The IOPS Visionary System is intended for the evaluation of vascular anatomy as captured via 3D modeling from previously-acquired scan data. It is intended for real time tip positioning and navigation using sensor-equipped compatible catheters and guidewires used in endovascular interventions in the peripheral, aortic and aortic side branch vasculature. The system is indicated for use as an adjunct to fluoroscopy. The IOPS does not make a diagnosis.
Device Story
IOPS Visionary™ is a surgical navigation system for endovascular interventions; utilizes electromagnetic tracking to display real-time position/orientation of sensor-equipped guidewires and catheters; inputs include previously-acquired CT scan data and live sensor signals; system processes data to generate 3D vascular maps and superimposes navigation onto these maps; includes capability to overlay live fluoroscopic imaging; operated by clinicians in clinical settings; provides visual feedback on onboard monitors to assist navigation; enables precise device placement; adjunct to fluoroscopy; does not perform diagnosis.
Clinical Evidence
No clinical or animal testing was conducted. Substantial equivalence is supported by non-clinical system and software verification and validation testing.
Technological Characteristics
Electromagnetic tracking system; consists of cart, interface module, system control unit, and field generator; accessories include sensorized catheters (6 Fr), guidewires, fiducial tracking pad, and guidewire handle; accessories sterilized via EtO; connectivity includes USB pass-through and video output; software performs vascular segmentation, centerline/surface analysis, and real-time sensor tracking.
Indications for Use
Indicated for patients undergoing endovascular interventions in peripheral, aortic, and aortic side branch vasculature requiring real-time navigation and tip positioning of sensor-equipped catheters and guidewires.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
Predicate Devices
Intra-Operative Positioning System (IOPS) (K243842)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 17, 2026
Centerline Biomedical, Inc.
Carroll Martin
Sr. Regulatory Affairs Manager
4535 Renaissance Pkwy.
Cleveland, Ohio 44128
Re: K254089
Trade/Device Name: IOPS Visionary™
Regulation Number: 21 CFR 870.1425
Regulation Name: Programmable Diagnostic Computer
Regulatory Class: Class II
Product Code: DQK, DQY, DQX
Dated: December 19, 2025
Received: December 19, 2025
Dear Carroll Martin:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254089 - Carroll Martin
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K254089 - Carroll Martin
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Aneesh S. Deoras -S
Aneesh Deoras
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K254089 | ? |
| Please provide the device trade name(s). | | ? |
| IOPS Visionary™ | | |
| Please provide your Indications for Use below. | | ? |
| The IOPS Visionary System is intended for the evaluation of vascular anatomy as captured via 3D modeling
from previously-acquired scan data. It is intended for real time tip positioning and navigation using sensor-
equipped compatible catheters and guidewires used in endovascular interventions in the peripheral, aortic
and aortic side branch vasculature. The system is indicated for use as an adjunct to fluoroscopy. The IOPS
does not make a diagnosis. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
Centerline Biomedical Traditional
510(k) PREMARKET NOTIFICATION
IOPS Visionary System
Centerline Biomedical
# 510(k) Summary for the IOPS Visionary System
## I. Device Name
Trade Name: IOPS Visionary™
Device Class: Class II
Common/Usual Name: Programmable Diagnostic Computer
Classification Name: Computer, diagnostic, programmable
Product Codes: DQK, DQY, DQX
Regulation Number: 870.1425
## II. Predicate Device
The predicate device to which substantial equivalence is claimed is the Intra-Operative Positioning System (IOPS), cleared under K243842.
## III. Device Description
The IOPS® Visionary system displays the position and orientation of sensor-equipped IOPS guidewires and IOPS catheters utilizing electromagnetic tracking technology. The system enables mapping of the patient's vascular system utilizing previously-acquired scan data (CT). IOPS Visionary tracks the location and orientation of the sensors in real time, superimposing navigation of the IOPS Catheter and IOPS Guidewire to the patient's vascular map.
The system consists of a surgical navigation technology and a number of associated accessories. The navigation technology is a non-contact, reusable, multi-patient use device. The associated accessories are single use devices provided sterile (EtO).
The IOPS Visionary System is indicated for the evaluation of vascular anatomy as captured via 3D modeling from previously-acquired scan data. It is intended for real time tip positioning and navigation using sensor-equipped compatible catheters and guidewires used in endovascular interventions in the peripheral, aortic, and aortic side branch vasculature. The system is indicated for use as an adjunct to fluoroscopy. The IOPS Visionary System does not make a diagnosis.
The associated accessories include:
- Guidewire
- Catheters
- Fiducial Tracking Pad
- Guidewire Handle
The system consists of three major sections: the cart, the tracking system (Interface Module, System Control Unit and Field Generator) and the accessories.
K254089
510(k) Summary
{5}
Centerline Biomedical Traditional
510(k) PREMARKET NOTIFICATION
IOPS Visionary System
Centerline
Biomedical
# IV. Indications for Use
The IOPS Visionary System is intended for the evaluation of vascular anatomy as captured via 3D modeling from previously acquired scan data. It is intended for real time tip positioning and navigation using sensor-equipped compatible catheters and guidewires used in endovascular interventions in the peripheral, aortic and aortic side branch vasculature. The system is indicated for use as an adjunct to fluoroscopy. The IOPS does not make a diagnosis.
# V. Substantial Equivalence
| Device Features | Predicate Device Intra-Operative Positioning System, K243842 | Subject Device IOPS Visionary System | Equivalence Discussion |
| --- | --- | --- | --- |
| Intended Use | The IOPS (Intra-Operative Positioning System is intended for the evaluation of vascular anatomy as captured via 3D modeling from previously acquired scan data. It is intended for real time tip positioning and navigation using sensor-equipped compatible catheters and guidewires used in endovascular interventions in the peripheral, aortic and aortic side branch vasculature. The system is indicated for use as an adjunct to fluoroscopy. The IOPS does not make a diagnosis. | The IOPS Visionary System is intended for the evaluation of vascular anatomy as captured via 3D modeling from previously acquired scan data. It is intended for real time tip positioning and navigation using sensor-equipped compatible catheters and guidewires used in endovascular interventions in the peripheral, aortic and aortic side branch vasculature. The system is indicated for use as an adjunct to fluoroscopy. The IOPS does not make a diagnosis. | Same |
| Onboard Monitor | Displays patient maps and superimposed images | Displays patient maps and superimposed images | Same |
| Computer | Provides an operating system | Provides an operating system | Same |
| Keyboard | Allows data input | Allows data input | Same |
| Mouse | Allows data input | Allows data input | Same |
K254089
510(k) Summary
Page 2 of 9
{6}
Centerline Biomedical Traditional
510(k) PREMARKET NOTIFICATION
IOPS Visionary System
^{}[]
Centerline
Biomedical
| Device Features | Predicate Device Intra-Operative Positioning System, K243842 | Subject Device IOPS Visionary System | Equivalence Discussion |
| --- | --- | --- | --- |
| Uninterruptible power supply (UPS) | Provides power when electricity is not available | Provides power when electricity is not available. | Same |
| Power Requirements | 120V, 60HZ | 120V, 50/60HZ | Different. The addition of 50Hz enables the device to work with European standard electrical frequencies. This difference does not have any impact to safety or effectiveness |
| Voltage Transformer | Not present | Present | Different. This component allows the operation of the device with European 230V power input. This difference does not have any impact on safety or effectiveness. |
| Video Splitter | The video out enables the Mobile Cart to simultaneously display the user interface on the onboard monitor and provide a video signal for an external display. | The video out enables the Mobile Cart to simultaneously display the user interface on up to two onboard monitors and provide a video signal for an external display. | Different. This allows more than one user to observe the procedure. This does not impact the safety or effectiveness of the device. |
| Cables | Allows data transfer | Allows data transfer | Same |
K254089
510(k) Summary
Page 3 of 9
{7}
Centerline Biomedical Traditional
510(k) PREMARKET NOTIFICATION
IOPS Visionary System
^{}[]
Centerline
Biomedical
| Device Features | Predicate Device Intra-Operative Positioning System, K243842 | Subject Device IOPS Visionary System | Equivalence Discussion |
| --- | --- | --- | --- |
| IOPS Software Application | Software which creates vascular map used to show location of sensorized catheter and guidewires in real time.
Vascular map is generated from a preoperative CT scan which is processed to segment the vasculature and perform centerline and surface analysis. | Software which creates vascular map used to show location of sensorized catheter and guidewires in real time.
Vascular map is generated from a preoperative CT scan which is processed to segment the vasculature and perform centerline and surface analysis. Allows for overlay of live fluoroscopic imaging to the rendered vascular map. | Different. The additional capability for allowing for the overlay of live fluoroscopic imaging is a convenience to the user. Live imaging allows the user to take into consideration any changes to the patient that have occurred since the original preoperative CT scan or the intra-operative conebeam scan was taken. |
| System Control Unit (SCU) | Collects information from the SIUs, calculates the position and orientation of each sensor and interfaces with the host computer. | Collects information from the SIUs, calculates the position and orientation of each sensor and interfaces with the host computer. | Same |
K254089
510(k) Summary
Page 4 of 9
{8}
Centerline Biomedical Traditional
510(k) PREMARKET NOTIFICATION
IOPS Visionary System
^{}[]
Centerline
Biomedical
| Device Features | Predicate Device Intra-Operative Positioning System, K243842 | Subject Device IOPS Visionary System | Equivalence Discussion |
| --- | --- | --- | --- |
| Sensor Interface Unit (SIU)
(Renamed unit Interface Module for IOPS Visionary System) | Amplifies and digitizes the electrical signals from the sensors and provides an increased distance between the SCU and sensors, while minimizing the potential for data noise | Amplifies and digitizes the electrical signals from the sensors and provides an increased distance between the SCU and sensors, while minimizing the potential for data noise. The SIU (renamed the Interface Module) also serves as USB pass-through connection for Tableside Controls Unit | Different.
This modification allows for communication between the Tableside Controls Unit and the IOPS device. This communication, in conjunction with the Tableside Controls Unit, allows the user to control the IOPS device from the Tableside Controls Unit. This modification does not affect the functionality, safety or effectiveness of the device. |
K254089
510(k) Summary
Page 5 of 9
{9}
Centerline Biomedical Traditional
510(k) PREMARKET NOTIFICATION
IOPS Visionary System
^{}[]
Centerline
Biomedical
| Device Features | Predicate Device Intra-Operative Positioning System, K243842 | Subject Device IOPS Visionary System | Equivalence Discussion |
| --- | --- | --- | --- |
| Tableside Controls Unit (MDC) | Not present | Present. | Different. The addition of the Tableside Control Unit allows the clinician to have direct control over IOPS software display via surgical tableside-mounted hardware as opposed to having another person operate the device. The functions that the Tableside unit controls are controlled in the predicate via the keyboard and mouse. In this proposed device, these functions can be controlled either utilizing the keyboard and mouse or the Tableside unit. This is a convenience for the user. This modification does not affect the functionality, safety or effectiveness of the device. |
| Electromagnetic Field Generator | Emits a low-intensity, varying electromagnetic field and establishes the position of the tracking volume | Emits a low-intensity, varying electromagnetic field and establishes the position of the tracking volume | Same |
K254089
510(k) Summary
Page 6 of 9
{10}
Centerline Biomedical Traditional
510(k) PREMARKET NOTIFICATION
IOPS Visionary System
^{}[]
Centerline
Biomedical
| Device Features | Predicate Device Intra-Operative Positioning System, K243842 | Subject Device Intra-Operative Positioning System | Equivalence Discussion |
| --- | --- | --- | --- |
| Mounting brackets | Affixes field generator to operating table | Affixes field generator to operating table | Same |
| USB Ports | Present. Allows the user to copy fluoroscan images from a flash drive that was obtained from the user’s fluoroscan. These images are used with IOPS to create vessel and bones models for use during procedures. | Present. Allows the user to copy fluoroscan images from a flash drive that was obtained from the user’s fluoroscan. These images are used with IOPS to create vessel and bones models for use during procedures. | Same |
| Functionality | The IOPS (MC-1) displays the position and orientation of sensor equipped catheters, guidewires, and tracking pad utilizing electromagnetic tracking technology. The system enables mapping of the patient’s vascular system utilizing previously acquired CT scan data. IOPS registers the location and orientation of the sensors in real time, superimposing navigation of the catheters and guidewires to the patient’s vascular map displayed on a monitor. The system is for use as an adjunct to fluoroscopy and does not make a diagnosis. | The IOPS (MC-3) displays the position and orientation of sensor equipped catheters, guidewires, and tracking pad utilizing electromagnetic tracking technology. The system enables mapping of the patient’s vascular system utilizing previously acquired CT scan data. IOPS registers the location and orientation of the sensors in real time, superimposing navigation of the catheters and guidewires to the patient’s vascular map displayed on a monitor. The system is for use as an adjunct to fluoroscopy and does not make a diagnosis. | Same |
K254089
510(k) Summary
Page 7 of 9
{11}
Centerline Biomedical Traditional
510(k) PREMARKET NOTIFICATION
IOPS Visionary System
^{}[]
Centerline
Biomedical
| Device Accessories | Predicate Device Intra-Operative Positioning System, K243842 | Subject Device Intra-Operative Positioning System | Equivalence Discussion |
| --- | --- | --- | --- |
| IOPS Viewpoint Simple Curve Catheters (C00751; C01251) and Double Curve Catheters (C00752; C01252) | 6 Fr catheters equipped with multiple tracking sensors allowing the IOPS to detect and visualize the catheter tip position and shape, in real time, on a 3D rendering of the patient’s vascular map. They are available in two tip shape configurations: simple curve and double curve. Each curve is offered in two different working lengths (75cm and 125cm). | 6 Fr catheters equipped with multiple tracking sensors allowing the IOPS to detect and visualize the catheter tip position and shape, in real time, on a 3D rendering of the patient’s vascular map. They are available in two tip shape configurations: simple curve and double curve. Each curve is offered in two different working lengths (75cm and 125cm). | Same |
| IOPS Guidewire (ATW-2) | Guidewire is a 0.035” diameter wire with a single sensor at the distal end used to navigate through vasculature to facilitate placement of a catheter. | Guidewire is a 0.035” diameter wire with a single sensor at the distal end used to navigate through vasculature to facilitate placement of a catheter. | Same |
K254089
510(k) Summary
Page 8 of 9
{12}
Centerline Biomedical Traditional
510(k) PREMARKET NOTIFICATION
IOPS Visionary System
Centerline Biomedical
| Device Accessories | Predicate Device Intra-Operative Positioning System, K243842 | Subject Device Intra-Operative Positioning System | Equivalence Discussion |
| --- | --- | --- | --- |
| IOPS Fiducial Tracking Pad (T02111) | Fiducial Tracking Pad is equipped with a single tracking sensor allowing IOPS to track gross patient motion to allow maintenance of patient registration during endovascular procedure. It will not track minor patient motion such as breathing or cardiac movement. The tracking pad is equipped with radiopaque beads which allow registration of a conebeam CT scan of the patient in their current position to a previously acquired CT scan. | Fiducial Tracking Pad is equipped with a single tracking sensor allowing IOPS to track gross patient motion to allow maintenance of patient registration during endovascular procedure. It will not track minor patient motion such as breathing or cardiac movement. The tracking pad is equipped with radiopaque beads which allow registration of a conebeam CT scan of the patient in their current position to a previously acquired CT scan. | Same |
| IOPS Guidewire Handle (H01035) | Guidewire Handle is a non-sensorized handle that connects to an IOPS sensorized guidewire to allow detection and visualization of the guidewire tip position, in real time, on a 3D rendering of the patient’s vascular map. It consists of a handle housing which encases a cable assembly that connects to a sensorized IOPS Guidewire. | Guidewire Handle is a non-sensorized handle that connects to an IOPS sensorized guidewire to allow detection and visualization of the guidewire tip position, in real time, on a 3D rendering of the patient’s vascular map. It consists of a handle housing which encases a cable assembly that connects to a sensorized IOPS Guidewire. | Same |
## VI. Performance Testing
System and software verification and validation testing was conducted to establish equivalency to the predicate device in safety and effectiveness and to ensure the IOPS Visionary System met established requirements.
## VII. Animal and Clinical Testing
No animal or clinical testing was conducted.
## VIII. Conclusions
The differences between the IOPS Visionary System and its predicate do not raise new or different questions of safety or effectiveness. The successful completion of non-clinical testing demonstrates that the IOPS Visionary System performs as intended and is substantially equivalent to the predicate device.
K254089
510(k) Summary
Page 9 of 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.