K254143 · EM-TECH Solutions, Inc. · ITX · Aug 10, 2026 · Radiology
Device Facts
Record ID
K254143
Device Name
EM-TECH Passive Ultrasound Probe Strap
Applicant
EM-TECH Solutions, Inc.
Product Code
ITX · Radiology
Decision Date
Aug 10, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1570
Device Class
Class 2
Indications for Use
The EM-TECH Passive Ultrasound Probe Strap is intended to passively secure an ultrasound transducer against a patient's body during diagnostic ultrasound examinations performed by trained medical professionals. The device is intended for use in adult patients only.
Device Story
Non-powered, reusable mechanical accessory; secures ultrasound transducer against patient body during diagnostic exams. Comprises adjustable elastic strap, probe holder pocket, hook-and-loop fasteners. Used in clinical environments by trained medical professionals (e.g., echocardiographers). Operator maintains continuous manual control of probe placement, pressure, angulation, and coupling. Device does not generate, process, or modify ultrasound signals; purely mechanical stabilization. Benefits include consistent contact pressure and probe retention during imaging. Healthcare providers use output (ultrasound images) from separate imaging system to inform clinical decisions.
Clinical Evidence
Performance and usability feasibility study conducted in 27 adult transthoracic echocardiography cases using a phased-array probe. Three echocardiographers performed imaging; three cardiologists independently reviewed 225 de-identified image/video clips using a 1-5 quality scale. 216 of 225 ratings (96.0%) were scored 3 or higher (adequate). No device slippage, malfunction, or use error reported. No animal testing performed.
Technological Characteristics
Non-powered, reusable mechanical accessory. Materials: elastic fabric, hook-and-loop fasteners. Form factor: adjustable body strap with probe holder pocket. No electronics, software, power sources, or energy-delivery mechanisms. Standalone device; does not interface with ultrasound system electronics. Sterilization: cleaned/disinfected between uses per instructions.
Indications for Use
Indicated for adult patients to passively secure an ultrasound transducer against the body during diagnostic ultrasound examinations performed by trained medical professionals.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
Predicate Devices
Transcranial and Vascular Doppler Diagnostic Ultrasound Transducers, Model EMS-9UA (K092164)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
EM-TECH Solutions, Inc.
Enayatullah Motahedy
Owner/President
25272 McIntyre Square
Chantilly, Virginia 20152
August 10, 2026
Re: K254143
Trade/Device Name: EM-TECH Passive Ultrasound Probe Strap
Regulation Number: 21 CFR 892.1570
Regulation Name: Diagnostic Ultrasonic Transducer
Regulatory Class: Class II
Product Code: ITX
Dated: July 7, 2026
Received: July 8, 2026
Dear Enayatullah Motahedy:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254143 - Enayatullah Motahedy
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K254143 - Enayatullah Motahedy
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Michael D.
O'hara -S
Date: 2026.08.10 14:36:11 -04'00' For
Yanna Kang, Ph.D.
Assistant Director
Mammography and Ultrasound Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K254143 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| EM-TECH PASSIVE ULTRASOUND PROBE STRAP | | |
| Please provide your Indications for Use below. | | ? |
| The EM-TECH Passive Ultrasound Probe Strap is intended to passively secure an ultrasound transducer against a patient's body during diagnostic ultrasound examinations performed by trained medical professionals. The device is intended for use in adult patients only. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K254143
510(k) Summary
Prepared on: 2026-07-11
# Contact Details
21 CFR 807.92(a)(1)
Applicant Name
EM-TECH SOLUTIONS INC
Applicant Address
25272 MCINTYRE SQUARE CHANTILLY VA 20152 United States
Applicant Contact Telephone
703-342-7125
Applicant Contact
Mr. ENAYATULLAH MOTAHEDY
Applicant Contact Email
EMOTAHEDY@GMAIL.COM
# Device Name
21 CFR 807.92(a)(2)
Device Trade Name
EM-TECH PASSIVE ULTRASOUND PROBE STRAP
Common Name
Diagnostic ultrasonic transducer
Classification Name
Transducer, Ultrasonic, Diagnostic
Regulation Number
892.1570
Product Code(s)
ITX
# Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
Predicate #
Predicate Trade Name (Primary Predicate is listed first)
Product Code
K092164
Transcranial and Vascular Doppler Diagnostic Ultrasound Transducers, Model EMS-9UA
ITX
K994353
Securline Umbilical Disposable Fetal Monitoring Abdominal Straps (reference device only)
HFM
# Device Description Summary
21 CFR 807.92(a)(4)
Device Description Summary
The EM-TECH Passive Ultrasound Probe Strap is a non-powered, reusable medical device accessory intended to passively secure an ultrasound transducer against a patient's body during diagnostic ultrasound examinations performed by trained medical professionals. The device is intended for use in adult patients only.
The device consists of an adjustable elastic body strap, a probe holder pocket, and a hook-and-loop fastening system. The elastic strap is wrapped around the patient's body at the desired imaging location and secured to maintain consistent contact pressure between the ultrasound transducer and the patient's skin. The probe holder pocket positions and retains the transducer without interfering with ultrasound transmission.
The trained operator maintains continuous manual control of probe placement, pressure, angulation, coupling, and image acquisition throughout the procedure.
The device operates through purely mechanical means and contains no electronic components, software, firmware, power sources, or energy-delivery mechanisms. All ultrasound signal generation, processing, and display functions are performed by the ultrasound imaging system, which is a separate device not included in this submission.
The EM-TECH Passive Ultrasound Probe Strap contacts intact skin for limited durations and is intended for use in professional clinical environments. The device is reusable and is cleaned and disinfected between uses in accordance with the provided Instructions for Use.
# Intended Use/Indications for Use
21 CFR 807.92(a)(5)
{5}
The EM-TECH Passive Ultrasound Probe Strap is intended to passively secure an ultrasound transducer against a patient's body during diagnostic ultrasound examinations performed by trained medical professionals. The device is intended for use in adult patients only.
## Indications for Use Comparison
21 CFR 807.92(a)(5)
FDA-cleared K092164, Transcranial and Vascular Doppler Diagnostic Ultrasound Transducers, Model EMS-9UA, is the proposed primary predicate under 21 CFR 892.1570. FDA's record lists primary Product Code OQQ with subsequent codes ITX and IYN. K092164 includes the EMS9UA Transcranial Doppler with Robotic Probe Headband; the cleared headband tracks the Doppler signal and mechanically positions its probe during monitoring. The subject device instead provides passive, non-powered transthoracic probe support and does not change the diagnostic ultrasound system's purpose, acoustic output, or imaging performance. K994353, Securline Umbilical Disposable Fetal Monitoring Abdominal Straps, is reference only for stretch-fabric construction and prolonged intact-skin contact; it is not the predicate.
## Technological Comparison
21 CFR 807.92(a)(6)
The subject device is a passive, non-powered mechanical accessory with no electronics, software, energy delivery, or acoustic function. It does not generate, transmit, receive, process, or modify ultrasound and does not alter frequency, power, MI/TI, or beam characteristics. K092164 is powered and uses robotic probe positioning for transcranial Doppler monitoring; the subject device is passive and used for transthoracic echocardiography. The submitted performance evidence addresses the subject device's probe retention, coupling, image quality, user handling, and patient tolerance. K994353 is not relied upon as the predicate.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
Performance and usability feasibility testing was conducted in 27 adult transthoracic echocardiography cases using the GE S70N phased-array transthoracic cardiac probe. Three RDCS echocardiographers with 2 to 20 years of cardiac-echo experience used the device under continuous manual probe control. The evaluation included standard manual/no-strap and strap-assisted image acquisition. Three board-certified cardiologists independently reviewed 225 de-identified image and video clips using a predefined 1-5 image-quality scale, with scores of 3 or higher considered adequate for the intended imaging task. Of the 225 ratings, 216 (96.0%) were scored 3 or higher. No device slippage, fall-off, device malfunction, use error, or manual fallback was documented in the reviewed completed source fields. Nine ratings were below 3 and were addressed through case-specific review, with most occurring in technically difficult studies. No animal testing was conducted or relied upon for this submission. Based on these results, the subject device performs its intended passive stabilization function, and the performance testing supports a conclusion that the device is substantially equivalent to the identified predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.