K254160 · VacHeal, LLC · OKO · Sep 18, 2026 · General, Plastic Surgery
Device Facts
Record ID
K254160
Device Name
VacHeal Negative Pressure Wound Therapy System
Applicant
VacHeal, LLC
Product Code
OKO · General, Plastic Surgery
Decision Date
Sep 18, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4683
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The VacHeal™ NPWT System is indicated for patients who would benefit from wound management via application of negative pressure to remove small amounts of exudate from closed surgical incisions.
Device Story
Non-powered, portable, single-patient-use, sterile, disposable NPWT system; operates via spring-based mechanical mechanism to generate and regulate negative pressure; maintains -75 mmHg to -125 mmHg nominal range; includes one-piece dressing with AQUAZONE® polyester foam and 30 mL pump-mounted reservoir; used for closed surgical incisions; clinician-managed; provides continuous therapy; pressure stability maintained via Lock/Unlock knob; visual indicators confirm operation; benefits patient by managing exudate and supporting wound healing; wear time up to 7 days.
Clinical Evidence
No clinical data. Evidence consists of non-clinical bench testing including pumping time, pressure decay, cycle testing, valve pressure, mechanical strength (drop/pull tests), indicator verification, 7-day pressure/exudate handling, shelf life, and ISO 10993-1 biocompatibility testing. A porcine model study for closed surgical incision wound healing was also conducted.
Technological Characteristics
Non-powered, spring-based mechanical suction; nominal pressure range -75 to -125 mmHg; AQUAZONE® polyester foam dressing; 30 mL reservoir; single-use; sterile; disposable. Patient-contacting materials comply with ISO 10993-1. No electrical components.
Indications for Use
Indicated for patients requiring wound management via negative pressure to remove small amounts of exudate from closed surgical incisions.
Regulatory Classification
Identification
A non-powered suction apparatus device intended for negative pressure wound therapy is a device that is indicated for wound management via application of negative pressure to the wound for removal of fluids, including wound exudate, irrigation fluids, and infectious materials. It is further indicated for management of wounds, burns, flaps, and grafts.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Non-powered Suction Apparatus Device Intended for Negative Pressure Wound Therapy (NPWT).” See § 878.1(e) for the availability of this guidance document.
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
September 18, 2026
VacHeal, LLC
Tiffany Morrison
President
40 E. Montgomery Ave.
4th Floor
Ardmore, Pennsylvania 19003
Re: K254160
Trade/Device Name: VacHeal Negative Pressure Wound Therapy System
Regulation Number: 21 CFR 878.4683
Regulation Name: Non-Powered suction apparatus device intended for negative pressure wound therapy
Regulatory Class: Class II
Product Code: OKO
Dated: December 19, 2025
Received: December 22, 2025
Dear Tiffany Morrison:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254160 - Tiffany Morrison
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K254160 - Tiffany Morrison
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
YU-CHIEH CHIU -S
Yu-Chieh Chiu, Ph.D.
Assistant Director
DHT4B: Division of Plastic and Reconstructive Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K254160 | ? |
| Please provide the device trade name(s). | | ? |
| VacHeal™ Negative Pressure Wound Therapy (NPWT) System | | |
| Please provide your Indications for Use below. | | ? |
| The VacHeal™ NPWT System is indicated for patients who would benefit from wound management via application of negative pressure to remove small amounts of exudate from closed surgical incisions. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K254160
## 510(K) Summary
K254160
### Contact Details
| Applicant Name | VacHeal, LLC |
| --- | --- |
| Applicant Address | 40 East Montgomery Ave. 4^{th} Floor Ardmore PA 19003 United States |
| Applicant Contact Telephone | 717-856-1202 |
| Applicant Contact | Ms. Tiffany Morrison |
| Applicant Contact Email | tiffany@parvizisurgical.com |
### Device Name
| Device Trade Name | VacHeal™ Negative Pressure Wound Therapy (NPWT) System |
| --- | --- |
| Common Name | Non-Powered suction apparatus device intended for negative pressure wound therapy |
| Classification Name | Negative Pressure Wound Therapy Non-Powered Suction Apparatus |
| Regulation Number | 878.4683 |
| Product Code(s) | OKO |
### Legally Marketed Predicate Devices
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| --- | --- | --- |
| K151710 | SNaP® Wound Care System | OKO |
### Device Description
The VacHeal™ Negative Pressure Wound Therapy (NPWT) System is a non-powered, portable, single patient use, sterile, disposable NPWT System that is intended for wound management via application of negative pressure to closed surgical incisions for removal of small amounts of exudate.
The VacHeal™ System maintains Negative Pressure Wound Therapy in the -75 mmHg to -125 mmHg nominal range and has a wear time up to 7 days.
### Intended Use/ Indications for Use
The VacHeal™ NPWT System is indicated for patients who would benefit from wound management via application of negative pressure to remove small amounts of exudate from closed surgical incisions.
Page 1 of 4
{5}
K254160
## Indications for Use Comparison
The VacHeal™ Negative Pressure Wound Therapy (NPWT) System and SNaP® Wound Care System share the same intended purpose of applying negative pressure to wounds to allow wound management through the removal of small amounts of exudate. While the SNaP® Wound Care System is intended for the management of wound types including chronic, acute, traumatic, subacute, and dehisced wounds; partial-thickness burns; ulcers (e.g., diabetic or pressure); surgically closed incisions; flaps; and grafts, VacHeal™ NPWT System is intended only for the use on closed surgical incisions.
The VacHeal™ NPWT System does not introduce new indications or expand upon the intended use of the predicate device.
Therefore, the VacHeal™ NPWT System is substantially equivalent to the SNaP® Wound Care System with respect to intended use and indications for use.
## Technical Comparison
The VacHeal™ NPWT System, the predicate SNaP® Wound Care System, are non-powered, portable, single-use, disposable Negative Pressure Wound Therapy (NPWT) systems intended for wound management via the application of negative pressure to remove exudate.
These devices utilize suction and mechanically generated negative pressure without the use of electrically powered components. In the VacHeal™ NPWT System and SNaP® System, negative pressure is generated and regulated through spring-based mechanical mechanisms within a closed system, providing continuous therapy. Maximum pressure limitation and over-pressure prevention are achieved through mechanical means. Each system incorporates visual indicators to confirm proper operation and locking mechanisms to prevent pressure loss and maintain pressure stability over time.
The VacHeal™ NPWT System and the SNaP® Wound Care System differ in certain technological characteristics. The VacHeal™ NPWT System delivers variable negative pressure within a nominal pressure range of -75 mmHg to -125 mmHg, whereas the SNaP® Wound Care System delivers fixed negative pressure levels (75, 100, or 125 mmHg) depending on cartridge size. Pressure stability in the VacHeal™ NPWT System is maintained via a Lock/Unlock knob, while the SNaP® Wound Care System utilizes an activation/reset key that is removed after activation.
Additional differences include dressing configuration and exudate management. The VacHeal™ NPWT System utilizes a one-piece dressing with AQUAZONE® polyester foam and a 30 mL pump-mounted reservoir/canister for fluid collection. The SNaP® Wound Care System utilizes a proprietary hydrocolloid dressing with reticulated open-cell foam and a gel-containing 60 mL cartridge for exudate containment. Dressing change frequency differs, with the VacHeal™ NPWT System, the dressing is intended for single use for up to 7 days, with frequency adjusted by the clinician as appropriate.
Page 2 of 4
{6}
K254160
All patient-contacting materials for both devices comply with ISO 10993-1. The identified differences in technological characteristics do not alter the fundamental principle of operation and do not raise new questions of safety or effectiveness. Performance and bench testing demonstrate that the VacHeal™ NPWT System performs as intended and is as safe and effective as the predicate device.
### **Non-Clinical and/or Clinical Tests Summary**
The following performance data were provided in support of the substantial equivalence determination.
#### ***Non-Clinical Testing***
Non-clinical testing was conducted on the VacHeal™ NPWT System according to the special controls guidance document "Non-powered Suction Apparatus Device Intended for Negative Pressure Wound Therapy (NPWT) - Class II Special Controls Guidance for Industry and FDA Staff" (issued June 16, 2026) and in accordance with the international standard ISO 10993-1 "Biological Evaluation of Medical Devices- Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. Non-clinical testing included:
- Pumping Time to Operating Pressure
- Pressure Decay Vs. Time
- Cycle Test Pressure System and Locking Mechanism
- Umbrella Relief Valve Pressure
- Bulb Press Force
- Cycle Test Bulb
- Drop Test
- Pull Test on Tubing Joints
- Flag Indicator Verification Testing
- Pressure and Exudate Handling Over Time Test (7-Day Wear)
- Shelf Life Test
Biocompatibility:
-Cytotoxicity
-Sensitization
-Irritation
-Material Medicated Pyrogenicity
-Acute and Subacute systemic toxicity
Page 3 of 4
{7}
K254160
-Closed Surgical Incision Wound Healing animal study in porcine model
### Clinical Testing
Not Applicable
### Conclusions
The test data submitted for the VacHeal™ NPWT System support the safety of the device and that the device shall perform as intended in the specified use conditions. The data demonstrate that the VacHeal™ NPWT System is substantially equivalent to the predicate device with respect to the intended use and technical comparison. The performance data demonstrates that any technological differences between the subject and predicate device do not raise new issues of safety and effectiveness and is therefore the VacHeal™ NPWT System is substantially equivalent to SNaP® Wound Care System.
Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.