← Product Code [ELW](/productcode/ELW) · K254152

# Dia-X Sil Light Body (K254152)

_DiaDent Group International · ELW · Sep 16, 2026 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K254152

## Device Facts

- **Applicant:** DiaDent Group International
- **Product Code:** [ELW](/productcode/ELW.md)
- **Decision Date:** Sep 16, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3660
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

Crown and bridge impressions - Inlay and onlay impressions - Functional impressions - Implant impressions - Denture and partial denture impressions

## Device Story

Dia-X Sil Light Body is a vinyl polysiloxane (VPS) dental impression material; supplied in cartridges for use by dental professionals in clinical settings. The material is a two-part base/catalyst system that cures to form an elastomeric impression of oral structures. It is used to capture precise anatomical details of teeth and gums for restorative or prosthetic procedures. The clinician applies the material to a tray or directly to the oral site; the material sets to create a negative mold, which is then used to produce a dental model. This process facilitates the fabrication of crowns, bridges, inlays, onlays, and dentures, benefiting the patient by ensuring accurate fit and function of dental restorations.

## Clinical Evidence

No clinical data was collected or provided. Substantial equivalence is supported by bench testing in accordance with ISO 4823:2021 (working time, consistency, detail reproduction, linear dimensional change, compatibility with gypsum, elastic recovery, and strain in compression) and biocompatibility testing per ISO 10993-5, ISO 10993-10, and ISO 10993-23.

## Technological Characteristics

Vinyl polysiloxane (VPS) impression material; Type 3 (ISO 4823). Composition includes siloxanes, silicones, silica, silicone polyethercopolymer, and platinum-based catalyst. Two-part base/catalyst handling system. Non-sterile. Physical properties: consistency ≥36mm, detail reproduction 20μm, elastic recovery >96.5%, strain in compression 2-20%.

## Regulatory Identification

Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.

## Special Controls

*Classification.* Class II (Special Controls).

## Predicate Devices

- Bright Impress - Light ([K214086](/device/K214086.md))
- DentMix VPS Impression Material ([K223892](/device/K223892.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

**FDA** U.S. FOOD & DRUG
ADMINISTRATION

September 16, 2026

DiaDent Group International
Jin An
RA Team
16, Osongsaengmyeong 4-Ro, Osong-Eup, Heungdeok-Gu
Cheongju-Si, 28186
REPUBLIC OF KOREA

Re: K254152

Trade/Device Name: Dia-X Sil Light Body
Regulation Number: 21 CFR 872.3660
Regulation Name: Impression Material
Regulatory Class: Class II
Product Code: ELW
Dated: August 18, 2026
Received: August 18, 2026

Dear Jin An:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K254152 - Jin An

Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K254152 - Jin An

Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

**MICHAEL E. ADJODHA -S**

Michael E. Adjodha, MChE, RAC, CQIA

Assistant Director

DHT1B: Division of Dental and
ENT Devices

OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices

Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K254152 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  Dia-X Sil Light Body (A Type)  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  - Crown and bridge impressions - Inlay and onlay impressions - Functional impressions - Implant impressions - Denture and partial denture impressions  |   |   |
|  Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |

{4}

DiaDent Group International

Product Name: Dia-X Sil Light Body

# 510(k) Summary

This summary of 510(k) substantial equivalence information is submitted in accordance with the requirements of 21 CFR 807.92.

1. Application Information

|  510(k) Number | K254152  |
| --- | --- |
|  Date Prepared | Aug 18, 2026  |
|  Company Name and Address | DiaDent Group International16, Osongsaengmyeong 4-ro, Osong-eup, Heungdeok-gu, Cheongju-si, Chungcheongbuk-do, 28161, Republic of Korea  |
|  Contact Person | Byung Chul KimQuality Assurance ManagerPhone: +82-43-266-2315FAX: +82-43-235-2315Email: diadent68@diadent.co.kr  |

2. Device Information

|  Common Name | Dental Impression Material  |
| --- | --- |
|  Classification Name | Material, Impression  |
|  Regulation Description | Impression material  |
|  Review Panel | Dental  |
|  Regulation Number | 21 CFR 872.3660  |
|  Product Code | ELW  |
|  Device Class | II  |
|  Device Trade Name | Dia-X Sil Light Body  |

3. Predicate Device

|   | Primary Predicate Device | Predicate Device  |
| --- | --- | --- |
|  510(k) Number | K214086 | K223892  |
|  Applicant | Genoss Co.,Ltd. | Innovative Product Brands, Inc.  |
|  Device Name | Bright Impress - Light | DentMix VPS Impression Material  |
|  Regulation Number | 21 CFR 872.3660 | 21 CFR 872.3660  |
|  Product Code | ELW | ELW  |
|  Device Class | II | II  |

## 4. Device Description

Dia-X Sil Light Body is a vinyl polysiloxane (VPS) impression material designed for impression taking. It complies with ISO 4823 standards and is classified as Type 3.

|  No. | Model Name | Composition  |
| --- | --- | --- |
|  1 | Dia-X Sil Light Body A Type | Dia-X Sil Light Body 50ml Cartridge 1ea  |
|  2 | Dia-X Sil Light Body | Dia-X Sil Light Body 50ml Cartridge 2ea  |

Page 1

{5}

DiaDent Group International

Product Name: Dia-X Sil Light Body

|   | C Type |   |
| --- | --- | --- |
|  3 | Dia-X Sil Light Body E Type | Dia-X Sil Light Body 50ml Cartridge 4ea  |
|  4 | Dia-X Sil Light Body G type | Dia-X Sil Light Body 50ml Cartridge 10ea  |
|  5 | Dia-X Sil Light Body Sample | Dia-X Sil Light Body 25ml Cartridge 1ea  |

### 5. Indications for Use

- Crown and bridge impressions
- Inlay and onlay impressions
- Functional impressions
- Implant impressions
- Denture and partial denture impressions

### 6. Technological Characteristics

The subject device, Dia-X Sil Light Body has similar characteristics to the primary predicate device, Bright Impress-Light, DentMix VPS Impression Material.

[Comparison table]

|   | Subject Device | Primary Predicate Device | Predicate Device | Discuss  |
| --- | --- | --- | --- | --- |
|  510(k) Number | K254152 | K214086 | K223892 | -  |
|  Product Code | ELW | ELW | ELW | Equivalent  |
|  Device Class | II | II | II | Equivalent  |
|  Manufacturer | DiaDent Group International | Genoss Co.,Ltd. | Innovative Product Brands, Inc. | -  |
|  Device Name | Dia-X Sil Light Body | Bright Impress - Light | DentMix VPS Impression Material | -  |
|  Indications for Use | -Crown and bridge impressions-Inlay and onlay impressions-Functional impressions-Implant impressions-Denture and partial denture impressions | Impression of inlay, onlay crown, and bridge preparation-Crown, bridge impression-Inlay and onlay impression-Functional impression-Denture impression-Study model impression | DentMix VPS Impression Material is intended for use with all crown and bridge, occlusal and implant impression techniques to reproduce the structure of a patient's teeth and gums. | Equivalent  |
|  Raw Materials | - Siloxanes and Silicones, di-Me Vinyl group-Terminated-Silica-Silicone Polyethercopolymer-Siloxanes and Silicones, Di-ME, Me Hydrogen | - Polyvinyl Siloxane | - Vinyl Polysiloxane-Filler-Pigments | * See the below table.  |

Page 2

{6}

DiaDent Group International

Product Name: Dia-X Sil Light Body

|   | - Platinum, 1,3-diethenyl 1,1,3,3 tetramethyldisiloxane complexes |  |  |   |
| --- | --- | --- | --- | --- |
|  Description | Dia-X Sil Light Body is a vinyl polysiloxane (VPS) impression material designed for impression taking. It complies with ISO 4823 standards and is classified as Type 3. | Bright Impress impression material is a fast-set of addition-reaction silicone elastomer (Vinyl polysiloxane)-based material for dental professionals, consisting of five types (Light, Medium, Heavy, Bite and Putty) with superior hydrophilicity, dimensional accuracy, high tensile strength, and resistant to deformation. It is designed for versatile impression techniques of crowns, bridges, orthodontics and implants. | DentMix VPS Impression Material is an addition-reaction base/catalyst polyvinylsiloxane dental impression material intended as an alternative to traditional alginate materials. It is available in regular set and fast set. Both are available in light body, regular body/monophase and heavy body. | Similarity  |
|  Performance Standard Conformance | ISO 4823 | ISO 4823 | ISO 4823 | Equivalent  |
|  Physical and Mechanical properties | - Consistency: ≥ 36mm - Detail reproduction: 20μm line - Linear dimensional change: < 1.5% - Compatibility with gypsum : 50μm line - Elastic Recovery: > 96.5% - Strain in compression: 2 to 20% | - Consistency: ≥ 36mm - Detail Reproduction: 20 μm reproduction without interruption - Compatibility with gypsum: 50 μm reproduction without interruption - Elastic Recovery: ≥ 96.5% - Strain in compression: 2.0~20% | - Consistency: 40mm - Linear Dimensional Change: 0.08% - Compatibility with Gypsum: Pass - Elastic Recovery: 98.9% - Strain in compression: 5.91% | Equivalent  |
|  Handling System | Two part base/catalyst system | Two part base/catalyst system | Two part base/catalyst system | Equivalent  |
|  Biocompatibility | Biocompatible | Biocompatible | Biocompatible | Equivalent  |
|  Sterility | Non-sterile | Non-sterile | Non-sterile | Equivalent  |

* Difference table

|  Subject Device | Predicate Device | Predicate Device | Discussion  |
| --- | --- | --- | --- |

Page 3

{7}

DiaDent Group International

Product Name: Dia-X Sil Light Body

|  - Siloxanes and Silicones, di-Me Vinyl group-Terminated - Silica - Silicone Polyethercopolymer - Siloxanes and Silicones, Di-ME, Me Hydrogen - Platinum, 1,3-diethenyl 1,1,3,3 tetramethyldisiloxane complexes | - Polyvinyl Siloxane | - Vinyl Polysiloxane - Filler - Pigments | Both the Subject Device and the Predicate Devices contain Polyvinyl Siloxane as the primary chemical constituent. These differences have been evaluated through biocompatibility testing and do not raise any new issues of safety or effectiveness.  |
| --- | --- | --- | --- |

### 7. Non-Clinical Performance Data

The performance and biological tests were conducted on the subject device; Dia-X Sil Light Body according to the following standards.

ISO 4823:2021, Dentistry — Elastomeric impression and bite registration materials

ISO 7405:2018, Dentistry — Evaluation of biocompatibility of medical devices used in dentistry

ISO 10993-1:2018, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process

ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity

ISO 10993-10:2021, Biological evaluation of medical devices - Part 10: Tests for skin sensitization

ISO 10993-23:2021, Biological evaluation of medical devices - Part 23: Tests for irritation

### Performance Testing (Bench Testing)

To demonstrate the physical properties of the subject device, the following specific bench tests were performed in accordance with ISO 4823:2021:

|  Item | Requirement | Pass/Fail  |
| --- | --- | --- |
|  Working-time | At least 1 minute | Pass  |
|  Consistency | 36mm or greater | Pass  |
|  Detail reproduction | The 20μm line without interruption | Pass  |
|  Linear dimensional change | 1.5% or less | Pass  |
|  Compatibility with gypsum | The 50μm line without interruption | Pass  |
|  Elastic recovery | At least 96.5% | Pass  |
|  Strain in compression | Within the range of 2% to 20% | Pass  |

Page 4

{8}

DiaDent Group International

Product Name: Dia-X Sil Light Body

### Biocompatibility Testing

To demonstrate the biological safety of the subject device, the following specific biocompatibility tests were performed:

|  Item | Standard  |
| --- | --- |
|  In vitro Cytotoxicity | ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity  |
|  Skin Sensitization | ISO 10993-10:2021 Biological evaluation of medical devices - Part 10: Tests for skin sensitization  |
|  Oral Mucosa Irritation | ISO 10993-23:2021 Biological evaluation of medical devices - Part 23: Tests for irritation  |

The test results demonstrated that the subject device, Dia-X Sil Light Body, met all the acceptance criteria of the applicable standards and performed as intended.

### 8. Clinical Performance Data

No clinical data was collected or provided to support substantial equivalence between the subject and predicate device.

### 9. Conclusions

Based on the above information and all data provided in this submission, the subject device is substantially equivalent to the legally marketed device identified in this submission.

Page 5

---

**Source:** [https://fda-staging.innolitics.com/device/K254152](https://fda-staging.innolitics.com/device/K254152)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).