CROSSNAV Navigation Enabled Instruments for Disc Preparation and Interbody
K254157 · Medos International SARL · OLO · May 26, 2026 · Neurology
Device Facts
Record ID
K254157
Device Name
CROSSNAV Navigation Enabled Instruments for Disc Preparation and Interbody
Applicant
Medos International SARL
Product Code
OLO · Neurology
Decision Date
May 26, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Navigation Enabled Instruments are reusable instruments indicated to be used during the preparation and placement of DePuy Synthes implants as well as discectomy and / or bony resection, during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or percutaneous procedures. The Navigation Enabled Instruments are designed for use with only the specific DePuy Synthes implant system(s) for which they are indicated and with the VELYS SPINE Navigation as well as with the Brainlab Navigation System and the Medtronic StealthStation Navigation System. The Navigation Enabled Instruments are indicated for use in surgical spinal procedures, in which: • the use of EXPEDIUM™, VIPER™, SYMPHONY™ OCT, TriALTIS™, and the CONDUIT™ Spine System is indicated, if used for implant placement, • the use of stereotactic surgery may be appropriate, and • reference to a rigid anatomical structure, such as the pelvis or a vertebra, can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy. These procedures include but are not limited to spinal fusion. The Navigation Enabled Instruments are also compatible with DePuy Synthes Power Systems and the Medtronic IPC™ POWEREASE System. The Navigation Enabled Instruments used in conjunction with the SYMPHONY™ OCT System are intended to support indicated cervical and thoracic polyaxial screw placement only. The Navigation Enabled Instruments indicated for discectomy and / or bony resection, as well as the CONDUIT implants placement are for use with Medtronic StealthStation System only.
Device Story
Reusable surgical instruments (drills, taps, screwdrivers, sizers, inserters) for spinal surgery; used in open or percutaneous procedures. Instruments attach to navigation tracking arrays (VELYS, Brainlab, or Medtronic StealthStation) to provide real-time intraoperative visualization of instrument position relative to patient anatomy. Surgeon uses navigation system to precisely locate anatomical structures and guide implant placement or bone resection. Compatible with DePuy Synthes and Medtronic power systems. Benefits include improved surgical precision and guidance during spinal procedures.
Clinical Evidence
No clinical data. Performance evaluation included dimensional comparisons, quantitative accuracy testing in a simulated clinical setting using cadaver specimens, and summative usability evaluation in a simulated clinical setting using cadaver specimens.
Technological Characteristics
Reusable stainless steel instruments. Includes drills, taps, screwdrivers, sizers, and inserters. Navigational tracking via external systems (VELYS, Brainlab, Medtronic StealthStation) using tracking arrays. Compatible with manual or power-driven operation. Sterilization required.
Indications for Use
Indicated for patients undergoing spinal surgery (e.g., spinal fusion) requiring discectomy, bony resection, or placement of specific DePuy Synthes implants (EXPEDIUM, VIPER, SYMPHONY OCT, TriALTIS, CONDUIT) where stereotactic navigation is appropriate and rigid anatomical structures (pelvis/vertebrae) can be referenced via CT, MR, or fluoroscopic imaging.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
CROSSNAV Navigation Enabled Instruments and UNAS Navigation Arrays (K241893)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 26, 2026
Medos International SARL
% Marina Minnock
Regulatory Affairs Specialist
Synthes GmbH
Eimattstrasse 3
Oberdorf, BL 4436
Switzerland
Re: K254157
Trade/Device Name: CROSSNAV Navigation Enabled Instruments for Disc Preparation and Interbody
Regulation Number: 21 CFR 882.4560
Regulation Name: Stereotaxic Instrument
Regulatory Class: Class II
Product Code: OLO
Dated: April 27, 2026
Received: April 27, 2026
Dear Marina Minnock:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254157 - Marina Minnock
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K254157 - Marina Minnock
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali -S
Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Page 1 of 1
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K254157 | ? |
| Please provide the device trade name(s). | | ? |
| CROSSNAV Navigation Enabled Instruments for Disc Preparation and Interbody | | |
| Please provide your Indications for Use below. | | ? |
| Navigation Enabled Instruments are reusable instruments indicated to be used during the preparation and placement of DePuy Synthes implants as well as discectomy and / or bony resection, during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or percutaneous procedures. The Navigation Enabled Instruments are designed for use with only the specific DePuy Synthes implant system(s) for which they are indicated and with the VELYS SPINE Navigation as well as with the Brainlab Navigation System and the Medtronic StealthStation Navigation System. The Navigation Enabled Instruments are indicated for use in surgical spinal procedures, in which: • the use of EXPEDIUM™, VIPER™, SYMPHONY™ OCT, TriALTIS™, and the CONDUIT™ Spine System is indicated, if used for implant placement, • the use of stereotactic surgery may be appropriate, and • reference to a rigid anatomical structure, such as the pelvis or a vertebra, can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy. These procedures include but are not limited to spinal fusion. The Navigation Enabled Instruments are also compatible with DePuy Synthes Power Systems and the Medtronic IPC™ POWEREASE System. The Navigation Enabled Instruments used in conjunction with the SYMPHONY™ OCT System are intended to support indicated cervical and thoracic polyaxial screw placement only. The Navigation Enabled Instruments indicated for discectomy and / or bony resection, as well as the CONDUIT implants placement are for use with Medtronic StealthStation System only. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K254157
# 510(k) SUMMARY
## A. Submitter Information
510(k) Sponsor: Medos International, SARL
Contact Person: Marina Minnock
Regulatory Affairs Specialist
Eimattstrasse 3
4436 Oberdorf BL
Switzerland
Telephone: +41 76 706 07 99
Email: MMinock@its.jnj.com
## B. Date Prepared
26 May 2026
## C. Device Name
Trade/Proprietary Name: CROSSNAV Navigation Enabled Instruments for Disc Preparation and Interbody
Common/Usual Name: Orthopedic Stereotaxic Instrument
Device Classification and Regulation: Class II
21 CFR §882.4560
Product Code: OLO
## D. Predicate Device Names
Primary Predicate Device: StealthStation S8 Spine Software (K251282) – OLO
Additional Predicate Device: CROSSNAV Navigation Enabled Instruments and UNAS Navigation Arrays (K241893) – OLO
Page 1 of 3
{5}
K254157
## E. Device Description
Navigation Enabled Instruments (CROSSNAV™ Instruments) are reusable instruments used for the preparation and placement of DePuy Synthes EXPEDIUM™, VIPER™, SYMPHONY™ OCT, TriALTIS™, and CONDUIT™ implants, in either open or percutaneous procedures. The Navigation Enabled Instruments include drills, taps and screwdrivers and can be operated manually or under power. They also include manual tissue and bone removal / manipulation devices as well as interbody sizers and interbody implant inserters. These instruments are designed for navigated and non-navigated use. Navigation Enabled Instruments also include the CROSSNAV™ Adaptor to facilitate navigation of the Navigation Enabled Instruments with the VELYS™ SPINE Navigation as well as with the Brainlab Navigation System. Navigation of these instruments is achieved using the VELYS™ SPINE Navigation as well as the Brainlab Navigation System and the Medtronic StealthStation™ Navigation System and their associated tracking arrays.
## F. Indications for Use
Navigation Enabled Instruments are reusable instruments indicated to be used during the preparation and placement of DePuy Synthes implants as well as discectomy and / or bony resection, during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or percutaneous procedures. The Navigation Enabled Instruments are designed for use with only the specific DePuy Synthes implant system(s) for which they are indicated and with the VELYS SPINE Navigation as well as with the Brainlab Navigation System and the Medtronic StealthStation Navigation System. The Navigation Enabled Instruments are indicated for use in surgical spinal procedures, in which:
- the use of EXPEDIUM™, VIPER™, SYMPHONY™ OCT, TriALTIS™, and the CONDUIT™ Spine System is indicated, if used for implant placement,
- the use of stereotactic surgery may be appropriate, and
- reference to a rigid anatomical structure, such as the pelvis or a vertebra, can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy.
These procedures include but are not limited to spinal fusion. The Navigation Enabled Instruments are also compatible with DePuy Synthes Power Systems and the Medtronic IPC™ POWEREASE System.
The Navigation Enabled Instruments used in conjunction with the SYMPHONY™ OCT System are intended to support indicated cervical and thoracic polyaxial screw placement only. The Navigation Enabled Instruments indicated for discectomy and / or bony resection, as well as the CONDUIT implants placement are for use with Medtronic StealthStation System only.
Page 2 of 3
{6}
K254157
## G. Summary of Similarities and Differences in Technological Characteristics, Performance, and Intended Use
The technological characteristics, including design, material and performance as well as intended use of CROSSNAV Navigation Enabled Instruments for Disc Preparation and Interbody are consistent with those of the predicate devices.
The subject devices represent an additional subset of Navigation Enabled Instruments designed for surgical site access, surgical site preparation, and interbody implant trialing and insertion. Similarly to predicate devices, CROSSNAV Navigation Enabled Instruments for Disc Preparation and Interbody are designed for Medtronic StealthStation Navigation System compatibility and utilize the same general characteristics and navigation workflow as previously cleared Navigation Enabled Instruments, including navigation tracker attachment, toolcard selection and intraoperative visualization of instrument position. Compatibility with the Medtronic StealthStation Navigation System is established via the existing DePuy Synthes coupling to NavLock trackers. This does not raise new questions of safety and effectiveness based on application of recognized consensus standards and design controls.
## H. Materials
The CROSSNAV Navigation Enabled Instruments for Disc Preparation and Interbody are manufactured from stainless steel alloys. The biocompatibility evaluation of the instruments leveraged equivalence in materials and manufacturing to previously cleared instruments (K241893).
## I. Performance Data
The performance data for the CROSSNAV Navigation Enabled Instruments for Disc Preparation and Interbody consists of the following evaluations:
- Dimensional comparisons between subject and predicate devices,
- Quantitative accuracy evaluation in a simulated clinical setting with cadaver specimens,
- Summative usability evaluation in a simulated clinical setting with cadaver specimens.
## J. Conclusion
The indications for use of CROSSNAV Navigation Enabled Instruments for Disc Preparation and Interbody are consistent with those of the predicate devices. The technological characteristics of CROSSNAV Navigation Enabled Instruments for Disc Preparation and Interbody in terms of design, materials and performance are consistent with those of the predicate devices. CROSSNAV Navigation Enabled Instruments for Disc Preparation and Interbody are substantially equivalent to the predicate devices.
Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.