MAXFRAME AUTOSTRUT™ with Bluetooth Multi-Axial Correction System
Applicant
Synthes (USA) Products, LLC
Product Code
KTT · Orthopedic
Decision Date
Jun 11, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The MAXFRAME AUTOSTRUT™ with Bluetooth Multi-Axial Correction System is indicated for the following treatments in adults, and in both children (3–12) and adolescents (12–21) in which the growth plates have fused or will not be crossed with hardware: • fracture fixation (open and closed) • pseudoarthrosis of long bones • limb lengthening (epiphyseal or metaphyseal distraction) • joint arthrodesis • infected fractures or nonunions • correction of bony or soft tissue deformities • correction of segmental defects
Device Story
External ring fixation system (Stewart Platform type) for orthopedic deformity correction; utilizes transfixion wires, Schanz screws, and pins attached to rings/plates connected by six telescoping struts. Surgeon assembles frame to address specific patient anatomy. Subject device adds Bluetooth functionality to existing MAXFRAME/AUTOSTRUT hardware; integrates with MAXFRAME 3 web-based software and mobile app. System enables automated adjustment of strut length to correct length, rotation, and angular deformity simultaneously. Used in clinical settings by surgeons to manage bone/soft tissue deformities; software/app facilitates frame configuration and adjustment workflows. Benefits include precise, multi-axial correction of complex orthopedic conditions.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical performance testing, including software verification/validation, cybersecurity penetration testing, EMC (IEC 60601), radiated susceptibility (RTCA DO-160), electrical safety (IEC 60601), IP 58 ingress protection, operational temperature, and system reliability testing.
Technological Characteristics
External ring fixation system; metallic bone fixation components. Features Bluetooth wireless connectivity for automated strut adjustment. System includes web-based software and mobile application. MR Unsafe. Complies with IEC 60601 (EMC/Electrical Safety) and RTCA DO-160 (Radiated Susceptibility). IP 58 rated for ingress protection.
Indications for Use
Indicated for adults, children (3-12), and adolescents (12-21) with fused growth plates or where hardware does not cross growth plates for fracture fixation, pseudoarthrosis, limb lengthening, joint arthrodesis, infected fractures/nonunions, and correction of bony/soft tissue deformities or segmental defects. Contraindicated for use in the spine.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 11, 2026
Synthes (USA) Products, LLC
Mitchel Bartko
Regulatory Affairs Specialist III
1301 Goshen Pkwy.
West Chester, Pennsylvania 19380
Re: K254087
Trade/Device Name: MAXFRAME AUTOSTRUT with Bluetooth Multi-Axial Correction System
Regulation Number: 21 CFR 888.3030
Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories
Regulatory Class: Class II
Product Codes: KTT, OSN
Dated: May 13, 2026
Received: May 13, 2026
Dear Mitchel Bartko:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254087 - Mitchel Bartko
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K254087 - Mitchel Bartko
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Peter G.
Allen -S
Digitally signed by
Peter G. Allen -S
Date: 2026.06.11
18:21:27 -04'00'
for Lixin Liu, Ph.D.
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K254087 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| MAXFRAME AUTOSTRUT™ with Bluetooth Multi-Axial Correction System | | |
| Please provide your Indications for Use below. | | ? |
| The MAXFRAME AUTOSTRUT™ with Bluetooth Multi-Axial Correction System is indicated for the following treatments in adults, and in both children (3–12) and adolescents (12–21) in which the growth plates have fused or will not be crossed with hardware: • fracture fixation (open and closed) • pseudoarthrosis of long bones • limb lengthening (epiphyseal or metaphyseal distraction) • joint arthrodesis • infected fractures or nonunions • correction of bony or soft tissue deformities • correction of segmental defects | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☑ Children (2 years old to < 12 years old) ☑ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{4}
K254087 (Page 1 of 2)
# **510(K) SUMMARY**
| Sponsor | Synthes (USA) Products, LLC 1301 Goshen Parkway West Chester, PA 19380 |
| --- | --- |
| Contact | Mitchel Bartko Regulatory Affairs Specialist III T: +1 908 808 6541 E: mbartko1@ITS.JNJ.com |
| Alternate Contact | Jeffrey Krawiec, PhD Senior Regulatory Affairs Program Lead T: +1 484 816 6858 E: jkrawiec@its.jnj.com |
| Date Prepared | May 7, 2026 |
| Proprietary Name | MAXFRAME AUTOSTRUT™ with Bluetooth Multi-Axial Correction System |
| Classification Name | Single/Multiple Component Metallic Bone Fixation Appliances And Accessories |
| Classification | Class II Regulation Numbers: 21 CFR §888.3030 Product Codes: KTT; OSN |
| Predicate Device | Traditional 510(k) submission with the following predicate devices: • K202810 - AutoStrut G2 • K211313 - MAXFRAME Multi-Axial Correction System (aka MAXFRAME) |
| Device Description | MAXFRAME AUTOSTRUT™ with Bluetooth Multi-Axial Correction System consists of an automated hexapod control system kit with Bluetooth functionality, accessories, MAXFRAME™ 3 Software (a web-based software) and MAXFRAME AUTOSTRUT™ Mobile Application (App) to build an automated external fixation frame for the treatment of soft tissue and bone deformities. Additionally, it is used with hardware components from the MAXFRAME Multi-Axial Correction System and the MAXFRAME AUTOSTRUT System which information on these systems can be found in IFU GP3026 and IFU GP3150, respectively. MAXFRAME AUTOSTRUT with Bluetooth Control System is MR Unsafe. The MAXFRAME AUTOSTRUT with Bluetooth Multi-Axial Correction System is only compatible with the DePuy Synthes MAXFRAME and MAXFRAME AUTOSTRUT Systems. The MAXFRAME AUTOSTRUT with Bluetooth Multi-Axial Correction System with Adult and Pediatric indications is an external ring fixation system that is combined with MAXFRAME 3 Software (web-based) and MAXFRAME AUTOSTRUT Mobile App in treatment of soft tissue and bone deformities. In the MAXFRAME AUTOSTRUT with Bluetooth Multi-Axial Correction System, implanted transfixion wires, Schanz screws and pins are attached to the rings and plates surrounding the deformity using bolts, nuts and connecting plates. Upper and lower rings are then connected to one another using six telescoping struts, creating a “Stewart Platform” type device. Adjusting the strut length |
Page 1 of 2
{5}
K254087 (Page 2 of 2)
| | allows for correction of length, rotation, and angular deformity at the same time. The modular nature of a ring fixation frame allows multiple frame options. A ring fixation frame is assembled individually by the surgeon to address the different characteristics of each case. |
| --- | --- |
| Indications for Use | The MAXFRAME AUTOSTRUT with Bluetooth Multi-Axial Correction System is indicated for the following treatments in adults, and in both children (3-12) and adolescents (12-21) in which growth plates have fused or will not be crossed with hardware: • fracture fixation (open and closed) • pseudoarthrosis of long bones • limb lengthening (epiphyseal or metaphyseal distraction) • joint arthrodesis • infected fractures or nonunions • correction of bony or soft tissue deformities • correction of segmental defects |
| Contraindications | The MAXFRAME AUTOSTRUT with Bluetooth Multi-Axial Correction System is not intended for use in the spine. |
| Non-Clinical Performance Testing | To demonstrate the safety and efficacy of the subject devices and support the substantial equivalence to their predicates, the following testing was performed: • Software Testing as Part of Verification & Validation • Cybersecurity Testing (e.g. penetration testing) • Electrical Magnetic Compatibility (EMC) Testing per IEC 60601 • Radiated Susceptibility (RS) Testing per RTCA DO-160, Section 20.5 • Electrical Safety Testing per IEC 60601 • IP 58 Testing (ingress protection) • Operational Temperature Testing • System Reliability Testing Note: Magnetic Resonance (MR) testing was not provided as the device is MR Unsafe which is unchanged from the predicate device. |
| Clinical Performance Data | Clinical testing was not necessary for the determination of substantial equivalence. |
| Substantial Equivalence | The subject devices have the same intended use and indications as their predicate devices. Additionally, the subject devices are an iteration on the predicate device to include wireless (Bluetooth) functionality and the development of software applications to facilitate that functionality. Therefore, the fundamental design, technology, materials, and use of the devices are similar. Non-clinical performance data for software, cybersecurity, EMC, electrical safety, ingress protection, operational temperature and system reliability is included in this premarket notification to demonstrate that the subject device do not raise any new questions of safety and effectiveness compared to the predicate device. The proposed devices are at least as safe and effective as the predicate devices. |
| Conclusion | It is concluded that the information provided demonstrate the substantial equivalence of the subject devices to their predicate devices. |
Page 2 of 2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.