K254142 · DiaDent Group International · ELW · Sep 16, 2026 · Dental
Device Facts
Record ID
K254142
Device Name
Dia-X Sil Heavy Body
Applicant
DiaDent Group International
Product Code
ELW · Dental
Decision Date
Sep 16, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3660
Device Class
Class 2
Indications for Use
Crown and bridge impressions - Inlay and onlay impressions - Functional impressions - Implant impressions - Denture and partial denture impressions
Device Story
Dia-X Sil Heavy Body is a vinyl polysiloxane (VPS) dental impression material; supplied in two-part base/catalyst cartridges for use with dispensing systems. Used by dental professionals in clinical settings to capture accurate anatomical impressions of teeth and gums for restorative procedures (crowns, bridges, inlays, onlays, implants, dentures). Material is mixed and applied to impression trays; sets to form a stable, elastic mold of oral structures. Output is a physical negative impression used by dental laboratories or clinicians to fabricate dental prosthetics. Benefits include high dimensional accuracy, detail reproduction, and elastic recovery, facilitating precise prosthetic fit.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing per ISO 4823:2021 (working time, consistency, detail reproduction, dimensional change, gypsum compatibility, elastic recovery, strain in compression) and biocompatibility testing per ISO 10993-5, 10993-10, and 10993-23.
Technological Characteristics
Vinyl polysiloxane (VPS) impression material; Type 1 per ISO 4823. Composition: Siloxanes/silicones (vinyl group-terminated), silica, silicone polyethercopolymer, platinum complex catalyst. Two-part base/catalyst handling system. Non-sterile. Physical properties: <35mm consistency, 50μm detail reproduction, <1.5% linear dimensional change, >96.5% elastic recovery, 0.8-20% strain in compression.
Indications for Use
Indicated for dental impression procedures including crown, bridge, inlay, onlay, functional, implant, and denture/partial denture impressions in patients requiring dental restorative or prosthetic treatment.
Regulatory Classification
Identification
Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
September 16, 2026
DiaDent Group International
Jin An
RA team
16, Osongsaengmyeong 4-Ro, Osong-Eup, Heungdeok-Gu
Cheongju-Si, 28186
REPUBLIC OF KOREA
Re: K254142
Trade/Device Name: Dia-X Sil Heavy Body
Regulation Number: 21 CFR 872.3660
Regulation Name: Impression Material
Regulatory Class: Class II
Product Code: ELW
Dated: August 18, 2026
Received: August 18, 2026
Dear Jin An:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254142 - Jin An
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K254142 - Jin An
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**MICHAEL E. ADJODHA -S**
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K254142 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| Dia-X Sil Heavy Body | | |
| Please provide your Indications for Use below. | | ? |
| - Crown and bridge impressions - Inlay and onlay impressions - Functional impressions - Implant impressions - Denture and partial denture impressions | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
DiaDent Group International
Product Name: Dia-X Sil Heavy Body
# 510(k) Summary
This summary of 510(k) substantial equivalence information is submitted in accordance with the requirements of 21 CFR 807.92.
# 1. Application Information
| 510(k) Number | K254142 |
| --- | --- |
| Date Prepared | Aug 18, 2026 |
| Company Name and Address | DiaDent Group International 16, Osongsaengmyeong 4-ro, Osong-eup, Heungdeok-gu, Cheongju-si, Chungcheongbuk-do, 28161, Republic of Korea |
| Contact Person | Byung Chul Kim Quality Assurance Manager Phone: +82-43-266-2315 FAX: +82-43-235-2315 Email: diadent68@diadent.co.kr |
# 2. Device Information
| Common Name | Dental Impression Material |
| --- | --- |
| Classification Name | Material, Impression |
| Regulation Description | Impression material |
| Review Panel | Dental |
| Regulation Number | 21 CFR 872.3660 |
| Product Code | ELW |
| Device Class | II |
| Device Trade Name | Dia-X Sil Heavy Body |
# 3. Predicate Device
| | Primary Predicate Device | Predicate Device |
| --- | --- | --- |
| 510(k) Number | K214086 | K223892 |
| Applicant | Genoss Co.,Ltd. | Innovative Product Brands, Inc. |
| Device Name | Bright Impress-Heavy | DentMix VPS Impression Material |
| Regulation Number | 21 CFR 872.3660 | 21 CFR 872.3660 |
| Product Code | ELW | ELW |
| Device Class | II | II |
# 4. Device Description
Dia-X Sil Heavy Body is a vinyl polysiloxane (VPS) impression material designed for impression taking. It complies with ISO 4823 standards and is classified as Type 1.
| No. | Model Name | Composition |
| --- | --- | --- |
| 1 | Dia-X Sil Heavy Body A Type | Dia-X Sil Heavy Body 50ml Cartridge 1ea |
| 2 | Dia-X Sil Heavy Body C Type | Dia-X Sil Heavy Body 50ml Cartridge 2ea |
Page 1
{5}
DiaDent Group International
Product Name: Dia-X Sil Heavy Body
| 3 | Dia-X Sil Heavy Body E Type | Dia-X Sil Heavy Body 50ml Cartridge 4ea |
| --- | --- | --- |
| 4 | Dia-X Sil Heavy Body G type | Dia-X Sil Heavy Body 50ml Cartridge 10ea |
| 5 | Dia-X Sil Heavy Body I type | Dia-X Sil Heavy Body 50ml Cartridge 24ea |
| 6 | Dia-X Sil Heavy Body Sample | Dia-X Sil Heavy Body 25ml Cartridge 1ea |
### 5. Indications for Use
- Crown and bridge impressions
- Inlay and onlay impressions
- Functional impressions
- Implant impressions
- Denture and partial denture impressions
### 6. Technological Characteristics
The subject device, Dia-X Sil Heavy Body has similar characteristics to the primary predicate device, Bright Impress-Heavy, DentMix VPS Impression Material.
[Comparison table]
| | Subject Device | Primary Predicate Device | Predicate Device | Discuss |
| --- | --- | --- | --- | --- |
| 510(k) Number | K254142 | K214086 | K223892 | - |
| Product Code | ELW | ELW | ELW | Equivalent |
| Device Class | II | II | II | Equivalent |
| Manufacturer | DiaDent Group International | Genoss Co.,Ltd. | Innovative Product Brands, Inc. | - |
| Device Name | Dia-X Sil Heavy Body | Bright Impress-Heavy | DentMix VPS Impression Material | - |
| Indications for Use | -Crown and bridge impressions-Inlay and onlay impressions-Functional impressions-Implant impressions-Denture and partial denture impressions | Impression of inlay, onlay crown, and bridge preparation-Crown, bridge impression-Inlay and onlay impression-Functional impression-Denture impression-Study model impression | DentMix VPS Impression Material is intended for use with all crown and bridge, occlusal and implant impression techniques to reproduce the structure of a patient's teeth and gums. | Equivalent |
| Raw Materials | - Siloxanes and Silicones, di-Me Vinyl Group-Terminated-Silica-Silicone Polyethercopolymer | - Polyvinyl Siloxane | - Vinyl Polysiloxane-Filler-Pigments | * See the below table. |
Page 2
{6}
DiaDent Group International
Product Name: Dia-X Sil Heavy Body
| | - Siloxanes and Silicones, Di-ME, Me Hydrogen - Platinum, 1,3-diethenyl 1,1,3,3 tetramethyldisiloxane complexes | | | |
| --- | --- | --- | --- | --- |
| Description | Dia-X Sil Heavy Body is a vinyl polysiloxane (VPS) impression material designed for impression taking. It complies with ISO 4823 standards and is classified as Type 1. | Bright Impress impression material is a fast-set of addition-reaction silicone elastomer (Vinyl polysiloxane)-based material for dental professionals, consisting of five types (Light, Medium, Heavy, Bite and Putty) with superior hydrophilicity, dimensional accuracy, high tensile strength, and resistant to deformation. It is designed for versatile impression techniques of crowns, bridges, orthodontics and implants. | DentMix VPS Impression Material is an addition-reaction base/catalyst polyvinylsiloxane dental impression material intended as an alternative to traditional alginate materials. It is available in regular set and fast set. Both are available in light body, regular body/monophase and heavy body. | Similarity |
| Performance Standard Conformance | ISO 4823 | ISO 4823 | ISO 4823 | Equivalent |
| Physical and Mechanical properties | - Consistency: < 35mm - Detail reproduction: 50μm line - Linear dimensional change: < 1.5% - Compatibility with gypsum: 50μm line - Elastic Recovery: > 96.5% - Strain in compression: 0.8% to 20% | - Consistency: ≤ 35mm - Detail Reproduction: 50 μm reproduction without interruption - Compatibility with gypsum: 50 μm reproduction without interruption - Elastic Recovery: ≥ 96.5% - Strain in compression: 0.8~20% | - Consistency: 32mm - Linear Dimensional Change: 0.14% - Compatibility with Gypsum: Pass - Elastic Recovery: 99.1% - Strain in compression: 2.54% | Equivalent |
| Handling System | Two part base/catalyst system | Two part base/catalyst system | Two part base/catalyst system | Equivalent |
| Biocompatibility | Biocompatible | Biocompatible | Biocompatible | Equivalent |
| Sterility | Non-sterile | Non-sterile | Non-sterile | Equivalent |
Page 3
{7}
DiaDent Group International
Product Name: Dia-X Sil Heavy Body
* Difference table
| Subject Device | Predicate Device | Predicate Device | Discussion |
| --- | --- | --- | --- |
| - Siloxanes and Silicones, di-Me Vinyl group-Terminated- Silica- Silicone Polyethercopolymer- Siloxanes and Silicones, Di-ME, Me Hydrogen- Platinum, 1,3-diethenyl 1,1,3,3 tetramethyldisiloxane complexes | - Polyvinyl Siloxane | - Vinyl Polysiloxane- Filler- Pigments | Both the Subject Device and the Predicate Devices contain Polyvinyl Siloxane as the primary chemical constituent. These differences have been evaluated through biocompatibility testing and do not raise any new issues of safety or effectiveness. |
### 7. Non-Clinical Performance Data
The performance and biological tests were conducted on the subject device; Dia-X Sil Heavy Body according to the following standards.
ISO 4823:2021, Dentistry — Elastomeric impression and bite registration materials
ISO 7405:2018, Dentistry — Evaluation of biocompatibility of medical devices used in dentistry
ISO 10993-1:2018, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2021, Biological evaluation of medical devices - Part 10: Tests for skin sensitization
ISO 10993-23:2021, Biological evaluation of medical devices - Part 23: Tests for irritation
#### Performance Testing (Bench Testing)
To demonstrate the physical properties of the subject device, the following specific bench tests were performed in accordance with ISO 4823:2021:
| Item | Requirement | Pass/Fail |
| --- | --- | --- |
| Working-time | At least 1 minute | Pass |
| Consistency | Within 35mm or less | Pass |
| Detail reproduction | The 50μm line without interruption | Pass |
| Linear dimensional change | 1.5% or less | Pass |
| Compatibility with gypsum | The 50μm line without interruption | Pass |
| Elastic recovery | At least 96.5% | Pass |
| Strain in compression | Within the range of 0.8% to 20.0% | Pass |
#### Biocompatibility Testing
To demonstrate the biological safety of the subject device, the following specific biocompatibility tests were performed:
Page 4
{8}
DiaDent Group International
Product Name: Dia-X Sil Heavy Body
| Item | Standard |
| --- | --- |
| In vitro Cytotoxicity | ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity |
| Skin Sensitization | ISO 10993-10:2021 Biological evaluation of medical devices - Part 10: Tests for skin sensitization |
| Oral Mucosa Irritation | ISO 10993-23:2021 Biological evaluation of medical devices - Part 23: Tests for irritation |
The test results demonstrated that the subject device, Dia-X Sil Heavy Body, met all the acceptance criteria of the applicable standards and performed as intended.
### 8. Clinical Performance Data
No clinical data was collected or provided to support substantial equivalence between the subject and predicate device.
### 9. Conclusions
Based on the above information and all data provided in this submission, the subject device is substantially equivalent to the legally marketed device identified in this submission.
Page 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.