Standalone analysis of 127 de-identified digital images from 4 US dental offices
—
Tooth labeling
Machine Learning
—
Accuracy: 98.6%
—
—
Standalone analysis of 127 de-identified digital images from 4 US dental offices
—
Indications for Use
VELMENI for DENTISTS (V4D) 3D is radiological, automated, concurrent read, computer-assisted detection software intended to assist in the clinical detection of anatomical structures (maxilla plus cranium, mandible, mandibular canal, temporomandibular joint, maxillary sinus ,teeth ,airway) in CBCT radiographs of permanent teeth in patients 15 years of age or older. This device provides additional information for licensed dental professionals in examining radiographs of patients' teeth. The device is not intended as a substitute for a complete examination by the dental professional or their clinical judgment that considers other relevant information from the patient's history or actual in-vivo clinical assessment. Final diagnoses and patient treatment plan are the responsibility of the dental professional.
Device Story
V4D 3D is a cloud-hosted, web-based software platform for dental professionals; it processes CBCT radiographs to perform automated detection and segmentation of anatomical structures. Input consists of DICOM scans (in .zip format) uploaded via a web interface; the backend API routes these to an AI-Worker engine for ML-based analysis. The system outputs visual detections of anatomical structures and a panoramic viewer with automated dental arch identification; it also supports STL file export for educational/research purposes. The software acts as a concurrent read tool, providing additional information to assist the clinician during radiographic examination. It does not replace clinical judgment or complete examination; final diagnosis and treatment planning remain the responsibility of the dental professional.
Clinical Evidence
Bench testing only. Performance evaluated using 127 de-identified digital CBCT images from 4 US dental offices. Metrics included DICE coefficient, sensitivity, specificity, and tooth labeling accuracy. Results: Airway (DICE 0.968, Sens 99.2%), Mandibular canal (DICE 0.914, Sens 96.8%), Mandible (DICE 0.952, Sens 100%), Maxilla/cranium (DICE 0.951, Sens 100%), Maxillary sinus (DICE 0.96, Sens 100%), TMJ (DICE 0.950, Sens 96.9%). Teeth assessment showed DICE 0.957, Sensitivity 96.9%, Specificity 99.4%. Overall tooth labeling accuracy was 98.6%.
Technological Characteristics
Cloud-hosted, web-based software (SaaS). Operates on standard web browsers (Safari, Edge, Chrome, Firefox). Utilizes Machine Learning algorithms for image segmentation and detection of anatomical structures in CBCT DICOM data. Data input via web interface; output includes visual overlays and STL exports. Software verification and validation conducted per FDA guidance.
Indications for Use
Indicated for licensed dental professionals to assist in the clinical detection of anatomical structures (maxilla, cranium, mandible, mandibular canal, TMJ, maxillary sinus, teeth, airway) in CBCT radiographs of permanent teeth in patients 15 years of age or older.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
July 23, 2026
Velmeni, Inc.
% Mini Suri
CEO
333 W. Maude Ave.
SUNNYVALE, CA 94085
Re: K254096
Trade/Device Name: VELMENI for DENTISTS (V4D) 3D
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: June 18, 2026
Received: June 22, 2026
Dear Mini Suri:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254096 - Mini Suri
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K254096 - Mini Suri
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K254096
Device Name
VELMENI for DENTISTS (V4D) 3D
Indications for Use (Describe)
VELMENI for DENTISTS (V4D) 3D is radiological, automated, concurrent read, computer-assisted detection software intended to assist in the clinical detection of anatomical structures (maxilla plus cranium, mandible, mandibular canal, temporomandibular joint, maxillary sinus ,teeth ,airway) in CBCT radiographs of permanent teeth in patients 15 years of age or older. This device provides additional information for licensed dental professionals in examining radiographs of patients' teeth. The device is not intended as a substitute for a complete examination by the dental professional or their clinical judgment that considers other relevant information from the patient's history or actual in-vivo clinical assessment. Final diagnoses and patient treatment plan are the responsibility of the dental professional.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
K254096
# VELMENI for DENTISTS (V4D) 3D
## 510(k) Summary
In accordance with 21 CFR 807.87(h) and 21 CFR 807.92, the following 510(k) Summary for VELMENI for DENTISTS (V4D) 3D is provided:
**Submitter Information**
| Submitter: | Velmeni Inc. 333 West Maude Avenue, STE 207 Sunnyvale, CA 94085 Phone: 201-289-3500 |
| --- | --- |
| Date Prepared: | July 22, 2026 |
| Contact Person: | Mini Suri, CEO Velmeni Inc. Phone: 201-289-3500 Email: mini@velmeni.com |
**Identification of the Device**
| Trade Name: | VELMENI for DENTISTS (V4D) 3D |
| --- | --- |
| Common Name: | Medical image management and processing system |
| Classification Name | Medical image management and processing system 21 CFR 892.2050 |
| Product Code: | QIH |
| Device Class: | Class II |
**Predicate Device**
Predicate Device: Pearl Second Opinion 3D (K243989)
**Device Description**
V4D 3D introduces 3D image processing to the VELMENI AI platform. The device comprises of the following key components
- Web Application Interface delivers front-end capabilities and is the point of interaction between the device and the user.
- Machine Learning (ML) Engine delivers V4D's core ML capabilities for 2D and 3D imaging.
- Backend API allows interaction between all the components, as defined in this section, in order to fulfill the user & requests on the web application interface.
- Queue receives and stores messages from Backend API to send to AI-Worker.
- AI-Worker accepts 3D scan analysis requests from Backend API via the Queue.
- For 3D, DICOM scans (in .zip format) are passed via Backend API in nifti format to the appropriate 3D ML engine. The ML analysis results are returned to the Backend API.
- Database and File Storage store critical information related to the application, including user data, patient profiles, analysis results, 3D scans and associated data.
{5}
The following non-medical interfaces are also available with VELMENI for DENTISTS (V4D) 3D
- VELMENI BRIDGE (VB) acts as a conduit enabling data and information exchange between Backend API and third-party software like Patient Management or Imaging Software
### Intended Use/ Indications for Use
VELMENI for DENTISTS (V4D) 3D is radiological, automated, concurrent read, computer-assisted detection software intended to assist in the clinical detection of anatomical structures (maxilla plus cranium, mandible, mandibular canal, temporomandibular joint, maxillary sinus ,teeth ,airway) in CBCT radiographs of permanent teeth in patients 15 years of age or older. This device provides additional information for licensed dental professionals in examining radiographs of patients' teeth. The device is not intended as a substitute for a complete examination by the dental professional or their clinical judgment that considers other relevant information from the patient's history or actual in-vivo clinical assessment. Final diagnoses and patient treatment plan are the responsibility of the dental professional.
### Technological Comparison
The proposed VELMENI for DENTISTS (V4D) 3D device is substantially equivalent to the primary predicate device; Pearl Second Opinion 3D K243989
| Specification | Proposed Device | Predicate Device |
| --- | --- | --- |
| Manufacturer | Velmeni Inc. | Pearl |
| Classification | 892.2050 | 892.2050 |
| Product Code | QIH | QIH |
| Image Modality | CBCT Radiograph | Radiograph |
| Intended Body Part | Dental Structures | Dental Structures |
| End User | Licensed dental professional | Dentist |
| Patient Population | Patients requiring dental services, all sexes, at least 15 years of age, and with permanent dentition. | Patients 12 years of age or older |
| Prescription or OTC | Prescription Use | Prescription Use |
{6}
| **Reader Workflow** | Concurrent Read | Concurrent Read |
| --- | --- | --- |
| **Image Source** | DICOM | Unknown |
| **Cloud Hosted Software** | Yes | Yes |
| **Platform** | Web – Safari, Edge, Chrome, Firefox | Web |
| **User Interface** | Mouse, keyboard, trackpad | Mouse, keyboard, trackpad |
| **Data Input** | CBCT Images | CBCT Images |
| **Data Output** | AI Detection of non-pathological (maxilla, mandible, mandibular canal, temporomandibular joint, maxillary sinus, teeth, airway), Panoramic viewer for automated dental arch and SLT file export “For educational and research purposes only; not intended for 3D printing or treatment planning” | Dentition, Maxilla, Mandible, Inferior Alveolar Canal and Mental Foramen (IAN), Maxillary Sinus, Nasal space, and airway |
| **Segmentation** | Segmentation on radiograph | Segmentation on radiograph |
| **Tooth Number Module** | Yes | No |
| **Model (Algorithm Type)** | Machine Learning | Computer vision and Machine Learning |
{7}
# **Performance Data**
# **Software Verification and Validation Tests:**
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submission for Device Software Functions." Verification of the software was conducted to ensure that the product works as designed. Validation was conducted to validate the design and the performance of the device to meet user needs and intended uses. VELMENI for DENTISTS (V4D) 3D passed all verification and validation software tests.
# **Bench Testing:**
The standalone analysis evaluating segmentation and correct labeling demonstrated the Velmeni 3D CBCT algorithm performance. 127 de-identified digital images were collected from 4 separate US dental offices representative of the general population. The data collected had representation from 3 manufacturers(INSTRUMENTARIUM DENTAL, Planmeca and Sirona) with 20 or more cases, constituting 87% of the test cases. A subgroup analysis was performed that included the following categories: scanner type, sex, and age categories (≤40, 41-64, >64 and 15-21, 22-49, and >= 50).
Based upon the bench testing performed it was determined that the proposed device performed similarly to the predicate
Table 1 Mean DICE and Sensitivity for V4D Software - Airway, Canal, Mandible, Maxilla, Sinus and TMJ
| Assessment | DICE Mean | Sensitivity(95% CI) |
| --- | --- | --- |
| Airway | 0.968 (95% CI: 0.96, 0.974) | 99.2%(97.6%, 100.0%) |
| Mandibular canal (Canal) | 0.914 (95% CI: 0.903, 0.924) | 96.8%(93.5%, 99.2%) |
| Mandible | 0.952 (95% CI: 0.949, 0.956) | 100.0%(97.1%, 100.0%) |
| Maxilla plus cranium (Maxilla) | 0.951 (95% CI: 0.947, 0.955) | 100.0%(97.1%, 100.0%) |
| Maxillary Sinus (Sinus) | 0.96 (95% CI: 0.9, 0.9) | 100.0%(97.1%, 100.0%) |
| Temporomandibular joint (TMJ) | 0.950 (95% CI: 0.937, 0.961) | 96.9%(94.2%, 99.1%) |
Table 2 Mean DICE,Sensitivity & Specificity for V4D Software - Teeth
| Assessment | DICE Mean | Sensitivity(95% CI) | Specificity(95% CI) |
| --- | --- | --- | --- |
| Teeth | 0.957 (95% CI: 0.952, 0.962) | 96.9%(95.8%, 98.0%) | 99.4%(98.7%, 99.9%) |
| Teeth (imputing 0 for missed teeth*) | 0.944 (95% CI: 0.935, 0.953) | - | - |
{8}
*This includes a small number of teeth (1.3% of all teeth) not identified by the algorithm but indicated by the ground truth which were imputed as 0 for the DICE analysis.
Table 3 Tooth Labeling Accuracy for V4D Software
| Assessment | Accuracy(95% CI) |
| --- | --- |
| Overall Correct Tooth Labeling | 98.6%(97.9%, 99.1%) |
| Images with 100% Overall Correct Tooth labeling | 77.0%(69.0%, 84.1%) |
## Conclusion
Both the predicate device and the proposed VELMENI for DENTISTS (V4D) 3D device share the same intended purpose. Despite minor technological differences, these differences in technological features do not raise concerns regarding safety or effectiveness. Following the information reviewed as part of this 510k, it can be concluded that the VELMENI for DENTISTS (V4D) 3D is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.