VDMS-1000S is intended for generating mammographic images that can be used for screening and diagnosis of breast cancer. It can be used to acquire 2D digital mammograms in the same clinical applications as traditional film/screen systems.
Device Story
Full-field digital mammography system comprising gantry (high-voltage generator, X-ray tube, collimator, digital detector), console table, foot switch, compression paddle, and magnification stand. Device acquires X-ray images, converts signals to digital format, and stores them in a workstation. Software applies image processing algorithms to optimize image quality. Used in clinical settings for breast cancer screening and diagnosis. Healthcare providers use the resulting 2D digital mammograms to inform clinical decision-making. Supports DICOM standard for PACS integration.
Clinical Evidence
Clinical image evaluation conducted per FDA 'Class II Special Controls Guidance Document: Full Field Digital, System, X-ray, Mammographic'. Study confirmed the subject device provides images of equivalent diagnostic capability to the predicate device (RMF-2000).
Technological Characteristics
Full-field digital mammography system; amorphous silicon TFT with Photo diode detector; CsI scintillator; high-frequency generator; semi-auto/auto/manual exposure modes; DICOM connectivity; complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-45 standards.
Indications for Use
Indicated for screening and diagnosis of breast cancer in adult patients, primarily females older than 40 years.
Regulatory Classification
Identification
A full-field digital mammography system is a device intended to produce planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). The special control for the device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Full-Field Digital Mammography System.”*See* § 892.1(e) for the availability of this guidance document.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 28, 2026
Vieworks Co., Ltd.
Seonghwan Park
Regulatory Affairs Assistant Manager
41-3, Burim-Ro 170beon-Gil, Dongan-Gu
ANYANG-SI, GYEONGGI-DO, 14055
REPUBLIC OF KOREA
Re: K254092
Trade/Device Name: Bellalun 2D (VDMS-1000S)
Regulation Number: 21 CFR 892.1715
Regulation Name: Full-Field Digital Mammography System
Regulatory Class: Class II
Product Code: MUE
Dated: April 29, 2026
Received: April 29, 2026
Dear Seonghwan Park:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254092 - Seonghwan Park
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K254092 - Seonghwan Park
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# YANNA S. KANG -S
Yanna Kang, Ph.D.
Assistant Director
Mammography and Ultrasound Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K254092 | |
| Device Name Bellalun 2D (VDMS-1000S) | |
| Indications for Use (Describe) VDMS-1000S is intended for generating mammographic images that can be used for screening and diagnosis of breast cancer. It can be used to acquire 2D digital mammograms in the same clinical applications as traditional film/screen systems. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
# 510(k) Summary (K254092)
This 510(k) summary information is prepared in accordance with 21 CFR 807.92
1. Date Prepared [21 CFR 807.92(a) (1)]
11/28/2025
2. Submitter's Information [21 CFR 807.92(a) (1)]
| Name of Sponsor: | Viewworks Co., Ltd. |
| --- | --- |
| Address: | 41-3, Burim-ro 170beon-gil, Dongan-gu, Anyang-si, Gyeonggi-do, 14055 Republic of Korea |
| Contact Name: | Seonghwan Park / Regulatory Affairs Assistant Manager |
| Telephone | +82-70-7751-9602 |
| Registration Number: | 3006013411 |
| Name of Manufacturer: | Same as Sponsor |
3. Trade Name, Common Name, Classification [21 CFR 807.92(a) (2)]
| Trade Name: | Bellalun 2D (VDMS-1000S) |
| --- | --- |
| Common Name: | Digital Mammography System |
| Classification Name: | Full Field Digital, System, X-Ray, Mammographic |
| Classification Panel: | Radiology |
| Classification Regulation: | 21 CFR 892.1715 |
| Product Code: | MUE |
| Device Class: | 2 |
4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]
| 510(k) Number: | K220073 |
| --- | --- |
| Applicant: | DRTECH Corporation |
| Trade Name: | RMF-2000 |
| Common Name: | Digital Mammography System |
| Classification Name: | Full-Field Digital Mammography System |
| Classification Panel: | Radiology |
| Classification Regulation: | 21 CFR 892.1715 |
| Product Code: | MUE |
| Device Class: | 2 |
| Decision Date: | 01/26/2023 |
| Type: | Traditional |
{5}
5. Description of the Device [21 CFR 807.92(a) (4)]
- General Description
The device is composed of a gantry—which includes a high-voltage generator, X-ray tube assembly, collimator, and digital detector—along with a console table and accessories such as a foot switch, compression paddle, and magnification stand. It is intended for use in diagnostic mammography.
The image signals generated by X-ray exposure are converted into digital signals and stored in the workstation (console table). Various image processing algorithms are applied through software to achieve optimal image quality. In addition, the device complies with the international medical imaging standard, DICOM (Digital Imaging and Communications in Medicine), to support integration with PACS.
6. Indications for Use [21 CFR 807.92(a)(5)]
VDMS-1000S is intended for generating mammographic images that can be used for screening and diagnosis of breast cancer. It can be used to acquire 2D digital mammograms in the same clinical applications as traditional film/screen systems.
7. Technological Characteristics [21 CFR 807.92(a) (6)]
Comparisons with the predicate, devices show the technological characteristics of the proposed VDMS-1000S device to be substantially equivalent to the predicate devices. The proposed devices are functionally identical to the predicate devices.
8. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92]
When compared to the predicate devices (K220073), the VDMS-1000S presented in this submission has the substantially equivalent:
- Indications for Use
- Patient Population
- Design
- Materials Scintillator
- Detector Type
- Power supply
- Exposure Mode
- Generator Type
- Breast Compression System
- SID
{6}
There are no significant difference between the VDMS-1000S and the predicate device that would adversely affect the use of the product. It is substantially equivalent to these devices in design, function, materials, operational principles and intended use.
| Parameter | Predicate Device | Subject Device | Equivalence |
| --- | --- | --- | --- |
| 510(k) Number | K220073 | - | - |
| Manufacturer | DRTECH Corporation | Viewworks Co., Ltd. | - |
| Model Name | RMF-2000 | VDMS-1000S | - |
| Classification Name | Full Field Digital, System, X-ray, Mammographic | | Equivalent |
| Classification Panel | Radiology | | Equivalent |
| Classification Regulation | 21 CFR 892.1715 | | Equivalent |
| Product Code | MUE | | Equivalent |
| Device Class | 2 | | Equivalent |
| Indications for Use | The RMF-2000 generates 2D digital mammography images. The RMF-2000 is intended to be used for screening and diagnosis of breast cancer. This unit is intended for use in the same clinical applications as traditional screen film mammography systems. | VDMS-1000S is intended for generating mammographic images that can be used for screening and diagnosis of breast cancer. It can be used to acquire 2D digital mammograms in the same clinical applications as traditional film/screen systems. | Substantially Equivalent |
| Patient Population | Age: Adult for the purpose of mammography screening | Age: VDMS-1000S is used to examine adult patients, mainly female patients older than 40 years old. | Equivalent |
| Design | Detector Size: 240 x 300 mm | Detector Size: 254.5 x 327.5 mm | Substantially Equivalent |
| | Pixel Pitch -RSM 2430UDP: 65um -RSM 2430TD: 76um | Pixel Pitch -FXMD-2430DAH: 75um | Substantially Equivalent |
| | Resolution -RSM 2430UDP: 4,608 x3,584 -RSM 2430TD: 3,840 x 3,072 | Resolution -FXMD-2430DAH: 3,840 x 3,072 | Substantially Equivalent |
{7}
| Materials Scintillator | RSM 2430UDP (Selenium) RSM 2430TD (CsI) | FXMD-2430DAH (CsI) | Substantially Equivalent |
| --- | --- | --- | --- |
| Detector Type | RSM 2430UDP: a-Se RSM 2430TD: amorphous silicon with CsI scintillator | FXMD-2430DAH: amorphous silicon TFT with Photo diode | Substantially Equivalent |
| Power supply | Single-Phase, Input voltage 200-240 Vac (±10%) | Single-Phase, Input voltage 220-240 Vac | Substantially Equivalent |
| Exposure Mode | AEC (Automatic Exposure Control), Manual | Semi-Auto, Auto, Manual | Substantially Equivalent |
| Generator Type | High Frequency | High Frequency | Equivalent |
| Breast Compression System | Maximum Compression that can be applied (N) Motor driven: 200 Manual: 30 | Maximum Compression that can be applied (N) Motor driven: 200 Manual: 50 | Substantially Equivalent |
| SID | 660 mm | 650 mm | Substantially Equivalent |
{8}
9. Summary of Non-Clinical Data
A comparison test was conducted between the subject devices VDMS-1000S and the predicate device (K220073) on the items such as DQE, MTF.
These detectors comply with the following international and FDA-recognized consensus standards:
- IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007) + A2: 2020
- Notification of MFDS No. 2025-06, Annex 1 (includes National Differences for Korea)
- IEC 60601-1-6: 2010 (Third Edition) + A2: 2020
- IEC 60601-1-2:2020
- IEC 60601-2-45:2011 + A2:2022
10. Summary of Clinical Data
A clinical image evaluation according to CDRH's 'Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Full Field Digital, System, X-ray, Mammographic was conducted, and the study confirmed that the subject device provides images of equivalent diagnostic capability to the predicate devices, RMF-2000 and its results demonstrate substantial equivalence.
11. Conclusion [21 CFR 807.92(b) (3)]
The VDMS-1000S Digital Mammography System is substantially equivalent to the currently marketed and predicate device (K220073) in terms of Indications for Use, Patient Population, Design, Materials Scintillator, Detector Type, Power supply, Exposure Mode, Generator Type, Breast Compression System, SID and safety and effectiveness.
Additionally, Substantial equivalence was demonstrated through the non-clinical performance, which complied with the requirements specified in the international and FDA-recognized consensus standards, IEC60601-1, IEC 60601-1-2, IEC 60601-2-45 and the clinical test, which complied with the requirements specified in the CDRH's 'Class II Special Controls Guidance Document: Full Field Digital, System, X-ray, Mammographic.
The results of these tests demonstrate that VDMS-1000S Digital Mammography System meets the acceptance criteria and is adequate for this intended use. The comparison of technological characteristics, non-clinical performance data, safety testing, and clinical image concurrence data demonstrates that the device is as safe, and effective as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.