K254114 · Baylis Medical Technologies, Inc. · DSA · Jan 18, 2026 · Cardiovascular
Device Facts
Record ID
K254114
Device Name
Baylis Connector Cable
Applicant
Baylis Medical Technologies, Inc.
Product Code
DSA · Cardiovascular
Decision Date
Jan 18, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2900
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Baylis Connector Cable is intended to connect Baylis Radiofrequency (RF) Puncture Generators to separately cleared compatible RF devices.
Device Story
Baylis Connector Cable is a sterile, single-use 10-foot cable; connects Baylis RF Puncture generators (e.g., PrecisePath, RFP-100A) to compatible RF delivery devices (e.g., RF guidewires). Device functions as a physical conduit for RF power transfer from generator to puncture device. Used in clinical settings by healthcare professionals. Output is the delivery of RF energy to the compatible device; enables clinical procedures involving RF puncture. Benefits include reliable power transmission for RF-based interventions.
Clinical Evidence
Bench testing only. Verification and validation activities included electrical safety, EMC, generator compatibility, mechanical integrity (connector strength, cable durability), and sterilization/packaging validation. Results demonstrate equivalence to the predicate device.
Technological Characteristics
10-foot single insulated conductor cable; multi-pin positive locking generator connector; spring-loaded wire connection connector. Sterile, EO sterilized. 4-year shelf life. Complies with IEC 60601-1 and IEC 60601-2-2. No software or active electronic components; passive electrical cable.
Indications for Use
Indicated for connecting Baylis RF Puncture generators to separately cleared compatible RF devices for use in clinical procedures requiring RF energy delivery.
Regulatory Classification
Identification
A patient transducer and electrode cable (including connector) is an electrical conductor used to transmit signals from, or power or excitation signals to, patient-connected electrodes or transducers.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
January 18, 2026
Baylis Medical Technologies, Inc.
Barb Boyce
Senior Regulatory Affairs Specialist
2645 Matheson Blvd. E, Mississauga, On L4w 5m2
Mississauga, ON L4W 5S4
Canada
Re: K254114
Trade/Device Name: Baylis Connector Cable
Regulation Number: 21 CFR 870.2900
Regulation Name: Patient Transducer And Electrode Cable (Including Connector)
Regulatory Class: Class II
Product Code: DSA
Dated: December 18, 2025
Received: December 19, 2025
Dear Barb Boyce:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254114 - Barb Boyce
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K254114 - Barb Boyce
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MARCO CANNELLA -S
for
- Aneesh Deoras
- Assistant Director
- Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices
- Office of Cardiovascular Devices
- Office of Product Evaluation and Quality
- Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K254114 | ? |
| Please provide the device trade name(s). | | ? |
| Baylis Connector Cable | | |
| Please provide your Indications for Use below. | | ? |
| The Baylis Connector Cable is intended to connect Baylis Radiofrequency (RF) Puncture Generators to separately cleared compatible RF devices. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K254114
Baylis Connector Cable
Baylis Medical Technologies Inc.
# 510(k) Summary
21 CFR 807.92
## I. Submitter Information
Company Name: Baylis Medical Technologies Inc.
Address: 2645 Matheson Blvd. East, Mississauga, Ontario L4W 5S4, Canada
Phone: (905) 948-5800
Contact Person: Barb Boyce
Summary Prepared On: 16-Dec-2025
## II. Device Identification
Device Trade Name: Baylis Connector Cable
Device Common Name: Cable, transducer and electrode, patient, including connector
Classification Name: 21 CFR 870.2900
Device Class: Class II
Product Code: DSA
## III. Identification of Predicate Device
Predicate Device: RFP-100A Connector Cable (Single Use), cleared under 510(K) number K230571
Predicate Device Manufacturer: Baylis Medical Technologies Inc.
Recall Status: Predicate device has not been the subject of a design-related recall.
## IV. Device Description
The Baylis Connector Cable is a sterile, single-use device designed to connect Baylis-approved Radiofrequency (RF) generators (including the PrecisePath RF Puncture Generator and RFP-100A Generator) to separately cleared compatible RF devices (e.g., RF guidewires).
Special 510(k)
Special 510(K) Summary
{5}
K254114
Baylis Connector Cable
Baylis Medical Technologies Inc.
The Baylis Connector Cable is 10 feet long with a multi-pin positive locking Generator Connector at one end and a spring-loaded wire connection connector at the other end which allows connection to the RF delivery device. Between the connector ends is a single insulated conductor cable.
Principle of Operation: Enables RF power transfer from generator to RF puncture device; no changes from predicate device.
# V. Indications for Use
The Baylis Connector Cable is intended to connect Baylis Radiofrequency (RF) Puncture generators to separately cleared compatible RF devices.
# VI. Comparison to Predicate Device
The Baylis Connector Cable and its predicate share the same intended use, principles of operation, and fundamental technology. Differences are minor and do not raise new questions of safety or effectiveness.
| Detail | Equivalence | Comment |
| --- | --- | --- |
| Legal Manufacturer | Identical | Baylis Medical Technologies Inc. |
| Indications for Use | Similar | Change in the name of the cable and updated list of compatible generators |
| Classification | Identical | 21 CFR 870.2900 |
| FDA Product Code | Identical | DSA |
| Sterility | Identical | Sterile, EO |
| Design | Similar | Both the predicate and subject device share the same fundamental design, the subject device differs as follows:
- Minor design change to connector ends |
| Materials | Similar | Both the predicate and subject device share the same fundamental design, the subject device differs in materials as follows:
- Minor change to insulation and connector material |
| Compatible Generators | Minor difference | Update to list of compatible generators to include PrecisePath Radiofrequency Puncture Generator (PRG-1) |
| Packaging | Identical | Tyvek/poly pouch in shelf box |
| Shelf-life | Identical | 4 years |
Special 510(k)
Special 510(K) Summary
{6}
K254114
Baylis Connector Cable
Baylis Medical Technologies Inc.
# VII. Performance Data
Verification and validation activities confirm that the Baylis Connector Cable meets the same requirements as the predicate device. Testing included:
- Electrical Safety, EMC, and Generator Compatibility: IEC 60601-1 and IEC 60601-2-2 applicable requirements
- Mechanical Integrity: Connector strength, cable durability
- Sterilization & Packaging: EO sterilization validation, shelf-life testing (leveraged from predicate device’s 510(K) since these parameters remain unchanged)
Results demonstrate that the subject device performs equivalently to the predicate device and does not introduce new risks.
# VIII. Conclusions
The Baylis Connector Cable is substantially equivalent to the predicate device (RFP-100A Connector Cable (Single Use), K230571). Both devices share the same intended use, principles of operation, and fundamental technology. Verification and validation of design and material changes to the Baylis Connector Cable further support substantial equivalence to the legally marketed predicate device.
Special 510(k)
Special 510(K) Summary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.