Over 900 WatchPAT night studies collected in clinical trials
>1 (sleep lab technicians/physicians)
Performance validation study of 246 patients
>1 (sleep scorers)
Indications for Use
The WatchPAT SW analyzes data recorded by WatchPAT devices (WP300, WP1, WP400) and automatically generates respiratory indices, sleep stages, snoring level, and body position. WatchPAT SW is intended for use under the direction of a healthcare professional to aid in the diagnosis of obstructive and central sleep related breathing disorders. Central indices are indicated for use in patients 17 years and older. All other outputs are indicated for 12 years and older.
Device Story
WatchPAT SW is a software-only update for analyzing sleep study data recorded by WatchPAT hardware (WP300, WP1, WP400). Input signals include PAT, pulse rate, oximetry, actigraphy, snoring, body position, and chest movement. The software uses machine learning (neural networks and classifiers) to process these inputs and automatically generate respiratory indices (AHI, RDI, AHIc, AI), sleep stages (N1, N2, N3, REM, Wake), snoring levels, body position, and cardiac flags (AFIB, premature beats). Used in professional settings (office/clinic/hospital) by healthcare professionals to view, edit, and generate sleep reports. The software is non-continuous learning (deterministic output for given input). It aids clinicians in diagnosing sleep-disordered breathing by providing automated analysis of home-recorded sleep data, facilitating clinical decision-making for sleep disorder management.
Clinical Evidence
Clinical validation involved 246 patients comparing WatchPAT SW outputs against manual PSG scoring (AASM guidelines). Results: AHI (TH≥15) sensitivity 0.91, specificity 0.93; AHIc (TH≥10) sensitivity 0.88, specificity 0.94; AI (TH≥5) sensitivity 0.79, specificity 0.97; ArI (TH≥30) sensitivity 0.86, specificity 0.82. Sleep stages Cohen's Kappa 0.65 (5-level). All outputs met performance goals, demonstrating substantial equivalence to PSG.
Technological Characteristics
Machine Learning Medical Device (MLMD) using neural networks and heuristic classifiers. Inputs: PAT, pulse, oximetry, actigraphy, snoring, body position, chest movement. Connectivity: Cloud-based (WP1/WP400) or USB (WP300). Standards: ISO 14971, IEC 62304. Deterministic, non-continuous learning architecture.
Indications for Use
Indicated for patients 12 years and older (central indices 17+) suspected of having obstructive or central sleep-related breathing disorders. Used under direction of healthcare professional.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
Predicate Devices
Watch-PAT 400 ("WP400") (Itamar Medical Ltd) (K250460)
{0}
FDA U.S. FOOD & DRUG
ADMINISTRATION
July 31, 2026
Itamar Medical, Ltd.
% John Smith
Partner
Hogan Lovells US LLP
555 13 St. NW
Washington, District of Columbia 20004
Re: K254042
Trade/Device Name: WatchPAT SW
Regulation Number: 21 CFR 868.2375
Regulation Name: Breathing Frequency Monitor
Regulatory Class: Class II
Product Code: MNR
Dated: July 2, 2026
Received: July 2, 2026
Dear John Smith:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K254042 - John Smith
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K254042 - John Smith
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Binoy J.
Mathews -S
Digitally signed by Binoy J. Mathews -S
Date: 2026.07.31 16:58:11 -04'00'
For
James J. Lee, Ph.D.
Director
DHT1C: Division of Anesthesia,
Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K254042
Device Name
WatchPAT SW
Indications for Use (Describe)
The WatchPAT SW analyzes data recorded by WatchPAT devices (WP300, WP1, WP400) and automatically generates respiratory indices, sleep stages, snoring level, and body position. WatchPAT SW is intended for use under the direction of a healthcare professional to aid in the diagnosis of obstructive and central sleep related breathing disorders. Central indices are indicated for use in patients 17 years and older. All other outputs are indicated for 12 years and older.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
# 510(k) SUMMARY
K254042
Applicant's Name: Itamar Medical ltd.
9 Halamish st.
Caesarea 3088900, Israel
Tel: +972 4 617 7000
Fax: +972 4 627 5598
Contact Person: John Smith, M.D., J.D.
Hogan Lovells Cadwalader US LLP
Columbia Square
555 Thirteenth Street, NW
Washington, DC 20004-1109
Tel: (202) 637-3638
Fax: (202) 637-5910
Email: john.smith@hoganlovells.com
Date Prepared: July 31, 2026
Trade Name: WatchPAT SW
Common or Usual Name: Ventilatory Effort Recorder
Classification Name: Breathing Frequency Monitor
Medical Specialty: Anesthesiology
Product Code: Ventilatory Effort Recorder, MNR
Device Class: Class II
Regulation Number: 868.2375
Panel: Anesthesiology
Predicate Device: Watch-PAT 400 ("WP400") (Itamar Medical Ltd), cleared under K250460; product code MNR
Reference Device: DeepRESP, NOX Medical, K252330
# Purpose of the 510(k) notice
ZOLL Itamar is introducing an update to the WatchPAT Software algorithm used for analysis of signals recorded from WatchPAT devices. The WatchPAT devices themselves remain unchanged - the current submission focuses on software modifications only (algorithm updates and new outputs). In the predicate device WP400, the WatchPAT SW (zzzPAT/WP Interface) was submitted as part of the WatchPAT device system.
{5}
# Indications for Use
The WatchPAT SW analyzes data recorded by WatchPAT devices (WP300, WP1, WP400) and automatically generates respiratory indices, sleep stages, snoring level, and body position. WatchPAT SW is intended for use under the direction of a healthcare professional to aid in the diagnosis of obstructive and central sleep related breathing disorders.
Central indices are indicated for use in patients 17 years and older. All other outputs are indicated for 12 years and older.
# Device Description
Similar to the WP Software in the WP400 predicate, the function of the WatchPAT SW is to register the device for a new study, download the recorded study data, view and analyze the data, and provide sleep reports with automatic analysis outputs for the HCP. The Sleep Reports generated by the updated WatchPAT SW are the same as the predicate, with the addition of two new outputs. Use of the WatchPAT SW with all three WatchPAT devices provides similar output information and sleep reports.
# Technological Characteristics
The principles of operation are the same between the updated WatchPAT SW and the predicate SW which was submitted with the WP400 (K250460). The input signals to the SW are the same as in the predicate device. The main differences are that there are two main algorithms modified in this submission, (1) sleep staging and (2) event detection (including two new outputs, Apnea Index and Arousal Index).
The WatchPAT SW provides the following outputs, same as the predicate device:
(1) Respiratory Disturbance Index "RDI"
(2) Apnea-Hypopnea index "AHI"
(3) Central Apnea-Hypopnea index "AHIC"
(4) Sleep stages
(5) Body position
(6) Snoring level (in decibels)
(7) Cheyne Stokes Respiration pattern \(\%\)CSR", and
(8) Atrial Fibrillation and Premature Beats (flagging only, as additional information to the sleep report).
Out of the above outputs, RDI, AHI, AHIc, and sleep stages algorithms are updated in this submission. Body Position, Snoring level, %CSR and Arrhythmia detection feature are unchanged.
In addition, WatchPAT SW provides the following new outputs:
(1) Apnea Index "AI" and
(2) Arousal Index "Arl".
The reference device DeepRESP SW K252330 is used to support evaluation of specific new outputs. DeepRESP SW (K252330) provides comparable outputs, including Apnea Index, Arousal
{6}
Index, and 5-level Sleep Staging including N1 and N2.
The clinical performance data submitted supports the inclusion of the updated and new outputs to WatchPAT SW.
## SW Design and Development
The updated WatchPAT SW is a Machine Learning Medical Device (MLMD) that does not perform continuous learning; for a given set of inputs, there will always be the same output. The input signals to the updated WatchPAT SW are the same as in the predicate device. The updated SW uses ML technologies (including several Neural Network models and other classifiers) to improve Events Detection and Sleep Stages algorithms. The Classifiers, combining Machine Learning and heuristics were used in the updated SW to achieve enhanced performance.
Training of the models was performed using over 900 WatchPAT night studies collected as part of clinical trials in which WatchPAT recordings were obtained simultaneously with in-laboratory polysomnography (PSG) in sleep laboratory settings. The data was collected across geographically diverse sites, including U.S. and non-U.S. centers. Training data included patients referred for sleep studies due to suspected sleep disorders and/or patients with cardiovascular comorbidities justifying referral for sleep assessment. The dataset includes a wide age range, a broad BMI distribution spanning underweight through morbidly obese and representation of both sexes consistent with proportions reported in the literature for sleep laboratory referred populations.
The PSG data was manually scored by experienced and qualified sleep lab technicians/physicians and served as the reference for the training process (supervised learning).
## Principles of Operation
The principles of operation are identical between the subject WatchPAT SW and the predicate SW which was submitted with the WP400 (K250460). First, the HCP registers the device to a specific patient. Once the patient performs the sleep study in their home, the data is either automatically uploaded via the cloud for the wireless configurations (WP400 and WP1) or manually uploaded to the WatchPAT SW via USB cable for WP300. Within WatchPAT SW, the HCP can then view the signals with automatic analysis, manually edit, and generate the sleep report.
## Performance and Safety Data
A series of safety and performance testing similar to the predicates were performed. These tests include non-clinical and clinical performance testing as listed below:
## Non-Clinical Performance
- The device was designed and tested under the following standards and guidelines:
- EN ISO 14971 Third Edition 2019-12 Medical devices - Application of risk management to medical devices
- IEC 62304 Edition 1.1 2015-06 Medical device software - Software life cycle processes
- Content of Premarket Submissions for Device Software Functions – FDA guidance issued June 2023.
- Software verification and validation testing was conducted to verify that the SW performs
{7}
according to its specifications as described in the Software Requirements Specifications (SRS). The Software Test Description (STD) demonstrates that the WatchPAT SW performs according to its specifications.
- The following FDA guidance were conformed to:
- Content of Premarket Submissions for Device Software Functions: June 2023
- Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions: September 2023
### Clinical Performance
Clinical testing was conducted to evaluate the performance of the WatchPAT device using the updated SW to aid in the diagnosis of sleep disordered breathing and sleep staging against polysomnography (PSG), considered the "gold standard". The manual scoring of the PSG, in accordance with the AASM sleep scoring guidelines, was performed for each recording by two independent sleep scorers. A total of 246 patients were included in the performance validation for device outputs.
Results: Across all evaluated outputs, the WatchPAT SW met the performance goals. The WatchPAT SW demonstrated diagnostic accuracy and agreement relative to the gold standard PSG scorers including AHI, RDI, AHIc and Sleep Stages. The results for new indices introduced, apnea index (AI) and arousal index (ArI), indicated their suitability for clinical use.
{8}
| Output | Results | |
| --- | --- | --- |
| AHI | AHI3% Sensitivity 0.91 (TH≥15) Specificity 0.93 (TH≥15) Cohen's Kappa κ=0.80 (TH≥15) ICC 0.95 | AHI4% Sensitivity 0.92 (TH≥15) Specificity 0.96 (TH≥15) Cohen's Kappa κ=0.88 (TH≥15) ICC 0.96 |
| RDI | RDI3% Sensitivity 0.87 (TH≥15) Specificity 0.92 (TH≥15) | RDI4% Sensitivity 0.90 (TH≥15) Specificity 0.93 (TH≥15) |
| AHIc | AHIc3% Sensitivity 0.88 (TH≥10) Specificity 0.94 (TH≥10) | AHIc4% Sensitivity 0.93 (TH≥10) Specificity 0.96 (TH≥10) |
| AI | Sensitivity 0.79 (TH≥5) Specificity 0.97 (TH≥5) | |
| ArI | Sensitivity 0.86 (TH≥30) Specificity 0.82 (TH≥30) | |
| Sleep Stages | Cohen's Kappa 5 levels κ=0.65 Sleep/Wake κ=0.77 REM/non-REM κ=0.72 | |
Results presented for indices (AHI, AHIc, AI, ArI) are the Updated SW comparison to the average of the two PSG scorers. For RDI, a single scorer was used. Results for Sleep Stages are presented as comparison to consensus of scorers.
## Performance Summary
Overall, the updated WatchPAT SW met all the objective performance goals. The results demonstrated substantially equivalent performance with an adequate degree of accuracy and agreement between the Updated WatchPAT SW and the 2 scorers for all outputs.
## Substantial Equivalence
The table below compares the technological characteristics of the Updated WatchPAT SW to the predicate SW submitted as part of the WatchPAT400 device.
Table 1. Substantial Equivalence Comparison table
| Properties | Subject Device: | Predicate Device: K250460 | Reference Device: K252330 | Comparison |
| --- | --- | --- | --- | --- |
| | WatchPAT SW | SW as part of WP400 | DeepRESP For specific technical characteristics | |
| Intended Use/Indications for use | The WatchPAT SW analyzes data recorded by WatchPAT devices (WP300, WP1, WP400) and automatically generates respiratory indices, sleep stages, snoring level, and body position. WatchPAT SW is intended for use under the direction of a healthcare | Intended use of the WP400 device: The WatchPAT device is a non-invasive home care device for use with patients suspected to have sleep related breathing disorders. The WatchPAT is a diagnostic aid for the detection of sleep related breathing disorders, sleep | N/A | Similar. The intended use is simplified for clarity relative to the WP400 device (predicate). |
{9}
| | professional to aid in the diagnosis of obstructive and central sleep related breathing disorders. Central indices are indicated for use in patients 17 years and older. All other outputs are indicated for 12 years and older. | staging (Rapid Eye Movement (REM) Sleep, Light Sleep, Deep Sleep and Wake), snoring level and body position. The WatchPAT generates a peripheral arterial tonometry ("PAT") Respiratory Disturbance Index ("PRDI"), Apnea-Hypopnea index ("PAHI"), Central ApneaHypopnea index ("PAHIC"), PAT sleep staging identification (PSTAGES) and snoring level and body position discrete states from an external integrated snoring and body position sensor. The WatchPAT's PSTAGES and snoring level and body position provide supplemental information to its PRDI/PAHI/PAHIC. The WatchPAT's PSTAGES and snoring level and body position are not intended to be used as the sole or primary basis for diagnosing any sleep related breathing disorder, prescribing treatment, or determining whether additional diagnostic assessment is warranted. PAHIC is indicated for use in patients 17 years and older. All other parameters are indicated for 12 years and older | | |
| --- | --- | --- | --- | --- |
| **Patient Population** | 12 years and above | 12 years and above | N/A | Same. |
| **Environment of Use** | Professional Use setting office/clinic/hospital | zzzPAT/WPInterface SW: Professional Use setting office/clinic/hospital | N/A | Same. |
| **Data Access** | For WP1 and WP400 (wireless configurations) download data from cloud server. For WP300 (wired configuration) download data from device wired USB connection | For WP1 and WP400 (wireless configurations) download data from cloud server. For WP300 (wired configuration) download data from device wired USB connection | N/A | Same. |
{10}
| Principle of operation | First, the HCP registers the device to a specific patient. Once the patient performs the sleep study in their home, the data is either automatically uploaded to the cloud for the wireless configurations (WP400 and WP1) or manually uploaded to the WatchPAT Analysis SW via USB cable for WP300. Within the WatchPAT Analysis SW the HCP downloads the sleep study data, and can then view the signals, run the automatic analysis, manually edit the signals, and generate the sleep report. | First, the HCP registers the device to a specific patient. Once the patient performs the sleep study in their home, the data is either automatically uploaded to the cloud for the wireless configurations (WP400 and WP1) or manually uploaded to the WatchPAT Analysis SW via USB cable for WP300. Within the WatchPAT Analysis SW the HCP downloads the sleep study data, and can then view the signals, run the automatic analysis, manually edit the signals, and generate the sleep report. | N/A | Same. |
| --- | --- | --- | --- | --- |
| Compatible Input Devices | WatchPAT devices: WP300, WP1, WP400 | WatchPAT devices: WP300, WP1, WP400 | N/A | Same. |
| Input signals | PAT, Pulse rate, Oximetry, Actigraphy, Snoring, Body Position, Chest Movement | PAT, Pulse rate, Oximetry, Actigraphy, Snoring, Body Position, Chest Movement | N/A | Same. |
| Outputs | AHI, AHIc, %CSR, RDI, AI, ArI Snoring level Body position – 5 discrete states 5 Sleep Stages, N1 N2 N3 REM Wake AFIB, PB Oxygen Saturation Statistics | pAHI, pAHIc, %CSR, pRDI Snoring level Body position – 5 discrete states 4 Sleep Stages, Light Deep REM Wake AFIB, PB Oxygen Saturation Statistics | AHI, CAHI, All Apnea, Central Apnea, Obstructive Apnea, Hypopnea, Arousal Events, Desaturation, Sleep Stages (N1 N2 N3 REM Wake), Position | Similar to predicate. Apnea Index (AI), Arousal Index (ArI) and differentiation of “Light” sleep stage into N1 and N2 are similar to Reference device. |
| Main Performance Results | AHI (threshold = 15) Sensitivity 91% Specificity 93% AHIc (threshold = 10) Sensitivity 88% Specificity 94% Sleep Stages Cohen’s Kappa 0.65 All apnea Sensitivity 0.79 (TH≥5) Specificity 0.97 (TH≥5) Arousal events Sensitivity 0.86 (TH≥30) Specificity 0.82 (TH≥30) | AHI (threshold = 15) Sensitivity 85% Specificity 88.2% AHIc (threshold = 10) Sensitivity 71.4% Specificity 98.6% Sleep Stages Cohen’s Kappa 0.46 | All apnea PPA%: 83.7 NPA%: 98.2 OPA%: 97.1 Arousal events PPA%: 62.1 NPA%: 89.1 OPA%: 81.5 | Similar. |
| Editing functionality | Automatic analysis for all outputs, with capability to manually edit | Automatic analysis for all outputs, with capability to manually edit | N/A | Same. |
{11}
| Sleep Report | Comprises of Outputs generated, graphics, histograms, patient information | Comprises of Outputs generated, graphics, histograms, patient information | N/A | Similar. Only difference is adaptation to add the new indices and sleep stages. |
| --- | --- | --- | --- | --- |
| Algorithmic Design | Non-continuous learning MLMD Deterministic | Non-continuous learning Deterministic | N/A | Similar. Subject WatchPAT SW utilizes modern ML technologies. Safety and effectiveness of algorithm has been verified and validated via non-clinical and clinical testing. |
| Cybersecurity | Encrypted HTTPS/TLS1.2 secure data transfer | Encrypted HTTPS/TLS1.2 secure data transfer | N/A | Same. |
The subject device (WatchPAT SW) is submitted separately from previously cleared hardware, while the primary predicate K250460 is the complete WatchPAT 400 system. This software-only submission approach is justified by precedent (Reference device: K241960 DeepRESP SW cleared separately from NOX Sleep System hardware) and reflects the common software platform used across all WatchPAT hardware configurations with independent software lifecycle management.
## Conclusion
The Updated WatchPAT SW has the same intended use and similar indications as the predicate SW (zzzPAT/WPInterface) submitted in K250460. Both are intended for use with patients suspected of having sleep related breathing disorders. The subject WatchPAT SW does not raise different questions of safety or effectiveness, as all new outputs are based on the same inputs that exist in the previous FDA-cleared products and use thereof under a diagnostic aid indication does not alter the clinical utility or risk of the device. Further, non-clinical and clinical performance testing were conducted, and all testing demonstrated substantially equivalent performance. Based on the comparison described above, the updated WatchPAT SW is found substantially equivalent to the predicate zzzPAT/WPInterface.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.