← Product Code [QJP](/productcode/QJP) · K254071

# p17 XR Microcatheter (K254071)

_Wallaby Medical · QJP · Aug 18, 2026 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K254071

## Device Facts

- **Applicant:** Wallaby Medical
- **Product Code:** [QJP](/productcode/QJP.md)
- **Decision Date:** Aug 18, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The p17 XR Microcatheter is indicated for use with compatible accessories in the delivery of interventional and therapeutic devices and infusion of diagnostic agents, such as contrast media, into the peripheral and neuro vasculature. It is not intended for use in coronary vasculature.

## Device Story

The p17 XR Microcatheter is a single-lumen, non-active, surgically invasive device used by trained physicians in healthcare facilities. It serves as a conduit for delivering interventional/therapeutic devices and infusing diagnostic agents into peripheral and neuro vasculature. The device features a PTFE-lined inner lumen, a stainless-steel coil, a nitinol braid in the proximal shaft, and a hydrophilic-coated distal section. It is operated manually by a physician under fluoroscopic guidance, with radiopaque marker bands providing visualization. The device is supplied sterile for single use. By providing a stable, navigable pathway through tortuous vasculature, it facilitates precise placement of therapeutic devices or agents, potentially improving clinical outcomes in neurovascular and peripheral interventions.

## Clinical Evidence

No clinical or animal studies were performed. Substantial equivalence is supported by non-clinical bench testing, including dimensional verification, navigability, kink resistance, coating lubricity/integrity, delivery/retrieval friction, torque, tip stiffness, stability, radiopacity, and compatibility with interventional devices and liquid embolic agents (DMSO).

## Technological Characteristics

Single-lumen microcatheter; 0.017" ID; 150/160 cm length. Materials: PTFE liner, stainless-steel coil, nitinol braid (proximal), Pebax/polyamide exterior, platinum-iridium marker bands, Trogamid hub. Hydrophilic coating on distal 60 cm. Sterilization: Ethylene Oxide. Standards: ISO 10555-1, ISO 80369-7/20, ISO 10993-1. Non-active, physical mode of action.

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- Headway 17 Advanced Microcatheter ([K101542](/device/K101542.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

**FDA** U.S. FOOD & DRUG
ADMINISTRATION

August 18, 2026

Wallaby Medical
Rachel McDaid
Senior Regulatory Affairs Specialist
22901 Mill Creek Drive
Laguna Hills, California 92653

Re: K254071

Trade/Device Name: p17 XR Microcatheter

Regulation Number: 21 CFR 870.1250

Regulation Name: Percutaneous Catheter

Regulatory Class: Class II

Product Code: QJP, DQY

Dated: July 16, 2026

Received: July 16, 2026

Dear Rachel McDaid:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K254071 - Rachel McDaid

Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K254071 - Rachel McDaid

Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

NAIRA MURADYAN -S

Naira Muradyan, PhD

Assistant Director

DHT5A: Division of Neurosurgical,

Neurointerventional, and

Neurodiagnostic Devices

OHT5: Office of Neurological and

Physical Medicine Devices

Office of Product Evaluation and Quality

Center for Devices and Radiological Health

Enclosure

{3}

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K254071 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  p17 XR Microcatheter  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  The p17 XR Microcatheter is indicated for use with compatible accessories in the delivery of interventional and therapeutic devices and infusion of diagnostic agents, such as contrast media, into the peripheral and neuro vasculature. It is not intended for use in coronary vasculature.  |   |   |
|  Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |

{4}

沃比医疗
Wallaby Medical

p17 XR Microcatheter Traditional 510(k)

### K254071 510(k) Summary (21 CFR 807.92)

#### I. SUBMITTER

Wallaby Medical

22901 Mill Creek Drive,

Laguna Hills, California 92653

Contact Person: Rachel McDaid

Phone: +353 (0)91 740128

Date Prepared: July 16th, 2026

#### II. DEVICE

Device Tradename: p17 XR Microcatheter

Common or Usual Name: Percutaneous Catheter

Classification: Class II device according to 21 CFR 870.1250 (Percutaneous catheter)

Product Codes and Review Panel:

|  Product Codes | QJP (primary), DQY (secondary)  |
| --- | --- |
|  Review Panel | Cardiovascular  |

#### III. PREDICATE DEVICE

Name of Predicate Device: Headway 17 Advanced Microcatheter

Name of Predicate Device Manufacturer: MicroVention, Inc.

510(k) number of Predicate Device: K101542

#### IV. DEVICE DESCRIPTION

##### Device Description:

The p17 XR Microcatheter is a 0.017" inner diameter (ID) single lumen microcatheter with working length of 150 cm or 160 cm. The inner lumen is lined with a polytetrafluoroethylene (PTFE) liner, and the outer layer consists of Pebax of varying durometers and polyamide. The outer distal 60 cm is coated with a hydrophilic coating. Between the polymer layers and the PTFE liner there is a stainless-steel coil which extends throughout the entire device shaft, and a nitinol braid over the stainless-steel coil in the proximal section of the device shaft. The distal tip of the microcatheter has either one or two radiopaque platinum-iridium marker bands depending on device configuration. The proximal end of the device contains an injection molded Trogamid hub and Pebax strain relief.

The p17 XR Microcatheter is a non-active, surgically invasive device intended for limited duration use within the vasculature.

Page 1 of 10

{5}

**沃比医疗**^{}[] Wallaby Medical

**p17 XR Microcatheter Traditional 510(k)**

# **Environment of Use:**

The p17 XR Microcatheter is intended to be used in a healthcare facility/hospital by trained physicians only.

# **V. INDICATIONS FOR USE**

**Table 1: Comparison of Indications for Use of the p17 XR Microcatheter with the predicate device**

|  Parameter | Predicate Device Headway 17 Advanced Microcatheter K101542 | Subject Device p17 XR Microcatheter K254071  |
| --- | --- | --- |
|  **Indications for Use** | The Headway 17 Advanced Microcatheter is intended for general intravascular use, including the peripheral, coronary and neuro vasculature for the infusion of diagnostic agents, such as contrast media, and therapeutic agents, such as occlusion coils. | The p17 XR Microcatheter is indicated for use with compatible accessories in the delivery of interventional and therapeutic devices and infusion of diagnostic agents, such as contrast media, into the peripheral and neuro vasculature. It is not intended for use in coronary vasculature.  |

The subject device has a subset of the predicate's intended use.

Page 2 of 10

{6}

![img-0.jpeg](img-0.jpeg)

沃比医疗

Wallaby Medical

p17 XR Microcatheter Traditional 510(k)

## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Table 2: Comparison of technological characteristics of the p17 XR Microcatheter and predicate device Headway 17 Advanced Microcatheter

|  Parameter | Predicate Device K101542 | Subject Device K254071  |
| --- | --- | --- |
|  Device Name | Headway 17 Advanced Microcatheter | p17 XR Microcatheter  |
|  Model # | MC172150SX | MC171501 MC171601 MC171502 MC171602  |
|  Product Classification | 21 CFR 870.1250 Class II | 21 CFR 870.1250 Class II  |
|  Product Code | DQY | QJP, DQY  |
|  Device Design/Materials | Stainless steel coil, polymeric exterior, two marker bands | Stainless steel coil, nitinol braid, polymeric exterior, 1 or 2 marker bands  |
|  Marker Band Material/ Radiopacity | Radiopaque Platinum/Iridium | Radiopaque Platinum/Iridium  |
|  Hub Material | Nylon (polyamide) | Polyamide  |
|  Coating | Hydrophilic coating Length: 100cm | Hydrophilic coating Length: 60cm  |
|  Mode of Action | Physical | Physical  |
|  Working Length | 150cm | 150cm, 160cm  |
|  Outer Diameter | 0.022" Distal 0.031" Proximal | 0.025" Distal Tip 0.029" Proximal  |
|  Inner Diameter | 0.017" | 0.017"  |
|  Use | Single Use Device | Single Use Device  |
|  Sterilization | Ethylene Oxide | Ethylene Oxide  |
|  Biocompatible | Yes | Yes  |
|  Shelf Life | 3 years | 12 months  |
|  Accessory Devices | Introducer Sheath, Shaping Mandrel | Introducer Sheath, Shaping Mandrel, Rotating Hemostasis Valve(RHV)  |
|  Packaging Components/ Configuration | Device is stored within a packaging hoop sealed within a Tyvek to Nylon pouch, which is then packaged in a product carton. | Device is stored within a packaging hoop sealed within a Tyvek to Nylon pouch, which is then packaged in a product carton.  |

Page 3 of 10

{7}

[Non-Text]

沃比医疗

Wallaby Medical

p17 XR Microcatheter Traditional 510(k)

## VII. PERFORMANCE DATA

### Biocompatibility

Biocompatibility testing for the p17 XR Microcatheter was conducted based on International Organization for Standardization (ISO) 10993-1:2018 "Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process," and the US FDA guidance document "Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process"" (2023). Non-clinical bench testing has been conducted to evaluate compatibility of the p17 XR Microcatheter with dimethyl sulfoxide (DMSO), Onyx LES and SQUID liquid embolic agents.

Table 3 below summarizes biocompatibility testing performed on the p17 XR Microcatheter and accessories included with the device.

Table 3: Biocompatibility Testing Results for p17 XR Microcatheter

|  Test | Test Description | Conclusion  |
| --- | --- | --- |
|  p17 XR Microcatheter (including Shaping Mandrel) Test Results  |   |   |
|  CytotoxicityISO 10993-5 | ISO MEM Elution Cytotoxicity | No cytotoxic effect.  |
|  Irritation or Intracutaneous ReactivityISO 10993-23 | ISO Intracutaneous Irritation | No irritation indicated.  |
|  SensitizationISO 10993-10 | ISO Guinea Pig Maximization Sensitization | No sensitization indicated.  |
|  Systemic ToxicityISO 10993-11 | ISO Acute Systemic Toxicity | No acute systemic toxicity indicated.  |
|   |  ISO Materials Mediated Rabbit Pyrogen | p17 XR is deemed non-pyrogenic.  |
|  HemocompatibilityISO 10993-4 | Complement Activation | No complement activation indicated.  |
|   |  ASTM Hemolysis - Direct Contact and Extract Method | No hemolysis indicated.  |
|   |  Thromboresistance Evaluation | p17 XR is deemed comparable to the predicate.  |
|   |  ISO Heparinized Blood Platelet and Leukocyte Count Assay with Comparison Article | p17 XR is deemed comparable to the predicate.  |
|   |  ISO Partial Thromboplastin Time (PTT) Assay with Comparison Article | p17 XR is deemed comparable to the predicate.  |
|  Accessory Test Results - Introducer Sheath  |   |   |
|  CytotoxicityISO 10993-5 | ISO MEM Elution Cytotoxicity | No cytotoxic effect.  |
|   |  MTT | No cytotoxic effect.  |
|  SensitizationISO 10993-10 | ISO Guinea Pig Maximization | No sensitization indicated.  |

Page 4 of 10

{8}

[LOGO]

沃比医疗

Wallaby Medical

p17 XR Microcatheter Traditional 510(k)

|   | Sensitization |   |
| --- | --- | --- |
|  Irritation or Intracutaneous Reactivity ISO 10993-10 | ISO Intracutaneous Reactivity | No irritation indicated.  |
|  Systemic Toxicity ISO 10993-11 | ISO Acute Systemic Toxicity | No acute systemic toxicity indicated.  |
|   |  USP Material Mediated Pyrogenicity test | Test articles are deemed non-pyrogenic.  |
|  Hemocompatibility ISO 10993-4 | ASTM Hemolysis - Direct Contact and Extract Method | No hemolysis indicated.  |
|  Accessory Test Results - Rotating Hemostasis Valve  |   |   |
|  Cytotoxicity ISO 10993-5 | ISO MEM elution Cytotoxicity | No cytotoxic effect.  |
|  Hemocompatibility ISO 10993-4 | ASTM Hemolysis - Direct Contact and Extract Method | No hemolysis indicated.  |
|  Irritation or Intracutaneous Reactivity ISO 10993-10 | ISO Intracutaneous Reactivity | No irritation indicated.  |
|  Systemic Toxicity ISO 10993-11 | ISO Acute Systemic Toxicity | No acute systemic toxicity indicated.  |
|   |  ISO Materials Mediated Rabbit Pyrogen | RHV is deemed non-pyrogenic.  |
|  Sensitization ISO 10993-10 | ISO Guinea Pig Maximization Sensitization | No sensitization indicated.  |

### Sterilization

The p17 XR Microcatheter is sterilized by Ethylene Oxide gas. Sterilization takes place in Steris Sterilization Technologies (Suzhou) Ltd., Jiangsu, China.

The p17 XR Microcatheter has been adopted into an existing validated cycle per EN ISO 11135:2014 and AAMI TIR28:2009.

Standards utilized in regards to Sterilization:

ISO 11135:2014 “Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices”

FDA Guidance Documents utilized in regards to Sterilization:

“Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labelled as Sterile” (2016)

The sterilization cycle has been validated via the overkill method.

Page 5 of 10

{9}

**沃比医疗**^{}[] Wallaby Medical

**p17 XR Microcatheter Traditional 510(k)**

# **Shelf Life**

Shelf life conditioning has been completed and testing has been performed on devices subjected to the accelerated aging (AA) process to represent 1 year of aged units. Devices subjected to 1 year AA were tested in non-clinical bench testing.

*FDA Guidance Documents utilized in regards to Shelf Life:*

“Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling” (2019)

“Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters - Class II Special Controls Guidance for Industry and FDA” (2010)

“Intravascular Catheters, Wires, and Delivery Systems with Lubricious Coatings - Labeling Considerations” (2019)

Page 6 of 10

{10}

**沃比医疗**^{}[] Walloby Medical

**p17 XR Microcatheter Traditional 510(k)**

# **Non-Clinical Performance Data**

The following non-clinical performance tests were performed to support the substantial equivalence determination. These are summarized below.

# **PERFORMANCE TESTING - BENCH**

A full suite of performance testing on the bench was carried out on p17 XR Microcatheter. The results of this testing demonstrated compliance to all the design attributes and that the device performs as intended. The results of these tests, in conjunction with the substantial equivalence claims as outlined in the premarket notification, demonstrate substantial equivalence to the cited predicate device. Table 4 below summarizes the tests and results.

**Table 4: Summary of Non-Clinical Bench Testing**

|  Test Attribute | Test Method Summary | Result  |
| --- | --- | --- |
|  Dimensional Verification (Working Length, Inner Diameter (ID), Outer Diameter(OD)) | The dimensions of the catheter, shaping mandrel and introducer sheath are measured. | All samples met the acceptance criteria, confirming the device meets requirements to ensure it performs as intended.  |
|  Navigability | The device is tested for its ability to reach target site in an anatomical model. | All samples met the acceptance criteria, confirming the device meets requirements to ensure it performs as intended.  |
|  Kink Resistance | The catheter is tested at different locations for its ability to bend to clinically relevant radii without kinking. | All samples met the acceptance criteria, confirming the device meets requirements to ensure it performs as intended.  |
|  Coating Lubricity | The catheter's hydrophilic coating lubricity is tested by applying a force to the coated section of the catheter. The friction force shall be equal to or below the specified maximum friction values. | All samples met the acceptance criteria, confirming the device meets requirements to ensure it performs as intended. This is comparable to the predicate device, thus supporting substantial equivalence.  |
|  Coating Integrity | The coated length of device is inspected for defects post simulated use. | All samples met the acceptance criteria, confirming the device meets requirements to ensure it performs as intended.  |

Page 7 of 10

{11}

**沃比医疗**^{}[] Wallaby Medical

**p17 XR Microcatheter Traditional 510(k)**

|  Test Attribute | Test Method Summary | Result  |
| --- | --- | --- |
|  Delivery and Retrieval Friction | The device is tested for its ability to deliver and retract a interventional device from a target site in an anatomical model below a specified force. | All samples met the acceptance criteria, confirming the device meets requirements to ensure it performs as intended. This is comparable with the predicate device, thus supporting substantial equivalence  |
|  Torque | The catheter was evaluated for torque strength by rotating the test sample within an anatomical model. The catheter must exceed a specified minimum number of rotations. | All samples met the acceptance criteria, confirming the device meets requirements to ensure it performs as intended. This is comparable with the predicate device, thus supporting substantial equivalence.  |
|  Tip Stiffness | The catheter tip is tested by bending it under a specified force. | All samples met the acceptance criteria, confirming the device meets requirements to ensure it performs as intended. This is comparable with the predicate device, thus supporting substantial equivalence  |
|  Stability | Device shall maintain placement in a tortuous model during the delivery of an interventional device. | All samples met the acceptance criteria, confirming the device meets requirements to ensure it performs as intended.  |
|  Atraumatic Tip | The microcatheter tip shall be smooth, rounded. | All samples met the acceptance criteria, confirming the device meets requirements to ensure it performs as intended.  |
|  Marker Band Radiopacity | The marker bands on the microcatheter shall be visible under fluoroscopy. | All samples met the acceptance criteria, confirming the device meets requirements to ensure it performs as intended.  |
|  Tip Shapeability | The distal tip of the microcatheter must be able to hold shape after tip shaping, and tip shaping should not damage the microcatheter. | All samples met the acceptance criteria, confirming the device meets requirements to ensure it performs as intended.  |
|  Peelable Introducer Sheath | Peelable introducer sheath must be able to fit over the microcatheter tip without damaging the microcatheter, and peelable sheath must be able to peel away. | All samples met the acceptance criteria, confirming the device meets requirements to ensure it performs as intended.  |
|  Compatibility – Interventional Devices | An appropriately sized guidewire is delivered and retrieved through the microcatheter. The microcatheter is delivered and retrieved through an appropriately sized guide catheter. | All samples met the acceptance criteria, confirming the device meets requirements to ensure it performs as intended.  |
|  Chemical Compatibility | The microcatheter shall resist damage when exposed to saline and contrast media. | All samples met the acceptance criteria, confirming the device meets requirements to ensure it performs as intended  |

Page 8 of 10

{12}

**沃比医疗**^{}[] Walloby Medical

**p17 XR Microcatheter Traditional 510(k)**

|  Test Attribute | Test Method Summary | Result  |
| --- | --- | --- |
|  Compatibility – DMSO | The microcatheter shall resist damage and maintain mechanical integrity after extended DMSO exposure. | All samples met the acceptance criteria, confirming the device meets requirements to ensure it performs as intended.  |
|  Surface Defects | The microcatheter is examined under magnification for extraneous matter. | All samples met the acceptance criteria, confirming the device meets requirements to ensure it performs as intended.  |
|  Peak Shaft and Hub Tensile | The catheter is tested at different locations under tensile force to meet a pre-determined specification. | All samples met the acceptance criteria, confirming the device meets requirements to ensure it performs as intended.  |
|  Liquid Leak | The catheter is tested for leakage under a specified static pressure for a specified time. | All samples met the acceptance criteria, confirming the device meets requirements to ensure it performs as intended  |
|  Air Leak | The catheter is tested for air leakage for a specified time. | All samples met the acceptance criteria, confirming the device meets requirements to ensure it performs as intended.  |
|  Hub Compatibility | The catheter hub is tested per ISO 80369-7. | All samples met the acceptance criteria, confirming the device meets requirements to ensure it performs as intended.  |
|  Static Burst | The catheter is tested to withstand a specified static pressure. | All samples met the acceptance criteria, confirming the device meets requirements to ensure it performs as intended.  |
|  Dynamic Burst | The catheter is tested to withstand a specified power injection pressure and to characterize flow rates and pressures during delivery of fluids. | All samples met the acceptance criteria, confirming the device meets requirements to ensure it performs as intended.  |
|  Corrosion | The device and shaping mandrel are visually inspected for signs of corrosion post exposure to required conditions. | All samples met the acceptance criteria, confirming the device meets requirements to ensure it performs as intended.  |
|  Particulate Testing | The catheter particulate matter is tested by calculating the size and number of particulates generated during simulated use of the device in a neurovascular model. | All samples met the acceptance criteria, confirming the device meets requirements to ensure it performs as intended. This was comparable to the predicate device, thus, supporting substantial equivalence.  |

Page 9 of 10

{13}

沃比医疗
Wallaby Medical

p17 XR Microcatheter Traditional 510(k)

Standards utilized in regards to Bench Testing:

ISO 10555-1:2023 “Intravascular catheters - Sterile and single-use catheters - Part 1: General requirements”

ISO 80369-20:2015(E) “Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods”

ISO 80369-7:2021(E) “Small-bore connectors for liquids and gases in healthcare application - Part 7: Connectors for intravascular or hypodermic applications”

IEC 62366-1: 2015/AMD1:2020 “Medical Devices- Part 1: Application of Usability Engineering to Medical Devices”

FDA Guidance Documents utilized in regard to Bench Testing:

“Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling” (2019)

“Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters - Class II Special Controls Guidance for Industry and FDA” (2010)

“Applying Human Factors and Usability Engineering to Medical Devices” (2016)

Clinical Testing

Clinical testing was not performed on the p17 XR Microcatheter as no human studies were deemed necessary to establish substantial equivalence.

Animal Testing

An animal study was not deemed necessary to support the substantial equivalence of the subject device to the predicate device. Bench testing was determined sufficient to support substantial equivalence.

VIII. CONCLUSIONS

Wallaby Medical has demonstrated that the p17 XR Microcatheter is substantially equivalent to the predicate Headway 17 Advanced Microcatheter (K101542). The subject device has the same intended use, similar technological characteristics, similar materials, and the same operating principle as the predicate device. The differences do not raise new questions of safety or effectiveness. The subject device has been demonstrated to perform as intended through successful non-clinical performance testing.

Page 10 of 10

---

**Source:** [https://fda-staging.innolitics.com/device/K254071](https://fda-staging.innolitics.com/device/K254071)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
