Never use titanium alloy and/or cobalt chromium with stainless steel in the same construct. No warranties, expressed or implied, are made. Implied warranties of merchantability and fitness for a particular purpose or use are specifically excluded.

To achieve best results, do not use any of the Infinity™ OCT System implant components with components from any other system or manufacturer unless specifically labeled to do so in this or another Medtronic document. As with all orthopedic and neurosurgical implants, Infinity™ OCT System components should never be reused under any circumstances.

**Medtronic Navigated Referencing Devices**

Medtronic navigated referencing devices provide rigid fixation to patient anatomy and enables the assembly of a patient reference frame. Various types of instruments are available including spine referencing clamps and associated connections to maintain compatibility with current spine reference frames.

Medtronic navigated referencing devices are compatible with various Medtronic spinal implant systems including Medtronic computer assisted surgery systems, including, but not limited to, StealthStation™ and/or Mazor™ systems, and Medtronic navigation reusable instruments. These instruments are also compatible with Medtronic’s IPC™ Powerease™ System when connected to the Powerease™ Driver.

Medtronic navigated referencing devices are supplied as non-sterile reusable instruments. All devices should be cleaned and sterilized before use.

Medtronic navigated referencing devices are manufactured from a variety of materials commonly used in orthopedic and neurological procedures, which meet applicable national and/or international

{7}

K254165
Page 3 of 6

|   | standards. These instruments may be shipped in accessories. Accessories include cases, trays, lids, caddies, and modules fabricated from a variety of materials, which meet national and international standards, and are commonly used to enclose, protect, and organize Medtronic orthopedic and neurological reusable instruments.

**Medtronic Reusable Instruments and Accessories**
Reusable instruments are manufactured from a variety of materials commonly used in orthopedic and neurological procedures, which meet applicable national and/or international standards. These instruments may be shipped in Medtronic Transportation/ Sterilization Cassettes. The Medtronic Transportation/Sterilization Cassettes are designed to be used in conjunction with a legally marketed, validated, and FDA-cleared sterilization wrap to allow steam sterilization of non-sterile devices.
Medtronic Transportation/Sterilization Cassettes include cases, trays, lids, caddies, modules, or brackets fabricated from a variety of materials, which meet national and international standards, and are commonly used to enclose, protect, and organize Medtronic orthopedic or neurological non-sterile devices. The Medtronic Transportation/Sterilization Cassettes are not intended on their own to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA-cleared sterilization wrap. The Medtronic Transportation Sterilization Cassettes are intended for transport of non-sterile loads.  |
| --- | --- |
|  V. Indications for Use | **Infinity™ OCT System**
The Infinity™ OCT System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7), and the thoracic spine from T1-T3. These devices are to be used with autograft and/or allograft.
• Traumatic spinal fractures and/or traumatic dislocations.
• Instability or deformity.
• Failed previous fusions (e.g. pseudarthrosis).
• Tumors involving the cervical spine.
• Degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability.  |

{8}

K254165
Page 4 of 6

The Infinity™ OCT System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

The Infinity™ OCT System may be used with PASS OCT Patient Specific UNiD OCT rods. In order to achieve additional levels of fixation, the Infinity™ OCT System may be connected to the CD Horizon™ Spinal System and Vertex™ Reconstruction System rods with the Infinity™ OCT System rod connectors. Transition rods with differing diameters may also be used to connect the Infinity™ OCT System to the CD Horizon™ Spinal System. Refer to the CD Horizon™ Spinal System package insert, Vertex™ Reconstruction System, and PASS OCT Spinal System package inserts for a list of the indications of use.

Note: the 3.0mm multi axial screw (MAS) requires the use of MAS Crosslink™ at each level in which the 3.0mm screw is intended to be used.

The lateral offset connectors and MAS extension connectors are intended to be used with 3.5mm and larger diameter multi axial screws. The lateral offset connectors and MAS extension connectors are not intended to be used with 3.0mm screws.

Note: segmental fixation is recommended for these constructs.

## Medtronic Navigated Reusable Instruments for use with StealthStation™ and IPCTM POWEREASE™ Systems

Medtronic Navigated Reusable Instruments are intended to be used during the preparation and placement of Medtronic screws during spinal surgery to assist in precisely locating anatomical structures in either open, or minimally invasive, procedures. Medtronic Navigated Reusable Instruments are specifically designed for use with the StealthStation™ System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra can be identified relative to a CT or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy.

## Stealth™ Spine Clamps

When used with Medtronic computer assisted surgery systems,

{9}

K254165
Page 5 of 6

|   | defined as including the Stealth™ System, the following indications of use are applicable:
- The spine referencing devices are intended to provide rigid fixation between patient and patient reference frame for the duration of the surgery. The devices are intended to be reusable.
- The navigated instruments are specifically designed for use with Medtronic computer assisted surgery systems, which are indicated for any medical condition in which the use of stereotactic surgery may be appropriate or vertebra can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.
- The Stealth™ spine clamps are indicated for skeletally mature patients.  |
| --- | --- |
|  VI. Comparison of Technological Characteristics with the Predicate Devices | The subject devices have the same intended use, indications for use, materials, similar overall design, fundamental technology, sterilization, and surgical technique as the Infinity™ OCT System, CD HORIZON® Spinal System, Stealth™ Spine Clamps, and Navigated Infinity™ Instruments predicates:
- K163375, S.E. 08/21/2017
- K132639, S.E. 11/25/2013
- K170679, S.E. 05/11/2017
- K242464, S.E. 06/02/2025
- K131425, S.E. 08/04/2023
- K123656, S.E. 02/25/2013 The subject and predicate implants and instruments have the same function and fundamental scientific technology.  |
|  VII. Performance Data | In accordance with the Guidance for Industry and FDA Staff – Spinal System 510(k)’s, Medtronic has evaluated the subject devices to demonstrate substantial equivalence to the predicate devices.
The subject devices have been tested or rationalized based on if Medtronic believes that testing is not warranted for the subject devices as they do not present a new worst case when compared to the predicates.
Testing and/or Rationale completed for the following:
- ASTM F2193: Static Four-Point Bend, Four-Point Bend Fatigue  |

{10}

K254165
Page 6 of 6
|   | For the tested subject devices, the pre-determined acceptance criteria was met for all tests. For subject devices that are rationalized, all existing predicate data previously provided in the predicate 510(k)s is still applicable. Design validation has also been performed and demonstrated that the subject devices performed as intended.

Therefore, Medtronic believes the design verification and validation testing demonstrated that the subject devices are substantially equivalent to the predicate devices.  |
| --- | --- |
|  VIII. Conclusion | Based on the supporting evidence provided in this premarket notification, Medtronic believes the subject devices are substantially equivalent to the predicate devices.  |

---

**Source:** [https://fda-staging.innolitics.com/device/K254165](https://fda-staging.innolitics.com/device/K254165)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
