K254129 · Neko Health AB · MUD · May 22, 2026 · Cardiovascular
Device Facts
Record ID
K254129
Device Name
Spectrum-2
Applicant
Neko Health AB
Product Code
MUD · Cardiovascular
Decision Date
May 22, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K254129 · May 22, 2026
Spectrum-2
Neko Health AB
Retrospective clinical study data from prior clinical investigations (Spectrum-1 and Dermaflow Alpha)
Retrospective data was used to validate the spatial frequency domain imaging (SFDI) technique and assess the impact of demographic factors (skin tone, age, sex) on measurement performance.
Retrospective validation study (Spectrum-1 and Dermaflow Alpha data); Retrospective analysis
Diverse populations (varied skin tone, age, and sex)
Not applicable for this study
Validation of SFDI technique and assessment of demographic factors
Indications for Use
The Spectrum-2 device is indicated for use to record and monitor changes in oxygenation levels in superficial tissues during circulatory or perfusion examinations for individuals with potential circulatory compromise.
Device Story
Spectrum-2 is a non-invasive, contactless optical imaging system using Spatial Frequency Domain Imaging (SFDI). It projects spatial frequency patterns (500-1000nm) onto skin via a light engine and digital micromirror device. Eight cameras with optical bandpass filters capture reflected light. The device maps measured absorption spectra to known chromophore spectra to calculate oxygen saturation (StO2), oxyhemoglobin (HbO2), and deoxyhemoglobin (HbR). Used exclusively in Neko Health clinical facilities by qualified personnel. Output includes color-coded oxygenation and blood volume maps. Data aids clinicians in assessing superficial microvasculature perfusion. A 655nm Class 2 laser is used solely for positioning distance measurement. Benefits include non-invasive, real-time monitoring of tissue oxygenation changes to support clinical decision-making regarding circulatory status.
Clinical Evidence
Two retrospective clinical studies were conducted. Study 1 compared Spectrum-2 to a research device during vascular occlusion, demonstrating sensitivity to circulatory compromise with statistically significant changes in StO2 (p-values not specified) and strong regression agreement. Study 2 analyzed data from 2 prior investigations (Spectrum-1 and Dermaflow Alpha) to validate the SFDI technique across diverse demographics (skin tone, age, sex), confirming validity and clinical relevance. No animal studies were performed.
Indicated for individuals with potential circulatory compromise requiring recording and monitoring of superficial tissue oxygenation levels during circulatory or perfusion examinations.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 22, 2026
Neko Health AB
% Brittany Valdez Nava
Senior Manager
Healthcare Innovation Catalysts, Inc.
8024 Summer Mill Court
Bethesda, Maryland 20817
Re: K254129
Trade/Device Name: Spectrum-2
Regulation Number: 21 CFR 870.2700
Regulation Name: Oximeter
Regulatory Class: Class II
Product Code: MUD
Dated: December 18, 2025
Received: December 22, 2025
Dear Brittany Valdez Nava:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254129 - Brittany Valdez Nava
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K254129 - Brittany Valdez Nava
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA
L. HITHE-S
Digitally signed by
TANISHA L. HITHE-S
Date: 2026.05.22
12:34:14 -04'00"
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K254129 | |
| Device Name Spectrum-2 | |
| Indications for Use (Describe) The Spectrum-2 device is indicated for use to record and monitor changes in oxygenation levels in superficial tissues during circulatory or perfusion examinations for individuals with potential circulatory compromise. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
NEKO
# 1.0 510(K) SUMMARY
The content of this 510(k) summary is provided in conformance with 21 CFR § 807.92.
# 1.1 Submitter's Information
Company Name: Neko Health AB
Address: Svärdvägen 19
182 33 Danderyd
Sweden
Company Contact: Per Sundström
Head of Regulatory
Neko Health AB
Svärdvägen 19
182 33 Danderyd
Sweden
Phone: +46 76 344 42 34
Email: per.sundstrom@nekohealth.com
Official Correspondent: Brittany Valdez Nava, MRSc
Senior Manager
Healthcare Innovation Catalysts, Inc.
8024 Summer Mill Court
Bethesda, MD 20817
Date prepared: May 5, 2025
# 1.2 Subject Device Name
Trade Name: Spectrum-2
Common Name: Spectrum-2
Regulation Name: Oximeter
Regulation Number: 21 CFR 870.2700
Device Class: Class 2
Product Code: MUD
Product Code Name: Oximeter, Tissue Saturation
510(k) Review Panel: General & Plastic Surgery
Spectrum-2 510(k) Summary - K254129
{5}
NEKO
## 1.3 Legally Marketed Predicate Device
### 1.3.1 Clarifi Imaging System
The primary predicate device for the Spectrum-2 is the Clarifi Imaging System cleared under K181623.
510(k) Number K181623
Trade Name: Clarifi Imaging System
Common Name: Clarifi Imaging System
Regulation Name: Oximeter
Regulation Number: 21 CFR 870.2700
Device Class: Class 2
Product Code: MUD
Product Code Name: Oximeter, Tissue Saturation
510(k) Review Panel: General & Plastic Surgery
## 1.4 Device Description
### 1.4.1 Brief Description of the Device
The Spectrum-2 device is designed for contactless, non-invasive recording and monitoring of changes in concentration of haemoglobin and blood oxygenation in superficial tissues. Utilizing Spatial Frequency Domain Imaging (SFDI) technology, it employs a projector that emits light across a wavelength range of approximately 500 to 1000nm, projecting spatial frequency patterns onto the skin.
The skin's reaction to this light, in terms of absorption, reflection, and scattering, is contingent on its molecular composition. Eight cameras, each equipped with optical bandpass filters, capture the skin's response at specific wavelengths, enabling optical spectroscopy based on chromophore absorption spectra. The device identifies specific chromophores, including oxygenated haemoglobin and deoxygenated haemoglobin by mapping known absorption spectra with the measured ones.
The Spectrum-2 device includes a light engine, a digital micromirror device, optics, and cameras equipped with bandpass filters and is intended to be used as a non-invasive tissue oxygenation measurement system to capture the reflectance from skin illuminated with projected patterns that enables recording and monitoring of changes in oxygenation levels in superficial tissue.
Such information, captured during imaging measurement, includes tracking changes of:
- Oxygen saturation (StO2),
- Oxyhemoglobin (HbO2), and
- Deoxyhemoglobin (HbR)
in superficial tissue.
Spectrum-2 510(k) Summary - K254129
{6}
NEKO
The device delivers electromagnetic radiation that is non-ionizing in the form of visible laser light from the 655nm Class 2 laser distance sensor (used exclusively for measuring the working distance during the positioning phase) and visible and NIR light for the projected SFDI patterns from LEDs.
- The laser is of class 2 and is considered safe for the skin and the eyes without any additional mitigations. One shall not stare into the laser. The laser source of radiation is well below limit values as stated by IEC 60825-1: Safety of laser products – Part 1.
- The LED light sources are well below limit values as stated by IEC 62471: Photobiological safety of lamps and lamp systems
- The laser is not part of the optical measurement system and does not contribute to tissue oxygenation calculations.
The device also fulfills the requirements for general safety of medical devices as described in ANSI/AAMI/IEC-60601-1: Medical electrical equipment – Part 1: General requirements for basic safety and essential performance.
## 1.4.2 Indications for Use
The Spectrum-2 device is indicated for use to record and monitor changes in oxygenation levels in superficial tissues during circulatory or perfusion examinations for individuals with potential circulatory compromise.
## 1.4.3 Environment of Use
The Spectrum-2 device is intended for use exclusively within Neko Health clinical facilities. The device is not intended for use in other healthcare facilities or general medical practices.
Neko Health facilities are clinical environments that fulfill specific requirements regarding cleanliness, controlled indoor environment, and technical infrastructure necessary for safe and effective device operation. Once transported and installed by qualified Neko Health personnel, the device shall not be relocated.
## 1.5 Key Performance Characteristics
The key performance specifications and characteristics of the Spectrum-2 are outlined in Table 1.
Table 1: Key Performance Characteristics
| Feature | Specification/Characteristic |
| --- | --- |
| Electrical Power Supply | Supply Source: Mains-powered Voltage: 100 - 240 VAC 50/60 Hz Power Consumption: 230 VA |
| Operating Conditions | Ambient Temperature: 15 to 30°C (59 to 86°F) Operating Humidity: 20 to 90% RH |
| Storage Conditions | Storage Temperature: -40°C to 70°C (-40 to 158°F) Storage Humidity: 10 to 100%, including condensation Barometric Pressure: 50 kPa to 110 kPa |
| Expected Service Life | 10 years |
| Measurement Technology | Optical, contactless SFDI (Spatial Frequency Domain Imaging) |
| Measurement Method | Structured illumination and spectral model-based analysis of light returned from target tissue |
Spectrum-2 510(k) Summary - K254129
{7}
NEKO
| Feature | Specification/Characteristic |
| --- | --- |
| Measurements Monitored | • Oxygen saturation (StO2) • Oxyhemoglobin (HbO2) • Deoxyhemoglobin (HbR) |
| Measurement Sensor | The camera head is essentially the sensor part of the device, which incorporates the following main components: • 8 x machine vision cameras, one for each wavelength. • Lenses and bandpass filters for the cameras. • LED based light engine with DMD projector. • Control boards. • On-board cooling solution. |
| Precision | <5% error |
| Measurement Stability | <1% drift |
| Image Homogeneity | <4% roll-of |
| Signal-to-Noise Ratio (SNR) | >40 |
| Measurement Time | Frame Rate: 0.5 Hz Acquisition Time for single frame: 2 seconds Recording duration interval: 2 seconds to 35 minutes |
| Location of Measurement | Two-dimensional area of superficial microvasculature |
| Laser Distance Sensor | Purpose: Working distance measurement for camera head positioning only Safety Classification: Class 2 Radiation Pattern/Beam Divergence: Collimated beam Wavelength: 655 nm Max. Power/Energy Output: 1mW Accuracy: ±1mm Pulse Durations and Repetition Rate: the laser distance measurement module is only active during the positioning phase. During this phase the laser is running continuously distance measurements, for which each measured distance consists of 64 pairs of pulses during a timeframe of about 400ms. |
| Data Display | • Oxygenation map (image with color coded blood oxygenation) Blood volume map (image with color coded total hemoglobin concentration) |
| Data Transfer Method | Ethernet over optical fiber to host computer |
| Data Storage | Local and cloud-based |
| External Connections | Interface for monitor, ethernet |
| USB | At least USB 3.0 or faster |
Spectrum-2 510(k) Summary - K254129
{8}
NEKO
## 1.6 Comparison of Key Technological Characteristics with the Predicate Device(s)
### 1.6.1 Clarifi Imaging System
The Spectrum-2 device and the Clarifi Imaging System are both non-invasive, non-contact optical imaging systems that utilize multispectral illumination and spectral analysis of reflected or remitted light to assess tissue oxygenation and hemoglobin-related parameters in superficial biological tissue, including oxygen saturation (StO₂), oxyhemoglobin (HbO₂), and deoxyhemoglobin (HbR).
Both systems utilize Spatial Frequency Domain Imaging (SFDI) principles, projecting spatially-modulated light patterns onto tissue and measuring the resulting reflectance to determine optical absorption and scattering properties. Functionally, both devices incorporate a light engine, projection module/DMD, optics, and cameras with wavelength-specific filters, which may enable quantitative assessment of tissue chromophores.
**Main Differences between Spectrum-2 and Clarifi Imaging System [K181623]**
The differences between the Spectrum-2 and the predicate device are limited to component-level implementation, including light source configuration, wavelength selection, and imaging hardware. Both devices rely on the same fundamental SFDI-based optical principles and analytical methods to derive tissue oxygenation parameters.
- Measurement and Optical Acquisition Approach
- The Spectrum-2 employs a projector that emits light of various wavelengths and patterns of various spatial frequencies with a dependency on the skin’s molecular composition and structure within the examined skin volume. Cameras are then used to image the reflected light per illuminated wavelength (approximately 500 to 1000nm), and the data is used to model the skin scattering and absorption properties for collecting oxygenation data.
- The Clarifi Imaging System uses discrete visible and near-infrared wavelengths between 450 and 1000 nm and a complementary metal oxide semiconductor (CMOS) camera for collecting hyperspectral images. Clarifi also uses a projector, a number of LEDs and camera technologies similar to that of Spectrum-2, being the difference between them the number of LEDs, the specific discrete wavelengths used, and the number of cameras.
Importantly, the Spectrum-2’s intended use is within the scope of the predicate device’s intended use, as it is limited to recording and monitoring changes in tissue oxygenation in superficial tissues during circulatory or perfusion assessments. The Spectrum-2 does not expand the clinical use beyond that of the predicate and does not introduce new types of outputs, diagnostic claims, or user populations that would alter the risk profile.
The differences in technological characteristics have been evaluated through performance bench testing. The results of these studies verify that these differences do not raise new or different questions of safety and effectiveness when the Spectrum-2 is used as intended.
Spectrum-2 510(k) Summary - K254129
{9}
NEKO
## 1.7 Performance Data
### 1.7.1 Performance – Bench Testing
Performance testing to support the determination of substantial equivalence of the Spectrum-2 is summarized in Table 2.
Table 2: Spectrum-2 Performance Testing Summary
| Test Conducted | Test Objective |
| --- | --- |
| System Level Performance Testing | Ensure the device performs in accordance with the product requirements and design inputs |
| Material Verification Testing | Demonstrate that material selection, control, and validation are scientifically justified and support overall device performance and safety. |
| Inter-scanner Variability | Evaluates whether multiple Spectrum-2 scanners produce consistent absorption-coefficient and chromophore-concentration measurements by comparing inter-device variability against predefined acceptance criteria |
| Resolution Verification | Evaluates the Spectrum-2 device’s precision, stability, homogeneity, and signal-to-noise ratio using diffuse reflectance scans of tissue-mimicking phantoms to confirm that its performance meets the acceptance criteria established for substantial equivalence to the predicate device |
| Servicing/Installation Testing | Confirm that the device can be correctly installed, configured, and serviced in accordance with the manufacturer’s instructions, and that these activities do not adversely affect device performance or safety |
| Labeling Verification | Confirm that all device labels, packaging labels, and Instructions for Use (IFU) meet applicable regulatory requirements, are legible, durable, and accurately convey the intended information to the user |
### 1.7.2 Human Factors Validation Testing
A Usability Evaluation performed on the Spectrum-2 found no impact on the safe and effective use of the device. Neko Health AB concludes that the system is safe and effective for the intended users, uses, and use environments.
### 1.7.3 Cybersecurity
Cybersecurity documentation has been provided with this application as recommended by the FDA's Guidance for Industry and FDA Staff, "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions". All software components underwent appropriate cybersecurity assessment and testing.
Spectrum-2 510(k) Summary - K254129
{10}
NEKO
## 1.7.4 Electrical Safety and Electromagnetic Compatibility (EMC)
The Spectrum-2 was evaluated for electromagnetic compatibility (EMC) in accordance with IEC 60601-1-2 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests.
Additional electrical safety testing was conducted in accordance with IEC 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
## 1.7.5 Radiation Safety
A comprehensive radiation safety evaluation was conducted for the device’s optical emission components, including all visible LEDs used for illumination and the laser distance sensor integrated into the system. The following testing was performed:
- IEC 62471:2006 - Photobiological safety of lamps and lamp systems
- IEC 60825-1:2020 - Safety of laser products - Part 1: Equipment classification and requirements
- System-level Radiation Hazard Assessment
Across all evaluations, the device was demonstrated to meet the applicable radiation safety requirements and does not pose hazardous levels of optical exposure under normal use.
## 1.7.6 Software Verification and Validation Testing
System level software verification testing were performed to demonstrate the efficacy of the software and to confirm operation of the machine. The following testing was performed:
- Functional and Performance Verification
- Regression Testing
- Code Reviews
Software verification information within this submission is provided in accordance with the following FDA guidance documents:
- Content of Premarket Submissions for Device Software Functions (14 June 2023)
- Guidance for Off-The-Shelf Software Use in Medical Devices (11 August 2023)
- Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions (26 June 2025)
## 1.7.7 Animal Studies
No animal studies were performed on this device.
Spectrum-2 510(k) Summary - K254129
{11}
NEKO
## 1.7.8 Clinical Studies
Two retrospective clinical studies were conducted to evaluate Spectrum-2’s performance and underlying measurement principles.
The first study compared Spectrum-2 to a research device to assess sensitivity to circulatory compromise. Spectrum-2 successfully detected reductions in tissue oxygen saturation (StO₂) during vascular occlusion and subsequent increases after release, meeting the primary effectiveness endpoint. Spectrum-2 demonstrated highly statistically significant changes between baseline and compromised states, confirming sensitivity to circulatory compromise.
Regression analysis showed strong agreement between Spectrum-2 and the research device, with slopes and intercepts consistent with equivalence expectations, satisfying the first secondary effectiveness endpoint.
The second retrospective study analyzed data from 2 clinical investigations, Spectrum-1 and Dermaflow Alpha, to validate the spatial frequency domain imaging (SFDI) technique and assess demographic factors such as skin tone, age, and sex. Results confirmed that the measurement method integral to Spectrum-2 remains valid and clinically relevant across diverse populations. Observed differences were consistent with literature expectations and did not raise new safety or effectiveness concerns.
## 1.8 Conclusions
Based on above discussions and enclosed sections regarding substantial equivalence, Neko Health AB concludes that the Spectrum-2 does not raise any new or different questions of safety or effectiveness and is therefore substantially equivalent to the Clarifi Imaging System.
Spectrum-2 510(k) Summary - K254129
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.