The BosCATH™ Support Catheter is indicated for general intravascular use, including the neuro and peripheral vasculature. The BosCATH™ Support Catheter can be used to facilitate introduction of diagnostic agents or therapeutic devices. The BosCATH™ Support Catheter is not intended for use in coronary arteries.
Device Story
Single-lumen support catheter; variable stiffness; hydrophilic coating; dual-layer coil/braid wire reinforcement; radiopaque marker bands; luer hub. Used in neuro and peripheral vasculature; operated by physicians. Facilitates navigation and introduction of diagnostic/therapeutic devices. Provides support and flexibility during intravascular procedures. Benefits patient by enabling access to target vasculature for intervention.
Clinical Evidence
No clinical data. Substantial equivalence based on bench performance testing, including physical attributes, simulated use, burst pressure, tensile strength, kink resistance, torque strength, corrosion resistance, coating lubricity/durability, and biocompatibility (ISO 10993 series).
Technological Characteristics
Single-lumen catheter; thermoplastic polyurethane (NEUsoft), Pebax, Grilamid outer layer; stainless-steel coil/braid and PTFE inner layer; polycarbonate hub; polyether-based thermoplastic polyurethane strain relief. Hydrophilic coating; radiopaque Platinum/Iridium marker bands. Sterilized via Ethylene Oxide. Complies with ISO 10555-1, ISO 80369-7/20.
Indications for Use
Indicated for general intravascular use in neuro and peripheral vasculature to facilitate introduction of diagnostic agents or therapeutic devices. Not for use in coronary arteries.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 21, 2026
Sonorous NV, Inc.
% Erik Mulchandani
Quality Assurance & Regulatory Affairs Consultant
Global MedTech, LLC
1302 Scholarship
Irvine, California 92612
Re: K254046
Trade/Device Name: BosCATH Support Catheter
Regulation Number: 21 CFR 870.1250
Regulation Name: Percutaneous Catheter
Regulatory Class: Class II
Product Code: QJP, DQY
Dated: March 20, 2026
Received: March 20, 2026
Dear Erik Mulchandani:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254046 - Erik Mulchandani
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K254046 - Erik Mulchandani
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
NAIRA MURADYAN -S
Naira Muradyan, Ph.D.
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K254046 | |
| Device Name BosCATH™ Support Catheter | |
| Indications for Use (Describe) The BosCATH™ Support Catheter is indicated for general intravascular use, including the neuro and peripheral vasculature. The BosCATH™ Support Catheter can be used to facilitate introduction of diagnostic agents or therapeutic devices. The BosCATH™ Support Catheter is not intended for use in coronary arteries. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
SONOROUS™
^{}[]®
K254046
# 510(k) Summary
| Date Prepared: | April 20, 2026 |
| --- | --- |
| Applicant: | Sonorous NV, Inc.
1 Spectrum Pointe Drive, Suite 150
Lake Forest, CA 92630 |
| Applicant Contact: | Erik Mulchandani
Quality Assurance & Regulatory Affairs Consultant
(202) 246-0678 |
| Device Name: | BosCATH™ Support Catheter |
| Regulation Number: | 21 CFR 870.1250 |
| Classification and Primary Product Code: | Class II, QJP (Catheter, Percutaneous, Neurovasculature) |
| Subsequent Product Code: | DQY (Catheter, Percutaneous) |
| Predicate Device: | SOFIA™ EX Intracranial Support Catheter (K230775) |
| Reference Device: | Navien™ Intracranial Support Catheter (K161152) |
## Device Description:
The BosCATH™ Support Catheter is a single-lumen catheter with variable stiffness and a hydrophilic coating to facilitate navigation through the vasculature. The catheter shaft is reinforced with a dual-layer construction of coil and braid wire, designed for providing enhanced support and flexibility. The distal end of the catheter incorporates one (1) or more radiopaque marker bands to aid in visualization under fluoroscopy. The inner lumen accommodates guidewires up to 0.038" in diameter, assisting in catheter placement. The proximal end of the catheter features a luer hub, allowing for the attachment of compatible accessories and the infusion of fluids as needed. The BosCATH™ Support Catheter is available in multiple lengths to accommodate physician preference and anatomical variability. The BosCATH™ Support Catheter is supplied sterile, non-pyrogenic, and intended for single use only. The BosCATH™ Support Catheter is supplied with a sterile introducer sleeve that is intended for single use only.
## Indications for Use:
The BosCATH™ Support Catheter is indicated for general intravascular use, including the neuro and peripheral vasculature. The BosCATH™ Support Catheter can be used to facilitate introduction of diagnostic agents or therapeutic devices. The BosCATH™ Support Catheter is not intended for use in coronary arteries.
{5}
K254046
SONOROUS
^{}[]®
HAZARDOUS INSTITUTE
| Table 1: Comparison of Subject, Predicate and Reference Devices | | | |
| --- | --- | --- | --- |
| Attribute | SOFIA™ EX Intracranial Support Catheter Predicate Device (K230775) | Navien™ Intracranial Support Catheter Reference Device (K161152) | BosCATH™ Support Catheter Subject Device (K254046) |
| Indications for Use | The SOFIA EX Intracranial Support Catheter is indicated for general intravascular use, including the neuro and peripheral vasculature. The SOFIA EX Intracranial Support Catheter can be used to facilitate introduction of diagnostic agents or therapeutic devices. The SOFIA EX Intracranial Support Catheter is not intended for use in coronary arteries. | The Navien™ Intracranial Support Catheter is indicated for the introduction of interventional devices into the peripheral and neurovasculature. | The BosCATH™ Support Catheter is indicated for general intravascular use, including the neuro and peripheral vasculature. The BosCATH™ Support Catheter can be used to facilitate introduction of diagnostic agents or therapeutic devices. The BosCATH™ Support Catheter is not intended for use in coronary arteries. |
| Primary Product Code | QJP | DQY | Same as K230775 |
| Subsequent Product Code | DQY | N/A | Same as K230775 |
| Regulation Number | 21 CFR 870.1250 | Same as K230775 | Same as K230775 |
| Design Feature(s) | | | |
| Tip | Straight | Same as K230775 | Same as K230775 |
| Dimensional Specifications | | | |
| Inner Diameter | 0.058" | 0.046" - 0.072" | Same as K230775 |
| Outer Diameter | 0.071" | 0.058" - 0.084" | 0.075" |
| Working Length (cm) | 115 | 90 - 130 | 105
115
125 |
| Materials | | | |
| Catheter Shaft | The outer layer consists of polyolefin elastomer, polyurethane elastomer (Pellethane), polyether | Polytetrafluoroethylene (PTFE) lined polymeric catheter | The outer layer consists of thermoplastic polyurethane (NEUsoft), polyether block amide (Pebax) and |
2
{6}
K254046
SONOROUS
INSURANCE COMPLI
| Table 1: Comparison of Subject, Predicate and Reference Devices | | | |
| --- | --- | --- | --- |
| Attribute | SOFIA™ EX Intracranial Support Catheter Predicate Device (K230775) | Navien™ Intracranial Support Catheter Reference Device (K161152) | BosCATH™ Support Catheter Subject Device (K254046) |
| | block amide (Pebax) and polyamide (Grilamid). The inner layer consists of stainless-steel braid, Nitinol coil, and PTFE. | | polyamide (Grilamid). The inner layer consists of stainless-steel coil/braid and PTFE. |
| Marker Band | Platinum/Iridium | Platinum | Same as K230775 |
| Coating | Hydrophilic | Same as K230775 | Same as K230775 |
| Hub | Nylon | Not Specified | Polycarbonate |
| Strain Relief | Polyurethane | Not Specified | Polyether-based thermoplastic polyurethane |
| Introducer Sheath | Pebax | Not Specified | Same as K230775 |
| Packaging | | | |
| Accessories | Introducer Sheath | Not Specified | Same as K230775 |
| Configuration | The catheter is placed into a high-density polyethylene (HDPE) dispenser tube. The dispenser tube and introducer sheath are placed on a polyethylene packaging card that is inserted into a Tyvek® pouch. The pouch and the instructions for use are placed in solid bleached sulfate carton. | The catheter is placed into a polyethylene hoop that is attached to packaging card inside a polyethylene terephthalate (PET)/polyethylene (PE) Tyvek® pouch inside a solid bleached sulfate carton. | The catheter is placed into a combination packaging card/dispensing tube with the hub clipped into the packaging card. The combination packaging card/dispensing tube with the catheter and the introducer sleeve are sealed inside a DuPont™ Tyvek® pouch. The sealed pouch and the instructions for use are placed into a solid bleached sulfate carton. |
| Stability | | | |
| Shelf Life | Not Specified | 24 Months | 12 Months |
| Sterility | | | |
| Sterilization Method | Ethylene Oxide | Same as K230775 | Same as K230775 |
3
{7}
K254046
Performance Testing:
Packaging:
| Test | Test Method Summary | Result |
| --- | --- | --- |
| Visual Inspection | The packaging and labeling of the BosCATH™ Support Catheter was evaluated per ASTM F1886/F1886M-16. | Pass |
| Bubble Leak | The BosCATH™ Support Catheter packaging was evaluated per ASTM F2096-11 (2019). | Pass |
| Seal Strength | The BosCATH™ Support Catheter packaging was evaluated per ASTM F88/F88M-23. | Pass |
Sterilization:
| Test | Test Method Summary | Result |
| --- | --- | --- |
| Ethylene Oxide Residual | The BosCATH™ Support Catheter was evaluated per ISO 10993-7:2008. | Pass |
| Ethylene Chlorohydrin Residual | The BosCATH™ Support Catheter was evaluated per ISO 10993-7:2008. | Pass |
| Bacterial Endotoxin Testing | The BosCATH™ Support Catheter was evaluated per United States Pharmacopeia, National Formulary, General Chapter <85> Bacterial Endotoxins Test. | Pass |
Biocompatibility:
| Biological Endpoint | Test | Test Method Summary | Results |
| --- | --- | --- | --- |
| Cytotoxicity | MEM Elution Cytotoxicity Assay (ISO) | The BosCATH™ Support Catheter was evaluated per ISO 10993-5:2009 | Pass: The BosCATH™ Support Catheter was considered to be non-cytotoxic. |
| Sensitization | Guinea Pig Maximization Test (ISO) | The BosCATH™ Support Catheter was evaluated per ISO 10993-10:2021 | Pass: The BosCATH™ Support Catheter was not considered a sensitizer. |
| Intracutaneous Reactivity | Intracutaneous Reactivity Test (ISO) | The BosCATH™ Support Catheter was evaluated per ISO 10993-23:2021 | Pass: The BosCATH™ Support Catheter was considered a non-irritant. |
| Acute Systemic Toxicity | Acute Systemic Toxicity Test (ISO) | The BosCATH™ Support Catheter was evaluated per ISO 10993-11:2017 | Pass: The BosCATH™ Support Catheter was |
{8}
K254046
| Biological Endpoint | Test | Test Method Summary | Results |
| --- | --- | --- | --- |
| | | | considered to be non-toxic. |
| Material-Mediated Pyrogenicity | Material Mediated Pyrogenicity Test (ISO/USP) | The BosCATH™ Support Catheter was evaluated per ISO 10993-11:2017 and USP-NF 2020: <151> | Pass: The BosCATH™ Support Catheter was considered to be non-pyrogenic. |
| Hemocompatibility | ASTM Hemolysis Assay (ISO) | The BosCATH™ Support Catheter was evaluated per ASTM F756-17 and ISO 10993-4:2017/A1: 2025 | Pass: The BosCATH™ Support Catheter was considered to be non-hemolytic. |
| | Complement Activation Assay (ISO) | The BosCATH™ Support Catheter was evaluated per ISO 10993-4:2017/A1:2025 | Pass: The BosCATH™ Support Catheter was considered to be a non-activator of the complement system. |
| | A Non-Anticoagulated Venous Thrombogenicity Study in Ovine | The BosCATH™ Support Catheter was evaluated per ISO 10993-4:2017 | Pass: Minimal thrombus formation was observed post-implantation of the BosCATH™ Support Catheter. |
Performance Data – Bench:
The BosCATH™ Support Catheter was evaluated per ISO 10555-1, ISO 80369-7, ISO 80369-20 and FDA guidance documents “Peripheral Percutaneous Transluminal Angioplasty (PTA) and Specialty Catheters - Premarket Notification (510(k)) Submissions” and “Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters - Class II Special Controls Guidance for Industry and FDA” for the following bench tests with passing results:
- Physical Attribute/Surface Contamination
- Simulated Use
- Dynamic Burst Pressure/Flow Rate
- Air Leakage
- Liquid Leakage
- Static Burst Pressure
- Tensile Strength
- Tip Flexibility
- Kink Resistance
- Torque Strength
- Corrosion Resistance
- Coating Lubricity
- Coating Durability
- Particulate
- Radiopacity
- Small-Bore Connectors
5
{9}
K254046
Performance Data – Animal:
The determination of substantial equivalence is based upon successful completion of bench performance testing.
Performance Data – Clinical:
The determination of substantial equivalence is based upon successful completion of bench performance testing.
Conclusion:
The subject BosCATH™ Support Catheter has the same intended use and similar technological characteristics as the predicate device SOFIA™ EX Intracranial Support Catheter. The differences between the subject device and predicate device do not raise new questions of safety and effectiveness. The conclusions drawn from device comparison and nonclinical testing conducted demonstrate that the subject device performs as intended and is substantially equivalent to the predicate SOFIA™ EX Intracranial Support Catheter.
6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.