K254112 · DiaDent Group International · ELW · Mar 19, 2026 · Dental
Device Facts
Record ID
K254112
Device Name
Dia-X Sil Bite
Applicant
DiaDent Group International
Product Code
ELW · Dental
Decision Date
Mar 19, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3660
Device Class
Class 2
Indications for Use
Indicated for bite registration in prosthodontics, orthodontics, and implantology. Used for recording the occlusal relationship to aid in the fabrication of crowns, bridges, dentures, and orthodontic appliances. Suitable of occlusal analysis and adjustment in clinical dental procedures.
Device Story
Dia-X Sil Bite is a vinyl polysiloxane (VPS)-based impression material; used by dental professionals in clinical settings to capture precise occlusal relationships. The material is supplied in cartridge form; applied to the patient's teeth/gums; hardens upon contact with oral moisture. The resulting physical impression serves as a template for the fabrication of dental prosthetics (crowns, bridges, dentures) and orthodontic appliances. It facilitates clinical occlusal analysis and adjustment. The device is non-sterile, permanent, and prescription-only.
Clinical Evidence
No clinical data was collected or provided. Substantial equivalence is supported by bench testing, including physical/mechanical property evaluation and biocompatibility testing (cytotoxicity, skin sensitization, and oral mucosa irritation) per ISO 10993 and ISO 4823 standards.
Technological Characteristics
Vinyl polysiloxane (VPS) silicone-based impression material; Type B per ISO 4823:2021. Supplied in 50ml cartridges. Non-sterile. Biocompatibility tested per ISO 10993-1, 5, 10, and 23. Mechanical properties include defined setting and working times.
Indications for Use
Indicated for patients requiring bite registration for prosthodontics, orthodontics, and implantology, including fabrication of crowns, bridges, dentures, and orthodontic appliances, and for occlusal analysis and adjustment.
Regulatory Classification
Identification
Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 19, 2026
DiaDent Group International
Jin An
RA team
16, Osongsaengmyeong 4-Ro, Osong-Eup, Heungdeok-Gu
Cheongju-Si, 28186
REPUBLIC OF KOREA
Re: K254112
Trade/Device Name: Dia-X Sil Bite
Regulation Number: 21 CFR 872.3660
Regulation Name: Impression Material
Regulatory Class: Class II
Product Code: ELW
Dated: December 16, 2025
Received: December 19, 2025
Dear Jin An:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254112 - Jin An
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K254112 - Jin An
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K254112
Device Name
Dia-X Sil Bite
Indications for Use (Describe)
- Indicated for bite registration in prosthodontics, orthodontics, and implantology.
- Used for recording the occlusal relationship to aid in the fabrication of crowns, bridges, dentures, and orthodontic appliances.
- Suitable of occlusal analysis and adjustment in clinical dental procedures.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
DiaDent Group International
Product Name: Dia-X Sil Bite
# 510(k) Summary
This summary of 510(k) substantial equivalence information is submitted in accordance with the requirements of 21 CFR 807.92.
## 1. Application Information
| 510(k) Number | K254112 |
| --- | --- |
| Date Prepared | Mar 19, 2026 |
| Company Name and Address | DiaDent Group International
16, Osongsaengmyeong 4-ro, Osong-eup, Heungdeok-gu,
Cheongju-si, Chungcheongbuk-do, 28161, Republic of Korea |
| Contact Person | Byung Chul Kim
Quality Assurance Manager
Phone: +82-43-266-2315
FAX: +82-43-235-2315
Email: diadent68@diadent.co.kr |
## 2. Device Information
| Device Type | Material, Impression |
| --- | --- |
| Regulation Description | Impression material |
| Review Panel | Dental |
| Regulation Number | 21 CFR 872.3660 |
| Product Code | ELW |
| Device Class | II |
| Device Name | Dia-X Sil Bite |
## 3. Primary Predicate Device
| 510(k) Number | K882690 |
| --- | --- |
| Applicant | 3M ESPE |
| Device Name | Imprint Bite |
| Regulation Number | 21 CFR 872.3660 |
| Product Code | ELW |
| Device Class | II |
## 4. Device Description
Dia-X Sil Bite is vinyl polysiloxane (VPS) – based bite registration material designed for occlusal capture. Classified as Type B under ISO 4823 standards.
| No. | Model Name | Composition |
| --- | --- | --- |
| 1 | Dia-X Sil Bite A Type | Dia-X Sil Bite 50ml Cartridge 1ea |
| 2 | Dia-X Sil Bite C Type | Dia-X Sil Bite 50ml Cartridge 2ea |
| 3 | Dia-X Sil Bite E Type | Dia-X Sil Bite 50ml Cartridge 4ea + Mixing Tip 10ea |
| 4 | Dia-X Sil Bite G type | Dia-X Sil Bite 50ml Cartridge 10ea |
Page 1
{5}
DiaDent Group International
Product Name: Dia-X Sil Bite
# 5. Indications for Use
- Indicated for bite registration in prosthodontics, orthodontics, and implantology.
- Used for recording the occlusal relationship to aid in the fabrication of crowns, bridges, dentures, and orthodontic appliances.
- Suitable of occlusal analysis and adjustment in clinical dental procedures.
# 6. Technological Characteristics
The subject device, Dia-X Sil Bite has similar characteristics to the primary predicate device, Imprint Bite.
[Comparison table]
| | Subject Device | Predicate Device | Discuss |
| --- | --- | --- | --- |
| 510(k) Number | - | K882690 | - |
| Product Code | ELW | ELW | Equivalent |
| Device Class | II | II | Equivalent |
| Manufacturer | DiaDent Group International | 3M ESPE | - |
| Device Name | Dia-X Sil Bite | Imprint Bite | - |
| Intended Use | to obtain the structure of a patient's teeth and gums | to obtain the structure of a patient's teeth and gums | Equivalent |
| Raw Materials | - Silicone Base | -Silicone Base | * See the below table. |
| Principle of operation | It is an elastic occlusal recording material in a convenient cartridge form designed to harden when in contact with oral moisture. It is used to capture precise occlusal relationships | It is an elastic occlusal recording material in a convenient cartridge form designed to harden when in contact with oral moisture. It is used to capture precise occlusal relationships. | Equivalent |
| Performance Standard Conformance | Conformed ISO 4823 | Conformed ISO 4823 | Equivalent |
| Physical and Mechanical properties | - Setting time - Working time - Hardness | - Setting time - Working time - Hardness | Equivalent |
| Biocompatibility | Biocompatible | Biocompatible | Equivalent |
| Use | Prescription / Hospital | Prescription / Hospital | Equivalent |
| Period of use | Permanent | Permanent | Equivalent |
| Sterility | Non-sterile | Non-sterile | Equivalent |
| Shelf-life | 2 years | 2 years | Equivalent |
{6}
DiaDent Group International
Product Name: Dia-X Sil Bite
* Difference table
| Subject Device | Predicate Device | Discussion |
| --- | --- | --- |
| - Silicone Base | -Silicone Base | Both this device and the existing device contain the main ingredient, silicone base. Through the biocompatibility test results, the difference does not raise any issue of safety and effectiveness. |
7. Non-Clinical Performance Data
The performance and biological tests were conducted on the subject device; Dia-X Sil Bite according to the following standards.
ISO 4823:2021, Dentistry — Elastomeric impression and bite registration materials
ISO 7405:2018, Dentistry — Evaluation of biocompatibility of medical devices used in dentistry
ISO 10993-1:2018, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2021, Biological evaluation of medical devices - Part 10: Tests for skin sensitization
ISO 10993-23:2021, Biological evaluation of medical devices - Part 23: Tests for irritation
To demonstrate equivalence in the biological safety of the subject device, the following specific biocompatibility tests were performed:
In vitro Cytotoxicity Test (per ISO 10993-5:2009)
Skin Sensitization Test (per ISO 10993-10:2021)
Oral Mucosa Irritation Test (per ISO 10993-23:2021)
The test results corresponded to the requirements of standards. Therefore, the subject device is substantially equivalent to the predicate device.
8. Clinical Performance Data
No clinical data was collected or provided to support substantial equivalence between the subject and predicate device.
9. Conclusions
Based on the above information and all data provided in this submission, the subject device is substantially equivalent to the legally marketed device identified in this submission.
Page 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.