K254036 · Datex Ohmeda, Inc. · HGM · Aug 10, 2026 · Obstetrics/Gynecology
Device Facts
Record ID
K254036
Device Name
Novii+ Wireless Patch System
Applicant
Datex Ohmeda, Inc.
Product Code
HGM · Obstetrics/Gynecology
Decision Date
Aug 10, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2740
Device Class
Class 2
Indications for Use
The Novii+ Pod is an antepartum and intrapartum Maternal/Fetal Monitor that non-invasively measures fetal heart rate (FHR), uterine activity (UA) and maternal heart rate (MHR). The Novii+ Pod acquires the FHR tracing from abdominal surface electrodes that pick up the fetal ECG (fECG) signal. Using the same surface electrodes, the Pod also acquires the UA tracing from the uterine electromyography (EMG) signal and the MHR tracing from the maternal ECG signal (mECG). The Pod is indicated for use on women who are at 34 weeks and 0/7 days and greater with singleton pregnancies, using surface electrodes on the maternal abdomen. The Novii Patch is an accessory to the Novii+ Pod that connects directly to the Novii+ Pod and contains the surface electrodes that attach to the abdomen. The Novii+ Interface is an accessory to the Novii+ Pod which provides a means of interfacing the wireless output of the Novii+ Pod to the transducer inputs of a Maternal/Fetal Monitor. The Novii+ Interface enables signals collected by the Novii+ Pod to be printed and displayed on a Maternal/Fetal Monitor and sent to a central network, if connected. The Novii+ Pod Maternal/Fetal Monitor and its accessories are intended for use by healthcare professionals in a clinical setting.
Device Story
Battery-powered maternal/fetal monitor; uses abdominal surface electrodes to acquire fetal ECG (fECG), uterine electromyography (EMG), and maternal ECG (mECG) signals. System consists of a wearable Novii+ Pod, a disposable Novii Patch, and a Novii+ Interface. Pod connects to patch via pogo pins; processes signals via analog front end with instrumentation amplifiers and bandpass filters. Wireless data transmission to interface via Bluetooth 4.0. Interface connects to standard maternal/fetal monitors for printing, display, and network integration. Used by healthcare professionals in clinical settings for fetal/maternal surveillance. Provides real-time continuous monitoring; alerts for signal loss, battery status, and coincidence. Benefits include non-invasive monitoring of fetal/maternal well-being.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including electrical performance, mechanical performance, electromagnetic compatibility (IEC 60601-1-2), electrical safety (IEC 60601-1), and biocompatibility (ISO 10993-1).
Technological Characteristics
System includes a battery-powered Pod and disposable patch. Patch features 5 Silver/Silver Chloride electrodes on PET substrate; latex/PVC free. Pod uses analog front end with 4 instrumentation amplifiers and bandpass filters. Connectivity via Bluetooth 4.0 (Class 1.5). Sterilization/reprocessing for reusable components (Pod/Interface) leveraged from predicate. Complies with IEC 60601-1 and IEC 60601-1-2.
Indications for Use
Indicated for women at 34 weeks and 0/7 days or greater gestation with singleton pregnancies for antepartum and intrapartum monitoring of fetal heart rate, maternal heart rate, and uterine activity.
Regulatory Classification
Identification
A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.
{0}
**U.S. FOOD & DRUG**
ADMINISTRATION
August 10, 2026
Datex Ohmeda, Inc.
Shiwani Zalpuri
Manger, Regulatory Affairs
3114 N Grandview Blvd.
Waukesha, Wisconsin 53188
Re: K254036
Trade/Device Name: Novii+ Wireless Patch System
Regulation Number: 21 CFR 884.2740
Regulation Name: Perinatal Monitoring System and accessories
Regulatory Class: II
Product Code: HGM
Dated: July 9, 2026
Received: July 9, 2026
Dear Shiwani Zalpuri:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.
{1}
K254036 - Shiwani Zalpuri
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K254036 - Shiwani Zalpuri
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Monica D. Garcia -S
Monica D. Garcia, Ph.D.
Assistant Director
DHT3B: Division of Reproductive,
Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K254036
Device Name
Novii+ Wireless Patch System
Indications for Use (Describe)
The Novii+ Pod is an antepartum and intrapartum Maternal/Fetal Monitor that non-invasively measures fetal heart rate (FHR), uterine activity (UA) and maternal heart rate (MHR). The Novii+ Pod acquires the FHR tracing from abdominal surface electrodes that pick up the fetal ECG (fECG) signal. Using the same surface electrodes, the Pod also acquires the UA tracing from the uterine electromyography (EMG) signal and the MHR tracing from the maternal ECG signal (mECG). The Pod is indicated for use on women who are at 34 weeks and 0/7 days and greater with singleton pregnancies, using surface electrodes on the maternal abdomen.
The Novii Patch is an accessory to the Novii+ Pod that connects directly to the Novii+ Pod and contains the surface electrodes that attach to the abdomen.
The Novii+ Interface is an accessory to the Novii+ Pod which provides a means of interfacing the wireless output of the Novii+ Pod to the transducer inputs of a Maternal/Fetal Monitor. The Novii+ Interface enables signals collected by the Novii+ Pod to be printed and displayed on a Maternal/Fetal Monitor and sent to a central network, if connected.
The Novii+ Pod Maternal/Fetal Monitor and its accessories are intended for use by healthcare professionals in a clinical setting
Type of Use (Select one or both, as applicable)
☑
Prescription Use (Part 21 CFR 801 Subpart D)
☐
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
{4}
Traditional 510(k)
510(k) Summary
Novii+ Wireless Patch System
Alternative Patch
K254036
Page 1 of 6
GE HealthCare
# 510(k) Summary K254036
In accordance with 21 CFR 807.92 the following summary information is provided:
| Date: | 10 August 2026 |
| --- | --- |
| Owner/Submitter: | GE HealthCare (dba Datex- Ohmeda, Inc.) 3114 N. Grandview Blvd Waukesha, WI 53188 |
| Primary Contact Person: | Shiwani Zalpuri Regulatory Affairs Manager GE HealthCare Phone: +91 9871090801 Email: shiwani.zalpuri@gehealthcare.com |
| Secondary Contact Person: | Isabel McGann Regulatory Affairs Director GE HealthCare Phone: 720-447-0747 Email: Isabel.mcgann@gehealthcare.com |
| Device Trade Name: | Novii+ Wireless Patch System |
| Common/Usual Name: | Perinatal monitoring system and accessories |
| Regulation Number: | 21 CFR 884.2740 |
| Regulation Name: | Perinatal monitoring system and accessories |
| Classification: | Class II |
| Product Code: | HGM (System, Monitoring, Perinatal) |
| Predicate Device: | Novii+ Wireless Patch system (K231964) The predicate device has not been subject to a design-related recall. |
| Device Description: | The Novii+ Wireless Patch System (Novii+ system) is battery-powered Maternal/Fetal Monitor that measures abdominal fetal heart rate (FHR), abdominal maternal heart rate (MHR), and abdominal uterine activity (UA) and sends data to a previously cleared Corometrics 259cx (K072976) or a Corometrics 174 maternal/fetal monitor (K993751) via Novii+ Pod. The monitor enables the abdominal electrophysiological signal to be picked up from three different positions on the maternal abdomen using the 5 electrodes on the Novii Patch. The Novii+ Pod acquires the FHR tracing from abdominal surface electrodes that pick up the fetal ECG (fECG) signal. Using the same surface electrodes, the Pod also acquires the UA tracing from the uterine electromyography (EMG) signal and the MHR tracing from the maternal ECG signal (mECG). The Novii+ Pod is a battery-powered device that the patient wears for surveillance of fetal and maternal well-being. The Novii+ Pod is indicated for use on women who are at 34 weeks and 0/7 days and greater with singleton pregnancies. This Traditional 510(k) is to add an alternate, equivalent patch to the previously cleared Novii+ Wireless Patch System (K231964). This alternate patch is interchangeable as it is equivalent in form, fit and function to the predicate patch. |
Page 1 of 6
-Public-
{5}
K254036
Page 2 of 6
**Traditional 510(k)**
510(k) Summary
**Novii+ Wireless Patch System**
Alternative Patch
![GE HealthCare logo]() GE HealthCare
| **Indications for Use:** | The Novii+ Pod is an antepartum and intrapartum Maternal/Fetal Monitor that non-invasively measures fetal heart rate (FHR), uterine activity (UA) and maternal heart rate (MHR). The Novii+ Pod acquires the FHR tracing from abdominal surface electrodes that pick up the fetal ECG (fECG) signal. Using the same surface electrodes, the Pod also acquires the UA tracing from the uterine electromyography (EMG) signal and the MHR tracing from the maternal ECG signal (mECG). The Pod is indicated for use on women who are at 34 weeks and 0/7 days and greater with singleton pregnancies, using surface electrodes on the maternal abdomen. The Novii Patch is an accessory to the Novii+ Pod that connects directly to the Novii+ Pod and contains the surface electrodes that attach to the abdomen. The Novii+ Interface is an accessory to the Novii+ Pod which provides a means of interfacing the wireless output of the Novii+ Pod to the transducer inputs of a Maternal/Fetal monitor. The Novii+ Interface enables signals collected by the Novii+ Pod to be printed and displayed on a Maternal/Fetal Monitor and sent to a central network, if connected. The Novii+ Pod Maternal/Fetal Monitor and its accessories are intended for use by healthcare professionals in a clinical setting. |
| --- | --- |
# **COMPARISON OF INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE:**
A comparison of the intended use and technological features of the subject and predicate device is provided in Table 1 below.
**Table 1: High-level Comparison of Subject Device to Predicate Device**
| Specification | Predicate Device Novii+ Wireless Patch System K231964 | Subject Device Novii+ Wireless Patch System K254036 | Discussion of Differences |
| --- | --- | --- | --- |
| **Indications for Use** | The Novii+ Pod is an antepartum and intrapartum Maternal/Fetal Monitor that non-invasively measures fetal heart rate (FHR), uterine activity (UA) and maternal heart rate (MHR). The Novii+ Pod acquires the FHR tracing from abdominal surface electrodes that pick up the fetal ECG (fECG) signal. Using the same surface electrodes, the Pod also acquires the UA tracing from the uterine electromyography (EMG) signal and the MHR tracing from the maternal ECG signal (mECG). The Pod is indicated for use on women who are at 34 weeks and 0/7 days and greater with singleton pregnancies, using surface electrodes on the maternal abdomen. | The Novii+ Pod is an antepartum and intrapartum Maternal/Fetal Monitor that non-invasively measures fetal heart rate (FHR), uterine activity (UA) and maternal heart rate (MHR). The Novii+ Pod acquires the FHR tracing from abdominal surface electrodes that pick up the fetal ECG (fECG) signal. Using the same surface electrodes, the Pod also acquires the UA tracing from the uterine electromyography (EMG) signal and the MHR tracing from the maternal ECG signal (mECG). The Pod is indicated for use on women who are at 34 weeks and 0/7 days and greater with singleton pregnancies, using surface electrodes on the maternal abdomen. | Identical |
Page 2 of 6
-Public-
{6}
K254036
Page 3 of 6
**Traditional 510(k)**
510(k) Summary
**Novii+ Wireless Patch System**
Alternative Patch
![GE HealthCare logo]() GE HealthCare
| Specification | Predicate Device Novii+ Wireless Patch System K231964 | Subject Device Novii+ Wireless Patch System K254036 | Discussion of Differences |
| --- | --- | --- | --- |
| | The Novii Patch is an accessory to the Novii+ Pod that connects directly to the Novii+ Pod and contains the surface electrodes that attach to the abdomen. The Novii+ Interface is an accessory to the Novii+ Pod which provides a means of interfacing the wireless output of the Novii+ Pod to the transducer inputs of a Maternal/Fetal Monitor. The Novii+ Interface enables signals collected by the Novii+ Pod to be printed and displayed on a Maternal/Fetal Monitor and sent to a central network, if connected. The Novii+ Pod Maternal/Fetal Monitor and its accessories are intended for use by healthcare professionals in a clinical setting. | Novii+ Pod and contains the surface electrodes that attach to the abdomen. The Novii+ Interface is an accessory to the Novii+ Pod which provides a means of interfacing the wireless output of the Novii+ Pod to the transducer inputs of a Maternal/Fetal Monitor. The Novii+ Interface enables signals collected by the Novii+ Pod to be printed and displayed on a Maternal/Fetal Monitor and sent to a central network, if connected. The Novii+ Pod Maternal/Fetal Monitor and its accessories are intended for use by healthcare professionals in a clinical setting. | |
| **Measured Parameters** | - Fetal Heart Rate (FHR) - Maternal Heart Rate (MHR) - Uterine Activity (UA) | - Fetal Heart Rate (FHR) - Maternal Heart Rate (MHR) - Uterine Activity (UA) | Identical |
| **Anatomical Sites** | Trans-abdominal | Trans-abdominal | Identical |
| **Environment of Use** | Healthcare professionals in a clinical setting | Healthcare professionals in a clinical setting | Identical |
| **Patient Population** | 34 weeks and 0/7 days and greater with singleton pregnancies | 34 weeks and 0/7 days and greater with singleton pregnancies | Identical |
| **Pod** | The Novii+ Pod connects to each electrode of the patch via Pogo pins. The pod has an analogue front end connected to each electrode that is made up of 4 instrumentation amplifiers and associated electronic bandpass filters. The Novii+ Pod has a dual tone color (Purple and white). | The Novii+ Pod connects to each electrode of the patch via Pogo pins. The pod has an analogue front end connected to each electrode that is made up of 4 instrumentation amplifiers and associated electronic bandpass filters. The Novii+ Pod has a dual tone color (Purple and white). | Identical |
Page 3 of 6
-Public-
{7}
K254036
Page 4 of 6
Traditional 510(k)
510(k) Summary
Novii+ Wireless Patch System
Alternative Patch
GE HealthCare
| Specification | Predicate Device Novii+ Wireless Patch System K231964 | Subject Device Novii+ Wireless Patch System K254036 | Discussion of Differences |
| --- | --- | --- | --- |
| | The Novii+ Pod is not backward compatible and cannot connect to original Novii Interface (K140862). | The Novii+ Pod is not backward compatible and cannot connect to original Novii Interface (K140862). | |
| **Novii Patch** | The Novii Patch is made of 5 Silver/silver chloride electrodes that are printed on Polyethylene terephthalate (PET) substrate. The Patch has a plastic clip with magnets to hold the Novii+ Pod in place. Patch: - Single use - 5 Silver/Silver Chloride electrodes on PET substrate with silver tracks and pads - Latex / PVC free The Novii Patch is backward compatible and can connect to original Novii Pod (K140862) and the Novii+ Pod. | The alternate Novii Patch is made of 5 Silver/silver chloride electrodes that are printed on Polyethylene terephthalate (PET) substrate. The Patch has a plastic clip with magnets to hold the Novii+ Pod in place. Patch: - Single use - 5 Silver/Silver Chloride electrodes on PET substrate with silver tracks and pads - Latex / PVC free The Novii Patch is backward compatible and can connect to original Novii Pod (K140862) and the Novii+ Pod. | **Different.** |
| **Supported Monitors** | GE Corometrics Maternal / Fetal Monitor | GE Corometrics Maternal / Fetal Monitor | Identical |
| **Alerts** | Interface provides the following alerts: - Patient out of Bluetooth range - Low Pod battery - MHR/FHR coincidence - Pod removed from patch - Pod not returned to charging bay - Patch electrode detached / broken pin - Patch not GEHC genuine | The Interface provides the following alerts: - Patient out of Bluetooth range - Low Pod battery - MHR/FHR coincidence - Pod removed from patch - Pod not returned to charging bay - Patch electrode detached / broken pin - Patch not GEHC genuine | Identical |
| **Data Interface (Wireless) Interoperability** | The Novii+ Pod and Interface connect wirelessly using Bluetooth version 4.0. Input: Bluetooth Class 1.5, V4.0 real/time continuous, from Novii™ Pod. Modified Series Protocol | The Novii+ Pod and Interface connect wirelessly using Bluetooth version 4.0. Input: Bluetooth Class 1.5, V4.0 real/time continuous, from Novii™ Pod. Modified Series Protocol Irda interface to initiate authentication and Bluetooth pairing | Identical |
Page 4 of 6
-Public-
{8}
Traditional 510(k)
510(k) Summary
Novii+ Wireless Patch System
Alternative Patch
K254036
Page 5 of 6
GE HealthCare
| Specification | Predicate Device Novii+ Wireless Patch System K231964 | Subject Device Novii+ Wireless Patch System K254036 | Discussion of Differences |
| --- | --- | --- | --- |
| | Irda interface to initiate authentication and Bluetooth pairing | | |
The subject and predicate device have identical indications for use statements and have the same intended use – non-invasive monitoring of fetal heart rate, maternal heart rate, and uterine activity. The subject device (Novii+ Wireless Patch System) includes an alternate patch that has similar fundamental scientific technology to the predicate device (Novii Wireless Patch system) . It maintains the predicate device's functionality, performance and clinical workflows. The inclusion of alternate patch does not raise new questions of safety and effectiveness and is evaluated through non-clinical performance testing as discussed below.
### PERFORMANCE TESTING
Summary of Non-Clinical Performance Testing:
Bench testing related to performance as part of the alternate patch qualification such as electrical performance, mechanical performance, electromagnetic compatibility, electrical safety, and biocompatibility was conducted on the Novii+ Wireless Patch System demonstrating the design meets the specifications as discussed below:
EMC/Electrical Safety/Wireless
- The Novii+ Wireless Patch System meets the EMC requirements described in IEC 60601-1-2 Edition 4.0 2014-02 "Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests" and 2022 FDA Guidance "Electromagnetic Compatibility (EMC) of Medical Devices".
- The Novii+ Wireless Patch System meets the electrical safety requirements of IEC 60601-1:2012 "Medical electrical equipment - Part 1: General requirements for basic safety and essential performance - Edition 3.1".
- The Novii+ wireless patch system (the Novii+ Pod and Patch patient contact parts at the time of defibrillation) was tested for misuse or emergency use of the defibrillator on patient per the requirements of IEC 60601-1: Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance clause 8.5.5 as applicable for the defibrillation.
- Wireless testing was leveraged from the predicate device as there is no change to Novii+ Pod.
Biocompatibility
The Novii+ Wireless Patch System follows the 2023 FDA Biocompatibility guidance document "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", Guidance for Industry and Food and Drug Administration Staff" and ISO 10993-1:2018 Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process. The testing data demonstrates that the subject device is not cytotoxic, non-sensitizing, and non-irritating.
Page 5 of 6
-Public-
{9}
Traditional 510(k)
510(k) Summary
Novii+ Wireless Patch System
Alternative Patch
K254036
Page 6 of 6
GE HealthCare
# Reprocessing
The Novii+ Wireless Patch System follows the 2015 FDA guidance “Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, Guidance for Industry and Food and Drug Administration Staff”. Since there is no change to the reusable components, Novii+ Pod and Interface, the reprocessing validation is leveraged from the predicate device.
# Software and Cybersecurity
Since there is no change to the device software and firmware in Novii+ Pod and Interface, the software and cybersecurity documentation is leveraged from the predicate device.
# Human Factors
Since there are no changes to the Novii+ Pod and Interface and the device use with inclusion of alternate patch, human factors testing is leveraged from the predicate device.
# CONCLUSION:
The results of the testing described above demonstrate that the Novii+ Wireless Patch System is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Page 6 of 6
-Public-
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.