# Flow Water-Based Personal Lubricant (K254011)

_LBDO Group, Inc. · NUC · Jun 22, 2026 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K254011

## Device Facts

- **Applicant:** LBDO Group, Inc.
- **Product Code:** [NUC](/productcode/NUC.md)
- **Decision Date:** Jun 22, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.5300
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

Flow Water-Based Personal Lubricant is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

## Device Story

Flow Water-Based Personal Lubricant is a clear, aqueous gel applied topically to penile and/or vaginal surfaces. It functions by forming a thin film to reduce friction, enhance comfort, and supplement natural lubrication during intimate sexual activity. The device is intended for over-the-counter use by consumers. It is not a contraceptive or spermicide. The product is formulated with purified water, humectants, plant-derived extracts, and stabilizing agents. It is compatible with natural rubber latex and polyisoprene condoms, but incompatible with polyurethane condoms. Clinical benefit is derived from improved comfort during sexual activity.

## Clinical Evidence

No clinical data. Evidence consists of bench testing: biocompatibility (cytotoxicity, sensitization, vaginal irritation, acute systemic toxicity per ISO 10993), condom compatibility (ASTM D7661-23), and stability testing (real-time and accelerated) confirming shelf-life and adherence to physical/microbial specifications.

## Technological Characteristics

Aqueous gel; pH 4.5–5.0; viscosity 3,000–5,000 cps; osmolality 720–904 mOsm/kg. Ingredients: water, aloe barbadensis leaf juice, propanediol, sodium hyaluronate, gluconolactone, plant extracts, xanthan gum, hydroxyethylcellulose, sodium benzoate, potassium sorbate, citric acid. Tested per USP <51> (antimicrobial effectiveness), USP <61>/<1111> (microbial limits). Biocompatibility per ISO 10993-1. Condom compatibility per ASTM D7661-23.

## Regulatory Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

## Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

## Predicate Devices

- The Sex Gel Personal Lubricant ([K181078](/device/K181078.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

**FDA** **U.S. FOOD & DRUG**  
ADMINISTRATION

June 22, 2026

LBDO Group, Inc.  
Joshua Benjamin  
Chief Financial Officer  
610 Eastern Pkwy.  
Brooklyn, New York 11225

Re: K254011  
Trade/Device Name: Flow Water-Based Personal Lubricant  
Regulation Number: 21 CFR 884.5300  
Regulation Name: Condom  
Regulatory Class: II  
Product Code: NUC  
Dated: December 15, 2025  
Received: December 15, 2025

Dear Joshua Benjamin:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device'

U.S. Food & Drug Administration  
10903 New Hampshire Avenue  
Silver Spring, MD 20903  
www.fda.gov

{1}

K254011 - Joshua Benjamin

Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See

{2}

K254011 - Joshua Benjamin

Page 3

the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Monica D. Garcia -S

Monica D. Garcia, Ph.D.

Assistant Director

DHT3B: Division of Reproductive,

Gynecology, and Urology Devices

OHT3: Office of Gastrorenal, ObGyn,

General Hospital, and Urology Devices

Office of Product Evaluation and Quality

Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.

# Indications for Use

510(k) Number (if known)
K254011

Device Name
Flow Water-Based Personal Lubricant

Indications for Use (Describe)

Flow Water-Based Personal Lubricant is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. It is not compatible with polyurethane condoms.

Type of Use (Select one or both, as applicable)

☐ Prescription Use (Part 21 CFR 801 Subpart D)

☑ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{4}

K254011
Page 1 of 5

# **510(k) Summary – K254011**
**Flow Water-Based Personal Lubricant**

**1. Submitter Information:**

LBDO Group Inc.
610 Eastern Parkway
Brooklyn, NY 11225, USA
TEL: +6465708150

**2. Submission Correspondent:**

Mr. Joshua Benjamin
Chief Financial Officer
LBDO Group Inc.
Email: josh@lbdo.com.au
+6465708150

**3. Date of Preparation:**

June 22, 2026

**4. Device Identification:**

Device Trade Name: Flow Water-Based Personal Lubricant
Common Name: Personal lubricant
Regulation Name: Condom
Regulation Number: 21 CFR 884.5300
Regulatory Class: Class II
Product Code: NUC (lubricant, personal)

**5. Predicate Device**

The Sex Gel Personal Lubricant (Necessaire, Inc.) - K181078

The predicate device has not been subject to a design-related recall.

**6. Device Description**

Flow Water-Based Personal Lubricant is a clear, aqueous gel formulation intended for penile and/or vaginal application to moisturize and lubricate during intimate activity. It is not a contraceptive or spermicide. The device functions by forming a thin, water-based film that reduces friction between body surfaces, enhancing comfort and ease of movement during intimate sexual activity.

Page 1 of 5

{5}

K254011  
Page 2 of 5

# **510(k) Summary – K254011**  
**Flow Water-Based Personal Lubricant**---

It is composed of purified water, humectants, plant-derived extracts, and stabilizing agents. The Flow Water-Based Personal Lubricant is compatible with natural rubber latex and polyisoprene condoms and is not compatible with polyurethane condoms. The device specifications are outlined in the table below.

**Table 1: Device Specifications for Flow Water-Based Personal Lubricant**

|  Property | Specification  |
| --- | --- |
|  Appearance | Clear, aqueous gel  |
|  Color | Colorless  |
|  Odor | Odorless  |
|  Viscosity | 3,000 cps – 5,000 cps  |
|  pH | 4.5 – 5.0  |
|  Osmolality | 720–904 mOsm/kg  |
|  Antimicrobial Effectiveness per USP <51> | Meets USP <51> Category 2 criteria. Bacteria show no less than 2.0 log reduction from initial count at 14 days and no increase from the 14-day count at 28 days. Yeast and molds show no increase from the initial calculated count at 14 and 28 days.  |
|  Total aerobic microbial count (TAMC) per USP <61> and <1111> | <100 cfu/g  |
|  Total yeast and mold count (TYMC) per USP <61> and <1111> | <10 cfu/g  |
|  Absence of pathogenic organisms (Staphylococcus aureus, Pseudomonas aeruginosa, Salmonella, Escherichia coli, and Candida albicans) | Absent  |

## 7. Indications for Use

Flow Water-Based Personal Lubricant is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

Page 2 of 5

{6}

K254011  
Page 3 of 5

# **510(k) Summary – K254011**  
**Flow Water-Based Personal Lubricant**---

# **8. Comparison of Intended Use and Technological Characteristics with the Predicate Device**

The table below compares the intended use and technological characteristics of the subject and predicate device.

|  **Characteristic/Feature** | **Flow Water-Based Personal Lubricant (Subject Device K254011)** | **The Sex Gel (Predicate Device K181078)**  |
| --- | --- | --- |
|  Indications for Use | Flow Water-Based Personal Lubricant is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms. | The Sex Gel Personal Lubricant is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex, polyisoprene, and polyurethane condoms.  |
|  Rx/OTC | OTC | OTC  |
|  Sterile | No | No  |
|  Formulation type | Water-based | Water-based  |
|  Primary ingredients | Water, aloe barbadensis leaf juice, propanediol, sodium hyaluronate, gluconolactone, plant extracts (Kakadu plum, Quandong), xanthan gum, hydroxyethylcellulose, sodium benzoate, potassium sorbate, and citric acid | Water, aloe barbadensis leaf juice, sorbitol, hydroxyethylcellulose, allantoin, lactic acid / tocopherols (vitamin E), sodium hyaluronate, sodium benzoate & potassium sorbate  |
|  Appearance | Clear, aqueous gel | Gel  |
|  pH | 4.5 – 5.0 | 4.0 – 5.0  |
|  Viscosity | 3,000 cps – 5,000 cps | 3,000 cps – 5,000 cps  |
|  Osmolality | 720 – 904 mOsm/kg | 435 – 535 mOsm/kg  |

Page 3 of 5

{7}

K254011

Page 4 of 5

# 510(k) Summary – K254011

# Flow Water-Based Personal Lubricant

|  Antimicrobial Effectiveness Tested per USP <62> | Yes | Yes  |
| --- | --- | --- |
|  Antimicrobial Effectiveness Tested per USP <51> | Yes | Yes  |
|  Total aerobic microbial count (TAMC) per USP <61> | <100 cfu/g | <10 cfu/g  |
|  Total mold/yeast count (TYMC) per USP <61> | <10 cfu/g | <10 cfu/g  |
|  Condom Compatibility | Compatible with Natural Rubber Latex and Polyisoprene Condoms. Not compatible with polyurethane condoms. | Compatible with Natural Rubber Latex, Polyisoprene, and Polyurethane Condoms.  |
|  Biocompatibility Tested | Yes | Yes  |
|  The product is not a contraceptive and does not contain spermicide | Yes | Yes  |
|  Shelf life | 6 months | 6 months  |

The subject device and predicate device have the same indications for use statements and intended use - to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. As shown in the table above, there are differences in formulation, osmolality, microbial specifications, and condom compatibility. These differences in technological characteristics do not raise different questions of safety and effectiveness.

### 9. Summary of Non-Clinical Performance Testing

#### Biocompatibility

Biocompatibility studies were performed in accordance with the 2023 FDA Guidance Document, Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices-Part 1: Evaluation and testing within a risk management process."

The following testing was conducted:

a. Cytotoxicity (ISO 10993-5:2009)

Page 4 of 5

{8}

K254011
Page 5 of 5

# **510(k) Summary – K254011**
**Flow Water-Based Personal Lubricant**

b. Sensitization (ISO 10993-10:2021)
c. Vaginal Irritation (ISO 10993-23:2021)
d. Acute Systemic Toxicity (ISO 10993-11:2017)

The test results demonstrate that the subject device is non-cytotoxic, non-sensitizing, non-irritating, and non-systemically toxic.

**Condom Compatibility:**

The subject device was tested for compatibility with natural rubber latex, polyisoprene, and polyurethane condoms using ASTM D7661-23, *Standard Test Method for Determining Compatibility of Personal Lubricants with Natural Rubber Latex Condoms*. The results show that Flow Water-Based Personal Lubricant is compatible with natural rubber latex and polyisoprene male condoms and is not compatible with lubricated polyurethane male condoms.

**Shelf Life:**

The subject device has a shelf-life of 6 months based on the results of real-time and accelerated stability testing. The shelf-life study evaluated all device specifications listed in Table 1 and met all device specifications throughout the stated shelf-life.

**10. Conclusions**

The results of the performance testing described above demonstrate that the Flow Water-Based Personal Lubricant is as safe and effective as the predicate device and supports a determination of substantial equivalence.

Page 5 of 5

---

**Source:** [https://fda-staging.innolitics.com/device/K254011](https://fda-staging.innolitics.com/device/K254011)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).