The LST – Life Saving Tourniquet is indicated for battlefield and trauma situations to: - Control difficult bleeds in the inguinal area. - Control difficult bleeds in the axilla area.
Device Story
Life Saving Tourniquet (LST) is a mechanical vascular clamp for emergency hemorrhage control in battlefield or trauma settings. Device consists of handle, inner strap, buckle, bridge, and pressure pin. Operator secures device via buckle; applies focal pressure to main artery of limb via manual winding of handle to drive pressure pin. Device maintains continuous compression during patient evacuation. Intended for prescription use by trained personnel. Benefits include effective hemorrhage control in junctional areas where standard tourniquets are ineffective.
Clinical Evidence
No clinical or animal testing performed. Substantial equivalence supported by bench testing, including usability, pressure retention, and effectiveness during evacuation.
Technological Characteristics
Mechanical vascular clamp; non-pneumatic. Components: handle, inner strap, buckle, bridge, pressure pin. Non-sterile, single-use. Materials meet ISO 10993-1 requirements for limited contact with intact skin. Manual operation; no energy source or software.
Indications for Use
Indicated for adults in battlefield and trauma situations to control difficult bleeds in the inguinal and axilla areas.
Regulatory Classification
Identification
A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 3, 2026
TW Medical
% Tyler Ting
Regulatory Director
Rook Quality Systems
1155 Mount Vernon Hwy
Dunwoody, Georgia 30338
Re: K254060
Trade/Device Name: Life Saving Tourniquet
Regulation Number: 21 CFR 870.4450
Regulation Name: Vascular clamp
Regulatory Class: Class II
Product Code: DXC
Dated: December 17, 2025
Received: December 17, 2025
Dear Tyler Ting:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254060 - Tyler Ting
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K254060 - Tyler Ting
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
RACHEL E. NEUBRANDER -S
for Katherine Trivedi
Assistant Director
DHT2B: Division of Circulatory Support, Structural, and Vascular Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K254060 | |
| Device Name Life Saving Tourniquet | |
| Indications for Use (Describe) The LST – Life Saving Tourniquet is indicated for battlefield and trauma situations to: - Control difficult bleeds in the inguinal area. - Control difficult bleeds in the axilla area. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K254060
In accordance with 21 CFR 807.92 the following summary information is provided:
## Contact Details
21 CFR 807.92(a)(1)
Tsadik-Weiser Medical, Ltd.
149 Revava Street, Ontiel, Israel
Applicant Contact: Yaakov Tsadik
CEO / Founder
149 Revava Street
Otniel, Israel 9040700
+972 54-7449946
ytsadik@t-w-medical.com
Correspondent: Tyler Ting, MS, RAC-Devices
Regulatory Director / Regulatory Consultant
Rook Quality Systems
1155 Mount Vernon Hwy, Suite 800
Dunwoody, GA 30338
970-404-1555
Tyler.ting@rookqs.com
Date Prepared: 11 DEC 2025
## Device
21 CFR 807-92(a)(2)
Trade Name: Life Saving Tourniquet (LST)
Common Name: Life Saving Tourniquet (LST)
Classification Name: Vascular Clamp
Regulatory Class: II
Product Code(s): DXC
Regulation Number: 870.4450
Classification Panel: Cardiology
## Predicate Device
21 CFR 807.92(a)(3)
{5}
Tsadik-Weiser Medical, Ltd. submits the following information to demonstrate that the subject device LST – Life Saving Tourniquet, is substantially equivalent to the following legally marketed predicate device:
510(k) Number
Name
Product Code
K131561
SAM Junctional Tourniquet
DXC
# Device Description
21 CFR 807.92(a)(4)
The LST – Life Saving Tourniquet is designed to effectively control severe bleeding in emergency situations by compressing the artery and maintaining continuous pressure during evacuation. The device is comprised of a durable handle, inner strap, sturdy buckle, dedicated bridge, pressure pin, and “cut here” section for easy application and evaluation. See Figure 1 below for device components. For junctional injuries, the LST is positioned so that the pressure pin applies force on the main artery leading to the limb. The strap is then tightened holding the pin firmly in place.
# The Israeli tourniquet:

Figure 1: LST - Life Saving Tourniquet component diagram
# Indications for Use
21 CFR 807.92 (a)(5)
{6}
The LST – Life Saving Tourniquet is indicated for battlefield and trauma situations to:
- Control difficult bleeds in the inguinal area.
- Control difficult bleeds in the axilla area.
# Comparison of Technological Characteristics with the Predicate Device
21 CFR 807.92 (a)(6)
The subject device has the same intended use, indications, principle of operation, similar technological and material characteristics as the predicate device, SAM Junctional Tourniquet (K131561). Comparisons of the devices can be found in Table 1:
Table 1 Comparison of the Subject, Primary Predicate, and Reference Devices
| | This 510(k) Notification The LST – Life Saving Tourniquet (Subject Device) | SAM Junctional Tourniquet (Predicate Device) K131561 | Discussion |
| --- | --- | --- | --- |
| Product Code, Device Risk Classification | DXC Class II | DXC Class II | Identical |
| Indications for Use | The LST - Life Saving Tourniquet is indicated for battlefield and trauma situations: • To control difficult bleeds in the inguinal area. • To control difficult bleeds in the axilla area | The SAM Junctional Tourniquet (SJT) is indicated for battlefield and trauma situations: • To control difficult bleeds in the inguinal area. • To control difficult bleeds in the axilla area. • To immobilize a pelvic fracture. | Substantially Equivalent – The subject device is indicated for junctional use in trauma and battlefield situations to control bleeding in the inguinal and axilla areas. The subject device does not include the additional pelvic fracture immobilization indication of the predicate. The subject device has a narrower indication for use. |
| Risk Classification | II | II | Identical |
| Regulation Number | 21 CFR 870.4450 | 21 CFR 870.4450 | Identical |
| Regulation Name | Vascular Clamp | Vascular Clamp | Identical |
{7}
| Intended Use Environment | Battlefield and trauma situations | Battlefield and trauma situations | Identical |
| --- | --- | --- | --- |
| Target Population | Adults | Adults | Identical |
| Rx or OTC | Rx only | Rx only | Identical |
| Mode of Operation | Buckle the tourniquet in place and use the handle to wind to apply additional pressure. | Buckle the tourniquet in place and use the hand pump to inflate the TCD until hemorrhage stops. | Substantially Equivalent - both the subject and predicate device use a buckle to secure the device and a manual mechanism for applying pressure. The subject device uses a handle to wind additional pressure, and the predicate uses a hand pump to inflate a TCD. |
| Pressure Point Mechanism | Manual, Non-pneumatic pressure pin | Manual, pneumatic | Substantially Equivalent - both the subject and predicate devices utilize a manual pressure point mechanism. The subject device is a mechanical pressure pin and the predicate is a manual pneumatic mechanism. |
| Shelf Life | N/A | N/A | Identical |
| Sterility | Nonsterile, single use | Nonsterile, single use | Identical |
## Biocompatibility
The device is intended for limited contact with intact skin. The device materials and intended use meet the requirements of Attachment G: Biocompatibility of Certain Devices in Contact with Intact Skin of FDA guidance document Use of International Standard ISO 10993-1 "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
## Performance Data
### Bench
Non-clinical validation testing was performed to demonstrate the substantial equivalence of the device to the predicate. The following tests were performed:
- Usability
- Pressure Retention Testing
- Effectiveness During Evacuation
{8}
## Animal
No animal or clinical testing was performed in support of this submission.
## Clinical
No animal or clinical testing was performed in support of this submission.
## Substantial Equivalence Conclusion
Based on the comparison of indication of use, material characteristics, technological features, and principle of operation, the LST – Life Saving Tourniquet is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.